Dai C, Li J, Li QM, Guo X, Fan YY, Qin HY. Effect of tunneled and nontunneled peripherally inserted central catheter placement: A randomized controlled trial.
J Vasc Access 2019;
21:511-519. [PMID:
31709895 DOI:
10.1177/1129729819888120]
[Citation(s) in RCA: 9] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/16/2022] Open
Abstract
PURPOSE
To compare the effect of tunneled and nontunneled peripherally inserted central catheter placement under B-mode ultrasound.
METHODS
A single center, randomized, controlled, nonblinded, prospective trial was conducted in Guangzhou, China, between July 2018 and May 2019. A total of 174 participants were randomized to the experimental group (tunneled peripherally inserted central catheter) or the control group (nontunneled peripherally inserted central catheter) and were followed until extubation. Basic characteristics, peripherally inserted central catheter characteristics, the incidence of complications, and the costs of peripherally inserted central catheter placement and maintenance were collected. Data were analyzed by intention-to-treat.
RESULTS
A total of 168 of the participants had successful peripherally inserted central catheter placements (85/87, 97.7% in the experimental group and 83/87, 95.4% in the control group, P = 0.682). Compared to the control group, the experimental group had a lower incidence of complications during the placement (18.4% vs 32.2%, P = 0.036), a lower incidence of wound oozing (27.6% vs 57.5%, P < 0.001), a lower incidence of medical adhesive-related skin injury (9.2% vs 25.3%, P = 0.005), a lower incidence of venous thrombosis (1.1% vs 9.2%, P = 0.034), a lower incidence of catheter dislodgement (1.1% vs 9.2%, P = 0.034), and lower costs of peripherally inserted central catheter maintenance at 1, 2, and 3 months (P < 0.05).
CONCLUSION
Tunneled peripherally inserted central catheter may be recommended for good effectiveness.