1
|
Zwan BJ, Caillet V, Booth JT, Colvill E, Fuangrod T, O'Brien R, Briggs A, O'Connor DJ, Keall PJ, Greer PB. Toward real-time verification for MLC tracking treatments using time-resolved EPID imaging. Med Phys 2021; 48:953-964. [PMID: 33354787 DOI: 10.1002/mp.14675] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/16/2020] [Revised: 11/29/2020] [Accepted: 12/03/2020] [Indexed: 12/25/2022] Open
Abstract
PURPOSE In multileaf collimator (MLC) tracking, the MLC positions from the original treatment plan are continuously modified to account for intrafraction tumor motion. As the treatment is adapted in real time, there is additional risk of delivery errors which cannot be detected using traditional pretreatment dose verification. The purpose of this work is to develop a system for real-time geometric verification of MLC tracking treatments using an electronic portal imaging device (EPID). METHODS MLC tracking was utilized during volumetric modulated arc therapy (VMAT). During these deliveries, treatment beam images were taken at 9.57 frames per second using an EPID and frame grabber computer. MLC positions were extracted from each image frame and used to assess delivery accuracy using three geometric measures: the location, size, and shape of the radiation field. The EPID-measured field location was compared to the tumor motion measured by implanted electromagnetic markers. The size and shape of the beam were compared to the size and shape from the original treatment plan, respectively. This technique was validated by simulating errors in phantom test deliveries and by comparison between EPID measurements and treatment log files. The method was applied offline to images acquired during the LIGHT Stereotactic Ablative Body Radiotherapy (SABR) clinical trial, where MLC tracking was performed for 17 lung cancer patients. The EPID-based verification results were subsequently compared to post-treatment dose reconstruction. RESULTS Simulated field location errors were detected during phantom validation tests with an uncertainty of 0.28 mm (parallel to MLC motion) and 0.38 mm (perpendicular), expressed as a root-mean-square error (RMSError ). For simulated field size errors, the RMSError was 0.47 cm2 and field shape changes were detected for random errors with standard deviation ≥ 2.5 mm. For clinical lung SABR deliveries, field location errors of 1.6 mm (parallel MLC motion) and 4.9 mm (perpendicular) were measured (expressed as a full-width-half-maximum). The mean and standard deviation of the errors in field size and shape were 0.0 ± 0.3 cm2 and 0.3 ± 0.1 (expressed as a translation-invariant normalized RMS). No correlation was observed between geometric errors during each treatment fraction and dosimetric errors in the reconstructed dose to the target volume for this cohort of patients. CONCLUSION A system for real-time delivery verification has been developed for MLC tracking using time-resolved EPID imaging. The technique has been tested offline in phantom-based deliveries and clinical patient deliveries and was used to independently verify the geometric accuracy of the MLC during MLC tracking radiotherapy.
Collapse
Affiliation(s)
- Benjamin J Zwan
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia.,Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia
| | - Vincent Caillet
- Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia.,ACRF Image X Institute, School of Health Sciences, University of Sydney, Sydney, NSW, Australia
| | - Jeremy T Booth
- Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia.,Institute of Medical Physics, School of Physics, University of Sydney, Sydney, Australia
| | - Emma Colvill
- Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia.,ACRF Image X Institute, School of Health Sciences, University of Sydney, Sydney, NSW, Australia
| | - Todsaporn Fuangrod
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia.,Faculty of Medicine and Public Health HRH Princess Chulabhorn College of Medical Science, Chulabhorn Royal Academy, Bangkok, Thailand
| | - Ricky O'Brien
- ACRF Image X Institute, School of Health Sciences, University of Sydney, Sydney, NSW, Australia
| | - Adam Briggs
- Northern Sydney Cancer Centre, Royal North Shore Hospital, St Leonards, NSW, Australia
| | - Daryl J O'Connor
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia
| | - Paul J Keall
- ACRF Image X Institute, School of Health Sciences, University of Sydney, Sydney, NSW, Australia
| | - Peter B Greer
- School of Mathematical and Physical Sciences, University of Newcastle, Newcastle, NSW, Australia.,Department of Radiation Oncology, Calvary Mater Hospital, Newcastle, NSW, Australia
| |
Collapse
|
2
|
Ma Y, Wang X, Mai R, Wang T, Pei Y, Liu S, Guo Y. An electronic portal image device (EPID)-based multiplatform rapid daily LINAC QA tool. J Appl Clin Med Phys 2021; 22:45-58. [PMID: 33410254 PMCID: PMC7856503 DOI: 10.1002/acm2.13055] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/07/2020] [Revised: 08/07/2020] [Accepted: 09/11/2020] [Indexed: 12/17/2022] Open
Abstract
PURPOSE To develop an efficient and economic daily quality research tool (DQRT) for daily check of multiplatform linear accelerators (LINACs) with flattening filter (FF) and flattening filter-free (FFF) photon beams by using an Electronic Portal Image Device (EPID). MATERIALS AND METHODS After EPID calibration, the monitored parameters were analyzed from a 10 cm × 10 cm open and 60° wedge portal images measured by the EPID with 100 MU exposure. Next, the repeatability of the EPID position accuracy, long-term stability, and linearity between image gray value and exposure were verified. Output and beam quality stability of the 6-MV FF and FFF beams measured by DQRT with the introduced setup errors of EPID were also surveyed. Besides, some test results obtained by DQRT were compared with those measured by FC65-G and Matrixx. At last, the tool was evaluated on three LINACs (Synergy, VersaHD, TrueBeam) for 2 months with two popular commercial QA tools as references. RESULTS There are no differences between repeatability tests for all monitored parameters. Image grayscale values obtained by EPID and exposure show good linearity. Either 6 MV FF or FFF photon beam shows minimal impact to the results. The differences between FC65-G, Matrixx and DQRT results are negligible. Monitor results of the two commercial tools are consistent with the DQRT results collected during the 2-month period. CONCLUSION With a shorter time and procedure, the DQRT is useful to daily QA works of LINACs, producing a QA result quality similarly to or more better than the traditional tools and giving richer contents to the QA results. For hospitals with limited QA time window available or lack of funding to purchase commercial QA tools, the proposed DQRT can provide an alternative and economic approach to accomplish the task of daily QA for LINACs.
Collapse
Affiliation(s)
- Yangguang Ma
- Department of Radiation Oncology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China
| | - Xuemin Wang
- Department of Radiotherapy Hospital Unit Radiation Therapy, Shaanxi Provincial Tumor Hospital, Xi'an, China
| | - Rizhen Mai
- Department of Medical Equipment, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China
| | - Tao Wang
- Department of Radiation Oncology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China
| | - Yuntong Pei
- Department of Radiation Oncology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China
| | - Shuaipeng Liu
- Department of Radiation Oncology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China
| | - Yuexin Guo
- Department of Radiation Oncology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China
| |
Collapse
|
3
|
Harris TC, Seco J, Ferguson D, Lehmann M, Huber P, Shi M, Jacobson M, Valencia Lozano I, Myronakis M, Baturin P, Fueglistaller R, Morf D, Berbeco R. Clinical translation of a new flat-panel detector for beam's-eye-view imaging. Phys Med Biol 2020; 65:225004. [PMID: 33284786 DOI: 10.1088/1361-6560/abb571] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/25/2022]
Abstract
Electronic portal imaging devices (EPIDs) lend themselves to beams-eye view clinical applications, such as tumor tracking, but are limited by low contrast and detective quantum efficiency (DQE). We characterize a novel EPID prototype consisting of multiple layers and investigate its suitability for use under clinical conditions. A prototype multi-layer imager (MLI) was constructed utilizing four conventional EPID layers, each consisting of a copper plate, a Gd2O2S:Tb phosphor scintillator, and an amorphous silicon flat panel array detector. We measured the detector's response to a 6 MV photon beam with regards to modulation transfer function, noise power spectrum, DQE, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and the linearity of the detector's response to dose. Additionally, we compared MLI performance to the single top layer of the MLI and the standard Varian AS-1200 detector. Pre-clinical imaging was done on an anthropomorphic phantom, and the detector's CNR, SNR and spatial resolution were assessed in a clinical environment. Images obtained from spine and liver patient treatment deliveries were analyzed to verify CNR and SNR improvements. The MLI has a DQE(0) of 9.7%, about 5.7 times the reference AS-1200 detector. Improved noise performance largely drives the increase. CNR and SNR of clinical images improved three-fold compared to reference. A novel MLI was characterized and prepared for clinical translation. The MLI substantially improved DQE and CNR performance while maintaining the same resolution. Pre-clinical tests on an anthropomorphic phantom demonstrated improved performance as predicted theoretically. Preliminary patient data were analyzed, confirming improved CNR and SNR. Clinical applications are anticipated to include more accurate soft tissue tracking.
Collapse
Affiliation(s)
- T C Harris
- Department of Radiation Oncology, Dana Farber/Brigham and Women's Cancer Center, Harvard Medical school, Boston, MA, United States of America. BioMedical Physics in Radiation Oncology, DKFZ, Heidelberg, Germany. Department of Physics, University of Heidelberg, Heidelberg, Germany
| | | | | | | | | | | | | | | | | | | | | | | | | |
Collapse
|
4
|
Bedford JL, Hanson IM. A method to verify sections of arc during intrafraction portal dosimetry for prostate VMAT. Phys Med Biol 2019; 64:205009. [PMID: 31553964 DOI: 10.1088/1361-6560/ab47c8] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
This study investigates the use of a running sum of images during segment-resolved intrafraction portal dosimetry for volumetric modulated arc therapy (VMAT), so as to alert the operator to an error before it becomes irremediable. At the time of treatment planning, predicted portal images were created for each segment of the VMAT arc, and at the time of delivery, intrafraction monitoring software polled the portal imager to read new images as they became available. The predicted and measured images were compared and displayed on a segment basis. In particular, a running sum of images from ten segments (a 'section') was investigated, with mean absolute difference between predicted and measured images being quantified. Images for 13 prostate patients were used to identify appropriate tolerance values for this statistic. Errors in monitor units of 2%-10%, field size of 2-10 mm, field position of 2-10 mm and path length of 10-50 mm were deliberately introduced into the treatment plans and delivered to a water-equivalent phantom and the sensitivity of the method to these errors was investigated. Gross errors were also considered for one case. The patient images show considerable variability from segment to segment, but when using a section of the arc the variability is reduced, so that the maximum value of mean absolute difference between predicted and measured images is reduced to below 12%, after excluding the first 10% of segments. This tolerance level is also found to be applicable for delivery of the plans to a water-equivalent phantom. Using this as a tolerance level for the error plans, a 10% increase in monitor units is detected, 4 mm increase or shift in multileaf collimator settings can be detected, and an air gap of dimensions 40 mm × 50 mm is detected. Gross errors can also be detected instantly after the first 10% of segments. The running difference between predicted and measured images over ten segments is able to identify errors at specific regions of the arc, as well as in the overall treatment.
Collapse
Affiliation(s)
- James L Bedford
- Joint Department of Physics, The Institute of Cancer Research and The Royal Marsden NHS Foundation Trust, London SM2 5PT, United Kingdom
| | | |
Collapse
|
5
|
Troy Teo P, Guo K, Ahmed B, Alayoubi N, Kehler K, Fontaine G, Sasaki D, Pistorius S. Evaluating a potential technique with local optical flow vectors for automatic organ-at-risk (OAR) intrusion detection and avoidance during radiotherapy. Phys Med Biol 2019; 64:145008. [PMID: 31252423 DOI: 10.1088/1361-6560/ab2db4] [Citation(s) in RCA: 4] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
Various techniques of deep inspiration breath hold (DIBH) have been used to mitigate the likelihood and risk of exposing the heart, an organ-at-risk (OAR) for unintended radiation during left breast radiotherapy. However, issues of reproducibility of these techniques warrant further investigation into the feasibility of detecting the intrusion of an OAR into the treatment field during intra-fractional treatment delivery. The increase of high-dose, low-fraction radiotherapy treatments makes it important to immediately adapt treatment once an OAR is detected in the treatment field. This proof-of-concept implementation includes an algorithm that detects and tracks the motion at the edges of a treatment field and a control algorithm that adapts the treatment aperture according to the motion detected. In accordance to the AAPM Task-Group (TG-132) report, image registration techniques should be verified with virtual and physical phantoms prior to clinical application. Since most OARs move as a result of respiration-induced motion, we have used a lung phantom to generate images of a generic OAR intruding into a treatment field with known velocity. The phantom was programmed to move with sinusoidal and lung patient tumor motion patterns and the accuracy of intrusion tracking and MLC adaptation were benchmarked with the ground truth-programmed motion of the OAR. The motions were recorded with an electronic portal imaging device (EPID). An optimal cluster size of 9 × 9 motion vectors was found to provide the smallest average absolute position error of 0.3 mm. A strong linear correlation between the adapted MLC leaves and the actual OAR position was observed. The algorithm had a mean position tracking error of -0.4 ± 0.3 mm and a precision of 1.1 mm. It is possible to adapt MLC leaves based on the motion detected at the edges of the irradiated field, and it would be feasible to shield an unplanned intrusion of an OAR into the treatment field.
Collapse
Affiliation(s)
- P Troy Teo
- Department of Physics and Astronomy, University of Manitoba, Winnipeg, Manitoba, R3T 2N2, Canada. Author to whom any correspondence should be addressed
| | | | | | | | | | | | | | | |
Collapse
|
6
|
Hiatt J, Mukwada G, Barnes M, Riis HL, Huynh D, Rowshanfarzad P. MLC positioning verification for small fields: a new investigation into automatic EPID-based verification methods. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE 2018; 41:945-955. [PMID: 30259333 DOI: 10.1007/s13246-018-0690-3] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/13/2018] [Accepted: 09/24/2018] [Indexed: 12/31/2022]
Abstract
Multileaf-collimator (MLC) defined small fields in radiotherapy are used in high dose, ultra-conformal techniques such as stereotactic radiotherapy and stereotactic radiosurgery. Proximity to critical structures and irreversible damage arising from inaccurate delivery mean that correct positioning of the MLC system is of the utmost importance. Some of the existing techniques for MLC positioning quality assurance make use of electronic portal imaging device (EPID) images. However, conventional collimation verification algorithms based on the full width at half maximum (FWHM) fail when applied to small field images acquired by an EPID due to overlapping aperture penumbrae, lateral electron disequilibrium and radiation source occlusion. The objective of this study was to investigate sub-pixel edge detection and other techniques with the aim of developing an automatic and autonomous EPID-based method suitable for MLC positional verification of small static fields with arbitrary shapes. Methods investigated included derivative interpolation, Laplacian of Gaussian (LoG) and an algorithm based on the partial area effect hypothesis. None of these methods were found to be suitable for MLC positioning verification in small field conditions. A method is proposed which uses a manufacturer-specific empirically modified FWHM algorithm which shows improvement over the conventional techniques in the small field size range. With a measured mean absolute difference from planned position for Varian linacs of 0.01 ± 0.26 mm, compared with the erroneous FWHM value of 0.70 ± 0.51 mm. For Elekta linacs the proposed algorithm returned 0.26 ± 0.25 mm, in contrast to the FWHM result of 1.79 ± 1.07 mm.
Collapse
Affiliation(s)
- Joshua Hiatt
- Department of Radiation Oncology, Liverpool & Macarthur Cancer Therapy Centres, Liverpool, NSW, 2170, Australia. .,School of Physics, Mathematics and Computing, Faculty of Engineering and Mathematical Sciences, The University of Western Australia, Crawley, WA, 6009, Australia.
| | - Godfrey Mukwada
- Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, 6009, Australia
| | - Michael Barnes
- Department of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, NSW, 2310, Australia.,University of Newcastle, Newcastle, NSW, 2308, Australia
| | | | - Du Huynh
- School of Physics, Mathematics and Computing, Faculty of Engineering and Mathematical Sciences, The University of Western Australia, Crawley, WA, 6009, Australia
| | - Pejman Rowshanfarzad
- School of Physics, Mathematics and Computing, Faculty of Engineering and Mathematical Sciences, The University of Western Australia, Crawley, WA, 6009, Australia
| |
Collapse
|
7
|
Zwan BJ, Hindmarsh J, Seymour E, Kandasamy K, Sloan K, David R, Lee C. The dosimetric impact of control point spacing for sliding gap MLC fields. J Appl Clin Med Phys 2016; 17:204-216. [PMID: 27929494 PMCID: PMC5690523 DOI: 10.1120/jacmp.v17i6.6345] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/29/2016] [Revised: 08/10/2016] [Accepted: 06/20/2016] [Indexed: 11/23/2022] Open
Abstract
Dynamic sliding gap multileaf collimator (MLC) fields are used to model MLC properties within the treatment planning system (TPS) for dynamic treatments. One of the key MLC properties in the Eclipse TPS is the dosimetric leaf gap (DLG) and precise determination of this parameter is paramount to ensuring accurate dose delivery. In this investigation, we report on how the spacing between control points (CPs) for sliding gap fields impacts the dose delivery, MLC positioning accuracy, and measurement of the DLG. The central axis dose was measured for sliding gap MLC fields with gap widths ranging from 2 to 40 mm. It was found that for deliveries containing two CPs, the central axis dose was underestimated by the TPS for all gap widths, with the maximum difference being 8% for a 2 mm gap field. For the same sliding gap fields containing 50 CPs, the measured dose was always within ±2% of the TPS dose. By directly measuring the MLC trajectories we show that this dose difference is due to a systematic MLC gap error for fields containing two CPs, and that the cause of this error is due to the leaf position offset table which is incorrectly applied when the spacing between CPs is too large. This MLC gap error resulted in an increase in the measured DLG of 0.5 mm for both 6 MV and 10 MV, when using fields with 2 CPs compared to 50 CPs. Furthermore, this change in DLG was shown to decrease the mean TPS‐calculated dose to the target volume by 2.6% for a clinical IMRT test plan. This work has shown that systematic MLC positioning errors occur for sliding gap MLC fields containing two CPs and that using these fields to model critical TPS parameters, such as the DLG, may result in clinically significant systematic dose calculation errors during subsequent dynamic MLC treatments. PACS number(s): 87.56.nk
Collapse
|
8
|
Zwan BJ, Barnes MP, Fuangord T, Stanton CJ, O'Connor DJ, Keall PJ, Greer PB. An EPID-based system for gantry-resolved MLC quality assurance for VMAT. J Appl Clin Med Phys 2016; 17:348-365. [PMID: 27685132 PMCID: PMC5874117 DOI: 10.1120/jacmp.v17i5.6312] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/31/2016] [Revised: 06/19/2016] [Accepted: 05/13/2016] [Indexed: 11/23/2022] Open
Abstract
Multileaf collimator (MLC) positions should be precisely and independently mea-sured as a function of gantry angle as part of a comprehensive quality assurance (QA) program for volumetric-modulated arc therapy (VMAT). It is also ideal that such a QA program has the ability to relate MLC positional accuracy to patient-specific dosimetry in order to determine the clinical significance of any detected MLC errors. In this work we propose a method to verify individual MLC trajectories during VMAT deliveries for use as a routine linear accelerator QA tool. We also extend this method to reconstruct the 3D patient dose in the treatment planning sys-tem based on the measured MLC trajectories and the original DICOM plan file. The method relies on extracting MLC positions from EPID images acquired at 8.41fps during clinical VMAT deliveries. A gantry angle is automatically tagged to each image in order to obtain the MLC trajectories as a function of gantry angle. This analysis was performed for six clinical VMAT plans acquired at monthly intervals for three months. The measured trajectories for each delivery were compared to the MLC positions from the DICOM plan file. The maximum mean error detected was 0.07 mm and a maximum root-mean-square error was 0.8 mm for any leaf of any delivery. The sensitivity of this system was characterized by introducing random and systematic MLC errors into the test plans. It was demonstrated that the system is capable of detecting random and systematic errors on the range of 1-2mm and single leaf calibration errors of 0.5 mm. The methodology developed in the work has potential to be used for efficient routine linear accelerator MLC QA and pretreatment patient-specific QA and has the ability to relate measured MLC positional errors to 3D dosimetric errors within a patient volume.
Collapse
|
9
|
Investigation of a real-time EPID-based patient dose monitoring safety system using site-specific control limits. Radiat Oncol 2016; 11:106. [PMID: 27520279 PMCID: PMC4983007 DOI: 10.1186/s13014-016-0682-y] [Citation(s) in RCA: 29] [Impact Index Per Article: 3.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/22/2015] [Accepted: 08/05/2016] [Indexed: 11/24/2022] Open
Abstract
Purpose The aim of this study is to investigate the performance and limitations of a real-time transit electronic portal imaging device (EPID) dosimetry system for error detection during dynamic intensity modulated radiation therapy (IMRT) treatment delivery. Sites studied are prostate, head and neck (HN), and rectal cancer treatments. Methods The system compares measured cumulative transit EPID image frames with predicted cumulative image frames in real-time during treatment using a χ comparison with 4 %, 4 mm criteria. The treatment site-specific thresholds (prostate, HN and rectum IMRT) were determined using initial data collected from 137 patients (274 measured treatment fractions) and a statistical process control methodology. These thresholds were then applied to data from 15 selected patients including 5 prostate, 5 HN, and 5 rectum IMRT treatments for system evaluation and classification of error sources. Results Clinical demonstration of real-time transit EPID dosimetry in IMRT was presented. For error simulation, the system could detect gross errors (i.e. wrong patient, wrong plan, wrong gantry angle) immediately after EPID stabilisation; 2 seconds after the start of treatment. The average rate of error detection was 7.0 % (prostate = 5.6 %, HN= 8.7 % and rectum = 6.7 %). The detected errors were classified as either clinical in origin (e.g. patient anatomical changes), or non-clinical in origin (e.g. detection system errors). Classified errors were 3.2 % clinical and 3.9 % non-clinical. Conclusion An EPID-based real-time error detection method for treatment verification during dynamic IMRT has been developed and tested for its performance and limitations. The system is able to detect gross errors in real-time, however improvement in system robustness is required to reduce the non-clinical sources of error detection.
Collapse
|
10
|
Persoon L, Podesta M, Nijsten S, Troost E, Verhaegen F. Time-Resolved Versus Integrated Transit Planar Dosimetry for Volumetric Modulated Arc Therapy. Technol Cancer Res Treat 2016; 15:NP79-NP87. [PMID: 26655145 DOI: 10.1177/1533034615617668] [Citation(s) in RCA: 21] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/07/2015] [Revised: 09/16/2015] [Accepted: 10/23/2015] [Indexed: 11/16/2022] Open
Abstract
Purpose: It is desirable that dosimetric deviations during radiation treatments are detected. Integrated transit planar dosimetry is commonly used to evaluate external beam treatments such as volumetric-modulated arc therapy. This work focuses on patient geometry changes which result in differences between the planned and the delivered radiation dose. Integrated transit planar dosimetry will average out some deviations. Novel time-resolved transit planar dosimetry compares the delivered dose of volumetric-modulated arc therapy to the planned dose at various time points. Four patient cases are shown where time-resolved transit planar dosimetry detects patient geometry changes during treatment. Methods: A control point to control point comparison between the planned dose and the treatment dose of volumetric-modulated arc therapy beams is calculated using the planning computed tomography and the kV cone-beam computed tomography of the day and evaluated with a time-resolved γ function. Results were computed for 4 patients treated with volumetric-modulated arc therapy, each showing an anatomical change: pleural effusion, rectal gas pockets, and tumor regression. Results: In all cases, the geometrical change was detected by time-resolved transit planar dosimetry, whereas integrated transit planar dosimetry showed minor or no indication of the dose discrepancy. Both tumor regression cases were detected earlier in the treatment with time-resolved planar dosimetry in comparison to integrated transit planar dosimetry. The pleural effusion and the gas pocket were detected exclusively with time-resolved transit planar dosimetry. Conclusions: Clinical cases were presented in this proof-of-principle study in which integrated transit planar dosimetry did not detect dosimetrically relevant deviations to the same extent time-resolved transit planar dosimetry was able to. Time-resolved transit planar dosimetry also provides results that can be presented as a function of arc delivery angle allowing easier interpretation compared to integrated transit planar dosimetry.
Collapse
Affiliation(s)
- L.C.G.G. Persoon
- Department of Radiation Oncology (MAASTRO), GROW—School for Oncology and Developmental Biology, Maastricht University Medical Centre, Maastricht, the Netherlands
| | - M. Podesta
- Department of Radiation Oncology (MAASTRO), GROW—School for Oncology and Developmental Biology, Maastricht University Medical Centre, Maastricht, the Netherlands
| | - S.M.J.J.G. Nijsten
- Department of Radiation Oncology (MAASTRO), GROW—School for Oncology and Developmental Biology, Maastricht University Medical Centre, Maastricht, the Netherlands
| | - E.G.C. Troost
- Department of Radiation Oncology (MAASTRO), GROW—School for Oncology and Developmental Biology, Maastricht University Medical Centre, Maastricht, the Netherlands
| | - F. Verhaegen
- Department of Radiation Oncology (MAASTRO), GROW—School for Oncology and Developmental Biology, Maastricht University Medical Centre, Maastricht, the Netherlands
- Medical Physics Unit, Department of Oncology, McGill University, Montréal, Québec, Canada
| |
Collapse
|
11
|
Visser R, Godart J, Wauben DJL, Langendijk JA, Van't Veld AA, Korevaar EW. Development of an iterative reconstruction method to overcome 2D detector low resolution limitations in MLC leaf position error detection for 3D dose verification in IMRT. Phys Med Biol 2016; 61:3843-56. [PMID: 27100169 DOI: 10.1088/0031-9155/61/10/3843] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
The objective of this study was to introduce a new iterative method to reconstruct multi leaf collimator (MLC) positions based on low resolution ionization detector array measurements and to evaluate its error detection performance. The iterative reconstruction method consists of a fluence model, a detector model and an optimizer. Expected detector response was calculated using a radiotherapy treatment plan in combination with the fluence model and detector model. MLC leaf positions were reconstructed by minimizing differences between expected and measured detector response. The iterative reconstruction method was evaluated for an Elekta SLi with 10.0 mm MLC leafs in combination with the COMPASS system and the MatriXX Evolution (IBA Dosimetry) detector with a spacing of 7.62 mm. The detector was positioned in such a way that each leaf pair of the MLC was aligned with one row of ionization chambers. Known leaf displacements were introduced in various field geometries ranging from -10.0 mm to 10.0 mm. Error detection performance was tested for MLC leaf position dependency relative to the detector position, gantry angle dependency, monitor unit dependency, and for ten clinical intensity modulated radiotherapy (IMRT) treatment beams. For one clinical head and neck IMRT treatment beam, influence of the iterative reconstruction method on existing 3D dose reconstruction artifacts was evaluated. The described iterative reconstruction method was capable of individual MLC leaf position reconstruction with millimeter accuracy, independent of the relative detector position within the range of clinically applied MU's for IMRT. Dose reconstruction artifacts in a clinical IMRT treatment beam were considerably reduced as compared to the current dose verification procedure. The iterative reconstruction method allows high accuracy 3D dose verification by including actual MLC leaf positions reconstructed from low resolution 2D measurements.
Collapse
Affiliation(s)
- R Visser
- Department of Radiation Oncology, University Medical Center Groningen, University of Groningen, Groningen, the Netherlands. Research group Healthy Ageing, Allied Health Care and Nursing, Hanze University of Applied Sciences, Groningen, the Netherlands
| | | | | | | | | | | |
Collapse
|
12
|
Ding A, Xing L, Han B. Development of an accurate EPID-based output measurement and dosimetric verification tool for electron beam therapy. Med Phys 2016; 42:4190-8. [PMID: 26133618 DOI: 10.1118/1.4922400] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/13/2022] Open
Abstract
PURPOSE To develop an efficient and robust tool for output measurement and absolute dose verification of electron beam therapy by using a high spatial-resolution and high frame-rate amorphous silicon flat panel electronic portal imaging device (EPID). METHODS The dosimetric characteristics of the EPID, including saturation, linearity, and ghosting effect, were first investigated on a Varian Clinac 21EX accelerator. The response kernels of the individual pixels of the EPID to all available electron energies (6, 9, 12, 16, and 20 MeV) were calculated by using Monte Carlo (MC) simulations, which formed the basis to deconvolve an EPID raw images to the incident electron fluence map. The two-dimensional (2D) dose distribution at reference depths in water was obtained by using the constructed fluence map with a MC simulated pencil beam kernel with consideration of the geometric and structural information of the EPID. Output factor measurements were carried out with the EPID at a nominal source-surface distance of 100 cm for 2 × 2, 3 × 3, 6 × 6, 10 × 10, and 15 × 15 cm(2) fields for all available electron energies, and the results were compared with that measured in a solid water phantom using film and a Farmer-type ion chamber. The dose distributions at a reference depth specific to each energy and the flatness and symmetry of the 10 × 10 cm(2) electron beam were also measured using EPID, and the results were compared with ion chamber array and water scan measurements. Finally, three patient cases with various field sizes and irregular cutout shapes were also investigated. RESULTS EPID-measured dose changed linearly with the monitor units and showed little ghosting effect for dose rate up to 600 MU/min. The flatness and symmetry measured with the EPID were found to be consistent with ion chamber array and water scan measurements. The EPID-measured output factors for standard square fields of 2 × 2, 3 × 3, 6 × 6, 10 × 10, 15 × 15 cm(2) agreed with film and ion chamber measurements. The average discrepancy between EPID and ion chamber/film measurements was 0.81% ± 0.60% (SD) and 1.34% ± 0.75%, respectively. For the three clinical cases, the difference in output between the EPID- and ion chamber array measured values was found to be 1.13% ± 0.11%, 0.54% ± 0.10%, and 0.74% ± 0.11%, respectively. Furthermore, the γ-index analysis showed an excellent agreement between the EPID- and ion chamber array measured dose distributions: 100% of the pixels passed the criteria of 3%/3 mm. When the γ-index was set to be 2%/2 mm, the pass rate was found to be 99.0% ± 0.07%, 98.2% ± 0.14%, and 100% for the three cases. CONCLUSIONS The EPID dosimetry system developed in this work provides an accurate and reliable tool for routine output measurement and dosimetric verification of electron beam therapy. Coupled with its portability and ease of use, the proposed system promises to replace the current film-based approach for fast and reliable assessment of small and irregular electron field dosimetry.
Collapse
Affiliation(s)
- Aiping Ding
- Department of Radiation Oncology, Stanford University School of Medicine, Stanford, Calilfornia 94305
| | - Lei Xing
- Department of Radiation Oncology, Stanford University School of Medicine, Stanford, Calilfornia 94305
| | - Bin Han
- Department of Radiation Oncology, Stanford University School of Medicine, Stanford, Calilfornia 94305
| |
Collapse
|
13
|
Rowshanfarzad P, Häring P, Riis HL, Zimmermann SJ, Ebert MA. Investigation of the mechanical performance of Siemens linacs components during arc: gantry, MLC, and electronic portal imaging device. MEDICAL DEVICES-EVIDENCE AND RESEARCH 2015; 8:457-66. [PMID: 26604840 PMCID: PMC4640401 DOI: 10.2147/mder.s89725] [Citation(s) in RCA: 5] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/13/2022] Open
Abstract
Background In radiotherapy treatments, it is crucial to monitor the performance of linac components including gantry, collimation system, and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method is suggested in conjunction with an algorithm to investigate the stability of these systems at various gantry angles with the aim of evaluating machine-related errors in treatments. Methods The EPID sag, gantry sag, changes in source-to-detector distance (SDD), EPID and collimator skewness, EPID tilt, and the sag in leaf bank assembly due to linac rotation were separately investigated by acquisition of 37 EPID images of a simple phantom with five ball bearings at various gantry angles. A fast and robust software package was developed for automated analysis of image data. Three Siemens linacs were investigated. Results The average EPID sag was within 1 mm for all tested linacs. Two machines showed >1 mm gantry sag. Changes in the SDD values were within 7.5 mm. EPID skewness and tilt values were <1° in all machines. The maximum sag in leaf bank assembly was <1 mm. Conclusion The method and software developed in this study provide a simple tool for effective investigation of the behavior of Siemens linac components with gantry rotation. Such a comprehensive study has been performed for the first time on Siemens machines.
Collapse
Affiliation(s)
| | - Peter Häring
- German Cancer Research Center (DKFZ), Medical Physics in Radiation Oncology, Heidelberg, Germany
| | - Hans L Riis
- Radiofysisk Laboratorium, Odense University Hospital, Odense C, Denmark
| | - Sune J Zimmermann
- Radiofysisk Laboratorium, Odense University Hospital, Odense C, Denmark
| | - Martin A Ebert
- School of Physics, The University of Western Australia, Crawley, WA, Australia ; Department of Radiation Oncology, Sir Charles Gairdner Hospital, Nedlands, WA, Australia
| |
Collapse
|