Zhang Y, Yan S, Cui Z, Wang Y, Li Z, Yin Y, Li B, Quan H, Zhu J. Out-of-field dose assessment for a 1.5 T MR-Linac with optically stimulated luminescence dosimeters. Med Phys 2021; 48:4027-4037. [PMID: 33714229 PMCID: PMC8360091 DOI: 10.1002/mp.14839] [Citation(s) in RCA: 6] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/18/2020] [Revised: 02/04/2021] [Accepted: 03/04/2021] [Indexed: 11/06/2022] Open
Abstract
PURPOSE To assess the out-of-field surface and internal dose of the 1.5 T MR-Linac compared to the conventional external beam linac using optically stimulated luminescence dosimeters (OSLDs), and evaluate the out-of-field dose calculation accuracy of the Monaco treatment planning system (TPS) of the 1.5T MR-Linac. METHODS A cubic solid water phantom, with OSLDs on the surface, was vertically irradiated by MR-Linac square fields with different sizes. In addition, OSLDs were arranged out of the beam edges in four directions. An anthropomorphic adult phantom, with 125 cm3 simulated volume, was irradiated in four orthogonal directions by both MR-Linac and conventional linac at the head, thoracic, and pelvic sites. Out-of-field doses were measured by OSLDs on both the surface and internal emulational organs at risk (OARs). The results were compared to the simulated dose from Monaco TPS. RESULTS At different field sizes (5 × 5 to 20 × 20 cm2 ) and distances (1 to 10 cm) to beam edge, the out-of-field surface dose measured on MR-Linac varied from 0.16 % (10 cm to 5 × 5 cm2 edge) to 7.02 % (1 cm to 20 × 20 cm2 edge) of the maximum dose laterally and from 0.14 % (10 cm to 5 × 5 cm2 edge) to 8.56 % (1 cm to 20 × 20 cm2 edge) of the maximum dose longitudinally. Compared to the OSLDs measured data, the Monaco TPS presented an overestimate of the out-of-field dose of OARs at 0-2 % isodose area on both surface and internal check points, and the overestimation gets greater as the distance increases. The underestimation was found to be 0-35% at 2-5% isodose area on both surface and internal check points. Compared to the conventional linac, MR-Linac delivered higher average values of out-of-field dose on surface check points (20%, 19%, 21%) and internal simulated OARs (42%, 37%, 9%) of the anthropomorphic phantom at head, thoracic, and pelvic irradiations, respectively. CONCLUSIONS Compared to the conventional linac, MR-Linac has the same out-of-field dose distribution. However, considering the absolute dose values, MR-Linac delivered relatively higher out-of-field doses on both surface and internal OARs. Additional radiation shielding to patients undergoing MR-Linac may provide protection from out-of-field exposure.
Collapse
Affiliation(s)
- Yan Zhang
- School of Physics and Technology, Wuhan University, Wuhan, P.R. China.,Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, P.R. China
| | - Shaojie Yan
- Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, P.R. China.,School of Nuclear Science and Technology, University of South China, Hengyang, P.R. China
| | - Zhen Cui
- Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, P.R. China
| | - Yungang Wang
- Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, P.R. China
| | - Zhenjiang Li
- Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, P.R. China
| | - Yong Yin
- Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, P.R. China
| | - Baosheng Li
- Department of Radiation Oncology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, P.R. China
| | - Hong Quan
- School of Physics and Technology, Wuhan University, Wuhan, P.R. China
| | - Jian Zhu
- Department of Radiation Oncology Physics and Technology, Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences, Jinan, P.R. China.,Shandong Medical Imaging and Radiotherapy Engineering Center, Jinan, P.R. China.,Shandong Key Laboratory of Digital Medicine and Computer Assisted Surgery, The Affiliated Hospital of Qingdao University, Qingdao, P.R. China
| |
Collapse
|