Yeung K, Eiberg JP, Collet-Billon A, Sandholt BV, Jessen ML, Sillesen HH, Eldrup N. 3-D Contrast-Enhanced Fusion Ultrasound for Accurate Volume Assessment of Vessel Lumen and Plaque in Carotid Artery Disease as Compared With Computed Tomography Angiography.
Ultrasound Med Biol 2024;
50:399-406. [PMID:
38171954 DOI:
10.1016/j.ultrasmedbio.2023.11.013]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 07/05/2023] [Revised: 11/20/2023] [Accepted: 11/28/2023] [Indexed: 01/05/2024]
Abstract
OBJECTIVE
Three-dimensional contrast-enhanced fusion ultrasound (CEFUS) of atherosclerotic carotid arteries provides spatial visualization of the vessel lumen, creating a lumenography. As in 3-D computed tomography angiography (CTA), 3-D CEFUS outlines the contrast-filled lumen. Plaque and vessel contours are distinguished in 3-D CEFUS, allowing plaque volume quantification as a valid estimate of carotid plaque burden. Three-dimensional CEFUS is unproven in intermodality studies, vindicating the assessment of 3-D CEFUS applicability and comparing 3-D CEFUS and 3-D CTA lumenography as a proof-of-concept study.
METHODS
Using an ultrasound system with magnetic tracking, a linear array transducer and SonoVue contrast agent, 3-D CEFUS acquisitions were generated by spatial stitching of serial 2-D images. From 3-D CEFUS and 3-D CTA imaging, the atherosclerotic carotid arteries were reconstructed with lumenography in an offline software program for lumen and plaque volume quantification. Bland-Altman analysis was used for inter-image modality agreement.
RESULTS
The study included 39 carotid arteries. Mean lumen and plaque volume in 3-D CEFUS were 0.63 cm3 (standard deviation [SD]: 0.26) and 0.62 cm3 (SD: 0.26), respectively. Lumen volume differences between 3-D CEFUS and 3-D CTA were non-significant, with a mean difference of 0.01 cm3 (SD: 0.02, p = 0.26) and limits of agreement (LoA) range of ±0.11 cm3. Mean plaque volume difference was -0.12 cm3 (SD: 0.19, p = 0.006) with a LoA range of ±0.39 cm3.
CONCLUSION
There was strong agreement in lumenography between 3-D CEFUS and 3-D CTA. The interimage modality difference in plaque volumes was substantial because of challenging vessel wall definition in 3-D CTA. Three-dimensional CEFUS is viable in quantifying carotid plaque volume burden and can potentially monitor plaque development over time.