Manu S, Suntharos P, Boyle GJ, Wang L, Prieto LR. Radiation Reduction in the Pediatric Catheterization Laboratory Using a Novel Imaging System.
J Invasive Cardiol 2018;
30:28-33. [PMID:
29035845]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Subscribe] [Scholar Register] [Indexed: 06/07/2023]
Abstract
OBJECTIVES
Radiation dose was compared between two modern imaging systems with different x-ray tube technology (Megalix vs Gigalix) and detector type (amorphous vs crystalline silicon) at the same institution.
BACKGROUND
Further reduction in radiation dose than currently reported may be achievable with advances in x-ray tube and detector technology.
METHODS
Radiation dose (air kerma, dose-area product [DAP]) was retrospectively compared for post-transplant pediatric patients undergoing right heart catheterization/biopsy (fluoroscopy only) or "annual" catheterization with coronary angiography in one of two imaging systems between January 2014 and December 2016. Comparisons were also made with published radiation doses.
RESULTS
A total of 122 right heart catheterizations with biopsy were performed in the Megalix/amorphous silicon (Si) lab and 168 in the Gigalix/crystalline Si lab. Age and weight were not statistically different for the two groups. There was a 50% decrease in median air kerma (2.2 mGy vs 1.1 mGy; P<.001) and 66% decrease in median DAP (52.2 μGy•m² vs 18.0 μGy•m²; P<.001) for the Gigalix/crystalline Si lab. A total of 24 "annual" catheterizations were performed in the Megalix/amorphous Si lab and 22 were performed in the Gigalix/crystalline Si lab. There was a 57% reduction in median air kerma (458.6 mGy vs 198.6 mGy; P<.001) and a 46% reduction in median DAP (2548.0 μGy•m² vs 1367.1 μGy•m²; P<.01) for the Gigalix/crystalline Si lab. Similar reductions were found on comparison with published doses.
CONCLUSION
The Gigalix tube and crystalline Si detector decrease radiation dose by 50%-60% for fluoroscopy and cine acquisition in pediatric patients.