[Citation(s) in RCA: 6] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Received: 03/23/2016] [Revised: 08/11/2016] [Accepted: 08/23/2016] [Indexed: 02/02/2023]
Abstract
INTRODUCTION
Little is known about pseudotumor frequency and risk factors for pseudotumor formation among different types of metal-on-metal (MoM) hip arthroplasties. A lower release of chromium and cobalt have been reported in MoM hip arthroplasties with a titanium sleeve compared to MoM designs without a titanium sleeve, but yet it is unknown whether a titanium sleeve reduces the pseudotumor frequency. We conducted a cross-sectional study to investigate: 1) pseudotumor frequency, 2) risk factors of pseudotumor formation 3) and correlations between pseudotumors, serum metal-ions, implant position, and clinical symptoms.
HYPOTHESIS
We expected a lower pseudotumor frequency in MoM hip articulation with a titanium sleeve than reported in MoM hip articulation designs using chromium-cobalt sleeve.
MATERIALS AND METHOD
A consecutive series of 41 patients/49 hips (31 males), mean age 52 (28-68) years, participated in a 5.5±0.5 (4-6.5) year follow-up study of their M2a_Magnum hip articulation (Biomet Inc., Warsaw, Indiana, USA). Patients were evaluated with magnetic resonance imaging (MRI), measurements of serum metal-ions, plain radiographs, and clinical outcome measures of Harris Hip Score (HHS) and Oxford Hip Score (OHS).
RESULTS
Eighteen of 47 hips (38%) had MRI-verified pseudotumors, all cystic, with a mean dimension of 10.6×25.6×41mm. Digital measurements on plain radiographs revealed a higher cup anteversion in patients with a pseudotumor of mean 28.4°±5.05° compared to mean 23.5°±6.5° in patients without a pseudotumor (P=0.009). Serum metal-ion concentrations, acetabular cup inclination and measures of HHS and OHS were similar between patients with and without a pseudotumor (P>0.46).
CONCLUSION
At 5.5±0.5years after surgery, MRI-verified cystic pseudotumors were frequently observed in M2a_Magnum hip articulations despite the use of titanium sleeves. The pseudotumors were related to high cup anteversion angles but not related to high serum metal-ions or clinical symptoms.
LEVEL OF EVIDENCE
IV: cross-sectional study.