Kretov Е, Naryshkin I, Baystrukov V, Grazhdankin I, Prokhorikhin A, Zubarev D, Biryukov A, Verin V, Boykov A, Malaev D, Pokushalov E, Romanov A, Bergmann MW. Three-months optical coherence tomography analysis of a biodegradable polymer, sirolimus-eluting stent.
J Interv Cardiol 2018;
31:442-449. [PMID:
29651802 DOI:
10.1111/joic.12510]
[Citation(s) in RCA: 8] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/23/2017] [Revised: 02/07/2017] [Accepted: 03/01/2018] [Indexed: 12/24/2022] Open
Abstract
OBJECTIVE
We aimed to assess early neointimal healing by optical coherence tomography (OCT) 3 months after implantation of the ultrathin Orsiro® sirolimus-eluting stent with biodegradable polymer.
BACKGROUND
New generations of drug-eluting stents with biodegradable polymer have been developed to avoid the continued vascular irritation of durable polymers.
METHODS
In this prospective, open-label study, 34 patients received an Orsiro® sirolimus-eluting stent with biodegradable polymer. In a subgroup of patients (n = 15), the intervention was performed under OCT guidance. All patients underwent OCT-examination at three months. The primary endpoint was 3-month neointimal healing (NIH) score, calculated by weighing the presence of filling defects, malapposed and uncovered struts. Secondary endpoint was maturity of tissue coverage at 3 months.
RESULTS
At 3 months, NIH score was 13.7 (5.4-22), covered struts per lesion were 90% (84-97%), malapposed struts were 2.7% (0.8-5.4%) and rate of mature tissue coverage was 47% (42-53%). No target lesion failure occurred up to 12 months. Patients with OCT-guided stent implantation demonstrated a trend toward earlier stent healing as demonstrated by superior NIH scores (angio guided: 17.6% [8.8-26.4]; OCT-guided: 9.8% [4.0-15.5]; mean difference -8, [95%CI: -18.7-2.9], P = 0.123). This group had significantly more covered struts per lesion (angio-guided: 86% [82-90]; 95% [92-99]; mean difference 9% [95%CI: 3-15], P = 0.001).
CONCLUSION
The Orsiro® sirolimus-eluting stent with biodegradable polymer shows early vascular healing with a high rate of strut coverage at 3-month follow-up. OCT guided stent implantation had a positive impact on early vascular healing.