Nacif MS, Turkbey EB, Gai N, Nazarian S, van der Geest RJ, Noureldin RA, Sibley CT, Ugander M, Liu S, Arai AE, Lima JAC, Bluemke DA. Myocardial T1 mapping with MRI: comparison of look-locker and MOLLI sequences.
J Magn Reson Imaging 2011;
34:1367-73. [PMID:
21954119 PMCID:
PMC3221792 DOI:
10.1002/jmri.22753]
[Citation(s) in RCA: 94] [Impact Index Per Article: 7.2] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/02/2011] [Accepted: 07/22/2011] [Indexed: 12/31/2022] Open
Abstract
PURPOSE
To evaluate the relationship between "Look-Locker" (LL) and modified Look-Locker Inversion recovery (MOLLI) approaches for T1 mapping of the myocardium.
MATERIALS AND METHODS
A total of 168 myocardial T1 maps using MOLLI and 165 maps using LL were obtained in human subjects at 1.5 Tesla. The T1 values of the myocardium were calculated before and at five time points after gadolinium administration. All time and heart rate normalizations were done. The T1 values obtained were compared to determine the absolute and bias agreement.
RESULTS
The precontrast global T1 values were similar when measured by the LL and by MOLLI technique (mean, 1004.9 ms ± 120.3 versus 1034.1 ms ± 53.1, respectively, P = 0.26). Postcontrast myocardial T1 time from LL was significantly longer than MOLLI from 5 to 25 min (mean difference, LL - MOLLI was +61.8 ± 46.4 ms, P < 0.001). No significant differences in T1 values were noted between long and short axis measurements for either MOLLI or LL.
CONCLUSION
Postcontrast LL and MOLLI showed very good agreement, although LL values are higher than MOLLI. Precontrast T1 values showed good agreement, however LL has greater limits of agreement. Short and long axis planes can reliably assess T1 values.