Gomez Bustamante T, Mercado Montoya M, Berjano E, González-Suárez A, Kulstad E. Proactive esophageal cooling during laser cardiac ablation: A computer modeling study. Lasers Surg Med 2024; 56:392-403. [PMID: 38436122 DOI: 10.1002/lsm.23774] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/28/2023] [Revised: 02/09/2024] [Accepted: 02/17/2024] [Indexed: 03/05/2024]
Abstract
BACKGROUND AND OBJECTIVES Laser ablation is increasingly used to treat atrial fibrillation (AF). However, atrioesophageal injury remains a potentially serious complication. While proactive esophageal cooling (PEC) reduces esophageal injury during radiofrequency ablation, the effects of PEC during laser ablation have not previously been determined. We aimed to evaluate the protective effects of PEC during laser ablation of AF by means of a theoretical study based on computer modeling. METHODS Three-dimensional mathematical models were built for 20 different cases including a fragment of atrial wall (myocardium), epicardial fat (adipose tissue), connective tissue, and esophageal wall. The esophagus was considered with and without PEC. Laser-tissue interaction was modeled using Beer-Lambert's law, Pennes' Bioheat equation was used to compute the resultant heating, and the Arrhenius equation was used to estimate the fraction of tissue damage (FOD), assuming a threshold of 63% to assess induced necrosis. We modeled laser irradiation power of 8.5 W over 20 s. Thermal simulations extended up to 250 s to account for thermal latency. RESULTS PEC significantly altered the temperature distribution around the cooling device, resulting in lower temperatures (around 22°C less in the esophagus and 9°C in the atrial wall) compared to the case without PEC. This thermal reduction translated into the absence of transmural lesions in the esophagus. The esophagus was thermally damaged only in the cases without PEC and with a distance equal to or shorter than 3.5 mm between the esophagus and endocardium (inner boundary of the atrial wall). Furthermore, PEC demonstrated minimal impact on the lesion created across the atrial wall, either in terms of maximum temperature or FOD. CONCLUSIONS PEC reduces the potential for esophageal injury without degrading the intended cardiac lesions for a variety of different tissue thicknesses. Thermal latency may influence lesion formation during laser ablation and may play a part in any collateral damage.
Collapse
Affiliation(s)
| | | | - Enrique Berjano
- Department of Electronic Engineering, BioMIT, Universitat Politècnica de València, Spain
| | - Ana González-Suárez
- Translational Medical Device Lab, School of Medicine, Lambe Institute for Translational Research, University of Galway, Ireland
- Valencian International University, Valencia, Spain
| | - Erik Kulstad
- University of Texas Southwestern Medical Center, Dallas, Texas, USA
| |
Collapse
|
2
|
Yamasaki T, Hattori T, Pak M, Kakita K. Left atrial roof region ablation using a visually guided laser balloon. Pacing Clin Electrophysiol 2024; 47:429-432. [PMID: 37221909 DOI: 10.1111/pace.14724] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 01/21/2023] [Revised: 05/06/2023] [Accepted: 05/08/2023] [Indexed: 05/25/2023]
Abstract
Balloon-based catheter ablation including visually guided laser balloon (VGLB) has been adopted a first line therapeutic strategy for the patients with atrial fibrillation (AF). Recently, the roof area ablation beyond pulmonary vein (PV) isolation (PVI) using cryoballoon has been described as an effective therapy for the patients with persistent AF. However, the roof area ablation performed with a VGLB remains unknown. In this case, we report the case of roof area ablation for the patient with persistent AF using a VGLB.
Collapse
Affiliation(s)
| | | | - Misen Pak
- Arrhythmia Care Center, Koseikai Takeda Hospital, Kyoto, Japan
| | - Ken Kakita
- Arrhythmia Care Center, Koseikai Takeda Hospital, Kyoto, Japan
| |
Collapse
|
3
|
Sasaki S, Shibuya Y, Minamiguchi H, Sakio T, Hamanaka Y, Kanda T, Ichibori Y, Makino N, Hirayama A, Higuchi Y. Endoscopic view of the vertebra during laser balloon ablation of paroxysmal atrial fibrillation. Clin Case Rep 2024; 12:e8463. [PMID: 38314190 PMCID: PMC10834377 DOI: 10.1002/ccr3.8463] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/02/2023] [Revised: 11/04/2023] [Accepted: 01/04/2024] [Indexed: 02/06/2024] Open
Abstract
Key Clinical Message Left atrial posterior wall on the vertebra is often difficult to obtain stable tissue contact with ablation-catheter. Laser balloon ablation is effective because the compression from the vertebra can be visualized through endoscopy. Abstract When performing pulmonary vein isolation (PVI) with radiofrequency, left atrial posterior wall on the vertebra is often difficult to obtain stable tissue contact with ablation-catheter because of the movement of the ablation point. Laser balloon ablation is effective for the achievement of durable PVI in cases with such anatomical characteristics because the compression from the vertebra can be visualized through endoscopy.
Collapse
Affiliation(s)
- Shun Sasaki
- Cardiovascular Division Osaka Police Hospital Osaka Japan
| | - Yuki Shibuya
- Cardiovascular Division Osaka Police Hospital Osaka Japan
| | | | | | - Yuma Hamanaka
- Cardiovascular Division Osaka Police Hospital Osaka Japan
| | - Takashi Kanda
- Cardiovascular Division Osaka Police Hospital Osaka Japan
| | | | | | | | | |
Collapse
|
4
|
Sun X, Zhao S, Yu S, Cui K. Cryoballoon vs. laser balloon ablation for atrial fibrillation: a meta-analysis. Front Cardiovasc Med 2023; 10:1278635. [PMID: 38169911 PMCID: PMC10761002 DOI: 10.3389/fcvm.2023.1278635] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/16/2023] [Accepted: 11/20/2023] [Indexed: 01/05/2024] Open
Abstract
Background Cryoballoon ablation (CBA) and laser balloon ablation (LBA) are two innovative ways for the treatment of atrial fibrillation (AF). This study aimed to evaluate the efficacy and safety of cryoballoon ablation and laser balloon ablation in patients with AF. Methods We searched Pubmed, Embase, Ovid, Web of Science and other databases for comparative trials comparing CB and LB ablation in the treatment of AF, from establishment of database to August, 2023. Results A total of 13 studies and 3,582 patients were included (CBA, n = 2,308; LBA, n = 1,274). There was no difference between CBA and LBA in acute PVI rate per vein, 12-months recurrence rate of AF, 12-months recurrence rate of atrial arrhythmia, occurrence rate of pericardial tamponade, occurrence rate of inguinal complications. LBA presented a lower acute PVI rate per patients (CBA 97.0% vs. LBA 93.4%, RR = 1.04, 95%CI: 1.01-1.07). Transient nerve palsy was more likely to occur after CBA (CBA 2.7% vs. LBA 0.7%, RR = 4.25, 95%CI: 2.06-8.76). However, the occurrence of persistent nerve palsy between CBA and LBA groups were similar (CB 1.4% vs. LB 1.0%, RR = 1.09, 95%CI: 0.55-2.14). In terms of procedural duration, the procedural time of CBA was shorter than that of LBA (WMD = -26.58, 95%CI: -36.71-16.46). Conclusions Compared with LBA, CBA had a shorter procedural duration. There was a higher incidence of transient but not persistent phrenic nerve palsy after CBA. Systematic Review Registration https://www.crd.york.ac.uk/prospero/display_record.php?RecordID=272607 Identifier (CRD42021272607).
Collapse
Affiliation(s)
- Xiaochi Sun
- Department of Cardiology, West China Hospital, Sichuan University, Chengdu, Sichuan, China
| | - Shenyu Zhao
- Department of Cardiology, West China Hospital, Sichuan University, Chengdu, Sichuan, China
| | - Simin Yu
- West China Medical School, Sichuan University, Chengdu, Sichuan, China
| | - Kaijun Cui
- Department of Cardiology, West China Hospital, Sichuan University, Chengdu, Sichuan, China
| |
Collapse
|
5
|
Sano M, Naruse Y, Kaneko Y, Sakakibara T, Narumi T, Urushida T, Maekawa Y. Feasibility of energy-guided short duration protocol of laser balloon based pulmonary vein isolation for atrial fibrillation (EG-Laser Study). Pacing Clin Electrophysiol 2023. [PMID: 37334754 DOI: 10.1111/pace.14760] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 07/30/2022] [Revised: 05/30/2023] [Accepted: 06/03/2023] [Indexed: 06/20/2023]
Abstract
BACKGROUND Laser balloon-based pulmonary vein isolation (LB-PVI) is available for atrial fibrillation (AF) ablation. The lesion size depends on laser energy; however, the default protocol is not an energy-based setting. We hypothesized that an energy-guided (EG) short-duration protocol may be an alternative to shorten the procedure time without affecting efficacy and safety. METHODS We evaluated the efficacy and safety of the EG short-duration protocol (EG group) (target energy 120 J/site [12W/10s; 10W/12s; 8.5W/14s; 5.5W/22s]) compared with the default protocol (control group) (12W/20s; 10W/20s; 8.5W/20s; 5.5W/30s). RESULTS A total of 52 consecutive patients (EG: n = 27 [103veins] and control: n = 25 [91veins]) undergoing LB-PVI (64 ± 10 years, 81% male, 77% paroxysmal) were enrolled. The EG group had a shorter total time in the pulmonary vein (PV) (43.0 ± 13.9 min vs. 61.1 ± 16.0 min, p < .0001), a shorter total laser application time (1348 ± 254 sec vs. 2032 ± 424 sec, p < .0001), and lower total laser energy (12455 ± 2284J vs. 18084 ± 3746J, p < .0001). There was no difference in the total number of laser applications (p = 0269) or first-pass isolation (p = .725). Acute reconduction was identified only in one vein in the EG. No significant differences were observed in the incidence of pinhole rupture (7.4% vs. 4%, p = 1.000) or phrenic nerve palsy (3.7% vs. 12%, p = .341). During a mean follow-up of 13.5 ± 6.1 months, Kaplan-Meier analysis revealed no significant difference in atrial tachyarrhythmia recurrence (p = .227). CONCLUSION LB-PVI with the EG short-duration protocol may be achieved in a shorter procedure time to avoid deterioration of efficacy and safety. The EG protocol is feasible as a novel point-by-point manual laser-application approach.
Collapse
Affiliation(s)
- Makoto Sano
- Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan
| | - Yoshihisa Naruse
- Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan
| | - Yutaro Kaneko
- Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan
| | | | - Taro Narumi
- Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan
| | - Tsuyoshi Urushida
- Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan
| | - Yuichiro Maekawa
- Division of Cardiology, Internal Medicine III, Hamamatsu University School of Medicine, Hamamatsu, Japan
| |
Collapse
|
6
|
Schildt S, Fredersdorf S, Jungbauer CG, Hauck C, Tarnowski D, Debl K, Neef S, Schach C, Sossalla S, Maier LS, Üçer E. A Pilot Trial to Compare the Long-Term Efficacy of Pulmonary Vein Isolation with High-Power Short-Duration Radiofrequency Versus Laser Energy with Rapid Ablation Mode. J Cardiovasc Dev Dis 2023; 10:98. [PMID: 36975862 PMCID: PMC10057338 DOI: 10.3390/jcdd10030098] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/09/2023] [Revised: 02/20/2023] [Accepted: 02/21/2023] [Indexed: 02/25/2023] Open
Abstract
BACKGROUND Pulmonary vein (PV) reconnection is the major cause of atrial fibrillation (AF) recurrence after pulmonary vein isolation (PVI). The probability of reconnection is higher if the primary lesion is not sufficiently effective, which can be unmasked with an adenosine provocation test (APT). High-power short-duration radiofrequency energy (HPSD) guided with ablation index (AI) and the third generation of the visually guided laser balloon (VGLB) are new methods for PVI. METHODS A total of 70 participants (35 in each group) who underwent a PVI with either AI-guided HPSD (50 W; AI 500 for the anterior and 400 for the posterior wall, respectively) or VGLB ablation were included in this observational pilot trial. Twenty minutes after each PVI, an APT was performed. The primary endpoint was the event-free survival from AF after three years. RESULTS A total of 137 (100%) PVs in the HPSD arm and 131 PVs (98.5%) in the VGLB arm were initially successfully isolated (p = 0.24). The overall procedure duration was similar in both arms (155 ± 39 in HPSD vs. 175 ± 58 min in VGLB, p = 0.191). Fluoroscopy time, left atrial dwelling time and duration from the first to the last ablation were longer in the VGLB arm (23 ± 8 vs. 12 ± 3 min, p < 0.001; 157 (111-185) vs. 134 (104-154) min, p = 0.049; 92(59-108) vs. 72 (43-85) min, p = 0.010). A total of 127 (93%) in the HPSD arm and 126 (95%) PVs in the VGLB arm remained isolated after APT (p = 0.34). The primary endpoint was met 1107 ± 68 days after ablation in 71% vs. 66% in the VGLB and HPSD arms, respectively (p = 0.65). CONCLUSIONS HPSD and VGLB did not differ with respect to long-term outcome of PVI. A large, randomized study should be conducted to compare clinical outcomes with respect to these new ablation techniques.
Collapse
Affiliation(s)
- Sönke Schildt
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Sabine Fredersdorf
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Carsten G Jungbauer
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Christian Hauck
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Daniel Tarnowski
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Kurt Debl
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Stefan Neef
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Christian Schach
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Samuel Sossalla
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Lars S Maier
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| | - Ekrem Üçer
- Internal Medicine II, Cardiology, University Hospital Regensburg, 93053 Regensburg, Germany
| |
Collapse
|
7
|
Kuroki K, Reddy VY, Iwasawa J, Kawamura I, Neuzil P, Estabrook B, Melsky G, Dukkipati SR, Koruth J. Preclinical evaluation of semi-automated laser ablation for pulmonary vein isolation: A comparative study. J Cardiovasc Electrophysiol 2023; 34:315-324. [PMID: 36511472 DOI: 10.1111/jce.15777] [Citation(s) in RCA: 1] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/23/2022] [Revised: 10/20/2022] [Accepted: 11/07/2022] [Indexed: 12/15/2022]
Abstract
INTRODUCTION Visually-guided laser balloon ablation (VGLA) currently requires careful manual rotation of the laser to create overlapping lesions. A novel semi-automated VGLA may reduce ablation times and lesion gaps. We aimed to compare semi-automated (SA) VGLA to that of manual (MN) VGLA. METHODS Acute: Nine swine underwent right superior pulmonary vein isolation (PVI) using either SA (n = 3, 13-18 W), MN (n = 3, 8.5-12 W), or radiofrequency (RF, n = 3, 25-40 W) and were killed acutely. Chronic: 16 swine, underwent PVI using either SA (n = 8, 15 W) or MN (n = 8, 10 W), and were survived for 1 month before being killed. All hearts were then submitted for pathological evaluation. RESULTS Acute: PVI was successful in all 9/9 swine with lesion counts significantly lower in the SA arm (5.3 ± 5.9, 33.7 ± 10.0, and 28.0 ± 4.4 in SA, MN, and RF arms; p = .007 for SA and MN). At necropsy, circumferentiality and transmurality were 98% and 94% in SA, 98% and 80% in MN, and 100% and 100% in RF arms. A single steam pop was noted on sectioning in the SA arm swine and occurred in the high dose (18 W) strategy. Chronic: PVI was acutely successful in 16/16 swine with no difference in PVI durability rates (62.5% vs. 75.0%), lesion transmurality (95.8 ± 17.4% vs.91.9 ± 25.9%), and circumferentiality (95.8 ± 6.6% vs. 94.8 ± 6.3%) between SA and MN arms. Catheter use time and lesion counts were lower in the SA arm compared to the MN arm (11.5 ± 12.7 vs. 21.8 ± 3.8 min, p = .046 and 4.8 ± 3.83 vs. 35.4 ± 4.4, p < .001). CONCLUSION Motor-assisted semi-automated laser balloon ablation can improve upon procedural efficiency by reducing ablation time.
Collapse
Affiliation(s)
- Kenji Kuroki
- Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA
| | - Vivek Y Reddy
- Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA
| | - Jin Iwasawa
- Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA
| | - Iwanari Kawamura
- Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA
| | - Petr Neuzil
- Department of Cardiology, Hospital Na Homolce, Prague, Czech Republic
| | | | | | - Srinivas R Dukkipati
- Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA
| | - Jacob Koruth
- Helmsley Electrophysiology Center, Icahn School of Medicine at Mount Sinai, New York, New York, USA
| |
Collapse
|
8
|
Tohoku S, Bordignon S, Bologna F, Chen S, Urbanek L, Operhalski F, Chun KJ, Schmidt B. Laser balloon in pulmonary vein isolation for atrial fibrillation: current status and future prospects. Expert Rev Med Devices 2021; 18:1083-1091. [PMID: 34618626 DOI: 10.1080/17434440.2021.1990754] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/20/2022]
Abstract
INTRODUCTION Visually guided laser balloon (LB) catheter has been an established modality dedicated for pulmonary vein (PV) isolation in patients with atrial fibrillation. The newly updated version of this novel device has technically evolved recent years. AREAS COVERED This review will summarize the contemporary technical evolution of LB catheter. Available efficacy outcomes and the historical change of ablation style will be evaluated. Furthermore, the future perspectives for clinical practice are discussed. EXPERT COMMENTARY The initial LB ablation system provided comparable clinical results in PV isolation with other technologies, but with a unique strategical concept enabling the direct visualization of the tissue to cauterize. With multigenerational development, the LB catheter has been equipped with more compliant balloon for favorable PV occlusion and a robotically motor driven continuous ablation mode (RAPID mode). These technical innovations changed the concept of the ablation strategy using LB catheter as 'point-by-point' into 'single-shot' fashion. The remaining tasks are further improvements such as equipping with real-time recording system of intracardiac electrogram, durable structured balloon and the instrument for visualizing the cauterization area in a 360-degree panoramic view, which includes potential possibilities to develop this novel device to the more optimal device for PV isolation.
Collapse
Affiliation(s)
- Shota Tohoku
- Academy for Arrhythmias (FAFA), Abteilung für Kardiologie, Medizinische Klinik III, Agaplesion Markus Krankenhaus, Cardioangiologisches Centrum Bethanien, Frankfurt, Germany
| | - Stefano Bordignon
- Academy for Arrhythmias (FAFA), Abteilung für Kardiologie, Medizinische Klinik III, Agaplesion Markus Krankenhaus, Cardioangiologisches Centrum Bethanien, Frankfurt, Germany
| | - Fabrizio Bologna
- Academy for Arrhythmias (FAFA), Abteilung für Kardiologie, Medizinische Klinik III, Agaplesion Markus Krankenhaus, Cardioangiologisches Centrum Bethanien, Frankfurt, Germany
| | - Shaojie Chen
- Academy for Arrhythmias (FAFA), Abteilung für Kardiologie, Medizinische Klinik III, Agaplesion Markus Krankenhaus, Cardioangiologisches Centrum Bethanien, Frankfurt, Germany.,Die Sektion Medizin, Universität zu Lübeck, Lübeck, Germany
| | - Lukas Urbanek
- Academy for Arrhythmias (FAFA), Abteilung für Kardiologie, Medizinische Klinik III, Agaplesion Markus Krankenhaus, Cardioangiologisches Centrum Bethanien, Frankfurt, Germany
| | - Felix Operhalski
- Universitätsklinikum Frankfurt, Medizinische Klinik 3- Klinik für Kardiologie, Frankfurt, Germany
| | - Kr Julian Chun
- Academy for Arrhythmias (FAFA), Abteilung für Kardiologie, Medizinische Klinik III, Agaplesion Markus Krankenhaus, Cardioangiologisches Centrum Bethanien, Frankfurt, Germany.,Die Sektion Medizin, Universität zu Lübeck, Lübeck, Germany
| | - Boris Schmidt
- Academy for Arrhythmias (FAFA), Abteilung für Kardiologie, Medizinische Klinik III, Agaplesion Markus Krankenhaus, Cardioangiologisches Centrum Bethanien, Frankfurt, Germany.,Universitätsklinikum Frankfurt, Medizinische Klinik 3- Klinik für Kardiologie, Frankfurt, Germany
| |
Collapse
|
9
|
Tohoku S, Bordignon S, Chen S, Zanchi S, Bianchini L, Trolese L, Operhalski F, Urbanek L, Chun KRJ, Schmidt B. Single-sweep pulmonary vein isolation using the new third-generation laser balloon-Evolution in ablation style using endoscopic ablation system. J Cardiovasc Electrophysiol 2021; 32:2923-2932. [PMID: 34535929 DOI: 10.1111/jce.15245] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/26/2021] [Revised: 08/25/2021] [Accepted: 09/08/2021] [Indexed: 01/24/2023]
Abstract
BACKGROUND The endoscopic ablation system (EAS) is an established ablation device for pulmonary vein isolation (PVI) in patients with atrial fibrillation (AF). The novel X3 EAS is now equipped with a contiguous circumferential ablation mode (RAPID mode). AIM To determine the feasibility of single-shot fashioned ablation using X3. METHODS Consecutive patients who underwent AF ablation using X3 were enrolled. We assessed the acute procedural data focusing on "Single-sweep PVI" defined as successful PVI with a single RAPID mode energy application, and on "first-pass isolation" defined as successful PVI after initial circular lesion set. RESULTS One hundred AF patients (56% male, age: 68 ± 10 years, 66% paroxysmal AF) were analyzed. A total of 379 of 383 PVs (99%) were isolated with X3. Single-sweep PVI and first-pass-isolation were achieved in 214 PVs (56%) and in 362 PVs (95%), respectively. Single-sweep PVI rates varied across PVs with higher rates at the superior PVs (61.2% vs. inferior PVs: 49.5%, p = .0239) and at PVs with maximal ostial diameter <24 mm (57.6% vs. >24 mm: 36.8%, p = .0151). The mean total procedure and fluoroscopy times were 43.0 ± 10 and 4.0 ± 2 min, respectively. In none of the patients an acute thromboembolic event (stroke or transient ischemic attack) or a pericardial effusion/tamponade occurred. A single transient phrenic nerve palsy was observed. CONCLUSION The new X3 EAS allows for single-shot fashioned ablation in terms of single-sweep PVI in half or more of PVs. The new RAPID ablation mode leads to an improved rate of first-pass isolation associated with very short procedure times without compromising safety.
Collapse
Affiliation(s)
- Shota Tohoku
- Cardioangiologisches Centrum Bethanien, Frankfurt, Germany
| | | | - Shaojie Chen
- Cardioangiologisches Centrum Bethanien, Frankfurt, Germany.,Die Sektion Medizin, Universität zu Lübeck, Lübeck, Germany
| | - Simone Zanchi
- Cardioangiologisches Centrum Bethanien, Frankfurt, Germany
| | | | - Luca Trolese
- Universitäts Herzzentrum Freiburg-Bad Krozingen, Klinik für Kardiologie und Angiologie I, Freiburg, Germany
| | - Felix Operhalski
- Universitätsklinikum Frankfurt, Medizinische Klinik 3-Klinik für Kardiologie, Frankfurt, Germany
| | - Lukas Urbanek
- Cardioangiologisches Centrum Bethanien, Frankfurt, Germany
| | - K R Julian Chun
- Cardioangiologisches Centrum Bethanien, Frankfurt, Germany.,Die Sektion Medizin, Universität zu Lübeck, Lübeck, Germany
| | - Boris Schmidt
- Cardioangiologisches Centrum Bethanien, Frankfurt, Germany.,Universitätsklinikum Frankfurt, Medizinische Klinik 3-Klinik für Kardiologie, Frankfurt, Germany
| |
Collapse
|
10
|
Nagase T, Kobori A, Inaba O, Sasaki Y, Tomizawa N, Asano S, Fukunaga H, Mabuchi K, Inoue K, Tanizaki K, Murai T, Iguchi N, Nitta J, Isobe M. Comparison of dragging ablation and point-by-point ablation with a laser balloon on linear lesion formation. J Cardiovasc Electrophysiol 2020; 31:2848-2856. [PMID: 32786049 DOI: 10.1111/jce.14714] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 06/28/2020] [Revised: 07/24/2020] [Accepted: 08/08/2020] [Indexed: 12/21/2022]
Abstract
INTRODUCTION Lesion size and continuity in dragging laser balloon (LB) ablation, which may enable fast and durable pulmonary vein isolation for atrial fibrillation, are unknown. We evaluated the differences in size and continuity of linear lesions formed by dragging ablation and conventional point-by-point ablation using an LB in vitro model. METHODS AND RESULTS Chicken muscles were cauterized using the first-generation LB in dragging and point-by-point fashion. Dragging ablation was manually performed with different dragging speeds (0.5-2°/s) using an overlap ratio of the beginning and last site during one application at 12 W/20 s and 8.5 W/30 s. Point-by-point ablation was performed with 25% and 50% overlap ratios at six energy settings (5.5 W/30 s to 12 W/20 s). Lesion depth, width, and continuity were compared. Lesion continuity was assessed by the surface and deep visible gap degree categorized from 1 (perfect) to 3 (poor). Twenty lesions were evaluated for each ablation protocol. Lesion depth and width in dragging ablation at high power (12 W) were comparable with most measurements in point-by-point ablation. Lesion depth and width were smaller at faster-dragging speed and lower power (8.5 W) in dragging ablation. The surface visible gap degree was better in dragging ablation at all dragging speeds than a 25% overlapped point-by-point ablation (p < .001). CONCLUSION Dragging LB ablation at high power provides deep and continuous linear lesion formation comparable with that of point-by-point LB ablation. However, lesion depth and width depending on the dragging speed and power.
Collapse
Affiliation(s)
- Takahiko Nagase
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Atsushi Kobori
- Department of Cardiovascular Medicine, Kobe City Medical Center General Hospital, Kobe, Hyogo, Japan
| | - Osamu Inaba
- Department of Cardiology, Saitama Red Cross Hospital, Saitama, Japan
| | - Yasuhiro Sasaki
- Department of Cardiovascular Medicine, Kobe City Medical Center General Hospital, Kobe, Hyogo, Japan
| | - Naoki Tomizawa
- Department of Cardiology, Saitama Red Cross Hospital, Saitama, Japan
| | - So Asano
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Hiroshi Fukunaga
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Kei Mabuchi
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Kanki Inoue
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Kohei Tanizaki
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Tatsuya Murai
- Department of Pathology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Nobuo Iguchi
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Junichi Nitta
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| | - Mitsuaki Isobe
- Department of Cardiology, Sakakibara Heart Institute, Fuchu, Tokyo, Japan
| |
Collapse
|
11
|
Huang HD, Rodriguez JM, Serafini NJ, Macias C, Winterfield J, Sharma PS, Larsen T, Krishnan K, Trohman RG. Comparison between minimal fluoroscopy and conventional approaches for visually guided laser balloon pulmonary vein isolation ablation. J Cardiovasc Electrophysiol 2020; 31:1608-1615. [PMID: 32406100 DOI: 10.1111/jce.14546] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 02/28/2020] [Revised: 04/17/2020] [Accepted: 05/10/2020] [Indexed: 11/28/2022]
Abstract
INTRODUCTION Although balloon-based techniques, such as the laser balloon (LB) ablation have simplified pulmonary vein isolation (PVI), procedural fluoroscopy usage remains higher in comparison to radiofrequency PVI approaches due to limited 3-dimensional mapping system integration. METHODS In this prospective study, 50 consecutive patients were randomly assigned in alternating fashion to a low fluoroscopy group (LFG; n = 25) or conventional fluoroscopy group (CFG; n = 25) and underwent de novo PVI procedures using visually guided LB technique. RESULTS There was no statistical difference in baseline characteristics or cross-overs between treatment groups. Acute PVI was accomplished in all patients. Mean follow up was 318 ± 69 days. Clinical recurrence of atrial fibrillation after PVI was similar between groups (CFG: 19% vs LFG: 15%; P = .72). Total fluoroscopy time was significantly lower in the LFG than the CFG (1.7 ± 1.4 vs 16.9 ± 5.9 minutes; P < .001) despite similar total procedure duration (143 ± 22 vs 148 ± 22 minutes; P = .42) and mean LA dwell time (63 ± 15 vs 59 ± 10 minutes; P = .28). Mean dose area product was significantly lower in the LFG (181 ± 125 vs 1980 ± 750 μGym2 ; P < .001). Fluoroscopy usage after transseptal access was substantially lower in the LFG (0.63 ± 0.43 vs 11.70 ± 4.32 minutes; P < .001). Complications rates were similar between both groups (4% vs 2%; P = .57). CONCLUSIONS This study demonstrates that LB PVI can be safely achieved using a novel low fluoroscopy protocol while also substantially reducing fluoroscopy usage and radiation exposure in comparison to conventional approaches for LB ablation.
Collapse
Affiliation(s)
- Henry D Huang
- Division of Cardiology, Rush University Medical Center, Chicago, Illinois
| | - Jason M Rodriguez
- Division of Cardiology, Rush University Medical Center, Chicago, Illinois
| | | | - Carlos Macias
- Division of Cardiology, UCLA Cardiac Arrhythmia Center, Ronald Reagan UCLA Medical Center, Los Angeles, California
| | - Jeffrey Winterfield
- Division of Cardiology, Medical University of South Carolina, Charleston, South Carolina
| | - Parikshit S Sharma
- Division of Cardiology, Rush University Medical Center, Chicago, Illinois
| | - Timothy Larsen
- Division of Cardiology, Rush University Medical Center, Chicago, Illinois
| | - Kousik Krishnan
- Division of Cardiology, Rush University Medical Center, Chicago, Illinois
| | - Richard G Trohman
- Division of Cardiology, Rush University Medical Center, Chicago, Illinois
| |
Collapse
|
12
|
Wissner E. The importance of antral pulmonary vein isolation during visually-guided laser balloon ablation. J Cardiovasc Electrophysiol 2020; 31:1606-1607. [PMID: 32367571 DOI: 10.1111/jce.14527] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 04/28/2020] [Accepted: 04/29/2020] [Indexed: 11/26/2022]
Affiliation(s)
- Erik Wissner
- Division of Cardiology, University of Illinois at Chicago, Chicago, Illinois
| |
Collapse
|
13
|
Huang HD, Serafini N, Rodriguez J, Sharma PS, Krishnan K, Trohman RG. Near-zero Fluoroscopic Approach for Laser Balloon Pulmonary Vein Isolation Ablation: A Case Study. J Innov Card Rhythm Manag 2020; 11:4069-4074. [PMID: 32368382 PMCID: PMC7192128 DOI: 10.19102/icrm.2020.110402] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/06/2019] [Accepted: 09/27/2019] [Indexed: 01/08/2023] Open
Abstract
Fluoroscopy remains a cornerstone imaging modality for catheter placement and positioning in electrophysiology device and ablation procedures. However, efforts are being made to reduce the cumulative exposure to radiation in the patient and physician alike. We present the case of a 59-year-old male patient with hypertension, chronic kidney disease, and paroxysmal atrial fibrillation who underwent successful near-fluoroless laser balloon (LB) pulmonary vein isolation (PVI) ablation. Though this case demonstrates the usage of a novel protocol for near-fluoroless LB ablation that resulted in successful, uncomplicated acute PVI, the feasibility and safety of this technique should be validated in a larger series or prospective comparative study.
Collapse
Affiliation(s)
- Henry D Huang
- Division of Internal Medicine, Section of Cardiology, Rush University Medical Center, Chicago, IL, USA
| | - Nicholas Serafini
- Division of Internal Medicine, Section of Cardiology, Rush University Medical Center, Chicago, IL, USA
| | - Jason Rodriguez
- Division of Internal Medicine, Section of Cardiology, Rush University Medical Center, Chicago, IL, USA
| | - Parikshit S Sharma
- Division of Internal Medicine, Section of Cardiology, Rush University Medical Center, Chicago, IL, USA
| | - Kousik Krishnan
- Division of Internal Medicine, Section of Cardiology, Rush University Medical Center, Chicago, IL, USA
| | - Richard G Trohman
- Division of Internal Medicine, Section of Cardiology, Rush University Medical Center, Chicago, IL, USA
| |
Collapse
|
14
|
Reynolds MR, Zheng Q, Doros G. Laser balloon ablation for AF: A systematic review and meta-analysis. J Cardiovasc Electrophysiol 2018; 29:1363-1370. [PMID: 30016008 DOI: 10.1111/jce.13698] [Citation(s) in RCA: 29] [Impact Index Per Article: 4.8] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 05/31/2018] [Revised: 06/26/2018] [Accepted: 07/09/2018] [Indexed: 11/30/2022]
Abstract
INTRODUCTION The HeartLight laser balloon ablation system was US Food and Drug Administration approved in 2016 for the treatment of paroxysmal atrial fibrillation (AF), but there have been numerous single-center and multicenter studies published reporting its outcomes, in addition to a few randomized trials. We aimed to systematically review and synthesize currently published outcome data on AF ablation using the laser balloon ablation system. METHODS AND RESULTS We performed a systematic review and meta-analysis of published studies of AF ablation performed using the laser balloon ablation system. Human studies reporting acute procedural results with a minimum of 6 months follow-up were included. Outcomes of interest included acute and 12-month procedural efficacy, safety, and procedure duration. Aggregated data were analyzed with random effects models, using a Bayesian hierarchical approach. We identified 17 published manuscripts comprising a sample of 1188 patients (mean age 61 years, 80% paroxysmal). At procedure end, 98.8% of targeted pulmonary veins were successfully isolated. The pooled estimate for 12-month freedom from atrial arrhythmia without use of antiarrhythmic drugs for patients with paroxysmal AF was 74.3% (95% confidence interval [CI], 59.9% to 86.4%), and for all AF types combined was 72.9% (65.3% to 79.9%). The most commonly reported procedural complication was phrenic nerve injury (pooled incidence 2.6%; 95% CI, 1.4% to 3.9%), which resolved during follow-up in most cases. CONCLUSION Laser balloon ablation is highly effective at achieving pulmonary vein isolation. Although comparisons are mainly indirect, safety and 12-month efficacy compare favorably with those observed using other currently used AF ablation technologies.
Collapse
Affiliation(s)
- Matthew R Reynolds
- Division of Cardiology, Lahey Hospital & Medical Center, Burlington, Massachusetts
- Baim Institute for Clinical Research, Boston, Massachusetts
| | - Qi Zheng
- Division of Cardiology, Brigham & Women's Hospital, Boston, Massachusetts
| | - Gheorghe Doros
- Baim Institute for Clinical Research, Boston, Massachusetts
- Department of Biostatistics, Boston University School of Public Health, Boston, Massachusetts
| |
Collapse
|
15
|
Kumar N, Pison L, Blaauw Y, Timmermans C, La Meir M, Gelsamino S, Crijns HJ, Maessen J. Pulmonary Vein Stenosis After Laser Balloon Ablation for Atrial Fibrillation. JACC Clin Electrophysiol 2015; 1:220-221. [PMID: 29759368 DOI: 10.1016/j.jacep.2015.04.001] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/12/2015] [Revised: 03/26/2015] [Accepted: 04/09/2015] [Indexed: 11/29/2022]
Affiliation(s)
- Narendra Kumar
- Department of Cardiology, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht (CARIM), Maastricht, the Netherlands.
| | - Laurent Pison
- Department of Cardiology, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht (CARIM), Maastricht, the Netherlands
| | - Yuri Blaauw
- Department of Cardiology, University Medical Center Groningen, Groningen, the Netherlands
| | - Carl Timmermans
- Department of Cardiology, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht (CARIM), Maastricht, the Netherlands
| | - Mark La Meir
- Department of Cardiac Surgery, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht (CARIM), Maastricht, the Netherlands
| | - Sandro Gelsamino
- Department of Cardiac Surgery, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht (CARIM), Maastricht, the Netherlands
| | - Harry J Crijns
- Department of Cardiology, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht (CARIM), Maastricht, the Netherlands
| | - Jos Maessen
- Department of Cardiac Surgery, Maastricht University Medical Centre and Cardiovascular Research Institute Maastricht (CARIM), Maastricht, the Netherlands
| |
Collapse
|