Ialongo C, D'alessandro A, Sapio M, Angeloni A, Porzio O. Evaluation of the interchangeability between the new fully-automated affinity chrome-mediated immunoassay (ACMIA) and the Quantitative Microsphere System (QMS) with a CE-IVD-certified LC-MS/MS assay for therapeutic drug monitoring of everolimus after solid organ transplantation.
Clin Chem Lab Med 2023;
61:245-250. [PMID:
36330751 DOI:
10.1515/cclm-2022-0699]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/19/2022] [Accepted: 10/16/2022] [Indexed: 11/06/2022]
Abstract
OBJECTIVES
This study aims to evaluate the interchangeability between the Siemens Healthineers' "EVRO" new affinity chrome-mediated immunoassay (ACMIA/EVRO) and Thermo Fisher Scientific's "EVER" Quantitative Microsphere System (QMS/EVER) with Chromsystems' CE-IVD-certified "MassTox" liquid-chromatography/tandem-mass spectrometry (LC-MS/MS) assay for the therapeutic drug monitoring of everolimus.
METHODS
A single lot of reagent, calibrators and controls were used for each assay. A total of 67 whole blood samples (n=67) from patients receiving solid organ transplant were analyzed (n=31 with kidney transplant and n=36 with liver transplant); Passing-Bablok regression and Bland-Altman difference plot were used to evaluate bias and individual agreement; LC-MS/MS analysis was used to measure the actual concentrations of calibrators and controls compared to the assigned value.
RESULTS
ACMIA/EVRO did not show any systematic bias compared to LC-MS/MS (intercept=0.244 ng/mL, 95% CI: -0.254 to 0.651 ng/mL). Nevertheless, significant proportional bias (slope=1.511, 95% CI: 1.420 to 1.619) associated to a combined bias of 44.8% (95% CI: 41.2-48.3%) was observed. Conversely, QMS/EVER did not show any bias at both systematic (intercept=-0.151 ng/mL, 95% CI: -0.671 to 0.256 ng/mL) and proportional level (slope=0.971, 95% CI: 0.895 to 1.074) with a non-statistically significant combined bias of -3.6% (95% CI: -8.4-1.1%). Based on a concentration of calibrators and controls above the assigned value for both the analytical methods, in the ACMIA/EVRO a correction which was approximately one-third of the correction for the QMS/EVER was observed.
CONCLUSIONS
ACMIA/EVRO but not QMS/EVER shows a lack of interchangeability with the CE-IVD-certified LC-MS/MS assay. We hypothesize that, as the ACMIA/EVRO uses an anti-sirolimus antibody, the under-corrected assigned value in the assay calibrators was not sufficient to reproduce the everolimus metabolites cross-reactivity occurring in real samples.