Gudino N, Sonmez M, Yao Z, Baig T, Nielles-Vallespin S, Faranesh AZ, Lederman RJ, Martens M, Balaban RS, Hansen MS, Griswold MA. Parallel transmit excitation at 1.5 T based on the minimization of a driving function for device heating. Med Phys 2015; 42:359-71. [PMID: 25563276 DOI: 10.1118/1.4903894] [Citation(s) in RCA: 17] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/07/2022] Open
Abstract
PURPOSE To provide a rapid method to reduce the radiofrequency (RF) E-field coupling and consequent heating in long conductors in an interventional MRI (iMRI) setup. METHODS A driving function for device heating (W) was defined as the integration of the E-field along the direction of the wire and calculated through a quasistatic approximation. Based on this function, the phases of four independently controlled transmit channels were dynamically changed in a 1.5 T MRI scanner. During the different excitation configurations, the RF induced heating in a nitinol wire immersed in a saline phantom was measured by fiber-optic temperature sensing. Additionally, a minimization of W as a function of phase and amplitude values of the different channels and constrained by the homogeneity of the RF excitation field (B1) over a region of interest was proposed and its results tested on the benchtop. To analyze the validity of the proposed method, using a model of the array and phantom setup tested in the scanner, RF fields and SAR maps were calculated through finite-difference time-domain (FDTD) simulations. In addition to phantom experiments, RF induced heating of an active guidewire inserted in a swine was also evaluated. RESULTS In the phantom experiment, heating at the tip of the device was reduced by 92% when replacing the body coil by an optimized parallel transmit excitation with same nominal flip angle. In the benchtop, up to 90% heating reduction was measured when implementing the constrained minimization algorithm with the additional degree of freedom given by independent amplitude control. The computation of the optimum phase and amplitude values was executed in just 12 s using a standard CPU. The results of the FDTD simulations showed similar trend of the local SAR at the tip of the wire and measured temperature as well as to a quadratic function of W, confirming the validity of the quasistatic approach for the presented problem at 64 MHz. Imaging and heating reduction of the guidewire were successfully performed in vivo with the proposed hardware and phase control. CONCLUSIONS Phantom and in vivo data demonstrated that additional degrees of freedom in a parallel transmission system can be used to control RF induced heating in long conductors. A novel constrained optimization approach to reduce device heating was also presented that can be run in just few seconds and therefore could be added to an iMRI protocol to improve RF safety.
Collapse
Affiliation(s)
- N Gudino
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio 44106 and National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892
| | - M Sonmez
- National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892
| | - Z Yao
- Department of Physics, Case Western Reserve University, Cleveland, Ohio 44106
| | - T Baig
- Department of Physics, Case Western Reserve University, Cleveland, Ohio 44106
| | - S Nielles-Vallespin
- National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892
| | - A Z Faranesh
- National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892
| | - R J Lederman
- National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892
| | - M Martens
- Department of Physics, Case Western Reserve University, Cleveland, Ohio 44106
| | - R S Balaban
- National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892
| | - M S Hansen
- National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892
| | - M A Griswold
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio 44106 and Department of Radiology, University Hospitals of Cleveland, Cleveland, Ohio 44106
| |
Collapse
|
2
|
Gudino N, Griswold MA. Multi-turn transmit coil to increase b1 efficiency in current source amplification. Magn Reson Med 2013; 69:1180-5. [PMID: 23401060 DOI: 10.1002/mrm.24612] [Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/09/2012] [Revised: 11/08/2012] [Accepted: 11/30/2012] [Indexed: 11/12/2022]
Abstract
PURPOSE A multi-turn transmit surface coil design was presented to improve B1 efficiency when used with current source amplification. METHODS Three different coil designs driven by an on-coil current-mode class-D amplifier with current envelope feedback were tested on the benchtop and through imaging in a 1.5 T scanner. Case temperature of the power field-effect transistor at the amplifier output stage was measured to evaluate heat dissipation for the different current levels and coil configurations. In addition, a lower power rated device was tested to exploit the potential gain in B1 obtained with the multi-turn coil. RESULTS As shown both on the benchtop and in a 1.5 T scanner, B1 was increased by almost 3-fold without increasing heat dissipation on the power device at the amplifier's output using a multi-turn surface coil. Similar gain was obtained when connecting a lower power rated field-effect transistor to the multi-turn coil. CONCLUSION In addition to reduce heat dissipation per B1 in the device, higher B1 per current efficiency allows the use of field-effect transistors with lower current ratings and lower port capacitances, which could improve the overall performance of the on-coil current source transmit system.
Collapse
Affiliation(s)
- N Gudino
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio 44106, USA
| | | |
Collapse
|
3
|
Gudino N, Heilman JA, Derakhshan JJ, Sunshine JL, Duerk JL, Griswold MA. Control of intravascular catheters using an array of active steering coils. Med Phys 2011; 38:4215-24. [PMID: 21859023 PMCID: PMC6961950 DOI: 10.1118/1.3600693] [Citation(s) in RCA: 25] [Impact Index Per Article: 1.9] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/16/2010] [Revised: 05/26/2011] [Accepted: 05/26/2011] [Indexed: 11/07/2022] Open
Abstract
PURPOSE To extend the concept of deflecting the tip of a catheter with the magnetic force created in an MRI system through the use of an array of independently controllable steering coils located in the catheter tip, and to present methods for visualization of the catheter and/or surrounding areas while the catheter is deflected. METHODS An array of steering coils made of 42-gauge wire was built over a 2.5 Fr (0.83 mm) fiber braided microcatheter. Two of the coils were 70 turn axial coils separated by 1 cm, and the third was a 15-turn square side coil that was 2 x 4 mm2. Each coil was driven independently by a pulse width modulation (PWM) current source controlled by a microprocessor that received commands from a MATLAB routine that dynamically set current amplitude and direction for each coil. The catheter was immersed in a water phantom containing 1% Gd-DTPA that was placed at the isocenter of a 1.5 T MRI scanner. Deflections of the catheter tip were measured from image-based data obtained with a real-time radio frequency (RF) spoiled gradient echo sequence (GRE). The small local magnetic fields generated by the steering coils were exploited to generate a hyperintense signal at the catheter tip by using a modified GRE sequence that did not include slice-select rewinding gradients. Imaging and excitation modes were implemented by synchronizing the excitation of the steering coil array with the scanner by ensuring that no current was driven through the coils during the data acquisition window; this allowed visualization of the surrounding tissue while not affecting the desired catheter position. RESULTS Deflections as large as 2.5 cm were measured when exciting the steering coils sequentially with a 100 mA maximum current per coil. When exciting a single axial coil, the deflection was half this value with 30% higher current. A hyperintense catheter tip useful for catheter tracking was obtained by imaging with the modified GRE sequence. Clear visualization of the areas surrounding the catheter was obtained by using the excitation and imaging mode even with a repetition time (TR) as small as 10 ms. CONCLUSIONS A new system for catheter steering is presented that allows large deflections through the use of an integrated array of steering coils. Additionally, two imaging techniques for tracking the catheter tip and visualization of surrounding areas, without interference from the active catheter, were shown. Together the demonstrated steerable catheter, control system and the imaging techniques will ultimately contribute to the development of a steerable system for interventional MRI procedures.
Collapse
Affiliation(s)
- N Gudino
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio 44106, USA
| | | | | | | | | | | |
Collapse
|