Shirasaka T, Kojima T, Yamane S, Mikayama R, Kawakubo M, Funatsu R, Kato T, Ishigami K, Funama Y. Effect of iodine concentration and body size on iodine subtraction in virtual non-contrast imaging: A phantom study. Radiography (Lond) 2023; 29:557-563. [PMID: 36965243 DOI: 10.1016/j.radi.2023.03.003] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/23/2022] [Revised: 02/19/2023] [Accepted: 03/05/2023] [Indexed: 03/27/2023]
Abstract
INTRODUCTION Dual-energy computed tomography (DECT) can generate virtual non-contrast (VNC) images. Herein, we sought to improve the accuracy of VNC images by identifying the optimal slope of contrast media (SCM) for VNC-image generation based on the iodine concentration and subject's body size. METHODS We used DECT to scan a multi-energy phantom including four iodine concentration rods (15, 10, 5, and 2 mg/mL), and 240 VNC images (eight SCM ranging from 0.49 to 0.56 × three body sizes × ten scans) that were generated by three-material decomposition. The CT number of each iodine and solid water rod part was measured in each VNC image. The difference in the CT number between the iodine and the solid water rod part was calculated and compared using paired t-test or repeated measures ANOVA. RESULTS The SCM that achieved an absolute value of the difference in CT number of <5.0 Hounsfield units (HU) for all body sizes simultaneously was greater at lower iodine concentration (SCM of 0.5, 0.51, and 0.53 at 10, 5, and 2 mg/mL iodine, respectively). At an iodine concentration of 15 mg/mL, no SCM achieved an absolute difference of <5.0 HU in CT number for all body sizes simultaneously. At all iodine concentrations, the SCM achieving the minimal difference in the CT number increased with the increase in body size. CONCLUSION By adjusting the SCM according to the iodine concentration and body size, it is possible to generate VNC images with an accuracy of <5.0 HU. IMPLICATIONS FOR PRACTICE Improving the accuracy of VNC images minimizing incomplete iodine subtraction would make it possible to replace true non-contrast (TNC) images with VNC images and reduce the radiation dose.
Collapse
Affiliation(s)
- T Shirasaka
- Graduate School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Kumamoto, 862-0976, Japan; Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan.
| | - T Kojima
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan; Department of Health Sciences, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan.
| | - S Yamane
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan.
| | - R Mikayama
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan.
| | - M Kawakubo
- Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan.
| | - R Funatsu
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan.
| | - T Kato
- Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan.
| | - K Ishigami
- Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi Ward, Fukuoka, 812-8582, Japan.
| | - Y Funama
- Department of Medical Radiation Sciences, Faculty of Life Sciences, Kumamoto University, 4-24-1 Kuhonji, Kumamoto, 862-0976, Japan.
| |
Collapse
|