Canham CD, Muradov PI, Simpson JB, Incavo SJ. Corrosion and adverse local tissue reaction after total hip arthroplasty with a
modular titanium alloy femoral neck.
Arthroplast Today 2017;
3:211-214. [PMID:
29204482 PMCID:
PMC5712034 DOI:
10.1016/j.artd.2017.03.003]
[Citation(s) in RCA: 12] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/27/2017] [Revised: 03/07/2017] [Accepted: 03/08/2017] [Indexed: 02/08/2023] Open
Abstract
This report describes a case of mechanically assisted crevice corrosion and secondary adverse local tissue reaction in a patient following a total hip arthroplasty, utilizing a modular neck (bi-modular) femoral component. Radiographic evaluation demonstrated a well-positioned, stable, cementless arthroplasty. Upon further evaluation, the patient had elevated serum cobalt and chromium levels, and magnetic resonance imaging demonstrated a periprosthetic pseudotumor. Corrosion of both the neck-stem and head-neck junctions was suspected. At the time of surgery, the neck-body junction was pristine; however, the head-neck junction of the implant demonstrated severe corrosive wear, a problem that has been reported only once previously with this particular bi-modular implant. This serves as a reminder that any modular junction may be susceptible to corrosion and not all bi-modular designs behave similarly.