1
|
Tuncel N, Mahdavi SR, Haghparast M, Karakus I, Haghparast A, Nazari V. The feasibility of the pretreatment verification of 2D dose distributions in radiation therapy with small fields using the electronic portal imaging device. J Cancer Res Ther 2023; 19:S815-S820. [PMID: 38087974 DOI: 10.4103/jcrt.jcrt_2398_22] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/13/2022] [Accepted: 02/06/2023] [Indexed: 02/23/2024]
Abstract
BACKGROUND The present study aims to evaluate the performance of an Electronic portal imaging device (EPID) for measuring dosimetric parameters and for verification of dose in small photon fields. MATERIAL AND METHODS In this study, the beam profiles were obtained using the amorphous silicon (a-Si) EPID for field sizes ranging from 1 × 1 to 10 × 10 cm 2 at energies 6 and 18 mega-voltage (MV). For comparison, the dosimetric parameters, including penumbra widths and field sizes, were measured with the pinpoint, diode, and Semiflex dosimeters. Finally, Rando Phantom was used to compare the two-dimensional (2D) Dose distribution between EPID and Treatment Planning System (TPS). RESULTS In both 5 cm and 10 cm depths, there were large differences between the measured doses obtained from TPS, Pinpoint detector, and Farmer detector in 1 × 1 field size. The differences become negligible as the field sizes increase and from 3 × 3 field size to 10 × 10 field size, the maximum observed differences are 2 cGy and 2.4 cGy for 5 cm and 10 cm depths, respectively. The results indicate that the penumbra widths are smaller in the Gantry-Target (GT) direction compared to the Right-Left (RL) direction. The maximum difference (47.6%) was observed for EPID in the 10 × 10 field size, and the minimum difference (16.6%) was observed for TPS in the 1 × 1 field size. Finally, 2D dose distributions obtained by EPID and TPS exhibit excellent agreement. CONCLUSION EPID is an excellent tool for the measurement of dosimetry parameters such as dose profiles, penumbra widths, field sizes, and pretreatment verification of 2D dose distributions, especially in small fields.
Collapse
Affiliation(s)
- Nina Tuncel
- Radiation Oncology Department, School of Medicine, Akdeniz University, Antalya, Turkey
| | - Seied Rabi Mahdavi
- Department of Medical Physics, School of Medicine, Iran University of Medical Sciences, Tehran, Iran
| | - Mohammad Haghparast
- Department of Radiology, Faculty of Paramedicine, Hormozgan University of Medical Sciences, Bandar Abbas, Iran
| | - Ismail Karakus
- Radiation Oncology Department, School of Medicine, Akdeniz University, Antalya, Turkey
| | - Abbas Haghparast
- Department of Medical Physics, Faculty of Medicine, Kermanshah University of Medical Sciences, Kermanshah, Iran
| | - Vahideh Nazari
- Department of Medical Physics, Hamadan University of Medical Sciences, Hamadan, Iran
| |
Collapse
|
2
|
George S, Ponmalar YR, Godson HF, Kumar AS, Ravindran BP. Influence of Jaw Setting in the Determination of Stereotactic Small-Field Output Factors with Different Detectors. J Med Phys 2022; 47:65-72. [PMID: 35548030 PMCID: PMC9084590 DOI: 10.4103/jmp.jmp_111_21] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/30/2021] [Revised: 11/14/2021] [Accepted: 11/15/2021] [Indexed: 11/08/2022] Open
Abstract
Background The experimental determination of relative output factors presents the greatest challenge, especially for small fields with different detectors. The aim of this study is to evaluate the influence of jaw positions on small-field output factors for the fields defined by micro-multileaf collimator and circular cones with different detectors. Materials and Methods The stereotactic output factors were measured on Primus linear accelerator with BrainLab micro-multileaf collimator (mMLC) and circular cones as add-on tertiary collimators. Square field sizes ranging from 0.6 cm × 0.6 cm to 9.8 cm × 9.8 cm and circular fields of diameter ranging from 1.0 cm to 4.0 cm were defined by mMLC and circular cones, respectively. The influence of jaw position on output factor was assessed for different geometric configurations with three different detectors. Results The values obtained with PinPoint ion chamber were consistent with microDiamond detector for fields greater than 24 mm × 24 mm, but an underestimation of 23.9% was noticed in 6 mm x 6 mm field size. For the mMLC defined field size of 6 mm × 6 mm, when the X-Y jaw was moved from 8 mm × 8 mm to 80 mm × 80 mm, an increase in the output by a factor of 1.7 was observed with both microDiamond and stereotactic radiosurgery diode, whereas an increase in output by a factor of 1.9 was noticed with PinPoint ion chamber. Conclusion Output factors obtained with different detectors show high differences in the smallest field size for all collimating systems. This study confirms that the position of X and Y jaw above the tertiary collimator significantly influences the small-field output factor.
Collapse
Affiliation(s)
- Seby George
- Department of Radiation Oncology, Christian Medical College, Vellore, Tamil Nadu, India
| | - Y. Retna Ponmalar
- Department of Radiation Oncology, Christian Medical College, Vellore, Tamil Nadu, India
| | - Henry Finlay Godson
- Department of Radiation Oncology, Christian Medical College, Vellore, Tamil Nadu, India
| | - A. Sathish Kumar
- Department of Radiation Oncology, Christian Medical College, Vellore, Tamil Nadu, India,Address for correspondence: Dr. Henry Finlay Godson, Department of Radiation Oncology, Christian Medical College, Vellore - 632 004, Tamil Nadu, India. E-mail:
| | - B. Paul Ravindran
- Department of Radiation Oncology, Christian Medical College, Vellore, Tamil Nadu, India
| |
Collapse
|
3
|
Gonod M, Chacon Avila C, Suarez MA, Crouzilles J, Laskri S, Vinchant JF, Aubignac L, Grosjean T. Miniaturized scintillator dosimeter for small field radiation therapy. Phys Med Biol 2021; 66. [PMID: 33971635 DOI: 10.1088/1361-6560/abffbb] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/31/2020] [Accepted: 05/10/2021] [Indexed: 11/11/2022]
Abstract
The concept of a miniaturized inorganic scintillator detector is demonstrated in the analysis of the small static photon fields used in external radiation therapy. Such a detector is constituted by a 0.25 mm diameter and 0.48 mm long inorganic scintillating cell (1.6 × 10-5cm3detection volume) efficiently coupled to a narrow 125μm diameter silica optical fiber using a tiny photonic interface (an optical antenna). The response of our miniaturized scintillator detector (MSD) under 6 MV bremsstrahlung beam of various sizes (from 1 × 1 cm2to 4 × 4 cm2) is compared to that of two high resolution reference probes, namely, a micro-diamond detector and a dedicated silicon diode. The spurious Cerenkov signal transmitted through our bare detector is rejected with a basic spectral filtering. The MSD shows a linear response regarding the dose, a repeatability within 0.1% and a radial directional dependence of 0.36% (standard deviations). Beam profiling at 5 cm depth with the MSD and the micro-diamond detector shows a mismatch in the measurement of the full widths at 80% and 50% of the maximum which does not exceed 0.25 mm. The same difference range is found between the micro-diamond detector and a silicon diode. The deviation of the percentage depth dose between the MSD and micro-diamond detector remains below 2.3% within the first fifteen centimeters of the decay region for field sizes of 1 × 1 cm2, 2 × 2 cm2and 3 × 3 cm2(0.76% between the silicon diode and the micro-diamond in the same field range). The 2D dose mapping of a 0.6 × 0.6 cm2photon field evidences the strong 3D character of the radiation-matter interaction in small photon field regime. From a beam-probe convolution theory, we predict that our probe overestimates the beam width by 0.06%, making our detector a right compromise between high resolution, compactness, flexibility and ease of use. The MSD overcomes problem of volume averaging, stem effects, and despite its water non-equivalence it is expected to minimize electron fluence perturbation due to its extreme compactness. Such a detector thus has the potential to become a valuable dose verification tool in small field radiation therapy, and by extension in Brachytherapy, FLASH-radiotherapy and microbeam radiation therapy.
Collapse
Affiliation(s)
- Mathieu Gonod
- Centre Georges François Leclerc (CGFL)-Dijon, France
| | - Carlos Chacon Avila
- FEMTO-ST Institute-Optics Department-UMR 6174-University of Bourgogne Franche-Comté-CNRS-Besançon, France
| | - Miguel Angel Suarez
- FEMTO-ST Institute-Optics Department-UMR 6174-University of Bourgogne Franche-Comté-CNRS-Besançon, France
| | - Julien Crouzilles
- SEDI-ATI Fibres Optiques, 8 Rue Jean Mermoz, F-91080 Évry-Courcouronnes, France
| | - Samir Laskri
- SEDI-ATI Fibres Optiques, 8 Rue Jean Mermoz, F-91080 Évry-Courcouronnes, France
| | | | | | - Thierry Grosjean
- FEMTO-ST Institute-Optics Department-UMR 6174-University of Bourgogne Franche-Comté-CNRS-Besançon, France
| |
Collapse
|
4
|
Msimang ZLM, Van Der Merwe D, Maphumulo N. Effect of Three-Dimensional Detector Orientation on Small-Field Output Factors. J Med Phys 2021; 45:256-260. [PMID: 33953502 PMCID: PMC8074723 DOI: 10.4103/jmp.jmp_50_20] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/10/2020] [Revised: 01/04/2021] [Accepted: 01/04/2021] [Indexed: 11/04/2022] Open
Abstract
The IAEA TRS 483 has recommended that the orientation for cylindrical ionization chambers be perpendicular to the beam for small-field output factor (OF) measurements. The recommendation was based on the unavailability of field output correction factor data for measurements using parallel orientation at the time of publication. Two three-dimensional (3D) air ionization chambers were used to perform measurements in parallel and perpendicular orientations and compared to data determined using a PTW 31018. The aim of the study was to establish whether the 3D detectors behaved as spherical or cylindrical devices. From the results, it was confirmed that the PTW 31016 and PTW 31021 detectors are suitable for OF measurements in both orientations for field sizes down to an equivalent square field of 1.8 cm and 0.96 cm, respectively, using the field output correction factor data published in the IAEA TRS 483. The preferred orientation is parallel to the beam to facilitate beam profile measurements and minimize the irradiation of the chamber stem and detector cable and decrease the volume averaging factor.
Collapse
Affiliation(s)
| | - Debbie Van Der Merwe
- Department of Physics, University of the Witwatersrand, Johannesburg, South Africa
| | | |
Collapse
|
5
|
Msimang ZLM, van der Merwe D, Maphumulo N. Repeatability of Small Field Output Factor Measurements with Various Detectors. J Med Phys 2021; 46:47-51. [PMID: 34267489 PMCID: PMC8240910 DOI: 10.4103/jmp.jmp_93_20] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/05/2020] [Revised: 01/08/2021] [Accepted: 01/24/2021] [Indexed: 11/05/2022] Open
Abstract
There are well established dosimetry reference standards for broad beams; however, there are no reference standards that can be used for both broad and small fields. The variation of the equivalent square fields and field output factors in small static photon fields when using a synthetic diamond, an electron diode, and ionization chambers (pin point, semiflex, and liquid filled) was investigated over time. Data from this study were compared to the data from other hospitals in the country and standard data sets, i.e., the British Journal of Radiology Supplement No. 25 of 1996 (BJR25) and the Radiological Physics Centre (RPC) 2012 data. The results showed that reliance on one detector and one measurement session, could yield incorrect field output factors (FOFs) for small fields. At least one of the detectors should be a solid state type with published field output correction factors and at least three measurement sessions should be performed for each FOF data point. Comparing measured data with published datasets, like RPC, will assist in verifying data. BJR25 datasets should not be used for Sclin ≤4 cm.
Collapse
Affiliation(s)
| | - Debbie van der Merwe
- Department of Physics, University of the Witwatersrand, Johannesburg, South Africa
| | | |
Collapse
|
6
|
Yani S, Budiansah I, Rhani MF, Haryanto F. Monte carlo model and output factors of elekta infinity™ 6 and 10 MV photon beam. Rep Pract Oncol Radiother 2020; 25:470-478. [PMID: 32494222 DOI: 10.1016/j.rpor.2020.03.021] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/19/2019] [Revised: 02/02/2020] [Accepted: 03/20/2020] [Indexed: 10/24/2022] Open
Abstract
Aim This study aimed to commission the Elekta Infinity™ working in 6 and 10 MV photon beam installed in Concord International Hospital, Singapore, and compare the OFs between MC simulation and measurement using PTW semiflex and microDiamond detector for small field sizes. Material and Methods There are two main steps in this study: modelling of Linac 6 and 10 MV photon beam and analysis of the output factors for field size 2 × 2-10 × 10 cm2. The EGSnrc/BEAMnrc-DOSXYZnrc code was used to model and characterize the Linac and to calculate the dose distributions in a water phantom. The dose distribution and OFs were compared to the measurement data in the same condition. Results The commissioning process was only conducted for a 10 × 10 cm2 field size. The PDD obtained from MC simulation showed a good agreement with the measurement. The local dose difference of PDDs was less than 2% for 6 and 10 MV. The initial electron energy was 5.2 and 9.4 MeV for 6 and 10 MV photon beam, respectively. This Linac model can be used for dose calculation in other situations and different field sizes because this Linac has been commissioned and validated using Monte Carlo simulation. The 10 MV Linac produces higher electron contamination than that of 6 MV. Conclusions The Linac model in this study was acceptable. The most important result in this work comes from OFs resulted from MC calculation. This value was more significant than the OFs from measurement using semiflex and microDiamond for all beam energy and field sizes because of the CPE phenomenon.
Collapse
Affiliation(s)
- Sitti Yani
- Department of Physics, Faculty of Mathematics and Natural Sciences, IPB University (Bogor Agricultural University), Babakan, Bogor, Indonesia.,Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung, Jalan Ganesa 10, Bandung, Indonesia
| | - Indra Budiansah
- Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung, Jalan Ganesa 10, Bandung, Indonesia
| | | | - Freddy Haryanto
- Department of Physics, Faculty of Mathematics and Natural Sciences, Institut Teknologi Bandung, Jalan Ganesa 10, Bandung, Indonesia
| |
Collapse
|
7
|
Evaluation of dosimetric parameters of small fields of 6 MV flattening filter free photon beam measured using various detectors against Monte Carlo simulation. JOURNAL OF RADIOTHERAPY IN PRACTICE 2020. [DOI: 10.1017/s1460396920000114] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
AbstractPurpose:This study aims to evaluate dosimetric parameters like percentage depth dose, dosimetric field size, depth of maximum dose surface dose, penumbra and output factors measured using IBA CC01 pinpoint chamber, IBA stereotactic field diode (SFD), PTW microDiamond against Monte Carlo (MC) simulation for 6 MV flattening filter-free small fields.Materials and Methods:The linear accelerator used in the study was a Varian TrueBeam® STx. All field sizes were defined by jaws. The required shift to effective point of measurement was given for CC01, SFD and microdiamond for depth dose measurements. The output factor of a given field size was taken as the ratio of meter readings normalised to 10 × 10 cm2 reference field size without applying any correction to account for changes in detector response. MC simulation was performed using PRIMO (PENELOPE-based program). The phase space files for MC simulation were adopted from the MyVarian Website.Results and Discussion:Variations were seen between the detectors and MC, especially for fields smaller than 2 × 2 cm2 where the lateral charge particle equilibrium was not satisfied. Diamond detector was seen as most suitable for all measurements above 1 × 1 cm2. SFD was seen very close to MC results except for under-response in output factor measurements. CC01 was observed to be suitable for field sizes above 2 × 2 cm2. Volume averaging effect for penumbra measurements in CC01 was observed. No detector was found suitable for surface dose measurement as surface ionisation was different from surface dose due to the effect of perturbation of fluence. Some discrepancies in measurements and MC values were observed which may suggest effects of source occlusion, shift in focal point or mismatch between real accelerator geometry and simulation geometry.Conclusion:For output factor measurement, TRS483 suggested correction factor needs to be applied to account for the difference in detector response. CC01 can be used for field sizes above 2 × 2 cm2 and microdiamond detector is suitable for above 1 × 1 cm2. Below these field sizes, perturbation corrections and volume averaging corrections need to be applied.
Collapse
|
8
|
Zhang M, Fan Q, Lei Y, Thapa B, Padula G. Assessment of an Elekta Versa HD linear accelerator for stereotactic radiosurgery with circular cone collimators. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY 2020; 28:71-82. [PMID: 31904001 DOI: 10.3233/xst-190580] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/10/2023]
Abstract
BACKGROUND Versa HD linear accelerators (linacs) are used for stereotactic radiosurgery treatment. However, the mechanical accuracy of such systems remains a concern. OBJECTIVE The purpose of this study was to evaluate the accuracy of an Elekta Versa HD linac. METHODS We performed measurements with a ball bearing phantom to calculate the rotational isocenter radii of the linac's gantry, collimator, and table, and determine the relative locations of those isocenters. We evaluated the accuracy of the cone-beam computed tomography (CBCT) guidance with a film-embedding head phantom and circular cone-collimated radiation beams. We also performed dosimetric simulations to study the effects of the linac mechanical uncertainties on non-coplanar cone arc delivery. RESULTS The mechanical uncertainty of the linac gantry rotation was 0.78 mm in radius, whereas that of the collimator and the table was <0.1 mm and 0.33 mm, respectively. The axes of rotation of the collimator and the table were coinciding with and 0.13 mm away from the gantry isocenter, respectively. Experiments with test plans demonstrated the limited dosimetric consequences on the circular arc delivery given the aforementioned mechanical uncertainties. End-to-end measurements determined that the uncertainty of the CBCT guidance was≤1 mm in each direction with respect to the reference CT image. CONCLUSIONS In arc delivery, the mechanical uncertainties associated with the gantry and the table do not require remarkable increases in geometric margins. If large enough, the residual setup errors following CBCT guidance will dominate the overall dosimetric consequence. Therefore, the Versa HD linac is a valid system for stereotactic radiosurgery using non-coplanar arc delivery.
Collapse
Affiliation(s)
- Mutian Zhang
- Radiation Therapy, Summa Health Cancer Institute, Akron, Ohio, USA
| | - Qiyong Fan
- Department of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA
| | - Yu Lei
- Department of Radiation Oncology, University of Nebraska Medical Center, Omaha, Nebraska, USA
| | - Bishnu Thapa
- Radiation Therapy, Summa Health Cancer Institute, Akron, Ohio, USA
| | - Gilbert Padula
- Radiation Therapy, Summa Health Cancer Institute, Akron, Ohio, USA
| |
Collapse
|
9
|
Casar B, Gershkevitsh E, Mendez I, Jurković S, Saiful Huq M. Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations - perpendicular and parallel. Med Phys 2020; 47:242-259. [PMID: 31677278 PMCID: PMC7003763 DOI: 10.1002/mp.13894] [Citation(s) in RCA: 18] [Impact Index Per Article: 4.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/29/2019] [Revised: 10/22/2019] [Accepted: 10/28/2019] [Indexed: 12/03/2022] Open
Abstract
PURPOSE The goal of the present work was to provide a large set of detector-specific output correction factors for seven small volume ionization chambers on two linear accelerators in four megavoltage photon beams utilizing perpendicular and parallel orientation of ionization chambers in the beam for nominal field sizes ranging from 0.5 cm2 × 0.5 cm2 to 10 cm2 × 10 cm2 . The present study is the second part of an extensive research conducted by our group. METHODS Output correction factors k Q clin , Q ref f clin , f ref were experimentally determined on two linacs, Elekta Versa HD and Varian TrueBeam for 6 and 10 MV beams with and without flattening filter for nine square fields ranging from 0.5 cm2 × 0.5 cm2 to 10 cm2 × 10 cm2 , for seven mini and micro ionization chambers, IBA CC04, IBA Razor, PTW 31016 3D PinPoint, PTW 31021 3D Semiflex, PTW 31022 3D PinPoint, PTW 31023 PinPoint, and SI Exradin A16. An Exradin W1 plastic scintillator and EBT3 radiochromic films were used as the reference detectors. RESULTS For all ionization chambers, values of output correction factors k Q clin , Q ref f clin , f ref were lower for parallel orientation compared to those obtained in the perpendicular orientation. Five ionization chambers from our study set, IBA Razor, PTW 31016 3D PinPoint, PTW 31022 3D PinPoint, PTW 31023 PinPoint, and SI Exradin A16, fulfill the requirement recommended in the TRS-483 Code of Practice, that is, 0.95 < k Q clin , Q ref f clin , f ref < 1.05 , down to the field size 0.8 cm2 × 0.8 cm2 , when they are positioned in parallel orientation; two of the ionization chambers, IBA Razor and PTW 31023 PinPoint, satisfy this condition down to the field size of 0.5 cm2 × 0.5 cm2 . CONCLUSIONS The present paper provides experimental results of detector-specific output correction factors for seven small volume ionization chambers. Output correction factors were determined in 6 and 10 MV photon beams with and without flattening filter down to the square field size of 0.5 cm2 × 0.5 cm2 for two orientations of ionization chambers - perpendicular and parallel. Our main finding is that output correction factors are smaller if they are determined in a parallel orientation compared to those obtained in a perpendicular orientation for all ionization chambers regardless of the photon beam energy, filtration, or linear accelerator being used. Based on our findings, we recommend using ionization chambers in parallel orientation, to minimize corrections in the experimental determination of field output factors. Latter holds even for field sizes below 1.0 cm2 × 1.0 cm2 , whenever necessary corrections remain within 5%, which was the case for several ionization chambers from our set. TRS-483 recommended perpendicular orientation of ionization chambers for the determination of field output factors. The present study presents results for both perpendicular and parallel orientation of ionization chambers. When validated by other researchers, the present results for parallel orientation can be considered as a complementary dataset to those given in TRS-483.
Collapse
Affiliation(s)
- Božidar Casar
- Department for Dosimetry and Quality of Radiological ProceduresInstitute of Oncology LjubljanaLjubljanaSlovenia
| | | | - Ignasi Mendez
- Department for Dosimetry and Quality of Radiological ProceduresInstitute of Oncology LjubljanaLjubljanaSlovenia
| | - Slaven Jurković
- Medical Physics DepartmentUniversity Hospital RijekaRijekaCroatia
- Department of Physics and BiophysicsFaculty of MedicineUniversity of RijekaRijekaCroatia
| | - M. Saiful Huq
- Department of Radiation OncologyUniversity of Pittsburgh School of Medicine and UPMC Hillman Cancer CenterPittsburghPAUSA
| |
Collapse
|
10
|
Smith CL, Montesari A, Oliver CP, Butler DJ. Evaluation of the IAEA-TRS 483 protocol for the dosimetry of small fields (square and stereotactic cones) using multiple detectors. J Appl Clin Med Phys 2019; 21:98-110. [PMID: 31886615 PMCID: PMC7021012 DOI: 10.1002/acm2.12792] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/23/2019] [Revised: 10/06/2019] [Accepted: 11/15/2019] [Indexed: 11/30/2022] Open
Abstract
The IAEA TRS 483 protocol1 for the dosimetry of small static fields in radiotherapy was used to calculate output factors for the Elekta Synergy linac at the Australian Radiation Protection and Nuclear Safety Agency (ARPANSA). Small field output factors for both square and circular fields were measured using nine different detectors. The “corrected” output factors (ratio of detector readings multiplied by the appropriate correction factor from the protocol) showed better consistency compared to the “uncorrected” output factors (ratio of detector readings only), with the relative standard deviation decreasing by approximately 1% after the application of the relevant correction factors. Comparisons relative to an arbitrarily chosen PTW 60019 microDiamond detector showed a reduction of maximal variation for the corrected values of approximately 3%. A full uncertainty budget was prepared to analyze the consistency of the output factors. Agreement within uncertainties between all detectors and field sizes was found, except for the 15 mm circular field. The results of this study show that the application of IAEA TRS 4831 when measuring small fields will improve the consistency of small field measurements when using multiple detectors contained within the protocol.
Collapse
Affiliation(s)
- Clare L Smith
- School of Health and Biomedical Sciences, RMIT University, Bundoora, VIC, Australia
| | | | - Christopher P Oliver
- Australian Radiation Protection and Nuclear Safety Agency, Yallambie, VIC, Australia
| | - Duncan J Butler
- Australian Radiation Protection and Nuclear Safety Agency, Yallambie, VIC, Australia
| |
Collapse
|
11
|
Evaluating small field dosimetry with the Acuros XB (AXB) and analytical anisotropic algorithm (AAA) dose calculation algorithms in the eclipse treatment planning system. JOURNAL OF RADIOTHERAPY IN PRACTICE 2019. [DOI: 10.1017/s1460396919000104] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
Abstract
AbstractBackground:An increasing number of external beam treatment modalities including intensity modulated radiation therapy, volumetric modulated arc therapy (VMAT) and stereotactic radiosurgery uses very small fields for treatment planning and delivery. However, there are major challenges in small photon field dosimetry, due to the partial occlusion of the direct photon beam source’s view from the measurement point, lack of lateral charged particle equilibrium, steep dose-rate gradient and volume averaging effect of the detector response and variation of the energy fluence in the lateral direction of the beam. Therefore, experimental measurements of dosimetric parameters such as percent depth doses (PDDs), beam profiles and relative output factors (ROFs) for small fields continue to be a challenge.Materials and Methods:In this study, we used a homogeneous water phantom and the heterogeneous anthropomorphic stereotactic end-to-end verification (STEEV) head phantom for all dose measurements and calculations. PDDs, lateral dose profiles and ROFs were calculated in the Eclipse Treatment Planning System version 13·6 using the Acuros XB (AXB) and the analytical anisotropic algorithms (AAAs) in a homogenous water phantom. Monte Carlo (MC) simulations and measurements using the Exradin W1 Scintillator were also accomplished for four photon energies: 6 MV, 6FFF, 10 MV and 10FFF. Two VMAT treatment plans were generated for two different targets: one located in the brain and the other in the neck (close to the trachea) in the head phantom (CIRS, Norfolk, VA, USA). A Varian Truebeam linear accelerator (Varian, Palo Alto, CA, USA) was used for all treatment deliveries. Calculated results with AXB and AAA were compared with MC simulations and measurements.Results:The average difference of PDDs between W1 Exradin Scintillator measurements and MC simulations, AAA and AXB algorithm calculations were 1·2, 2·4 and 3·2%, respectively, for all field sizes and energies. AXB and AAA showed differences in ROF of about 0·3 and 2·9%, respectively, compared with W1 Exradin Scintillator measured values. For the target located in the brain in the head phantom, the average dose difference between W1 Exradin Scintillator and the MC simulations, AAA and AXB were 0·2, 3·2 and 2·7%, respectively, for all field sizes. Similarly, for the target located in the neck, the respective dose differences were 3·8, 5·7 and 3·5%.Conclusion:In this study, we compared dosimetric parameters such as PDD, beam profile and ROFs in water phantom and isocenter point dose measurements in an anthropomorphic head phantom representing a patient. We observed that measurements using the W1 Exradin scintillator agreed well with MC simulations and can be used efficiently for dosimetric parameters such as PDDs and dose profiles and patient-specific quality assurance measurements for small fields. In both homogenous and heterogeneous media, the AXB algorithm dose prediction agrees well with MC and measurements and was found to be superior to the AAA algorithm.
Collapse
|
12
|
Commissioning and evaluation of a radiochromic EBT3 film dosimetry system. JOURNAL OF RADIOTHERAPY IN PRACTICE 2018. [DOI: 10.1017/s1460396918000444] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
AbstractPurposeThis work reports our study to commission a radiochromic film dosimetry system using the timely EBT3 film. We carried out dosimetric evaluations on different characteristics of photon beams (e.g., flatness, symmetry and penumbra) in radiation dose delivery.Materials and MethodsA Varian linear accelerator producing 6 and 15 MV photon beams with 120 multi-leaf collimator was used in this study. PTW ionisation chamber was used to measure the beam characteristics such as symmetry, flatness and penumbra and these measurements were used to commission the radiochormic EBT3 film dosimetry system. The results of irradiated films were analysed using the radiochromic film QA Pro software 2016.ResultsThe measured film doses were analysed at two different colour channels (green and red) using two scanning geometries (i.e., upper or lower side of film facing the scanner light source) at two dose levels (10 and 40 Gy). The difference between the ionisation chamber and film results was found insignificant and within the acceptable range as per the World Health Organisation standard.ConclusionResults of the comparison between the ionisation chamber and film measurements show that our radiochormic EBT3 film dosimetry system is reliable and cost-effective in the output measurement of a linear accelerator. Our measurements confirm that our EBT3 film dosimetry agreed well with the ionisation chamber, and can be used as a re-validation tool for linear accelerator quality control.
Collapse
|
13
|
Total scatter factor for small fields in radiotherapy: a dosimetric comparison. JOURNAL OF RADIOTHERAPY IN PRACTICE 2018. [DOI: 10.1017/s1460396917000681] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/06/2022]
Abstract
AbstractPurposeSmall field dosimetry is complicated and accuracy in the measurement of total scatter factor (TSF) is crucial for dosimetric calculations, in making optimum intensity-modulated radiotherapy plans for treating small target volumes. In this study, we intended to determine the TSF measuring properties of CC01 and CC04 detectors for field sizes ranging from sub-centimetre to the centimetre fields.Material and methodsCC01 and CC04 chamber detectors were used to measure TSF for 6 and 18 MV photon beam delivered from the linear accelerator, through small fields in a water phantom. Small fields were created by collimator jaws and multi-leaf collimators separately, with field sizes ranging from 0·6 to 10 cm2and 0·5 to 20 cm2, respectively.ResultsCC01 measured TSF at all the given field sizes created by jaws and multi-leaf collimators for both 6 and 18 MV beams whereas CC04 could not measure TSF for field sizes <1 cm2due to volume averaging and perturbation effects.ConclusionCC01 was shown to be effective for measurement of TSF in sub-centimetre field sizes. CC01 can be employed to measure other dosimetric quantities in small fields using different energy beams.
Collapse
|
14
|
Fagerstrom JM, Bender ET, Lawless MJ, Culberson WS. Design of a modulated orthovoltage stereotactic radiosurgery system. Med Phys 2017; 44:3776-3787. [PMID: 28498612 DOI: 10.1002/mp.12336] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/06/2016] [Revised: 05/01/2017] [Accepted: 05/01/2017] [Indexed: 11/11/2022] Open
Abstract
PURPOSE To achieve stereotactic radiosurgery (SRS) dose distributions with sharp gradients using orthovoltage energy fluence modulation with inverse planning optimization techniques. METHODS A pencil beam model was used to calculate dose distributions from an orthovoltage unit at 250 kVp. Kernels for the model were derived using Monte Carlo methods. A Genetic Algorithm search heuristic was used to optimize the spatial distribution of added tungsten filtration to achieve dose distributions with sharp dose gradients. Optimizations were performed for depths of 2.5, 5.0, and 7.5 cm, with cone sizes of 5, 6, 8, and 10 mm. In addition to the beam profiles, 4π isocentric irradiation geometries were modeled to examine dose at 0.07 mm depth, a representative skin depth, for the low energy beams. Profiles from 4π irradiations of a constant target volume, assuming maximally conformal coverage, were compared. Finally, dose deposition in bone compared to tissue in this energy range was examined. RESULTS Based on the results of the optimization, circularly symmetric tungsten filters were designed to modulate the orthovoltage beam across the apertures of SRS cone collimators. For each depth and cone size combination examined, the beam flatness and 80-20% and 90-10% penumbrae were calculated for both standard, open cone-collimated beams as well as for optimized, filtered beams. For all configurations tested, the modulated beam profiles had decreased penumbra widths and flatness statistics at depth. Profiles for the optimized, filtered orthovoltage beams also offered decreases in these metrics compared to measured linear accelerator cone-based SRS profiles. The dose at 0.07 mm depth in the 4π isocentric irradiation geometries was higher for the modulated beams compared to unmodulated beams; however, the modulated dose at 0.07 mm depth remained <0.025% of the central, maximum dose. The 4π profiles irradiating a constant target volume showed improved statistics for the modulated, filtered distribution compared to the standard, open cone-collimated distribution. Simulations of tissue and bone confirmed previously published results that a higher energy beam (≥ 200 keV) would be preferable, but the 250 kVp beam was chosen for this work because it is available for future measurements. CONCLUSIONS A methodology has been described that may be used to optimize the spatial distribution of added filtration material in an orthovoltage SRS beam to result in dose distributions with decreased flatness and penumbra statistics compared to standard open cones. This work provides the mathematical foundation for a novel, orthovoltage energy fluence-modulated SRS system.
Collapse
Affiliation(s)
- Jessica M Fagerstrom
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53705, USA
| | | | - Michael J Lawless
- Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53705, USA
| | - Wesley S Culberson
- Department of Medical Physics, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, WI, 53705, USA
| |
Collapse
|