In-hospital outcomes and predictors of paravalvular leak and deep implantation with the Evolut-R 34 mm device: A comparison with smaller Evolut-R sizes.
CARDIOVASCULAR REVASCULARIZATION MEDICINE 2021;
35:19-26. [PMID:
33933374 DOI:
10.1016/j.carrev.2021.03.024]
[Citation(s) in RCA: 4] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/15/2020] [Revised: 03/06/2021] [Accepted: 03/30/2021] [Indexed: 11/16/2022]
Abstract
PURPOSE
To compare in-hospital outcome of Evolut-R 34 mm vs. smaller Evolut-R devices and to identify predictors of paravalvular leak (PVL) and deep implantation specific for Evolut-R 34 mm.
METHODS
This single-center retrospective study included 359 consecutive patients undergoing transcatheter aortic valve replacement (TAVR) with Evolut-R 34 mm (N = 84,23.4%) and Evolut-R 23/26/29 mm (N = 275,76.6%) between 2016 and 2019.
RESULTS
Patients in Evolut-R 34 mm group were more frequently males, had lower STS score, ejection fraction, and mean aortic gradient compared to the Evolut-R 23/26/29 mm group. Horizontal aorta and large LVOT were more frequent findings in the Evolut-R 34 mm group, whereas calcium volume was comparable among the groups. During TAVR, mean implantation depth and contrast volume were greater in the Evolut-R 34 mm group, compared to the Evolut 23/26/29 mm group. Post-procedurally, 30-day mortality, ≥moderate PVL, device success and pacemaker implantation (PM) rates were comparable between groups. Among independent predictors of ≥moderate PVL, calcium volume (OR:1.04; p < 0.001) was predictive with different thresholds in both groups, whereas aortic angulation (OR:1.40; p = 0.005) was predictive only in Evolut-R 34 mm group at a cutoff of 60° (AUC:0.73; p = 0.043). Body weight (OR:1.03; p = 0.027), left ventricular outflow tract (LVOT) diameter (OR:1.34; p = 0.001), and mean aortic gradient (OR:0.96; p = 0.006) were independent predictors of deep implantation (mean depth ≥ 6 mm), with LVOT>27 mm being predictive specifically for Evolut-R 34 mm (AUC:0.66; p = 0.024).
CONCLUSIONS
TAVR with Evolut-R 34 mm and Evolut-R 23/26/29 mm showed comparable in-hospital outcome. Aortic angulation >60° and LVOT >27 mm were predictive respectively of ≥moderate PVL and deep implantation specifically in Evolut-R 34 mm patients.
Collapse