1
|
Zhang A, Tharwani K, Wang J, Seilo GK, Atie MA, Potkay JA. Roll-to-roll manufacturing of large surface area PDMS devices, and application to a microfluidic artificial lung. LAB ON A CHIP 2024; 24:4357-4370. [PMID: 39148312 PMCID: PMC11327552 DOI: 10.1039/d4lc00339j] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 04/20/2024] [Accepted: 07/16/2024] [Indexed: 08/17/2024]
Abstract
The ability to cost-effectively produce large surface area microfluidic devices would bring many small-scale technologies such as microfluidic artificial lungs (μALs) from the realm of research to clinical and commercial applications. However, efforts to scale up these devices, such as by stacking multiple flat μALs have been labor intensive and resulted in bulky devices. Here, we report an automated manufacturing system, and a series of cylindrical multi-layer lungs manufactured with the system and tested for fluidic fidelity and function. A roll-to-roll (R2R) system to engrave multiple-layer devices was assembled. Unlike typical applications of R2R, the rolling process is synchronized to achieve consistent radial positioning. This allows the fluidics in the final device to be accessed without being unwrapped. To demonstrate the capabilities of the R2R manufacturing system, this method was used to manufacture multi-layer μALs. Gas and blood are engraved in alternating layers and routed orthogonally to each other. The proximity of gas and blood separated by gas permeable PDMS permits CO2 and O2 exchange via diffusion. After manufacturing, they were evaluated using water for pressure drop and CO2 gas exchange. The best performing device was tested with fresh whole bovine blood for O2 exchange. Three μALs were successfully manufactured and passed leak testing. The top performing device had 15 alternating blood and gas layers. It oxygenated blood from 70% saturation to 95% saturation at a blood flow of 3 mL min-1 and blood side pressure drop of 234 mmHg. This new roll-to-roll manufacturing system is suitable for the automated construction of multi-layer microfluidic devices that are difficult to manufacture by conventional means. With some upgrades and improvements, this technology should allow for the automatic creation of large surface area microfluidic devices that can be employed for various applications including large-scale membrane gas exchange such as clinical-scale microfluidic artificial lungs.
Collapse
Affiliation(s)
- Andrew Zhang
- ECLS Laboratory, Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA.
- Ann Arbor Veteran Affairs Healthcare System, Ann Arbor, MI 48105, USA
| | - Kartik Tharwani
- ECLS Laboratory, Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA.
| | - Jennifer Wang
- ECLS Laboratory, Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA.
| | - Gabriele K Seilo
- ECLS Laboratory, Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA.
| | - Michael A Atie
- ECLS Laboratory, Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA.
- Ann Arbor Veteran Affairs Healthcare System, Ann Arbor, MI 48105, USA
| | - Joseph A Potkay
- ECLS Laboratory, Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA.
- Ann Arbor Veteran Affairs Healthcare System, Ann Arbor, MI 48105, USA
| |
Collapse
|
2
|
Calzuola ST, Newman G, Feaugas T, Perrault CM, Blondé JB, Roy E, Porrini C, Stojanovic GM, Vidic J. Membrane-based microfluidic systems for medical and biological applications. LAB ON A CHIP 2024; 24:3579-3603. [PMID: 38954466 DOI: 10.1039/d4lc00251b] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 07/04/2024]
Abstract
Microfluidic devices with integrated membranes that enable control of mass transport in constrained environments have shown considerable growth over the last decade. Membranes are a key component in several industrial processes such as chemical, pharmaceutical, biotechnological, food, and metallurgy separation processes as well as waste management applications, allowing for modular and compact systems. Moreover, the miniaturization of a process through microfluidic devices leads to process intensification together with reagents, waste and cost reduction, and energy and space savings. The combination of membrane technology and microfluidic devices allows therefore magnification of their respective advantages, providing more valuable solutions not only for industrial processes but also for reproducing biological processes. This review focuses on membrane-based microfluidic devices for biomedical science with an emphasis on microfluidic artificial organs and organs-on-chip. We provide the basic concepts of membrane technology and the laws governing mass transport. The role of the membrane in biomedical microfluidic devices, along with the required properties, available materials, and current challenges are summarized. We believe that the present review may be a starting point and a resource for researchers who aim to replicate a biological phenomenon on-chip by applying membrane technology, for moving forward the biomedical applications.
Collapse
Affiliation(s)
- Silvia Tea Calzuola
- UMR7646 Laboratoire d'hydrodynamique (LadHyX), Ecole Polytechnique, Palaiseau, France.
- Eden Tech, Paris, France
| | - Gwenyth Newman
- Eden Tech, Paris, France
- Department of Medicine and Surgery, Università degli Studi di Milano-Bicocca, Milan, Italy
| | - Thomas Feaugas
- Eden Tech, Paris, France
- Department of Medicine and Surgery, Università degli Studi di Milano-Bicocca, Milan, Italy
| | | | | | | | | | - Goran M Stojanovic
- Faculty of Technical Sciences, University of Novi Sad, T. D. Obradovića 6, 21000 Novi Sad, Serbia
| | - Jasmina Vidic
- Micalis Institute, INRAE, AgroParisTech, Université Paris-Saclay, Jouy-en-Josas, France
| |
Collapse
|
3
|
Saraei N, Dabaghi M, Fusch G, Rochow N, Fusch C, Selvaganapathy PR. Scaled-up Microfluidic Lung Assist Device for Artificial Placenta Application with High Gas Exchange Capacity. ACS Biomater Sci Eng 2024; 10:4612-4625. [PMID: 38904210 DOI: 10.1021/acsbiomaterials.3c01635] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 06/22/2024]
Abstract
Premature neonates with underdeveloped lungs experience respiratory issues and need respiratory support, such as mechanical ventilation or extracorporeal membrane oxygenation (ECMO). The "artificial placenta" (AP) is a noninvasive approach that supports their lungs and reduces respiratory distress, using a pumpless oxygenator connected to the systemic circulation, and can address some of the morbidity issues associated with ECMO. Over the past decade, microfluidic blood oxygenators have garnered significant interest for their ability to mimic physiological conditions and incorporate innovative biomimetic designs. Achieving sufficient gas transfer at a low enough pressure drop for a pumpless operation without requiring a large volume of blood to prime such an oxygenator has been the main challenge with microfluidic lung assist devices (LAD). In this study, we improved the gas exchange capacity of our microfluidic-based artificial placenta-type LAD while reducing its priming volume by using a modified fabrication process that can accommodate large-area thin film microfluidic blood oxygenator (MBO) fabrication with a very high gas exchange surface. Additionally, we demonstrate the effectiveness of a LAD assembled by using these scaled-up MBOs. The LAD based on our artificial placenta concept effectively increases oxygen saturation levels by 30% at a flow rate of 40 mL/min and a pressure drop of 23 mmHg in room air, which is sufficient to support partial oxygenation for 1 kg preterm neonates in respiratory distress. When the gas ambient environment was changed to pure oxygen at atmospheric pressure, the LAD would be able to support premature neonates weighing up to 2 kg. Furthermore, our experiments reveal that the LAD can handle high blood flow rates of up to 150 mL/min and increase oxygen saturation levels by ∼20%, which is equal to an oxygen transfer of 7.48 mL/min in an enriched oxygen environment and among the highest for microfluidic AP type devices. Such performance makes this LAD suitable for providing essential support to 1-2 kg neonates in respiratory distress.
Collapse
Affiliation(s)
| | | | | | - Niels Rochow
- Nuremberg Department of Pediatrics, Paracelsus Medical University, University Hospital, Nuremberg 90419, Germany
| | - Christoph Fusch
- Nuremberg Department of Pediatrics, Paracelsus Medical University, University Hospital, Nuremberg 90419, Germany
| | | |
Collapse
|
4
|
Roberts TR, Persello A, Harea GT, Vedula EM, Isenberg BC, Zang Y, Santos J, Borenstein JT, Batchinsky AI. First 24-Hour-Long Intensive Care Unit Testing of a Clinical-Scale Microfluidic Oxygenator in Swine: A Safety and Feasibility Study. ASAIO J 2024; 70:535-544. [PMID: 38165978 DOI: 10.1097/mat.0000000000002127] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/04/2024] Open
Abstract
Microfluidic membrane oxygenators are designed to mimic branching vasculature of the native lung during extracorporeal lung support. To date, scaling of such devices to achieve clinically relevant blood flow and lung support has been a limitation. We evaluated a novel multilayer microfluidic blood oxygenator (BLOx) capable of supporting 750-800 ml/min blood flow versus a standard hollow fiber membrane oxygenator (HFMO) in vivo during veno-venous extracorporeal life support for 24 hours in anesthetized, mechanically ventilated uninjured swine (n = 3/group). The objective was to assess feasibility, safety, and biocompatibility. Circuits remained patent and operated with stable pressures throughout 24 hours. No group differences in vital signs or evidence of end-organ damage occurred. No change in plasma free hemoglobin and von Willebrand factor multimer size distribution were observed. Platelet count decreased in BLOx at 6 hours (37% dec, P = 0.03), but not in HFMO; however, thrombin generation potential was elevated in HFMO (596 ± 81 nM·min) versus BLOx (323 ± 39 nM·min) at 24 hours ( P = 0.04). Other coagulation and inflammatory mediator results were unremarkable. BLOx required higher mechanical ventilator settings and showed lower gas transfer efficiency versus HFMO, but the stable device performance indicates that this technology is ready for further performance scaling and testing in lung injury models and during longer use conditions.
Collapse
Affiliation(s)
- Teryn R Roberts
- From the Autonomous Reanimation and Evacuation Research Program, The Geneva Foundation, San Antonio, Texas
| | - Antoine Persello
- From the Autonomous Reanimation and Evacuation Research Program, The Geneva Foundation, San Antonio, Texas
| | - George T Harea
- From the Autonomous Reanimation and Evacuation Research Program, The Geneva Foundation, San Antonio, Texas
| | - Else M Vedula
- Bioengineering Division, Draper, Cambridge, Massachusetts
| | | | - Yanyi Zang
- From the Autonomous Reanimation and Evacuation Research Program, The Geneva Foundation, San Antonio, Texas
| | - Jose Santos
- Bioengineering Division, Draper, Cambridge, Massachusetts
| | | | - Andriy I Batchinsky
- From the Autonomous Reanimation and Evacuation Research Program, The Geneva Foundation, San Antonio, Texas
| |
Collapse
|
5
|
Newman G, Leclerc A, Arditi W, Calzuola ST, Feaugas T, Roy E, Perrault CM, Porrini C, Bechelany M. Challenge of material haemocompatibility for microfluidic blood-contacting applications. Front Bioeng Biotechnol 2023; 11:1249753. [PMID: 37662438 PMCID: PMC10469978 DOI: 10.3389/fbioe.2023.1249753] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/29/2023] [Accepted: 08/07/2023] [Indexed: 09/05/2023] Open
Abstract
Biological applications of microfluidics technology is beginning to expand beyond the original focus of diagnostics, analytics and organ-on-chip devices. There is a growing interest in the development of microfluidic devices for therapeutic treatments, such as extra-corporeal haemodialysis and oxygenation. However, the great potential in this area comes with great challenges. Haemocompatibility of materials has long been a concern for blood-contacting medical devices, and microfluidic devices are no exception. The small channel size, high surface area to volume ratio and dynamic conditions integral to microchannels contribute to the blood-material interactions. This review will begin by describing features of microfluidic technology with a focus on blood-contacting applications. Material haemocompatibility will be discussed in the context of interactions with blood components, from the initial absorption of plasma proteins to the activation of cells and factors, and the contribution of these interactions to the coagulation cascade and thrombogenesis. Reference will be made to the testing requirements for medical devices in contact with blood, set out by International Standards in ISO 10993-4. Finally, we will review the techniques for improving microfluidic channel haemocompatibility through material surface modifications-including bioactive and biopassive coatings-and future directions.
Collapse
Affiliation(s)
- Gwenyth Newman
- Department of Medicine and Surgery, Università degli Studi di Milano-Bicocca, Milan, Italy
- Eden Tech, Paris, France
| | - Audrey Leclerc
- Institut Européen des Membranes, IEM, UMR 5635, Univ Montpellier, ENSCM, Centre National de la Recherche Scientifique (CNRS), Place Eugène Bataillon, Montpellier, France
- École Nationale Supérieure des Ingénieurs en Arts Chimiques et Technologiques, Université de Toulouse, Toulouse, France
| | - William Arditi
- Eden Tech, Paris, France
- Centrale Supélec, Gif-sur-Yvette, France
| | - Silvia Tea Calzuola
- Eden Tech, Paris, France
- UMR7648—LadHyx, Ecole Polytechnique, Palaiseau, France
| | - Thomas Feaugas
- Department of Medicine and Surgery, Università degli Studi di Milano-Bicocca, Milan, Italy
- Eden Tech, Paris, France
| | | | | | | | - Mikhael Bechelany
- Institut Européen des Membranes, IEM, UMR 5635, Univ Montpellier, ENSCM, Centre National de la Recherche Scientifique (CNRS), Place Eugène Bataillon, Montpellier, France
- Gulf University for Science and Technology (GUST), Mubarak Al-Abdullah, Kuwait
| |
Collapse
|
6
|
Ma LJ, Akor EA, Thompson AJ, Potkay JA. A Parametric Analysis of Capillary Height in Single-Layer, Small-Scale Microfluidic Artificial Lungs. MICROMACHINES 2022; 13:822. [PMID: 35744436 PMCID: PMC9229210 DOI: 10.3390/mi13060822] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 05/02/2022] [Revised: 05/23/2022] [Accepted: 05/23/2022] [Indexed: 02/04/2023]
Abstract
Microfluidic artificial lungs (μALs) are being investigated for their ability to closely mimic the size scale and cellular environment of natural lungs. Researchers have developed μALs with small artificial capillary diameters (10-50 µm; to increase gas exchange efficiency) and with large capillary diameters (~100 µm; to simplify design and construction). However, no study has directly investigated the impact of capillary height on μAL properties. Here, we use Murray's law and the Hagen-Poiseuille equation to design single-layer, small-scale μALs with capillary heights between 10 and 100 µm. Each µAL contained two blood channel types: capillaries for gas exchange; and distribution channels for delivering blood to/from capillaries. Three designs with capillary heights of 30, 60, and 100 µm were chosen for further modeling, implementation and testing with blood. Flow simulations were used to validate and ensure equal pressures. Designs were fabricated using soft lithography. Gas exchange and pressure drop were tested using whole bovine blood. All three designs exhibited similar pressure drops and gas exchange; however, the μAL with 60 µm tall capillaries had a significantly higher wall shear rate (although physiologic), smaller priming volume and smaller total blood contacting surface area than the 30 and 100 µm designs. Future μAL designs may need to consider the impact of capillary height when optimizing performance.
Collapse
Affiliation(s)
- Lindsay J. Ma
- Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA; (L.J.M.); (E.A.A.); (A.J.T.)
- Veterans Affairs Ann Arbor Healthcare System, Ann Arbor, MI 48109, USA
| | - Emmanuel A. Akor
- Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA; (L.J.M.); (E.A.A.); (A.J.T.)
- Veterans Affairs Ann Arbor Healthcare System, Ann Arbor, MI 48109, USA
| | - Alex J. Thompson
- Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA; (L.J.M.); (E.A.A.); (A.J.T.)
- Veterans Affairs Ann Arbor Healthcare System, Ann Arbor, MI 48109, USA
| | - Joseph A. Potkay
- Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA; (L.J.M.); (E.A.A.); (A.J.T.)
- Veterans Affairs Ann Arbor Healthcare System, Ann Arbor, MI 48109, USA
| |
Collapse
|
7
|
Astor TL, Borenstein JT. The microfluidic artificial lung: Mimicking nature's blood path design to solve the biocompatibility paradox. Artif Organs 2022; 46:1227-1239. [PMID: 35514275 DOI: 10.1111/aor.14266] [Citation(s) in RCA: 5] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/29/2022] [Revised: 04/03/2022] [Accepted: 04/04/2022] [Indexed: 11/28/2022]
Abstract
The increasing prevalence of chronic lung disease worldwide, combined with the emergence of multiple pandemics arising from respiratory viruses over the past century, highlights the need for safer and efficacious means for providing artificial lung support. Mechanical ventilation is currently used for the vast majority of patients suffering from acute and chronic lung failure, but risks further injury or infection to the patient's already compromised lung function. Extracorporeal membrane oxygenation (ECMO) has emerged as a means of providing direct gas exchange with the blood, but limited access to the technology and the complexity of the blood circuit have prevented the broader expansion of its use. A promising avenue toward simplifying and minimizing complications arising from the blood circuit, microfluidics-based artificial organ support, has emerged over the past decade as an opportunity to overcome many of the fundamental limitations of the current standard for ECMO cartridges, hollow fiber membrane oxygenators. The power of microfluidics technology for this application stems from its ability to recapitulate key aspects of physiological microcirculation, including the small dimensions of blood vessel structures and gas transfer membranes. An even greater advantage of microfluidics, the ability to configure blood flow patterns that mimic the smooth, branching nature of vascular networks, holds the potential to reduce the incidence of clotting and bleeding and to minimize reliance on anticoagulants. Here, we summarize recent progress and address future directions and goals for this potentially transformative approach to artificial lung support.
Collapse
Affiliation(s)
- Todd L Astor
- Biomembretics, Inc., Boston, Massachusetts, USA.,Department of Medicine, Massachusetts General Hospital, Boston, Massachusetts, USA
| | | |
Collapse
|
8
|
|
9
|
Lachaux J, Hwang G, Arouche N, Naserian S, Harouri A, Lotito V, Casari C, Lok T, Menager JB, Issard J, Guihaire J, Denis CV, Lenting PJ, Barakat AI, Uzan G, Mercier O, Haghiri-Gosnet AM. A compact integrated microfluidic oxygenator with high gas exchange efficiency and compatibility for long-lasting endothelialization. LAB ON A CHIP 2021; 21:4791-4804. [PMID: 34309615 DOI: 10.1039/d1lc00356a] [Citation(s) in RCA: 10] [Impact Index Per Article: 3.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/13/2023]
Abstract
We have developed and tested a novel microfluidic device for blood oxygenation, which exhibits a large surface area of gas exchange and can support long-term sustainable endothelialization of blood microcapillaries, enhancing its hemocompatibility for clinical applications. The architecture of the parallel stacking of the trilayers is based on a central injection for blood and a lateral injection/output for gas which allows significant reduction in shear stress, promoting sustainable endothelialization since cells can be maintained viable for up to 2 weeks after initial seeding in the blood microchannel network. The circular design of curved blood capillaries allows covering a maximal surface area at 4 inch wafer scale, producing high oxygen uptake and carbon dioxide release in each single unit. Since the conventional bonding process based on oxygen plasma cannot be used for surface areas larger than several cm2, a new "wet bonding" process based on soft microprinting has been developed and patented. Using this new protocol, each 4 inch trilayer unit can be sealed without a collapsed membrane even at reduced 15 μm thickness and can support a high blood flow rate. The height of the blood channels has been optimized to reduce pressure drop and enhance gas exchange at a high volumetric blood flow rate up to 15 ml min-1. The simplicity of connecting different units in the stacked architecture is demonstrated for 3- or 5-unit stacked devices that exhibit remarkable performance with low primary volume, high oxygen uptake and carbon dioxide release and high flow rate of up to 80 ml min-1.
Collapse
Affiliation(s)
- Julie Lachaux
- Université Paris-Saclay, CNRS, Centre de Nanosciences et Nanotechnologies C2N, UMR9001, Palaiseau 91120, France.
| | - Gilgueng Hwang
- Université Paris-Saclay, CNRS, Centre de Nanosciences et Nanotechnologies C2N, UMR9001, Palaiseau 91120, France.
| | - Nassim Arouche
- Université Paris-Saclay, INSERM, UMR-S-MD 1197, Hôpital Paul Brousse, Villejuif, France
| | - Sina Naserian
- Université Paris-Saclay, INSERM, UMR-S-MD 1197, Hôpital Paul Brousse, Villejuif, France
| | - Abdelmounaim Harouri
- Université Paris-Saclay, CNRS, Centre de Nanosciences et Nanotechnologies C2N, UMR9001, Palaiseau 91120, France.
| | - Valeria Lotito
- Université Paris-Saclay, CNRS, Centre de Nanosciences et Nanotechnologies C2N, UMR9001, Palaiseau 91120, France.
| | - Caterina Casari
- Université Paris-Saclay, INSERM, UMR S1176, Le Kremin-Bicêtre, France
| | - Thevy Lok
- LadHyX, CNRS, Ecole polytechnique, Institut polytechnique de Paris, Palaiseau 91120, France
| | - Jean Baptiste Menager
- Université Paris-Saclay, INSERM UMR_S 999 "Pulmonary Hypertension: Pathophysiology and Novel Therapies", Hôpital Marie Lannelongue, Le Plessis-Robinson, France
| | - Justin Issard
- Université Paris-Saclay, INSERM UMR_S 999 "Pulmonary Hypertension: Pathophysiology and Novel Therapies", Hôpital Marie Lannelongue, Le Plessis-Robinson, France
| | - Julien Guihaire
- Université Paris-Saclay, INSERM UMR_S 999 "Pulmonary Hypertension: Pathophysiology and Novel Therapies", Hôpital Marie Lannelongue, Le Plessis-Robinson, France
| | - Cécile V Denis
- Université Paris-Saclay, INSERM, UMR S1176, Le Kremin-Bicêtre, France
| | - Peter J Lenting
- Université Paris-Saclay, INSERM, UMR S1176, Le Kremin-Bicêtre, France
| | - Abdul I Barakat
- LadHyX, CNRS, Ecole polytechnique, Institut polytechnique de Paris, Palaiseau 91120, France
| | - Georges Uzan
- Université Paris-Saclay, INSERM, UMR-S-MD 1197, Hôpital Paul Brousse, Villejuif, France
| | - Olaf Mercier
- Université Paris-Saclay, INSERM UMR_S 999 "Pulmonary Hypertension: Pathophysiology and Novel Therapies", Hôpital Marie Lannelongue, Le Plessis-Robinson, France
| | - Anne-Marie Haghiri-Gosnet
- Université Paris-Saclay, CNRS, Centre de Nanosciences et Nanotechnologies C2N, UMR9001, Palaiseau 91120, France.
| |
Collapse
|
10
|
Santos JA, Gimbel AA, Peppas A, Truslow JG, Lang DA, Sukavaneshvar S, Solt D, Mulhern TJ, Markoski A, Kim ES, Hsiao JCM, Lewis DJ, Harjes DI, DiBiasio C, Charest JL, Borenstein JT. Design and construction of three-dimensional physiologically-based vascular branching networks for respiratory assist devices. LAB ON A CHIP 2021; 21:4637-4651. [PMID: 34730597 DOI: 10.1039/d1lc00287b] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/13/2023]
Abstract
Microfluidic lab-on-a-chip devices are changing the way that in vitro diagnostics and drug development are conducted, based on the increased precision, miniaturization and efficiency of these systems relative to prior methods. However, the full potential of microfluidics as a platform for therapeutic medical devices such as extracorporeal organ support has not been realized, in part due to limitations in the ability to scale current designs and fabrication techniques toward clinically relevant rates of blood flow. Here we report on a method for designing and fabricating microfluidic devices supporting blood flow rates per layer greater than 10 mL min-1 for respiratory support applications, leveraging advances in precision machining to generate fully three-dimensional physiologically-based branching microchannel networks. The ability of precision machining to create molds with rounded features and smoothly varying channel widths and depths distinguishes the geometry of the microchannel networks described here from all previous reports of microfluidic respiratory assist devices, regarding the ability to mimic vascular blood flow patterns. These devices have been assembled and tested in the laboratory using whole bovine or porcine blood, and in a porcine model to demonstrate efficient gas transfer, blood flow and pressure stability over periods of several hours. This new approach to fabricating and scaling microfluidic devices has the potential to address wide applications in critical care for end-stage organ failure and acute illnesses stemming from respiratory viral infections, traumatic injuries and sepsis.
Collapse
Affiliation(s)
- Jose A Santos
- Bioengineering Division, Draper, Cambridge, MA, USA.
| | - Alla A Gimbel
- Bioengineering Division, Draper, Cambridge, MA, USA.
| | | | | | - Daniel A Lang
- Bioengineering Division, Draper, Cambridge, MA, USA.
| | | | | | | | - Alex Markoski
- Bioengineering Division, Draper, Cambridge, MA, USA.
| | - Ernest S Kim
- Bioengineering Division, Draper, Cambridge, MA, USA.
| | | | - Diana J Lewis
- Bioengineering Division, Draper, Cambridge, MA, USA.
| | | | | | | | | |
Collapse
|
11
|
Santos J, Vedula EM, Lai W, Isenberg BC, Lewis DJ, Lang D, Sutherland D, Roberts TR, Harea GT, Wells C, Teece B, Karandikar P, Urban J, Risoleo T, Gimbel A, Solt D, Leazer S, Chung KK, Sukavaneshvar S, Batchinsky AI, Borenstein JT. Toward Development of a Higher Flow Rate Hemocompatible Biomimetic Microfluidic Blood Oxygenator. MICROMACHINES 2021; 12:888. [PMID: 34442512 PMCID: PMC8398684 DOI: 10.3390/mi12080888] [Citation(s) in RCA: 8] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 05/29/2021] [Revised: 07/18/2021] [Accepted: 07/24/2021] [Indexed: 01/05/2023]
Abstract
The recent emergence of microfluidic extracorporeal lung support technologies presents an opportunity to achieve high gas transfer efficiency and improved hemocompatibility relative to the current standard of care in extracorporeal membrane oxygenation (ECMO). However, a critical challenge in the field is the ability to scale these devices to clinically relevant blood flow rates, in part because the typically very low blood flow in a single layer of a microfluidic oxygenator device requires stacking of a logistically challenging number of layers. We have developed biomimetic microfluidic oxygenators for the past decade and report here on the development of a high-flow (30 mL/min) single-layer prototype, scalable to larger structures via stacking and assembly with blood distribution manifolds. Microfluidic oxygenators were designed with biomimetic in-layer blood distribution manifolds and arrays of parallel transfer channels, and were fabricated using high precision machined durable metal master molds and microreplication with silicone films, resulting in large area gas transfer devices. Oxygen transfer was evaluated by flowing 100% O2 at 100 mL/min and blood at 0-30 mL/min while monitoring increases in O2 partial pressures in the blood. This design resulted in an oxygen saturation increase from 65% to 95% at 20 mL/min and operation up to 30 mL/min in multiple devices, the highest value yet recorded in a single layer microfluidic device. In addition to evaluation of the device for blood oxygenation, a 6-h in vitro hemocompatibility test was conducted on devices (n = 5) at a 25 mL/min blood flow rate with heparinized swine donor blood against control circuits (n = 3). Initial hemocompatibility results indicate that this technology has the potential to benefit future applications in extracorporeal lung support technologies for acute lung injury.
Collapse
Affiliation(s)
- Jose Santos
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Else M. Vedula
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Weixuan Lai
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Brett C. Isenberg
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Diana J. Lewis
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Dan Lang
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - David Sutherland
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Teryn R. Roberts
- Autonomous Reanimation and Evacuation (AREVA) Research Program, The Geneva Foundation, Brooks City Base, San Antonio, TX 78006, USA; (T.R.R.); (G.T.H.); (A.I.B.)
| | - George T. Harea
- Autonomous Reanimation and Evacuation (AREVA) Research Program, The Geneva Foundation, Brooks City Base, San Antonio, TX 78006, USA; (T.R.R.); (G.T.H.); (A.I.B.)
| | - Christian Wells
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Bryan Teece
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Paramesh Karandikar
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Joseph Urban
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Thomas Risoleo
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Alla Gimbel
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| | - Derek Solt
- Thrombodyne, Inc., Salt Lake City, UT 84103, USA; (D.S.); (S.S.)
| | - Sahar Leazer
- Department of Medicine, F. Edward Hebert School of Medicine, Uniformed Services University of the Health Sciences, Bethesda, MD 20814, USA; (S.L.); (K.K.C.)
| | - Kevin K. Chung
- Department of Medicine, F. Edward Hebert School of Medicine, Uniformed Services University of the Health Sciences, Bethesda, MD 20814, USA; (S.L.); (K.K.C.)
| | | | - Andriy I. Batchinsky
- Autonomous Reanimation and Evacuation (AREVA) Research Program, The Geneva Foundation, Brooks City Base, San Antonio, TX 78006, USA; (T.R.R.); (G.T.H.); (A.I.B.)
| | - Jeffrey T. Borenstein
- Draper, Cambridge, MA 02139, USA; (J.S.); (W.L.); (B.C.I.); (D.J.L.); (D.L.); (D.S.); (C.W.); (B.T.); (P.K.); (J.U.); (T.R.); (A.G.)
| |
Collapse
|
12
|
Wang L, Li F, Feng Z, Shi Y. Comment on "A Pumpless Microfluidic Neonatal Lung Assist Device for Support of Preterm Neonates in Respiratory Distress". ADVANCED SCIENCE (WEINHEIM, BADEN-WURTTEMBERG, GERMANY) 2021; 8:2004382. [PMID: 34194930 PMCID: PMC8224411 DOI: 10.1002/advs.202004382] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 11/14/2020] [Revised: 12/24/2020] [Indexed: 06/13/2023]
Affiliation(s)
- Li Wang
- Department of Pediatrics University-Town Hospital of Chongqing Medical University Chongqing 401331 China
- Department of Pediatrics Daping Hospital Army Medical University Chongqing 400042 China
| | - Fang Li
- Department of Neonatology Ministry of Education Key Laboratory of Child Development and Disorders National Clinical Research Center for Child Health and Disorders China International Science and Technology Cooperation Base of Child Development and Critical Disorders Children's Hospital of Chongqing Medical University Chongqing Key Laboratory of Pediatrics Chongqing 400014 China
| | - Zhichun Feng
- Affiliated BaYi Children's Hospital General Hospital of the People's Liberation Army Beijing 100007 China
| | - Yuan Shi
- Department of Neonatology Ministry of Education Key Laboratory of Child Development and Disorders National Clinical Research Center for Child Health and Disorders China International Science and Technology Cooperation Base of Child Development and Critical Disorders Children's Hospital of Chongqing Medical University Chongqing Key Laboratory of Pediatrics Chongqing 400014 China
| |
Collapse
|
13
|
Gimbel AA, Hsiao JC, Kim ES, Lewis DJ, Risoleo TF, Urban JN, Borenstein JT. A high gas transfer efficiency microfluidic oxygenator for extracorporeal respiratory assist applications in critical care medicine. Artif Organs 2021; 45:E247-E264. [PMID: 33561881 DOI: 10.1111/aor.13935] [Citation(s) in RCA: 11] [Impact Index Per Article: 3.7] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/30/2020] [Revised: 01/10/2021] [Accepted: 02/05/2021] [Indexed: 12/15/2022]
Abstract
Advances in microfluidics technologies have spurred the development of a new generation of microfluidic respiratory assist devices, constructed using microfabrication techniques capable of producing microchannel dimensions similar to those found in human capillaries and gas transfer films in the same thickness range as the alveolar membrane. These devices have been tested in laboratory settings and in some cases in extracorporeal animal experiments, yet none have been advanced to human clinical studies. A major challenge in the development of microfluidic oxygenators is the difficulty in scaling the technology toward high blood flows necessary to support adult humans; such scaling efforts are often limited by the complexity of the fabrication process and the manner in which blood is distributed in a three-dimensional network of microchannels. Conceptually, a central advantage of microfluidic oxygenators over existing hollow-fiber membrane-based configurations is the potential for shallower channels and thinner gas transfer membranes, features that reduce oxygen diffusion distances, to result in a higher gas transfer efficiency defined as the ratio of the volume of oxygen transferred to the blood per unit time to the active surface area of the gas transfer membrane. If this ratio is not significantly higher than values reported for hollow fiber membrane oxygenators (HFMO), then the expected advantage of the microfluidic approach would not be realized in practice, potentially due to challenges encountered in blood distribution strategies when scaling microfluidic designs to higher flow rates. Here, we report on scaling of a microfluidic oxygenator design from 4 to 92 mL/min blood flow, within an order of magnitude of the flow rate required for neonatal applications. This scaled device is shown to have a gas transfer efficiency higher than any other reported system in the literature, including other microfluidic prototypes and commercial HFMO cartridges. While the high oxygen transfer efficiency is a promising advance toward clinical scaling of a microfluidic architecture, it is accompanied by an excessive blood pressure drop in the circuit, arising from a combination of shallow gas transfer channels and equally shallow distribution manifolds. Therefore, next-generation microfluidic oxygenators will require novel design and fabrication strategies to minimize pressure drops while maintaining very high oxygen transfer efficiencies.
Collapse
Affiliation(s)
| | | | - Ernest S Kim
- Bioengineering Division, Draper, Cambridge, MA, USA
| | | | | | | | | |
Collapse
|
14
|
Dabaghi M, Rochow N, Saraei N, Fusch G, Monkman S, Da K, Shahin‐Shamsabadi A, Brash JL, Predescu D, Delaney K, Fusch C, Selvaganapathy PR. A Pumpless Microfluidic Neonatal Lung Assist Device for Support of Preterm Neonates in Respiratory Distress. ADVANCED SCIENCE (WEINHEIM, BADEN-WURTTEMBERG, GERMANY) 2020; 7:2001860. [PMID: 33173732 PMCID: PMC7610273 DOI: 10.1002/advs.202001860] [Citation(s) in RCA: 11] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/18/2020] [Revised: 07/16/2020] [Indexed: 05/19/2023]
Abstract
Premature neonates suffer from respiratory morbidity as their lungs are immature, and current supportive treatment such as mechanical ventilation or extracorporeal membrane oxygenation causes iatrogenic injuries. A non-invasive and biomimetic concept known as the "artificial placenta" (AP) would be beneficial to overcome complications associated with the current respiratory support of preterm infants. Here, a pumpless oxygenator connected to the systemic circulation supports the lung function to relieve respiratory distress. In this paper, the first successful operation of a microfluidic, artificial placenta type neonatal lung assist device (LAD) on a newborn piglet model, which is the closest representation of preterm human infants, is demonstrated. This LAD has high oxygenation capability in both pure oxygen and room air as the sweep gas. The respiratory distress that the newborn piglet is put under during experimentation, repeatedly and over a significant duration of time, is able to be relieved. These findings indicate that this LAD has a potential application as a biomimetic artificial placenta to support the respiratory needs of preterm neonates.
Collapse
Affiliation(s)
| | - Niels Rochow
- Department of PediatricsMcMaster UniversityHamiltonONCanada
- Paracelsus Medical UniversityDepartment of PediatricsUniversity Hospital NurembergNurembergGermany
| | - Neda Saraei
- Department of Mechanical EngineeringMcMaster UniversityHamiltonONCanada
| | - Gerhard Fusch
- Department of PediatricsMcMaster UniversityHamiltonONCanada
| | | | - Kevin Da
- Department of Chemical EngineeringMcMaster UniversityHamiltonONCanada
| | | | - John L. Brash
- School of Biomedical EngineeringMcMaster UniversityHamiltonONCanada
- Department of Chemical EngineeringMcMaster UniversityHamiltonONCanada
| | | | - Kathleen Delaney
- Central Animal Facility DepartmentMcMaster UniversityHamiltonONCanada
| | - Christoph Fusch
- School of Biomedical EngineeringMcMaster UniversityHamiltonONCanada
- Department of PediatricsMcMaster UniversityHamiltonONCanada
- Paracelsus Medical UniversityDepartment of PediatricsUniversity Hospital NurembergNurembergGermany
| | - P. Ravi Selvaganapathy
- School of Biomedical EngineeringMcMaster UniversityHamiltonONCanada
- Department of Mechanical EngineeringMcMaster UniversityHamiltonONCanada
| |
Collapse
|
15
|
Swol J, Shigemura N, Ichiba S, Steinseifer U, Anraku M, Lorusso R. Artificial lungs--Where are we going with the lung replacement therapy? Artif Organs 2020; 44:1135-1149. [PMID: 33098217 DOI: 10.1111/aor.13801] [Citation(s) in RCA: 4] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/26/2020] [Revised: 08/03/2020] [Accepted: 08/04/2020] [Indexed: 12/12/2022]
Abstract
Lung transplantation may be a final destination therapy in lung failure, but limited donor organ availability creates a need for alternative management, including artificial lung technology. This invited review discusses ongoing developments and future research pathways for respiratory assist devices and tissue engineering to treat advanced and refractory lung disease. An overview is also given on the aftermath of the coronavirus disease 2019 pandemic and lessons learned as the world comes out of this situation. The first order of business in the future of lung support is solving the problems with existing mechanical devices. Interestingly, challenges identified during the early days of development persist today. These challenges include device-related infection, bleeding, thrombosis, cost, and patient quality of life. The main approaches of the future directions are to repair, restore, replace, or regenerate the lungs. Engineering improvements to hollow fiber membrane gas exchangers are enabling longer term wearable systems and can be used to bridge lung failure patients to transplantation. Progress in the development of microchannel-based devices has provided the concept of biomimetic devices that may even enable intracorporeal implantation. Tissue engineering and cell-based technologies have provided the concept of bioartificial lungs with properties similar to the native organ. Recent progress in artificial lung technologies includes continued advances in both engineering and biology. The final goal is to achieve a truly implantable and durable artificial lung that is applicable to destination therapy.
Collapse
Affiliation(s)
- Justyna Swol
- Department of Respiratory Medicine, Allergology and Sleep Medicine, Intensive Care Medicine, Paracelsus Medical University Nuremberg, General Hospital Nuremberg, Nuremberg, Germany
| | - Norihisa Shigemura
- Division of Cardiovascular Surgery, Temple University Health System Inc., Philadelphia, PA, USA
| | - Shingo Ichiba
- Department of Surgical Intensive Care Medicine, Nippon Medical School Hospital, Bunkyo-ku, Japan
| | - Ulrich Steinseifer
- Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Aachen, Germany
| | - Masaki Anraku
- Department of Thoracic Surgery, The University of Tokyo Graduate School of Medicine Faculty of Medicine, Bunkyo-ku, Japan
| | - Roberto Lorusso
- Cardio-Thoracic Surgery Department - Heart & Vascular Centre, Maastricht University Medical Hospital, Maastricht, The Netherlands
| |
Collapse
|
16
|
De Bie FR, Davey MG, Larson AC, Deprest J, Flake AW. Artificial placenta and womb technology: Past, current, and future challenges towards clinical translation. Prenat Diagn 2020; 41:145-158. [PMID: 32875581 DOI: 10.1002/pd.5821] [Citation(s) in RCA: 33] [Impact Index Per Article: 8.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/01/2020] [Revised: 08/24/2020] [Accepted: 08/29/2020] [Indexed: 01/12/2023]
Abstract
Extreme prematurity remains a major cause of neonatal mortality and severe long-term morbidity. Current neonatal care is associated with significant morbidity due to iatrogenic injury and developmental immaturity of extreme premature infants. A more physiologic approach, replacing placental function and providing a womb-like environment, is the foundational principle of artificial placenta (AP) and womb (AW) technology. The concept has been studied during the past 60 years with limited success. However, recent technological advancements and a greater emphasis on mimicking utero-placental physiology have improved the success of experimental models, bringing the technology closer to clinical translation. Here, we review the rationale for and history of AP and AW technology, discuss the challenges that needed to be overcome, and compare recent successful models. We conclude by outlining some remaining challenges to be addressed on the path towards clinical translation and opportunities for future research.
Collapse
Affiliation(s)
- Felix R De Bie
- Center for Fetal Research, Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA.,Department of Development and Regeneration, KU Leuven, Leuven, Belgium
| | - Marcus G Davey
- Center for Fetal Research, Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA
| | - Abby C Larson
- Center for Fetal Research, Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA
| | - Jan Deprest
- Department of Development and Regeneration, KU Leuven, Leuven, Belgium
| | - Alan W Flake
- Center for Fetal Research, Children's Hospital of Philadelphia, Philadelphia, Pennsylvania, USA
| |
Collapse
|
17
|
Thompson AJ, Ma LJ, Major T, Jeakle M, Lautner-Csorba O, Goudie MJ, Handa H, Rojas-Peña A, Potkay JA. Assessing and improving the biocompatibility of microfluidic artificial lungs. Acta Biomater 2020; 112:190-201. [PMID: 32434076 PMCID: PMC10168296 DOI: 10.1016/j.actbio.2020.05.008] [Citation(s) in RCA: 12] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/04/2020] [Revised: 05/05/2020] [Accepted: 05/07/2020] [Indexed: 02/08/2023]
Abstract
Microfluidic artificial lungs (µALs) have the potential to improve the treatment and quality of life for patients with acute or chronic lung injury. In order to realize the full potential of this technology (including as a destination therapy), the biocompatibility of these devices needs to be improved to produce long-lasting devices that are safe for patient use with minimal or no systemic anticoagulation. Many studies exist which probe coagulation and thrombosis on polydimethyl siloxane (PDMS) surfaces, and many strategies have been explored to improve surface biocompatibility. As the field of µALs is young, there are few studies which investigate biocompatibility of functioning µALs; and even fewer which were performed in vivo. Here, we use both in vitro and in vivo models to investigate two strategies to improve µAL biocompatibility: 1) a hydrophilic surface coating (polyethylene glycol, PEG) to prevent surface fouling, and 2) the addition of nitric oxide (NO) to the sweep gas to inhibit platelet activation locally within the µAL. In this study, we challenge µALs with clottable blood or platelet-rich plasma (PRP) and monitor the resistance to blood flow over time. Device lifetime (the amount of time the µAL remains patent and unobstructed by clot) is used as the primary indicator of biocompatibility. This study is the first study to: 1) investigate the effect of NO release on biocompatibility in a microfluidic network; 2) combine a hydrophilic PEG coating with NO release to improve blood compatibility; and 3) perform extended in vivo biocompatibility testing of a µAL. We found that µALs challenged in vitro with PRP remained patent significantly longer when the sweep gas contained NO than without NO. In the in vivo rabbit model, neither approach alone (PEG coating nor NO sweep gas) significantly improved biocompatibility compared to controls (though with larger sample size significance may become apparent); while the combination of a PEG coating with NO sweep gas resulted in significant improvement of device lifetime. STATEMENT OF SIGNIFICANCE: The development of microfluidic artificial lungs (µALs) can potentially have a massive impact on the treatment of patients with acute and chronic lung impairments. Before these devices can be deployed clinically, the biocompatibility of µALs must be improved and more comprehensively understood. This work explores two strategies for improving biocompatibility, a hydrophilic surface coating (polyethylene glycol) for general surface passivation and the addition of nitric oxide (NO) to the sweep gas to quell platelet and leukocyte activation. These two strategies are investigated separately and as a combined device treatment. Devices are challenged with clottable blood using in vitro testing and in vivo testing in rabbits. This is the first study to our knowledge that allows statistical comparisons of biocompatible µALs in animals, a key step towards eventual clinical use.
Collapse
Affiliation(s)
- Alex J Thompson
- VA Ann Arbor Healthcare System, 2215 Fuller Road, Ann Arbor, MI, USA, 48105; University of Michigan, 1150 W. Medical Center Drive, Ann Arbor, MI, USA, 48109.
| | - Lindsay J Ma
- VA Ann Arbor Healthcare System, 2215 Fuller Road, Ann Arbor, MI, USA, 48105; University of Michigan, 1150 W. Medical Center Drive, Ann Arbor, MI, USA, 48109
| | - Terry Major
- University of Michigan, 1150 W. Medical Center Drive, Ann Arbor, MI, USA, 48109
| | - Mark Jeakle
- University of Michigan, 1150 W. Medical Center Drive, Ann Arbor, MI, USA, 48109
| | | | - Marcus J Goudie
- University of Georgia, College of Engineering, 220 Riverbend Road, Athens, GA, USA, 30602
| | - Hitesh Handa
- University of Georgia, College of Engineering, 220 Riverbend Road, Athens, GA, USA, 30602
| | - Alvaro Rojas-Peña
- University of Michigan, 1150 W. Medical Center Drive, Ann Arbor, MI, USA, 48109
| | - Joseph A Potkay
- VA Ann Arbor Healthcare System, 2215 Fuller Road, Ann Arbor, MI, USA, 48105; University of Michigan, 1150 W. Medical Center Drive, Ann Arbor, MI, USA, 48109
| |
Collapse
|
18
|
Dabaghi M, Saraei N, Fusch G, Rochow N, Brash JL, Fusch C, Selvaganapathy PR. Microfluidic blood oxygenators with integrated hollow chambers for enhanced air exchange from all four sides. J Memb Sci 2020. [DOI: 10.1016/j.memsci.2019.117741] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/22/2022]
|
19
|
Evseev AK, Zhuravel SV, Alentiev AY, Goroncharovskaya IV, Petrikov SS. Membranes in Extracorporeal Blood Oxygenation Technology. MEMBRANES AND MEMBRANE TECHNOLOGIES 2019. [DOI: 10.1134/s2517751619040024] [Citation(s) in RCA: 21] [Impact Index Per Article: 4.2] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/23/2022]
|
20
|
Lakshmanan HHS, Shatzel JJ, Olson SR, McCarty OJT, Maddala J. Modeling the effect of blood vessel bifurcation ratio on occlusive thrombus formation. Comput Methods Biomech Biomed Engin 2019; 22:972-980. [PMID: 31066295 DOI: 10.1080/10255842.2019.1610744] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/21/2022]
Abstract
Vascular geometry is a major determinant of the hemodynamics that promote or prevent unnecessary vessel occlusion from thrombus formation. Bifurcations in the vascular geometry are repeating structures that introduce flow separation between parent and daughter vessels. We modelled the blood flow and shear rate in a bifurcation during thrombus formation and show that blood vessel bifurcation ratios determine the maximum shear rate on the surface of a growing thrombus. We built an analytical model that may aid in predicting microvascular bifurcation ratios that are prone to occlusive thrombus formation. We also observed that bifurcation ratios that adhere to Murray's law of bifurcations may be protected from occlusive thrombus formation. These results may be useful in the rational design of diagnostic microfluidic devices and microfluidic blood oxygenators.
Collapse
Affiliation(s)
- Hari Hara Sudhan Lakshmanan
- a Biomedical Engineering, School of Medicine , Oregon Health & Science University , Portland , OR , USA.,b Chemical and Biomedical Engineering , West Virginia University , Morgantown , WV , USA
| | - Joseph J Shatzel
- a Biomedical Engineering, School of Medicine , Oregon Health & Science University , Portland , OR , USA.,c Division of Hematology-Oncology, School of Medicine , Oregon Health & Science University , Portland , OR , USA
| | - Sven R Olson
- c Division of Hematology-Oncology, School of Medicine , Oregon Health & Science University , Portland , OR , USA
| | - Owen J T McCarty
- a Biomedical Engineering, School of Medicine , Oregon Health & Science University , Portland , OR , USA.,c Division of Hematology-Oncology, School of Medicine , Oregon Health & Science University , Portland , OR , USA
| | - Jeevan Maddala
- a Biomedical Engineering, School of Medicine , Oregon Health & Science University , Portland , OR , USA.,b Chemical and Biomedical Engineering , West Virginia University , Morgantown , WV , USA
| |
Collapse
|
21
|
Dabaghi M, Saraei N, Fusch G, Rochow N, Brash JL, Fusch C, Ravi Selvaganapathy P. An ultra-thin, all PDMS-based microfluidic lung assist device with high oxygenation capacity. BIOMICROFLUIDICS 2019; 13:034116. [PMID: 31263515 PMCID: PMC6597343 DOI: 10.1063/1.5091492] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 02/02/2019] [Accepted: 06/11/2019] [Indexed: 05/06/2023]
Abstract
Preterm neonates with immature lungs require a lung assist device (LAD) to maintain oxygen saturation at normal levels. Over the last decade, microfluidic blood oxygenators have attracted considerable interest due to their ability to incorporate unique biomimetic design and to oxygenate in a physiologically relevant manner. Polydimethylsiloxane (PDMS) has become the main material choice for these kinds of devices due to its high gas permeability. However, fabrication of large area ultrathin microfluidic devices that can oxygenate sufficient blood volumes at clinically relevant flow rates, entirely made of PDMS, have been difficult to achieve primarily due to failure associated with stiction of thin PDMS membranes to each other at undesired locations during assembly. Here, we demonstrate the use of a modified fabrication process to produce large area ultrathin oxygenators entirely made of PDMS and robust enough to withstand the hydraulic conditions that are encountered physiologically. We also demonstrate that a LAD assembled from these ultrathin double-sided microfluidic blood oxygenators can increase the oxygen saturation level by 30% at a flow rate of 30 ml/min and a pressure drop of 21 mm Hg in room air which is adequate for 1 kg preterm neonates. In addition, we demonstrated that our LAD could withstand high blood flow rate of 150 ml/min and increase oxygen saturation by 26.7% in enriched oxygen environment which is the highest gas exchange reported so far by any microfluidic-based blood oxygenators. Such performance makes this LAD suitable to provide support to 1 kg neonate suffering from respiratory distress syndrome.
Collapse
Affiliation(s)
| | - Neda Saraei
- Department of Mechanical Engineering, McMaster University, Hamilton, Ontario L8S 4L7, Canada
| | - Gerhard Fusch
- Department of Pediatrics, McMaster University, Hamilton, Ontario L8S 4K1, Canada
| | - Niels Rochow
- Department of Pediatrics, McMaster University, Hamilton, Ontario L8S 4K1, Canada
| | | | | | | |
Collapse
|
22
|
Thompson AJ, Ma LJ, Plegue TJ, Potkay JA. Design Analysis and Optimization of a Single-Layer PDMS Microfluidic Artificial Lung. IEEE Trans Biomed Eng 2019; 66:1082-1093. [DOI: 10.1109/tbme.2018.2866782] [Citation(s) in RCA: 20] [Impact Index Per Article: 4.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022]
|
23
|
Ashammakhi N, Elkhammas E, Hasan A. Translating advances in organ‐on‐a‐chip technology for supporting organs. J Biomed Mater Res B Appl Biomater 2018; 107:2006-2018. [DOI: 10.1002/jbm.b.34292] [Citation(s) in RCA: 18] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/08/2018] [Revised: 09/24/2018] [Accepted: 10/07/2018] [Indexed: 12/11/2022]
Affiliation(s)
- Nureddin Ashammakhi
- Division of Plastic Surgery, Department of SurgeryOulu University Hospital Oulu Finland
- Department of BioengineeringUniversity of California Los Angeles Los Angeles California
- School of Technology and InnovationsUniversity of Vaasa Vaasa Finland
- Biotechnology Research CenterAuthority for Natural Sciences Research and Technology Tripoli Libya
| | - Elmahdi Elkhammas
- Division of Transplantation Surgery, Department of SurgeryThe Ohio State University Wexner Medical Center, Comprehensive Transplant Center Columbus Ohio
| | - Anwarul Hasan
- Department of Mechanical and Industrial EngineeringQatar University Doha Qatar
- Biomedical Research CenterQatar University Doha Qatar
| |
Collapse
|
24
|
Dabaghi M, Saraei N, Fusch G, Rochow N, Brash JL, Fusch C, Selvaganapathy PR. An ultra-thin highly flexible microfluidic device for blood oxygenation. LAB ON A CHIP 2018; 18:3780-3789. [PMID: 30421770 DOI: 10.1039/c8lc01083h] [Citation(s) in RCA: 12] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/22/2023]
Abstract
Many neonates who are born premature suffer from respiratory distress syndrome (RDS) for which mechanical ventilation and an extracorporeal membrane oxygenation (ECMO) device are used in treatment. However, the use of these invasive techniques results in higher risk of complications like bronchopulmonary dysplasia or requires surgery to gain vascular access. An alternative biomimetic approach is to use the umbilical cord as a vascular access and to connect a passive device to the baby that functions like a placenta. This concept, known as the artificial placenta, provides enough oxygenation and causes minimal distress or complications. Herein, we have developed a new artificial placenta-type microfluidic blood oxygenator (APMBO) with high gas exchange, low priming volume and low hydraulic resistance such that it can be operated only by pressure differential provided by the baby's heart. Mimicking the placenta, we have made our new device ultra-thin and flexible so that it can be folded into a desired shape without losing its capability for gas exchange and achieve a compact form factor. The ability to fold allowed optimization of connectors and reduced the overall priming volume to the sub-milliliter range while achieving a high oxygen uptake which would be sufficient for preterm neonates with a birth-weight of around 0.5 kg.
Collapse
|
25
|
Dabaghi M, Fusch G, Saraei N, Rochow N, Brash JL, Fusch C, Ravi Selvaganapathy P. An artificial placenta type microfluidic blood oxygenator with double-sided gas transfer microchannels and its integration as a neonatal lung assist device. BIOMICROFLUIDICS 2018; 12:044101. [PMID: 30867861 PMCID: PMC6404930 DOI: 10.1063/1.5034791] [Citation(s) in RCA: 24] [Impact Index Per Article: 4.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 04/12/2018] [Accepted: 06/05/2018] [Indexed: 05/22/2023]
Abstract
Preterm neonates suffering from respiratory distress syndrome require assistive support in the form of mechanical ventilation or extracorporeal membrane oxygenation, which may lead to long-term complications or even death. Here, we describe a high performance artificial placenta type microfluidic oxygenator, termed as a double-sided single oxygenator unit (dsSOU), which combines microwire stainless-steel mesh reinforced gas permeable membranes on both sides of a microchannel network, thereby significantly reducing the diffusional resistance to oxygen uptake as compared to the previous single-sided oxygenator designs. The new oxygenator is designed to be operated in a pumpless manner, perfused solely due to the arterio-venous pressure difference in a neonate and oxygenate blood through exposure directly to ambient atmosphere without any air or oxygen pumping. The best performing dsSOUs showed up to ∼343% improvement in oxygen transfer compared to a single-sided SOU (ssSOU) with the same height. Later, the dsSOUs were optimized and integrated to build a lung assist device (LAD) that could support the oxygenation needs for a 1-2 kg neonate under clinically relevant conditions for the artificial placenta, namely, flow rates ranging from 10 to 60 ml/min and a pressure drop of 10-60 mmHg. The LAD provided an oxygen uptake of 0.78-2.86 ml/min, which corresponded to the increase in oxygen saturation from 57 ± 1% to 93%-100%, under pure oxygen environment. This microfluidic lung assist device combines elegant design with new microfabrication methods to develop a pumpless, microfluidic blood oxygenator that is capable of supporting 30% of the oxygen needs of a pre-term neonate.
Collapse
Affiliation(s)
| | - Gerhard Fusch
- Department of Pediatrics, McMaster University, Hamilton, Ontario L8S 4L7, Canada
| | - Neda Saraei
- Department of Mechanical Engineering, McMaster University, Hamilton, Ontario L8S 4L7, Canada
| | - Niels Rochow
- Department of Pediatrics, McMaster University, Hamilton, Ontario L8S 4L7, Canada
| | | | | | | |
Collapse
|
26
|
Peng J, Rochow N, Dabaghi M, Bozanovic R, Jansen J, Predescu D, DeFrance B, Lee SY, Fusch G, Ravi Selvaganapathy P, Fusch C. Postnatal dilatation of umbilical cord vessels and its impact on wall integrity: Prerequisite for the artificial placenta. Int J Artif Organs 2018; 41:393-399. [PMID: 29562805 DOI: 10.1177/0391398818763663] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
INTRODUCTION A lung assist device, which acts as an artificial placenta, can provide additional gas exchange for preterm and term newborns with respiratory failure. The concept of the lung assist device requires a large bore access via umbilical vessels to allow pumpless extracorporeal blood flow rates up to 30 mL/kg/min. After birth, constricted umbilical vessels need to be reopened for vascular access. The objective is to study the impact of umbilical vessel expansion on vessel integrity for achieving large bore access. METHODS Umbilical cords from healthy term deliveries were cannulated and dilatated with percutaneous transluminal angioplasty catheters in 1 mm increments from 4 to 8 mm for umbilical artery and from 4 to 15 mm for umbilical vein, n = 6 per expansion diameter. Paraffin-embedded transverse sections of dilated and control samples were HE & Van Gieson stained. Effects of dilatation, shown by splitting, were measured. RESULTS Umbilical vessel expansion led to concentric splitting, shown by areas devoid of extracellular matrix and nuclei in the tunica intima and media. No radial splitting was observed. Results suggest an expansion threshold of umbilical artery at 6 mm and umbilical vein at 7 mm, while maximal splitting was observed above this threshold (3.6 ± 0.8%, p = 0.043 for umbilical artery 7 mm and 6.3 ± 1.8%, p = 0.048 for umbilical vein 8 mm). Endothelial cell sloughing was present in all dilated samples but not in the control samples. CONCLUSION The suggested thresholds for safe expansions are similar to in utero umbilical vessel diameters and demonstrate a proof of concept for attaining large bore access for the lung assist device.
Collapse
Affiliation(s)
- Jenny Peng
- 1 Division of Neonatology, Department of Pediatrics, McMaster University, Hamilton, ON, Canada.,2 Faculty of Medicine, University of Ottawa, Ottawa, ON, Canada
| | - Niels Rochow
- 1 Division of Neonatology, Department of Pediatrics, McMaster University, Hamilton, ON, Canada
| | | | - Radenka Bozanovic
- 4 Department of Pathology and Molecular Medicine, Pediatric Pathology, McMaster University, Hamilton, ON, Canada
| | - Jan Jansen
- 4 Department of Pathology and Molecular Medicine, Pediatric Pathology, McMaster University, Hamilton, ON, Canada
| | - Dragos Predescu
- 5 Division of Cardiology, Department of Pediatrics, McMaster University, Hamilton, ON, Canada
| | - Bryon DeFrance
- 6 Division of Maternal-Fetal Medicine, Department of Obstetrics & Gynecology, McMaster University, Hamilton, ON, Canada
| | - Sau-Young Lee
- 1 Division of Neonatology, Department of Pediatrics, McMaster University, Hamilton, ON, Canada
| | - Gerhard Fusch
- 1 Division of Neonatology, Department of Pediatrics, McMaster University, Hamilton, ON, Canada
| | - Ponnambalam Ravi Selvaganapathy
- 3 Department of Mechanical Engineering, McMaster University, Hamilton, ON, Canada.,7 School of Biomedical Engineering, McMaster University, Hamilton, ON, Canada
| | - Christoph Fusch
- 1 Division of Neonatology, Department of Pediatrics, McMaster University, Hamilton, ON, Canada.,8 Department of Pediatrics, Paracelsus Medical School, General Hospital of Nuremberg, Nuremberg, Germany
| |
Collapse
|
27
|
Matharoo H, Dabaghi M, Rochow N, Fusch G, Saraei N, Tauhiduzzaman M, Veldhuis S, Brash J, Fusch C, Selvaganapathy PR. Steel reinforced composite silicone membranes and its integration to microfluidic oxygenators for high performance gas exchange. BIOMICROFLUIDICS 2018; 12:014107. [PMID: 29375728 PMCID: PMC5764751 DOI: 10.1063/1.5014028] [Citation(s) in RCA: 16] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 11/14/2017] [Accepted: 01/02/2018] [Indexed: 05/19/2023]
Abstract
Respiratory distress syndrome (RDS) is one of the main causes of fatality in newborn infants, particularly in neonates with low birth-weight. Commercial extracorporeal oxygenators have been used for low-birth-weight neonates in neonatal intensive care units. However, these oxygenators require high blood volumes to prime. In the last decade, microfluidics oxygenators using enriched oxygen have been developed for this purpose. Some of these oxygenators use thin polydimethylsiloxane (PDMS) membranes to facilitate gas exchange between the blood flowing in the microchannels and the ambient air outside. However, PDMS is elastic and the thin membranes exhibit significant deformation and delamination under pressure which alters the architecture of the devices causing poor oxygenation or device failure. Therefore, an alternate membrane with high stability, low deformation under pressure, and high gas exchange was desired. In this paper, we present a novel composite membrane consisting of an ultra-thin stainless-steel mesh embedded in PDMS, designed specifically for a microfluidic single oxygenator unit (SOU). In comparison to homogeneous PDMS membranes, this composite membrane demonstrated high stability, low deformation under pressure, and high gas exchange. In addition, a new design for oxygenator with sloping profile and tapered inlet configuration has been introduced to achieve the same gas exchange at lower pressure drops. SOUs were tested by bovine blood to evaluate gas exchange properties. Among all tested SOUs, the flat design SOU with composite membrane has the highest oxygen exchange of 40.32 ml/min m2. The superior performance of the new device with composite membrane was demonstrated by constructing a lung assist device (LAD) with a low priming volume of 10 ml. The LAD was achieved by the oxygen uptake of 0.48-0.90 ml/min and the CO2 release of 1.05-2.27 ml/min at blood flow rates ranging between 8 and 48 ml/min. This LAD was shown to increase the oxygen saturation level by 25% at the low pressure drop of 29 mm Hg. Finally, a piglet was used to test the gas exchange capacity of the LAD in vivo. The animal experiment results were in accordance with in-vitro results, which shows that the LAD is capable of providing sufficient gas exchange at a blood flow rate of ∼24 ml/min.
Collapse
Affiliation(s)
- Harpreet Matharoo
- Department of Mechanical Engineering, McMaster University, Hamilton, Ontario L8S 4L7, Canada
| | | | - Niels Rochow
- Department of Pediatrics, McMaster University, Hamilton, Ontario L8S 4K1, Canada
| | - Gerhard Fusch
- Department of Pediatrics, McMaster University, Hamilton, Ontario L8S 4K1, Canada
| | - Neda Saraei
- Department of Mechanical Engineering, McMaster University, Hamilton, Ontario L8S 4L7, Canada
| | - Mohammed Tauhiduzzaman
- Department of Mechanical Engineering, McMaster University, Hamilton, Ontario L8S 4L7, Canada
| | - Stephen Veldhuis
- Department of Mechanical Engineering, McMaster University, Hamilton, Ontario L8S 4L7, Canada
| | | | | | | |
Collapse
|
28
|
Thompson AJ, Marks LH, Goudie MJ, Rojas-Pena A, Handa H, Potkay JA. A small-scale, rolled-membrane microfluidic artificial lung designed towards future large area manufacturing. BIOMICROFLUIDICS 2017; 11:024113. [PMID: 28798849 PMCID: PMC5533476 DOI: 10.1063/1.4979676] [Citation(s) in RCA: 18] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Abstract] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 09/23/2016] [Accepted: 03/22/2017] [Indexed: 05/22/2023]
Abstract
Artificial lungs have been used in the clinic for multiple decades to supplement patient pulmonary function. Recently, small-scale microfluidic artificial lungs (μAL) have been demonstrated with large surface area to blood volume ratios, biomimetic blood flow paths, and pressure drops compatible with pumpless operation. Initial small-scale microfluidic devices with blood flow rates in the μl/min to ml/min range have exhibited excellent gas transfer efficiencies; however, current manufacturing techniques may not be suitable for scaling up to human applications. Here, we present a new manufacturing technology for a microfluidic artificial lung in which the structure is assembled via a continuous "rolling" and bonding procedure from a single, patterned layer of polydimethyl siloxane (PDMS). This method is demonstrated in a small-scale four-layer device, but is expected to easily scale to larger area devices. The presented devices have a biomimetic branching blood flow network, 10 μm tall artificial capillaries, and a 66 μm thick gas transfer membrane. Gas transfer efficiency in blood was evaluated over a range of blood flow rates (0.1-1.25 ml/min) for two different sweep gases (pure O2, atmospheric air). The achieved gas transfer data closely follow predicted theoretical values for oxygenation and CO2 removal, while pressure drop is marginally higher than predicted. This work is the first step in developing a scalable method for creating large area microfluidic artificial lungs. Although designed for microfluidic artificial lungs, the presented technique is expected to result in the first manufacturing method capable of simply and easily creating large area microfluidic devices from PDMS.
Collapse
Affiliation(s)
| | - L H Marks
- VA Ann Arbor Healthcare System, Ann Arbor, Michigan 48105, USA
| | - M J Goudie
- College of Engineering, University of Georgia, Athens, Georgia 30602, USA
| | - A Rojas-Pena
- Department of Surgery, University of Michigan, Ann Arbor, Michigan 48109, USA
| | - H Handa
- College of Engineering, University of Georgia, Athens, Georgia 30602, USA
| | | |
Collapse
|
29
|
Khalid N, Kobayashi I, Nakajima M. Recent lab-on-chip developments for novel drug discovery. WILEY INTERDISCIPLINARY REVIEWS-SYSTEMS BIOLOGY AND MEDICINE 2017; 9. [DOI: 10.1002/wsbm.1381] [Citation(s) in RCA: 42] [Impact Index Per Article: 6.0] [Reference Citation Analysis] [Track Full Text] [Subscribe] [Scholar Register] [Received: 09/26/2016] [Revised: 12/11/2016] [Accepted: 12/20/2016] [Indexed: 12/11/2022]
Affiliation(s)
- Nauman Khalid
- School of Food and Agricultural Sciences; University of Management and Technology; Lahore Pakistan
- Centre for Chemistry and Biotechnology, School of Life and Environmental Sciences; Deakin University; Waurn Ponds Australia
- Graduate School of Life and Environmental Sciences; University of Tsukuba; Tsukuba Japan
| | - Isao Kobayashi
- Graduate School of Life and Environmental Sciences; University of Tsukuba; Tsukuba Japan
- Food Research Institute; NARO; Tsukuba Japan
| | - Mitsutoshi Nakajima
- Graduate School of Life and Environmental Sciences; University of Tsukuba; Tsukuba Japan
- Food Research Institute; NARO; Tsukuba Japan
| |
Collapse
|
30
|
Yeager T, Roy S. Evolution of Gas Permeable Membranes for Extracorporeal Membrane Oxygenation. Artif Organs 2017; 41:700-709. [DOI: 10.1111/aor.12835] [Citation(s) in RCA: 39] [Impact Index Per Article: 5.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/11/2016] [Revised: 08/01/2016] [Accepted: 08/03/2016] [Indexed: 11/28/2022]
Affiliation(s)
- Torin Yeager
- Department of Bioengineering and Therapeutic Sciences; University of California; San Francisco CA USA
| | - Shuvo Roy
- Department of Bioengineering and Therapeutic Sciences; University of California; San Francisco CA USA
| |
Collapse
|
31
|
Bird SD. Artificial placenta: Analysis of recent progress. Eur J Obstet Gynecol Reprod Biol 2016; 208:61-70. [PMID: 27894031 DOI: 10.1016/j.ejogrb.2016.11.005] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/19/2016] [Revised: 10/18/2016] [Accepted: 11/10/2016] [Indexed: 12/18/2022]
Abstract
The artificial placenta (AP) has for many decades captured the imagination of scientists and authors with popular fiction including The Matrix and Aldous Huxley's "Brave New World", depicting a human surviving ex-utero in an artificial uterine environment (AUE). For scientists this has fascinated as a way forward for extremely preterm infants (EPIs) born less than 28 weeks of gestation. Early successes with mechanical ventilation (MV) for infants born above 28 weeks of gestation meant that AP research lost momentum. More recently, the gestational age limit for survival now borders on 23 weeks and corresponds to the biological milestone of lung development marked by the early canalicular stage of lung morphogenesis. The so called greyzone of 23-25 weeks represents a steep increase in mortality with decreasing gestational age and current options in neonatal care are on the fringes of efficacy for this population. A shift in thinking recognizes the vitality of EPIs as a fetus rather than a 37-40 week neonate and this has reinvigorated the concept of the AP. This review will discuss the scale of extreme preterm birth with special reference to previable infants born in the greyzone. Recent AP studies using sheep models are compared, technical obstacles discussed and future research themes identified.
Collapse
Affiliation(s)
- Stephen D Bird
- Department of Obstetrics and Gynaecology, The University of Melbourne, Australia.
| |
Collapse
|
32
|
Gimbel AA, Flores E, Koo A, García-Cardeña G, Borenstein JT. Development of a biomimetic microfluidic oxygen transfer device. LAB ON A CHIP 2016; 16:3227-34. [PMID: 27411972 PMCID: PMC4987252 DOI: 10.1039/c6lc00641h] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/22/2023]
Abstract
Blood oxygenators provide crucial life support for patients suffering from respiratory failure, but their use is severely limited by the complex nature of the blood circuit and by complications including bleeding and clotting. We have fabricated and tested a multilayer microfluidic blood oxygenation prototype designed to have a lower blood prime volume and improved blood circulation relative to current hollow fiber cartridge oxygenators. Here we address processes for scaling the device toward clinically relevant oxygen transfer rates while maintaining a low prime volume of blood in the device, which is required for clinical applications in cardiopulmonary support and ultimately for chronic use. Approaches for scaling the device toward clinically relevant gas transfer rates, both by expanding the active surface area of the network of blood microchannels in a planar layer and by increasing the number of microfluidic layers stacked together in a three-dimensional device are addressed. In addition to reducing prime volume and enhancing gas transfer efficiency, the geometric properties of the microchannel networks are designed to increase device safety by providing a biomimetic and physiologically realistic flow path for the blood. Safety and hemocompatibility are also influenced by blood-surface interactions within the device. In order to further enhance device safety and hemocompatibility, we have demonstrated successful coating of the blood flow pathways with human endothelial cells, in order to confer the ability of the endothelium to inhibit coagulation and thrombus formation. Blood testing results provide confirmation of fibrin clot formation in non-endothelialized devices, while negligible clot formation was documented in cell-coated devices. Gas transfer testing demonstrates that the endothelial lining does not reduce the transfer efficiency relative to acellular devices. This process of scaling the microfluidic architecture and utilizing autologous cells to line the channels and mitigate coagulation represents a promising avenue for therapy for patients suffering from a range of acute and chronic lung diseases.
Collapse
Affiliation(s)
- A A Gimbel
- Department of Biomedical Engineering, The Charles Stark Draper Laboratory, Inc., Cambridge, MA 02139, USA.
| | - E Flores
- Department of Biomedical Engineering, The Charles Stark Draper Laboratory, Inc., Cambridge, MA 02139, USA.
| | - A Koo
- Laboratory for Systems Biology, Center for Excellence in Vascular Biology, Department of Pathology, Brigham and Women's Hospital and Harvard Medical School, Boston, MA 02115, USA
| | - G García-Cardeña
- Laboratory for Systems Biology, Center for Excellence in Vascular Biology, Department of Pathology, Brigham and Women's Hospital and Harvard Medical School, Boston, MA 02115, USA
| | - J T Borenstein
- Department of Biomedical Engineering, The Charles Stark Draper Laboratory, Inc., Cambridge, MA 02139, USA.
| |
Collapse
|
33
|
Ambalavanan N, Aschner JL. Management of hypoxemic respiratory failure and pulmonary hypertension in preterm infants. J Perinatol 2016; 36 Suppl 2:S20-7. [PMID: 27225961 DOI: 10.1038/jp.2016.45] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 10/15/2015] [Revised: 02/10/2016] [Accepted: 02/16/2016] [Indexed: 01/26/2023]
Abstract
While diagnoses of hypoxemic respiratory failure (HRF) and pulmonary hypertension (PH) in preterm infants may be based on criteria similar to those in term infants, management approaches often differ. In preterm infants, HRF can be classified as 'early' or 'late' based on an arbitrary threshold of 28 postnatal days. Among preterm infants with late HRF, the pulmonary vascular abnormalities associated with bronchopulmonary dysplasia (BPD) represent a therapeutic challenge for clinicians. Surfactant, inhaled nitric oxide (iNO), sildenafil, prostacyclin and endothelin receptor blockers have been used to manage infants with both early and late HRF. However, evidence is lacking for most therapies currently in use. Chronic oral sildenafil therapy for BPD-associated PH has demonstrated some preliminary efficacy. A favorable response to iNO has been documented in some preterm infants with early PH following premature prolonged rupture of membranes and oligohydramnios. Management is complicated by a lack of clear demarcation between interventions designed to manage respiratory distress syndrome, prevent BPD and treat HRF. Heterogeneity in clinical phenotype, pathobiology and genomic underpinnings of BPD pose challenges for evidence-based management recommendations. Greater insight into the spectrum of disease phenotypes represented by BPD can optimize existing therapies and promote development of new treatments. In addition, better understanding of an individual's phenotype, genotype and biomarkers may suggest targeted personalized interventions. Initiatives such as the Prematurity and Respiratory Outcomes Program provide a framework to address these challenges using genetic, environmental, physiological and clinical data as well as large repositories of patient samples.
Collapse
Affiliation(s)
- N Ambalavanan
- Division of Neonatology, Department of Pediatrics, University of Alabama at Birmingham, Birmingham, AL, USA
| | - J L Aschner
- Department of Pediatrics, and Obstetrics, Gynecology and Women's Health, Albert Einstein College of Medicine; Children's Hospital at Montefiore, Bronx, NY, USA
| |
Collapse
|
34
|
Metelo-Coimbra C, Roncon-Albuquerque R. Artificial placenta: Recent advances and potential clinical applications. Pediatr Pulmonol 2016; 51:643-9. [PMID: 26915478 DOI: 10.1002/ppul.23401] [Citation(s) in RCA: 15] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 09/27/2015] [Revised: 12/02/2015] [Accepted: 01/21/2016] [Indexed: 12/12/2022]
Abstract
Lung immaturity remains a major cause of morbidity and mortality in extremely premature infants. Positive-pressure mechanical ventilation, the method of choice for respiratory support in premature infants, frequently promotes by itself lung injury and a negative impact in the circulatory function. Extracorporeal lung support has been proposed for more than 50 years as a potential alternative to mechanical ventilation in the treatment of severe respiratory failure of extremely premature infants. Recent advances in this field included the development of miniaturized centrifugal pumps and polymethylpentene oxygenators, as well as the successful use of pump-assisted veno-venous extracorporeal gas exchange systems in experimental artificial placenta models. This review, which includes studies published from 1958 to 2015, presents an update on the artificial placenta concept and its potential clinical applications. Special focus will be devoted to the milestones achieved so far and to the limitations that must be overcome before its clinical application. Notwithstanding, the artificial placenta stands as a promising alternative to mechanical ventilation in extremely premature infants. Pediatr Pulmonol. 2016;51:643-649. © 2016 Wiley Periodicals, Inc.
Collapse
Affiliation(s)
- Catarina Metelo-Coimbra
- Department of Physiology and Cardiothoracic Surgery, Faculty of Medicine of Porto, Porto, Portugal
| | - Roberto Roncon-Albuquerque
- Department of Physiology and Cardiothoracic Surgery, Faculty of Medicine of Porto, Porto, Portugal.,Department of Emergency and Intensive Care Medicine, Hospital de S.João, Porto, Portugal
| |
Collapse
|
35
|
Potkay JA. Reply to the 'Comment on "The promise of microfluidic artificial lungs"' by G. Wagner, A. Kaesler, U. Steinseifer, T. Schmitz-Rode and J. Arens, Lab Chip, 2016, 16. LAB ON A CHIP 2016; 16:1274-1277. [PMID: 26957040 DOI: 10.1039/c6lc00221h] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/05/2023]
Abstract
This response explores and discusses the critiques of Wagner et al. in their "Comment on 'The promise of microfluidic artificial lungs' by Joseph A. Potkay, Lab Chip, 2014, 14, 4122-4138".
Collapse
Affiliation(s)
- Joseph A Potkay
- VA Ann Arbor Healthcare System, Department of Surgery, University of Michigan, Ann Arbor, MI 48109, USA.
| |
Collapse
|
36
|
Wagner G, Kaesler A, Steinseifer U, Schmitz-Rode T, Arens J. Comment on "The promise of microfluidic artificial lungs" by J. A. Potkay, Lab Chip, 2014, 14, 4122-4138. LAB ON A CHIP 2016; 16:1272-1273. [PMID: 26956695 DOI: 10.1039/c5lc01508a] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/05/2023]
Abstract
This comment on an article that appeared in this journal (Potkay, Lab Chip, 2014, 14, 4122-4138) presents an alternative view on the feasibility and clinical application of current microfluidic artificial lungs.
Collapse
Affiliation(s)
- Georg Wagner
- Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, Pauwelsstraße 20, 52074 Aachen, Germany.
| | - Andreas Kaesler
- Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, Pauwelsstraße 20, 52074 Aachen, Germany.
| | - Ulrich Steinseifer
- Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, Pauwelsstraße 20, 52074 Aachen, Germany.
| | - Thomas Schmitz-Rode
- Institute of Applied Medical Engineering, Helmholtz Institute, Pauwelsstraße 20, 52074 Aachen, Germany
| | - Jutta Arens
- Department of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, Pauwelsstraße 20, 52074 Aachen, Germany.
| |
Collapse
|
37
|
Microfluidic Organ/Body-on-a-Chip Devices at the Convergence of Biology and Microengineering. SENSORS 2015; 15:31142-70. [PMID: 26690442 PMCID: PMC4721768 DOI: 10.3390/s151229848] [Citation(s) in RCA: 84] [Impact Index Per Article: 9.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 09/22/2015] [Revised: 11/16/2015] [Accepted: 12/04/2015] [Indexed: 12/24/2022]
Abstract
Recent advances in biomedical technologies are mostly related to the convergence of biology with microengineering. For instance, microfluidic devices are now commonly found in most research centers, clinics and hospitals, contributing to more accurate studies and therapies as powerful tools for drug delivery, monitoring of specific analytes, and medical diagnostics. Most remarkably, integration of cellularized constructs within microengineered platforms has enabled the recapitulation of the physiological and pathological conditions of complex tissues and organs. The so-called “organ-on-a-chip” technology, which represents a new avenue in the field of advanced in vitro models, with the potential to revolutionize current approaches to drug screening and toxicology studies. This review aims to highlight recent advances of microfluidic-based devices towards a body-on-a-chip concept, exploring their technology and broad applications in the biomedical field.
Collapse
|
38
|
Leung JM, Berry LR, Atkinson HM, Cornelius RM, Sandejas D, Rochow N, Selvaganapathy PR, Fusch C, Chan AKC, Brash JL. Surface modification of poly(dimethylsiloxane) with a covalent antithrombin-heparin complex for the prevention of thrombosis: use of polydopamine as bonding agent. J Mater Chem B 2015; 3:6032-6036. [PMID: 32262660 DOI: 10.1039/c5tb00808e] [Citation(s) in RCA: 14] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/21/2022]
Abstract
A modified poly(dimethyl siloxane) (PDMS) material is under development for use in an extracorporeal microfluidic blood oxygenator designed as an artificial placenta to treat newborn infants suffering from severe respiratory insufficiency. To prevent thrombosis triggered by blood-material contact, an antithrombin-heparin (ATH) covalent complex was coated on PDMS surface using polydopamine (PDA) as a "bioglue". Experiments using radiolabelled ATH showed that the ATH coating on PDA-modified PDMS remained substantially intact after incubation in plasma, 2% SDS solution, or whole blood over a three day period. The anticoagulant activity of the ATH-modified surfaces was also demonstrated: in contact with plasma the ATH-coated PDMS was shown to bind antithrombin (AT) selectively from plasma and to inhibit clotting factor Xa. It is concluded that modification of PDMS with polydopamine and ATH shows promise as a means of improving the blood compatibility of PDMS and hence of the oxygenator device.
Collapse
Affiliation(s)
- Jennifer M Leung
- School of Biomedical Engineering, McMaster University, Hamilton, Ontario, Canada.
| | | | | | | | | | | | | | | | | | | |
Collapse
|
39
|
|
40
|
Kovach KM, LaBarbera MA, Moyer MC, Cmolik BL, van Lunteren E, Sen Gupta A, Capadona JR, Potkay JA. In vitro evaluation and in vivo demonstration of a biomimetic, hemocompatible, microfluidic artificial lung. LAB ON A CHIP 2015; 15:1366-75. [PMID: 25591918 DOI: 10.1039/c4lc01284d] [Citation(s) in RCA: 29] [Impact Index Per Article: 3.2] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 05/22/2023]
Abstract
Despite the promising potential of microfluidic artificial lungs, current designs suffer from short functional lifetimes due to surface chemistry and blood flow patterns that act to reduce hemocompatibility. Here, we present the first microfluidic artificial lung featuring a hemocompatible surface coating and a biomimetic blood path. The polyethylene-glycol (PEG) coated microfluidic lung exhibited a significantly improved in vitro lifetime compared to uncoated controls as well as consistent and significantly improved gas exchange over the entire testing period. Enabled by our hemocompatible PEG coating, we additionally describe the first extended (3 h) in vivo demonstration of a microfluidic artificial lung.
Collapse
Affiliation(s)
- K M Kovach
- Advanced Platform Technology Center (APT Center), Louis Stokes Cleveland VA Medical Center, Cleveland, OH 44106, USA
| | | | | | | | | | | | | | | |
Collapse
|
41
|
Abstract
RATIONALE Much recent interest in lung bioengineering by pulmonary investigators, industry and the organ transplant field has seen a rapid growth of bioreactor development ranging from the microfluidic scale to the human-sized whole lung systems. A comprehension of the findings from these models is needed to provide the basis for further bioreactor development. OBJECTIVE The goal was to comprehensively review the current state of bioreactor development for the lung. METHODS A search using PubMed was done for published, peer-reviewed papers using the keywords "lung" AND "bioreactor" or "bioengineering" or "tissue engineering" or "ex vivo perfusion". MAIN RESULTS Many new bioreactors ranging from the microfluidic scale to the human-sized whole lung systems have been developed by both academic and commercial entities. Microfluidic, lung-mimic and lung slice cultures have the advantages of cost-efficiency and high throughput analyses ideal for pharmaceutical and toxicity studies. Perfused/ventilated rodent whole lung systems can be adapted for mid-throughput studies of lung stem/progenitor cell development, cell behavior, understanding and treating lung injury and for preliminary work that can be translated to human lung bioengineering. Human-sized ex vivo whole lung bioreactors incorporating perfusion and ventilation are amenable to automation and have been used for whole lung decellularization and recellularization. Clinical scale ex vivo lung perfusion systems have been developed for lung preservation and reconditioning and are currently being evaluated in clinical trials. CONCLUSIONS Significant advances in bioreactors for lung engineering have been made at both the microfluidic and the macro scale. The most advanced are closed systems that incorporate pressure-controlled perfusion and ventilation and are amenable to automation. Ex vivo lung perfusion systems have advanced to clinical trials for lung preservation and reconditioning. The biggest challenges that lie ahead for lung bioengineering can only be overcome by future advances in technology that solve the problems of cell production and tissue incorporation.
Collapse
Affiliation(s)
- Angela Panoskaltsis-Mortari
- Departments of Pediatrics and Medicine; Blood and Marrow Transplant Program; Pulmonary, Allergy, Critical Care and Sleep Medicine, University of Minnesota, Minneapolis, MN, 55455, U.S.A
| |
Collapse
|
42
|
Abstract
Microfluidic or microchannel artificial lungs promise to enable a new class of truly portable, therapeutic artificial lungs through feature sizes and blood channel designs that closely mimic those found in their natural counterpart. These new artificial lungs could potentially: 1) have surface areas and priming volumes that are a fraction of current technologies thereby decreasing device size and reducing the foreign body response; 2) contain blood flow networks in which cells and platelets experience pressures, shear stresses, and branching angles that copy those in the human lung thereby improving biocompatibility; 3) operate efficiently with room air, eliminating the need for gas cylinders and complications associated with hyperoxemia; 4) exhibit biomimetic hydraulic resistances, enabling operation with natural pressures and eliminating the need for blood pumps; and, 5) provide increased gas exchange capacity enabling respiratory support for active patients. This manuscript reviews recent research efforts in microfluidic artificial lungs targeted at achieving the advantages above, investigates the ultimate performance and scaling limits of these devices using a proven mathematical model, and discusses the future challenges that must be overcome in order for microfluidic artificial lungs to be applied in the clinic. If all of these promising advantages are realized and the remaining challenges are met, microfluidic artificial lungs could revolutionize the field of pulmonary rehabilitation.
Collapse
|