Jefferies JL, Spar DS, Chaouki AS, Khoury PR, Casson P, Czosek RJ. Continuous Arrhythmia Monitoring in Pediatric and Adult Patients With Left Ventricular Noncompaction.
Tex Heart Inst J 2022;
49:479861. [PMID:
35395088 DOI:
10.14503/thij-20-7497]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/23/2022]
Abstract
Patients with left ventricular noncompaction (LVNC) are at risk of clinically significant arrhythmias and sudden death. We evaluated whether implantable loop recorders could detect significant arrhythmias that might be missed in these patients during annual Holter monitoring. Selected pediatric and adult patients with LVNC who consented to implantable loop recorder placement were monitored for 3 years (study duration, 10 April 2014-9 December 2019). Fourteen subjects were included (age range, 6.5-36.4 yr; 8 males). Of 13 patients who remained after one device extrusion, one underwent implantable cardioverter-defibrillator placement. Four patients (31%) had significant arrhythmias: atrial tachycardia (n=2), nonsustained ventricular tachycardia (n=1), and atrial fibrillation (n=1). All 4 events were clinically asymptomatic and not associated with left ventricular ejection fraction. In addition, a high frequency of benign arrhythmic patterns was detected. Implantable loop recorders enable continuous, long-term detection of important subclinical arrhythmias in selected patients who have LVNC. These devices may prove to be most valuable in patients who have LVNC and moderate or greater ventricular dysfunction.
Collapse