Jia S, Zhou J, Zhang Q, Zhou S, Wang Z, Ye X, Wu J. Clinical research of fibroscan ‒ TE-CAP at noninvasive diagnosis of hepatic steatosis in children.
Clinics (Sao Paulo) 2024;
79:100387. [PMID:
38805982 PMCID:
PMC11152890 DOI:
10.1016/j.clinsp.2024.100387]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/12/2024] [Accepted: 04/29/2024] [Indexed: 05/30/2024] Open
Abstract
BACKGROUND & AIMS
The authors assess the diagnostic accuracy of the Transient Elastography-Controlled Attenuation Parameter (TE-CAP) in children of Southern China.
METHODS
105 obese or overweight children and adolescents were enrolled in the diagnostic test of TE-CAP assessment of hepatic steatosis using MRI-PDFF. Hepatic steatosis grades S0-S3 were classified. Statistical correlation, agreement and consistency between methods were evaluated. The diagnostic efficiency of TE-CAP was evaluated. The authors used the cutoff value of TE-CAP to detect hepatic steatosis in another 356 children.
RESULTS
The Area Under Curve (AUC) of TE-CAP for grade ≥ S1, ≥ S2, and ≥ S3 steatosis were 0.975, 0.984, and 0.997, respectively. For detecting ≥ S1 steatosis, TE-CAP had a sensitivity of 96 % and a specificity of 97 %. For detecting ≥ S2 steatosis, TE-CAP had a sensitivity of 97 % and a specificity of 93 %. For detecting ≥ S3 steatosis, TE-CAP had a sensitivity of 1 and a specificity of 94 %. TE-CAP and MRI-PDFF had a linear correlation (r = 0. 0.87, p < 0.001). The hepatic steatosis was identified in 40.2 % (143/356) of children in which the obesity and overweight were 69.8 % (113/162) and 40.0 % (18/45).
CONCLUSION
TE-CAP showed excellent diagnostic accuracy in pediatric hepatic steatosis.
Collapse