Damyanovich AZ, Tadic T, Foltz WD, Jelveh S, Bissonnette JP. Time-course assessment of 3D-image distortion on the 1.5 T Marlin/Elekta Unity MR-LINAC. Phys Med 2022; 100:90-98. [PMID: 35777256 DOI: 10.1016/j.ejmp.2022.05.009] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/05/2022] [Revised: 04/04/2022] [Accepted: 05/25/2022] [Indexed: 11/29/2022] Open
Abstract
PURPOSE The efficacy of MR-guided radiotherapy on a MR-LINAC (MR-L) is dependent on the geometric accuracy of its MR images over clinically relevant Fields-of-View (FOVs). Our objectives were to: evaluate gradient non-linearity (GNL) on the Elekta Unity MR-L across time via 76 weekly measurements of 3D-distortion over concentrically larger diameter spherical volumes (DSVs); quantify distortion measurement error; and assess the temporal stability of spatial distortion using statistical process control (SPC). METHODS MR-image distortion was assessed using a large-FOV 3D-phantom containing 1932 markers embedded in seven parallel plates, spaced 25 mm × 25 mm in- and 55 mm through-plane. Automatically analyzed T1 images yielded distortions in 200, 300, 400 and 500 mm concentric DSVs. Distortion measurement error was evaluated using median absolute difference analysis of imaging repeatability tests. RESULTS Over the measurement period absolute time-averaged distortion varied between: dr = 0.30 - 0.49 mm, 0.53 - 0.80 mm, 1.0 - 1.4 mm and 2.28 - 2.37 mm, for DSVs 200, 300, 400 and 500 mm at the 98th percentile level. Repeatability tests showed that imaging/repositioning introduces negligible error: mean ≤ 0.02 mm (max ≤ 0.3 mm). SPC analysis showed image distortion was stable across all DSVs; however, noticeable changes in GNL were observed following servicing at the one-year mark. CONCLUSIONS Image distortion on the MR-L is in the sub-millimeter range for DSVs ≤ 300 mm and stable across time, with SPC analysis indicating all measurements remain within control for each DSV.
Collapse
Affiliation(s)
- Andrei Z Damyanovich
- Department of Medical Physics, Princess Margaret Cancer Center, University Health Network, Toronto, Ontario, Canada; Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada; Techna Institute, Toronto, Ontario, Canada.
| | - Tony Tadic
- Department of Medical Physics, Princess Margaret Cancer Center, University Health Network, Toronto, Ontario, Canada; Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada; Techna Institute, Toronto, Ontario, Canada
| | - Warren D Foltz
- Department of Medical Physics, Princess Margaret Cancer Center, University Health Network, Toronto, Ontario, Canada; Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada; Techna Institute, Toronto, Ontario, Canada
| | - Salomeh Jelveh
- Department of Medical Physics, Princess Margaret Cancer Center, University Health Network, Toronto, Ontario, Canada
| | - Jean-Pierre Bissonnette
- Department of Medical Physics, Princess Margaret Cancer Center, University Health Network, Toronto, Ontario, Canada; Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada; Techna Institute, Toronto, Ontario, Canada; Department of Medical Biophysics, University of Toronto, Toronto, Ontario, Canada
| |
Collapse
|