Bell KJL, Li Y, Medcalf E, Ackermann D. COVID-19 rapid antigen tests approved for self-testing in Australia: published diagnostic test accuracy studies and manufacturer-supplied information. A systematic review.
Med J Aust 2023;
219:551-558. [PMID:
37903650 PMCID:
PMC10952141 DOI:
10.5694/mja2.52151]
[Citation(s) in RCA: 2] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/20/2022] [Accepted: 08/09/2023] [Indexed: 11/01/2023]
Abstract
OBJECTIVES
To review evaluations of the diagnostic accuracy of coronavirus disease 2019 (COVID-19) rapid antigen tests (RATs) approved by the Therapeutic Goods Administration (TGA) for self-testing by ambulatory people in Australia; to compare these estimates with values reported by test manufacturers.
STUDY DESIGN
Systematic review of publications in any language that reported cross-sectional, case-control, or cohort studies in which the participants were ambulatory people in the community or health care workers in hospitals in whom severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection was suspected, and the results of testing self-collected biological samples with a TGA-approved COVID-19 RAT were compared with those of reverse transcription-polymerase chain reaction (RT-PCR) testing for SARS-CoV-2. Estimates of diagnostic accuracy (sensitivity, specificity) were checked and compared with manufacturer estimates published on the TGA website.
DATA SOURCES
Publications (to 1 September 2022) identified in the Cochrane COVID-19 Study Register and the World Health Organization COVID-19 research database. Information on manufacturer diagnostic accuracy evaluations was obtained from the TGA website.
DATA SYNTHESIS
Twelve publications that reported a total of eighteen evaluations of eight RATs approved by the TGA for self-testing (manufacturers: All Test, Roche, Flowflex, MP Biomedicals, Clungene, Panbio, V-Chek, Whistling) were identified. Five studies were undertaken in the Netherlands, two each in Germany and the United States, and one each in Denmark, Belgium, and Canada; test sample collection was unsupervised in twelve studies, and supervised by health care workers or researchers in six. Estimated sensitivity with unsupervised sample collection ranged from 20.9% (MP Biomedicals) to 74.3% (Roche), and with supervised collection from 7.7% (V-Chek) to 84.4% (Panbio); the estimates were between 8.2 and 88 percentage points lower than the values reported by the manufacturers. Test specificity was high for all RATs (97.9-100%).
CONCLUSIONS
The risk of false negative results when using COVID-19 RATs for self-testing may be considerably higher than apparent in manufacturer reports on the TGA website, with implications for the reliability of these tests for ruling out infection.