Issaoui M, Miloro P, Balandraud X, Rivens I, Grédiac M, Blaysat B, Ouchchane L, Delabaere A, Sauvant-Rochat MP, Lemery D. Temperature Elevation in an Instrumented Phantom Insonated by B-Mode Imaging, Pulse Doppler and Shear Wave Elastography. Ultrasound Med Biol 2020; 46:3317-3326. [PMID: 32962891 DOI: 10.1016/j.ultrasmedbio.2020.08.021] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 04/02/2020] [Revised: 08/19/2020] [Accepted: 08/21/2020] [Indexed: 06/11/2023]
Abstract
Diagnostic ultrasound is the gold standard for obstetric scanning and one of the most important imaging techniques for perinatal and neonatal monitoring and diagnosis. Ultrasound provides detailed real-time anatomic information, including blood flow measurements and tissue elasticity. The latter is provided through various techniques including shear wave elastography (SWE). SWE is increasingly used in many areas of medicine, especially in detection and diagnosis of breast, thyroid and prostate cancers and liver disease. More recently, SWE has found application in gynaecology and obstetrics. This method mimics manual palpation, revealing the elastic properties of soft biological tissues. Despite its rising potential and expanding clinical interest in its use in obstetrics and gynaecology (such as for assessment of cervical ripening or organ development and structure during pregnancy), its effects on and potential risks to the developing fetus remain unknown. Risks should be evaluated by regulatory bodies before recommendations are made on the use of SWE. Because ultrasound is known to produce thermal and mechanical effects, this study measured the temperature increase caused by B-mode, pulse Doppler (PD) and SWE, using an instrumented phantom with 11 embedded thermocouples. Experiments were performed with an Aixplorer diagnostic ultrasound system (Supersonic Imagine, Aix-en-Provence, France). As expected, the greatest heating was detected by the thermocouple closest to the surface in contact with the transducer (2.9°C for SWE, 1.2°C for PD, 0.7°C for B-mode after 380-s excitation). Both conduction from the transducer face and direct heating owing to ultrasound waves contribute to temperature increase in the phantom with SWE associated with a larger temperature increase than PD and B-mode. This article offers a methodological approach and reference data for future safety studies, as well as initial recommendations about SWE safety in obstetrics and gynaecology.
Collapse
Affiliation(s)
- Maha Issaoui
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France.
| | - Piero Miloro
- Ultrasound and Underwater Acoustics, National Physical Laboratory, Teddington, UK
| | - Xavier Balandraud
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France
| | - Ian Rivens
- Department of Physics at The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, London, UK
| | - Michel Grédiac
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France
| | - Benoit Blaysat
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France
| | - Lemlih Ouchchane
- Université Clermont Auvergne, CHU Clermont-Ferrand, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France; Département de Santé Publique, Unité de Biostatistique et Informatique Médicale, CHU de Clermont-Ferrand, Clermont-Ferrand, France
| | - Amélie Delabaere
- Université Clermont Auvergne, CHU Clermont-Ferrand, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France; Pôle Femme et Enfant, CHU de Clermont-FerrandClermont-Ferrand, France
| | - Marie-Pierre Sauvant-Rochat
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France; Département de Santé Publique et Environnement, Université Clermont-Auvergne, UFR Pharmacie, Clermont-Ferrand, France
| | - Didier Lemery
- Université Clermont Auvergne, CHU Clermont-Ferrand, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France; Pôle Femme et Enfant, CHU de Clermont-FerrandClermont-Ferrand, France
| |
Collapse
|
2
|
Issaoui M, Balandraud X, Grédiac M, Blaysat B, Ouchchane L, Delabaere A, Sauvant-Rochat MP, Lemery D. Temperature Rise Caused by Shear Wave Elastography, Pulse Doppler and B-Mode in Biological Tissue: An Infrared Thermographic Approach. Ultrasound Med Biol 2020; 46:325-335. [PMID: 31735465 DOI: 10.1016/j.ultrasmedbio.2019.10.008] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/15/2019] [Revised: 10/05/2019] [Accepted: 10/11/2019] [Indexed: 06/10/2023]
Abstract
The aim of this study was to determine the interest in and relevance of the use of infrared thermography, which is a non-invasive full-field surface temperature measurement technique, to characterize the heterogeneous heating caused by ultrasound in biological tissue. Thermal effects of shear wave elastography, pulse Doppler and B-mode were evidenced in porcine tissue. Experiments were performed using a high-frequency echography Aixplorer system (Supersonic Imagine, Aix-en-Provence, France). For all three modes, ultrasound was applied continuously for 360 s while the temperature at the sample surface was recorded with a Cedip Jade III-MWIR infrared camera (Flir, Torcy, France). Temperature changes were detected for the three modes. In particular, "heat tunnels" crossing the sample were visualized from the early stages of the experiment. Heat conduction from the transducer was also involved in the global warming of the sample. The study widens the prospects for studies on tolerability, potentially in addition to classic approaches such as those using thermocouples.
Collapse
Affiliation(s)
- Maha Issaoui
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France.
| | - Xavier Balandraud
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France
| | - Michel Grédiac
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France
| | - Benoit Blaysat
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France
| | - Lemlih Ouchchane
- Université Clermont Auvergne, CHU Clermont-Ferrand, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France; Département de Santé Publique, Unité de Biostatistique et Informatique Médicale, CHU de Clermont-Ferrand, Clermont-Ferrand, France
| | - Amelie Delabaere
- Université Clermont Auvergne, CHU Clermont-Ferrand, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France; Pôle Femme et Enfant, Centre Hospitalier Universitaire de Clermont-Ferrand, Clermont-Ferrand, France
| | - Marie-Pierre Sauvant-Rochat
- Université Clermont Auvergne, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France; Département de Santé Publique et Environnement, Université Clermont-Auvergne, UFR Pharmacie, Clermont-Ferrand, France
| | - Didier Lemery
- Université Clermont Auvergne, CHU Clermont-Ferrand, CNRS, SIGMA Clermont, Institut Pascal, Clermont-Ferrand, France; Pôle Femme et Enfant, Centre Hospitalier Universitaire de Clermont-Ferrand, Clermont-Ferrand, France
| |
Collapse
|