Bianchi SD, Giovannetti P, Albrito F. [Niobium filtration in dental radiology. Effects on image quality and on dosage].
Radiol Med 1997;
93:681-5. [PMID:
9411513]
[Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Subscribe] [Scholar Register] [Indexed: 02/05/2023]
Abstract
UNLABELLED
The necessity of reducing the radiation dose to the patient in diagnostic radiology according to the ALARA guideline established by the ICRP has stimulated the research on additional filtration systems capable of removing the low-energy photons increasing the dose and worsening image quality. Very few literature studies deal with the effects of niobium filtration on image quality in dental radiography with the use of modulation transfer function (MTF) and square wave response function (SWRF). Only one study has considered those effects measuring dose absorption in an anthropomorphic phantom.
THE AIMS OF OUR STUDY WERE
1) to study the effects of a 30 microns additional niobium filter on image quality using the SWRF; 2) to compare the doses absorbed in vivo during a complete radiographic survey of the mouth, both with and without niobium filtration. Qualitative studies led us to conclude that niobium filtration does not significantly worsen radiographic image quality. The following doses were measured in the exposures with niobium filtration: 1678 microGy to 6000 microGy (intraoral doses) and 75 microGy to 3643 microGy (skin doses). The comparison with the doses measured during the exposures made with conventional filtration indicates that dose reduction is not significantly advantageous relative to risk reduction. In conclusion, additional niobium filtration is not advisable in dental radiology, also because of the filter cost and of the increased wear of the unit.
Collapse