Vossou M, Spiliopoulos S, Palialexis K, Antonakos J, Efstathopoulos EP. DIAGNOSTIC REFERENCE LEVELS AND COMPLEXITY INDICES IN INTERVENTIONAL RADIOLOGY.
Radiat Prot Dosimetry 2023;
199:ncac267-261. [PMID:
36566495 DOI:
10.1093/rpd/ncac267]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 06/07/2022] [Revised: 11/11/2022] [Accepted: 11/14/2022] [Indexed: 06/17/2023]
Abstract
The establishment of typical diagnostic reference levels (DRLs) values according to the complexity indices (CIs) for hepatic chemoembolisation (HC), iliac stent placement (ISP) and femoropopliteal revascularisation (FR) is reported in this study. To estimate patients' stochastic effects, effective dose was calculated through dose area product (DAP) values of this study and E/DAP conversion factors derived from the literature. Data for DAP, Reference Air Kerma (Ka,r) and fluoroscopy time (FT) were collected for 218 patients and CIs were assigned to each procedure to extract DRLs. To estimate effective dose, conversion factors and DAP values were used for seven IR procedures. DRL values for DAP were 141, 130 and 28 Gy*cm2 for HC, ISP, and FR, respectively. The corresponding DRL values for Ka,r were 634.6, 300.1 and 112.0 mGy, and for FT were 15.3, 12.4 and 17.9 min, respectively. CIs in interventional radiology are a useful tool for the optimisation of DRLs since they contribute to patient's doses.
Collapse