Malmberg MA, Odéen H, Hofstetter LW, Hadley JR, Parker DL. Validation of single reference variable flip angle (SR-VFA) dynamic T
1 mapping with T
2 * correction using a novel rotating phantom.
Magn Reson Med 2024;
91:1419-1433. [PMID:
38115639 PMCID:
PMC10872756 DOI:
10.1002/mrm.29944]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/27/2023] [Revised: 10/12/2023] [Accepted: 11/09/2023] [Indexed: 12/21/2023]
Abstract
PURPOSE
To validate single reference variable flip angle (SR-VFA) dynamic T1 mapping with and without T2 * correction against inversion recovery (IR) T1 measurements.
METHODS
A custom cylindrical phantom with three concentric compartments was filled with variably doped agar to produce a smooth spatial gradient of the T1 relaxation rate as a function of angle across each compartment. IR T1 , VFA T1 , and B1 + measurements were made on the phantom before rotation, and multi-echo stack-of-radial dynamic images were acquired during rotation via an MRI-compatible motor. B1 + -corrected SR-VFA and SR-VFA-T2 * T1 maps were computed from the sliding window reconstructed images and compared against rotationally registered IR and VFA T1 maps to determine the percentage error.
RESULTS
Both VFA and SR-VFA-T2 * T1 maps fell within 10% of IR T1 measurements for a low rotational speed, with a mean accuracy of 2.3% ± 2.6% and 2.8% ± 2.6%, respectively. Increasing rotational speed was found to decrease the accuracy due to increasing temporal smoothing over ranges where the T1 change had a nonconstant slope. SR-VFA T1 mapping was found to have similar accuracy as the SR-VFA-T2 * and VFA methods at low TEs (˜<2 ms), whereas accuracy degraded strongly with later TEs. T2 * correction of the SR-VFA T1 maps was found to consistently improve accuracy and precision, especially at later TEs.
CONCLUSION
SR-VFA-T2 * dynamic T1 mapping was found to be accurate against reference IR T1 measurements within 10% in an agar phantom. Further validation is needed in mixed fat-water phantoms and in vivo.
Collapse