Molinero-Mourelle P, Roccuzzo A, Yilmaz B, Lam WYH, Pow EHN, Del Río Highsmith J, Gómez-Polo M. Microleakage assessment of CAD-CAM Cobalt-Chrome and Zirconia abutments on a conical connection dental implant: A comparative in vitro study.
Clin Oral Implants Res 2022;
33:945-952. [PMID:
35818785 PMCID:
PMC9544167 DOI:
10.1111/clr.13973]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/28/2021] [Revised: 06/19/2022] [Accepted: 06/22/2022] [Indexed: 11/28/2022]
Abstract
Objective
To assess the marginal and bacterial microleakage in zirconia and CAD‐CAM or cast Co‐Cr implant abutments.
Methods
Sixty‐four conical connection implants with their respective abutments were divided into four groups (Co‐Cr (milled, laser‐sintered, and cast) and Zirconia (milled)). All specimens were subjected to a chewing simulation and thermocycling. After aging process, specimens were submerged in a 0.2% methylene blue solution with Porphyromonas gingivalis (P.g) for 48 h. The marginal microleakage was measured using a 40× optical microscopy at the internal part of the implant, and when positive microleakage was observed, a DNA isolation with a polymerase chain reaction (PCR) test was used. The microbiological assessment was based on colony forming units (CFUs).
Results
Thirty (47%) implant‐abutments presented microleakage and the PCR was performed on those specimens (1 Zirconia, 1 Co‐Cr milled, 14 Co‐Cr laser‐sintered and 14 cast). Seven specimens (1 Co‐Cr laser‐sintered and six cast) presented values below the PCR detection limit (< 100 CFUs). The lowest CFUs count occurred in the Co‐Cr milled group (5.17E+02 CFUs/ml) followed by zirconia (7.70E+03 CFUs/ml). The Co‐Cr cast (9.39E+03 CFUs/ml) and laser‐sintered (2.4E+05 CFUs/ml) groups had higher bacterial count. The CFU count comparison performed between Co‐Cr cast and laser‐sintered resulted in a statistically significant differences in favor of Co‐CrCL (p < .05).
Conclusions
The abutment material and fabrication technique affected the implant‐abutment microleakage. Although the CAD‐CAM abutments presented favorable results, all tested groups presented microleakage.