Gislason-Lee AJ, Keeble C, Malkin CJ, Egleston D, Bexon J, Kengyelics SM, Blackman D, Davies AG. Impact of latest generation cardiac interventional X-ray equipment on patient image quality and radiation dose for trans-catheter aortic valve implantations.
Br J Radiol 2016;
89:20160269. [PMID:
27610932 PMCID:
PMC5124836 DOI:
10.1259/bjr.20160269]
[Citation(s) in RCA: 16] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/23/2016] [Revised: 08/31/2016] [Accepted: 09/08/2016] [Indexed: 01/13/2023] Open
Abstract
OBJECTIVES
This study aimed to determine the impact on radiation dose and image quality of a new cardiac interventional X-ray system for trans-catheter aortic valve implantation (TAVI) patients compared to the previously-used cardiac X-ray system.
METHODS
Patient dose and image data were retrospectively collected from a Philips AlluraClarity (new) and Siemens Axion Artis (reference) X-ray system. Patient dose area product (DAP) and fluoroscopy duration of 41 patient cases from each X-ray system were compared using a Wilcoxon test. Ten patient aortograms from each X-ray system were scored by 32 observers on a continuous scale to assess the clinical image quality at the given phase of the TAVI procedure. Scores were dichotomised by acceptability and analysed using a Chi-squared test.
RESULTS
Significant reductions in patient dose (p << 0.001) were found for the new system with no significant change in fluoroscopy duration (p = 0.052); procedure DAP reduced by 55%, fluoroscopy DAP by 48% and "cine" acquisition DAP by 61%. There was no significant difference between image quality scores of the two X-ray systems (p = 0.06).
CONCLUSIONS
The new cardiac X-ray system demonstrated a very significant reduction in patient dose with no loss of clinical image quality. Advances in Knowledge: The huge growth of TAVI may impact on the radiation exposure of cardiac patients and particularly on operators including anaesthetists; cumulative exposure of interventional cardiologists performing high volume TAVI over 30-40 years may be harmful. The Phillips Clarity upgrade including improved image enhancement and optimised X-ray settings significantly reduced radiation without reducing clinically acceptable image quality.