1
|
di Franco F, Rosuel N, Gallin-Martel L, Gallin-Martel ML, Ghafooryan-Sangchooli M, Keshmiri S, Motte JF, Muraz JF, Pellicioli P, Ruat M, Serduc R, Verry C, Dauvergne D, Adam JF. Monocrystalline diamond detector for online monitoring during synchrotron microbeam radiotherapy. JOURNAL OF SYNCHROTRON RADIATION 2023; 30:1076-1085. [PMID: 37815374 PMCID: PMC10624038 DOI: 10.1107/s160057752300752x] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/24/2023] [Accepted: 08/28/2023] [Indexed: 10/11/2023]
Abstract
Microbeam radiation therapy (MRT) is a radiotherapy technique combining spatial fractionation of the dose distribution on a micrometric scale, X-rays in the 50-500 keV range and dose rates up to 16 × 103 Gy s-1. Nowadays, in vivo dosimetry remains a challenge due to the ultra-high radiation fluxes involved and the need for high-spatial-resolution detectors. The aim here was to develop a striped diamond portal detector enabling online microbeam monitoring during synchrotron MRT treatments. The detector, a 550 µm bulk monocrystalline diamond, is an eight-strip device, of height 3 mm, width 178 µm and with 60 µm spaced strips, surrounded by a guard ring. An eight-channel ASIC circuit for charge integration and digitization has been designed and tested. Characterization tests were performed at the ID17 biomedical beamline of the European Synchrotron Radiation Facility (ESRF). The detector measured direct and attenuated microbeams as well as interbeam fluxes with a precision level of 1%. Tests on phantoms (RW3 and anthropomorphic head phantoms) were performed and compared with simulations. Synchrotron radiation measurements were performed on an RW3 phantom for strips facing a microbeam and for strips facing an interbeam area. A 2% difference between experiments and simulations was found. In more complex geometries, a preliminary study showed that the absolute differences between simulated and recorded transmitted beams were within 2%. Obtained results showed the feasibility of performing MRT portal monitoring using a microstriped diamond detector. Online dosimetric measurements are currently ongoing during clinical veterinary trials at ESRF, and the next 153-strip detector prototype, covering the entire irradiation field, is being finalized at our institution.
Collapse
Affiliation(s)
- Francesca di Franco
- Université Grenoble-Alpes, CNRS, Grenoble INP, LPSC UMR5821, 38000 Grenoble, France
| | - Nicolas Rosuel
- Université Grenoble-Alpes, CNRS, Grenoble INP, LPSC UMR5821, 38000 Grenoble, France
| | | | | | | | - Sarvenaz Keshmiri
- Université Grenoble-Alpes, UGA/INSERM UA7 STROBE, 2280 Rue de la Piscine, 38400 Saint-Martin d’Hères, France
| | - Jean-François Motte
- Université Grenoble-Alpes, Institut Néel, CNRS, Grenoble-INP, Grenoble, France
| | - Jean-François Muraz
- Université Grenoble-Alpes, CNRS, Grenoble INP, LPSC UMR5821, 38000 Grenoble, France
| | | | | | - Raphael Serduc
- Université Grenoble-Alpes, UGA/INSERM UA7 STROBE, 2280 Rue de la Piscine, 38400 Saint-Martin d’Hères, France
- Centre Hospitalier Universitaire Grenoble-Alpes, CS10217, 38043 Grenoble, France
| | - Camille Verry
- Centre Hospitalier Universitaire Grenoble-Alpes, CS10217, 38043 Grenoble, France
| | - Denis Dauvergne
- Université Grenoble-Alpes, CNRS, Grenoble INP, LPSC UMR5821, 38000 Grenoble, France
| | - Jean-François Adam
- Université Grenoble-Alpes, UGA/INSERM UA7 STROBE, 2280 Rue de la Piscine, 38400 Saint-Martin d’Hères, France
- Centre Hospitalier Universitaire Grenoble-Alpes, CS10217, 38043 Grenoble, France
| |
Collapse
|
2
|
Comparison of the dosimetric response of two Sr salts irradiated with 60Co γ-rays and synchrotron X-rays at ultra-high dose rate. Radiat Phys Chem Oxf Engl 1993 2023. [DOI: 10.1016/j.radphyschem.2023.110923] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 03/22/2023]
|
3
|
Ventura JA, Donoghue JF, Nowell CJ, Cann LM, Day LRJ, Smyth LML, Forrester HB, Rogers PAW, Crosbie JC. The γH2AX DSB marker may not be a suitable biodosimeter to measure the biological MRT valley dose. Int J Radiat Biol 2021; 97:642-656. [PMID: 33617395 DOI: 10.1080/09553002.2021.1893854] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/31/2020] [Revised: 12/31/2020] [Accepted: 02/11/2021] [Indexed: 10/22/2022]
Abstract
PURPOSE γH2AX biodosimetry has been proposed as an alternative dosimetry method for microbeam radiation therapy (MRT) because conventional dosimeters, such as ionization chambers, lack the spatial resolution required to accurately measure the MRT valley dose. Here we investigated whether γH2AX biodosimetry should be used to measure the biological valley dose of MRT-irradiated mammalian cells. MATERIALS AND METHODS We irradiated human skin fibroblasts and mouse skin flaps with synchrotron MRT and broad beam (BB) radiation. BB doses of 1-5 Gy were used to generate a calibration curve in order to estimate the biological MRT valley dose using the γH2AX assay. RESULTS Our key finding was that MRT induced a non-linear dose response compared to BB, where doses 2-3 times greater showed the same level of DNA DSB damage in the valley in cell and tissue studies. This indicates that γH2AX may not be an appropriate biodosimeter to estimate the biological valley doses of MRT-irradiated samples. We also established foci yields of 5.9 ± 0 . 04 and 27.4 ± 2 . 5 foci/cell/Gy in mouse skin tissue and human fibroblasts respectively, induced by BB. Using Monte Carlo simulations, a linear dose response was seen in cell and tissue studies and produced predicted peak-to-valley dose ratios (PVDRs) of ∼30 and ∼107 for human fibroblasts and mouse skin tissue respectively. CONCLUSIONS Our report highlights novel MRT radiobiology, attempts to explain why γH2AX may not be an appropriate biodosimeter and suggests further studies aimed at revealing the biological and cellular communication mechanisms that drive the normal tissue sparing effect, which is characteristic of MRT.
Collapse
Affiliation(s)
- Jessica A Ventura
- Department of Obstetrics and Gynaecology, Royal Women's Hospital, University of Melbourne, Parkville, Australia
| | - Jacqueline F Donoghue
- Department of Obstetrics and Gynaecology, Royal Women's Hospital, University of Melbourne, Parkville, Australia
| | - Cameron J Nowell
- Monash Institute of Pharmaceutical Sciences, Monash University, Parkville, Australia
| | - Leonie M Cann
- Department of Obstetrics and Gynaecology, Royal Women's Hospital, University of Melbourne, Parkville, Australia
| | - Liam R J Day
- School of Science, RMIT University, Melbourne, Australia
| | - Lloyd M L Smyth
- Department of Obstetrics and Gynaecology, Royal Women's Hospital, University of Melbourne, Parkville, Australia
| | - Helen B Forrester
- School of Science, RMIT University, Melbourne, Australia
- Centre for Innate Immunity and Infectious Diseases, Hudson Institute of Medical Research, Monash University, Clayton, Australia
- Department of Molecular and Translational Sciences, Monash University, Clayton, Australia
| | - Peter A W Rogers
- Department of Obstetrics and Gynaecology, Royal Women's Hospital, University of Melbourne, Parkville, Australia
| | | |
Collapse
|
4
|
Pellicioli P, Donzelli M, Davis JA, Estève F, Hugtenburg R, Guatelli S, Petasecca M, Lerch MLF, Bräuer-Krisch E, Krisch M. Study of the X-ray radiation interaction with a multislit collimator for the creation of microbeams in radiation therapy. JOURNAL OF SYNCHROTRON RADIATION 2021; 28:392-403. [PMID: 33650550 DOI: 10.1107/s1600577520016811] [Citation(s) in RCA: 6] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 06/16/2020] [Accepted: 12/31/2020] [Indexed: 06/12/2023]
Abstract
Microbeam radiation therapy (MRT) is a developing radiotherapy, based on the use of beams only a few tens of micrometres wide, generated by synchrotron X-ray sources. The spatial fractionation of the homogeneous beam into an array of microbeams is possible using a multislit collimator (MSC), i.e. a machined metal block with regular apertures. Dosimetry in MRT is challenging and previous works still show differences between calculated and experimental dose profiles of 10-30%, which are not acceptable for a clinical implementation of treatment. The interaction of the X-rays with the MSC may contribute to the observed discrepancies; the present study therefore investigates the dose contribution due to radiation interaction with the MSC inner walls and radiation leakage of the MSC. Dose distributions inside a water-equivalent phantom were evaluated for different field sizes and three typical spectra used for MRT studies at the European Synchrotron Biomedical beamline ID17. Film dosimetry was utilized to determine the contribution of radiation interaction with the MSC inner walls; Monte Carlo simulations were implemented to calculate the radiation leakage contribution. Both factors turned out to be relevant for the dose deposition, especially for small fields. Photons interacting with the MSC walls may bring up to 16% more dose in the valley regions, between the microbeams. Depending on the chosen spectrum, the radiation leakage close to the phantom surface can contribute up to 50% of the valley dose for a 5 mm × 5 mm field. The current study underlines that a detailed characterization of the MSC must be performed systematically and accurate MRT dosimetry protocols must include the contribution of radiation leakage and radiation interaction with the MSC in order to avoid significant errors in the dose evaluation at the micrometric scale.
Collapse
Affiliation(s)
- P Pellicioli
- ID17 Biomedical Beamline, ESRF - The European Synchrotron, 71 avenue des Martyrs, Grenoble, France
| | - M Donzelli
- ID17 Biomedical Beamline, ESRF - The European Synchrotron, 71 avenue des Martyrs, Grenoble, France
| | - J A Davis
- School of Physics, University of Wollongong, Wollongong, Australia
| | - F Estève
- STROBE - Synchrotron Radiation for Biomedicine, Grenoble, France
| | - R Hugtenburg
- Swansea University Medical School, Singleton Park, Swansea, United Kingdom
| | - S Guatelli
- School of Physics, University of Wollongong, Wollongong, Australia
| | - M Petasecca
- School of Physics, University of Wollongong, Wollongong, Australia
| | - M L F Lerch
- School of Physics, University of Wollongong, Wollongong, Australia
| | - E Bräuer-Krisch
- ID17 Biomedical Beamline, ESRF - The European Synchrotron, 71 avenue des Martyrs, Grenoble, France
| | - M Krisch
- ID17 Biomedical Beamline, ESRF - The European Synchrotron, 71 avenue des Martyrs, Grenoble, France
| |
Collapse
|
5
|
Day LRJ, Donzelli M, Pellicioli P, Smyth LML, Barnes M, Bartzsch S, Crosbie JC. A commercial treatment planning system with a hybrid dose calculation algorithm for synchrotron radiotherapy trials. Phys Med Biol 2021; 66:055016. [PMID: 33373979 DOI: 10.1088/1361-6560/abd737] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/31/2022]
Abstract
Synchrotron Radiotherapy (SyncRT) is a preclinical radiation treatment which delivers synchrotron x-rays to cancer targets. SyncRT allows for novel treatments such as Microbeam Radiotherapy, which has been shown to have exceptional healthy tissue sparing capabilities while maintaining good tumour control. Veterinary trials in SyncRT are anticipated to take place in the near future at the Australian Synchrotron's Imaging and Medical Beamline (IMBL). However, before veterinary trials can commence, a computerised treatment planning system (TPS) is required, which can quickly and accurately calculate the synchrotron x-ray dose through patient CT images. Furthermore, SyncRT TPS's must be familiar and intuitive to radiotherapy planners in order to alleviate necessary training and reduce user error. We have paired an accurate and fast Monte Carlo (MC) based SyncRT dose calculation algorithm with EclipseTM, the most widely implemented commercial TPS in the clinic. Using EclipseTM, we have performed preliminary SyncRT trials on dog cadavers at the IMBL, and verified calculated doses against dosimetric measurement to within 5% for heterogeneous tissue-equivalent phantoms. We have also performed a validation of the TPS against a full MC simulation for constructed heterogeneous phantoms in EclipseTM, and showed good agreement for a range of water-like tissues to within 5%-8%. Our custom EclipseTM TPS for SyncRT is ready to perform live veterinary trials at the IMBL.
Collapse
Affiliation(s)
- L R J Day
- School of Science, RMIT University, Melbourne, Australia
| | - M Donzelli
- The European Synchrotron Radiation Facility, ID17 Biomedical Beamline, Grenoble, France.,Institute of Cancer Research, London, United Kingdom
| | - P Pellicioli
- The European Synchrotron Radiation Facility, ID17 Biomedical Beamline, Grenoble, France.,Inserm UA7 STROBE, Grenoble Alps University, Grenoble, France.,Swansea University Medical School, Singleton Park, Swansea, United Kingdom
| | - L M L Smyth
- Department of Obstetrics and Gynaecology, University of Melbourne, Royal Women's Hospital, Melbourne, Australia
| | - M Barnes
- School of Science, RMIT University, Melbourne, Australia.,Physical Sciences, Peter MacCallum Cancer Centre, Melbourne, Australia.,The Australian Synchrotron, Imaging and Medical Beamline, Melbourne, Australia
| | - S Bartzsch
- Institute of Cancer Research, London, United Kingdom.,Technical University of Munich, Munich, Germany
| | - J C Crosbie
- School of Science, RMIT University, Melbourne, Australia
| |
Collapse
|