Perez MÂF, Daniel D, Padula M, do Lago CL, Bottoli CBG. Determination of primary aromatic amines from cooking utensils by capillary electrophoresis-tandem mass spectrometry.
Food Chem 2021;
362:129902. [PMID:
34175690 DOI:
10.1016/j.foodchem.2021.129902]
[Citation(s) in RCA: 8] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/25/2020] [Revised: 03/02/2021] [Accepted: 03/28/2021] [Indexed: 11/18/2022]
Abstract
This paper describes a fast, sensitive, environment-friendly method for the determination of 19 primary aromatic amines (PAAs) in cooking utensils by capillary zone electrophoresis coupled with tandem mass spectrometry. The best electrophoretic separation of PAAs was obtained in 0.1 mol l-1 formic acid (pH 2.4) as the background electrolyte, fused silica capillary (67 cm) with a run time below 6 min. The proposed method presented a linear calibration with correlation coefficients higher than 0.99 and reproducibility in a range of 1-25%. Limits of detection were in the range of 0.2-1.3 μg kg-1 and recoveries were in a range of 85-120% for all the PAAs. The validated method was employed to determine PAAs on 36 samples of cooking utensils using acetic simulant. The results showed that 4,4'-diaminodiphenylmethane and aniline being the most frequently found PAAs in these samples and 28% of cooking utensils were not compliant.