Monte Carlo simulations and phantom validation of low-dose radiotherapy to the lungs using an interventional radiology C-arm fluoroscope.
Phys Med 2021;
94:24-34. [PMID:
34979431 DOI:
10.1016/j.ejmp.2021.12.014]
[Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/29/2021] [Revised: 11/30/2021] [Accepted: 12/27/2021] [Indexed: 11/26/2022] Open
Abstract
PURPOSE
To use MC simulations and phantom measurements to investigate the dosimetry of a kilovoltage x-ray beam from an IR fluoroscope to deliver low-dose (0.3-1.0 Gy) radiotherapy to the lungs.
MATERIALS AND METHODS
PENELOPE was used to model a 125 kV, 5.94 mm Al HVL x-ray beam produced by a fluoroscope. The model was validated through depth-dose, in-plane/cross-plane profiles and absorbed dose at 2.5-, 5.1-, 10.2- and 15.2-cm depths against the measured beam in an acrylic phantom. CT images of an anthropomorphic phantom thorax/lungs were used to simulate 0.5 Gy dose distributions for PA, AP/PA, 3-field and 4-field treatments. DVHs were generated to assess the dose to the lungs and nearby organs. Gafchromic film was used to measure doses in the phantom exposed to PA and 4-field treatments, and compared to the MC simulations.
RESULTS
Depth-dose and profile results were within 3.2% and 7.8% of the MC data uncertainty, respectively, while dose gamma analysis ranged from 0.7 to 1.0. Mean dose to the lungs were 1.1-, 0.8-, 0.9-, and 0.8- Gy for the PA, AP/PA, 3-field, and 4-field after isodose normalization to cover ∼ 95% of each lung volume. Skin dose toxicity was highest for the PA and lowest for the 4-field, and both arrangements successfully delivered the treatment on the phantom. However, the dose distribution for the PA was highly non-uniform and produced skin doses up to 4 Gy. The dose distribution for the 4-field produced a uniform 0.6 Gy dose throughout the lungs, with a maximum dose of 0.73 Gy. The average percent difference between experimental and Monte Carlo values were -0.1% (range -3% to +4%) for the PA treatment and 0.3% (range -10.3% to +15.2%) for the 4-field treatment.
CONCLUSION
A 125 kV x-ray beam from an IR fluoroscope delivered through two or more fields can deliver an effective low-dose radiotherapy treatment to the lungs. The 4-field arrangement not only provides an effective treatment, but also significant dose sparing to healthy organs, including skin, compared to the PA treatment. Use of fluoroscopy appears to be a viable alternative to megavoltage radiation therapy equipment for delivering low-dose radiotherapy to the lungs.