1
|
Abstract
BACKGROUND High-flow nasal cannula (HFNC) therapy is used for patients with respiratory failure. Recently, HFNC therapy with very high gas flows (ie, gas flows of 60-100 L/min) was reported to generate higher positive airway pressure and an associated decrease in breathing frequency. However, the humidification of HFNC therapy with very high gas flow remains to be clarified. METHODS We evaluated 3 heated humidifier systems: a single MR850, the Hummax2, and parallel MR850s. The MR850 is a pass-over humidifier system, and the Hummax2 works with a porous hollow polyethylene fiber membrane. The parallel MR850 system included 2 MR850s connected in parallel to the lung with a 22 mm Y-piece. Gas flow was set at 40-90 L/min in increments of 10 L/min, and FIO2 was set at 0.21. Heated humidifiers in the MR850 systems were set in invasive mode (40°C/-3), and with the Hummax2 the vapor temperature was set at 39°C. The simulated external nares were connected to a test lung via a standard ventilator circuit. One-way valves prevented mixing of inspired and expired gases. Compliance of the test lung was 0.05 L/cm H2O and resistance 5 cm H2O/L/s. Simulated tidal volumes (VT) were 300, 500, and 700 mL, with a breathing frequency of 10 or 20 breaths/min and an inspiratory time of 1.0 s. Temperature, relative humidity, and absolute humidity (AH) of inspired gas downstream of the external nares were measured using a hygrometer for 1 min, and results for the last 3 breaths were extracted. RESULTS With the single MR850, when gas flow was > 80 L/min, AH decreased as gas flow increased (P < .001). With the Hummax2, as gas flow increased, AH decreased (P < .001). With the parallel MR850s, regardless of gas flow, AH was constant. As breathing frequency increased, AH increased in all systems. CONCLUSIONS During HFNC therapy with very high gas flows in this bench study, conventional heated humidifiers did not provide adequate humidification. Caution is advised when using HFNC therapy with very high gas flows with conventional heated humidifiers.
Collapse
Affiliation(s)
- Yusuke Chikata
- Medical Equipment Center, Tokushima University Hospital, Tokushima, Japan
| | - Keisuke Morinishi
- Medical Equipment Center, Tokushima University Hospital, Tokushima, Japan
| | - Masaji Nishimura
- Emergency and Critical Care Medicine, Tokushima University Graduate School, Tokushima, Japan.
| |
Collapse
|
2
|
Shiozaki M, Inoue K, Ishiura J, Chikata Y, Kimura Y, Fukuda K, Tamura H, Fujiwara Y, Suwa S, Sumiyoshi M, Daida H. P4691The utility of a 0-hour/1-hour algorithm in patients with suspected non-ST elevation myocardial infarction in Japan. Eur Heart J 2017. [DOI: 10.1093/eurheartj/ehx504.p4691] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/15/2022] Open
|
3
|
Chikata Y, Ohnishi S, Nishimura M. Humidity and Inspired Oxygen Concentration During High-Flow Nasal Cannula Therapy in Neonatal and Infant Lung Models. Respir Care 2017; 62:532-537. [PMID: 28174331 DOI: 10.4187/respcare.05319] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022]
Abstract
BACKGROUND High-flow nasal cannula therapy (HFNC) for neonate/infants can deliver up to 10 L/min of heated and humidified gas, and FIO2 can be adjusted to between 0.21 and 1.0. With adults, humidification and actual FIO2 are known to vary according to inspiratory and HFNC gas flow, tidal volume (VT), and ambient temperature. There have been few studies focused on humidification and FIO2 in HFNC settings for neonates/infants, so we performed a bench study to investigate the influence of gas flow, ambient temperature, and respiratory parameters on humidification and actual FIO2 in a neonate/infant simulation. METHODS HFNC gas flow was set at 3, 5, and 7 L/min, and FIO2 was set at 0.3, 0.5, and 0.7. Spontaneous breathing was simulated using a 2-bellows-in-a-box model of a neonate lung. Tests were conducted with VT settings of 20, 30, and 40 mL and breathing frequencies of 20 and 30 breaths/min. Inspiratory time was 0.8 s with decelerating flow waveform. The HFNC tube was placed in an incubator, which was either set at 37°C or turned off. Absolute humidity (AH) and actual FIO2 were measured for 1 min using a hygrometer and an oxygen analyzer, and data for the final 3 breaths were extracted. RESULTS At all settings, when the incubator was turned on, AH was greater than when it was turned off (P < .001). When the incubator was turned off, as gas flow increased, AH increased (P < .001); however, VT did not affect AH (P = .16). As gas flow increased, actual FIO2 more closely corresponded to set FIO2 . When gas flow was 3 L/min, measured FIO2 decreased proportionally more at each FIO2 setting increment (P < .001). CONCLUSIONS AH was affected by ambient temperature and HFNC gas flow. Actual FIO2 depended on VT when gas flow was 3 L/min.
Collapse
Affiliation(s)
- Yusuke Chikata
- Medical Equipment Center, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima 770-8503, Japan
| | - Saki Ohnishi
- Faculty of Medicine, Tokushima University, 3-18-15 Kuramoto, Tokushima 770-8503, Japan
| | - Masaji Nishimura
- Critical Care and Emergency Medicine, Tokushima University Graduate School, 3-18-15 Kuramoto, Tokushima 770-8503, Japan.
| |
Collapse
|
4
|
|
5
|
Abstract
BACKGROUND High-flow nasal cannula (HFNC) therapy provides better humidification than conventional oxygen therapy. To allay loss of vapor as condensation, a servo-controlled heating wire is incorporated in the inspiratory tube, but condensation is not completely avoidable. We investigated factors that might affect condensation: thermal characteristics of the inspiratory tube, HFNC flow, and ambient temperature. METHODS We evaluated 2 types of HFNC tubes, SLH Flex 22-mm single tube and RT202. Both tubes were connected to a heated humidifier with water reservoir. HFNC flow was set at 20, 40, and 60 L/min, and FIO2 was set at 0.21. Air conditioning was used maintain ambient temperature at close to either 20 or 25°C. We weighed the tubes on a digital scale before (0 h) and at 3, 6, and 24 h after, turning on the heated humidifier, and calculated the amount of condensation by simple subtraction. The amount of distilled water used during 24 h was also recorded. RESULTS At 25°C, there was little condensation, but at 20°C and HFNC flow of 20, 40, and 60 L/min for 24 h, the amount of condensation with the SLH was 50.2 ± 10.7, 44.3 ± 17.7, and 56.6 ± 13.9 mg, and the amount with the RT202 was 96.0 ± 35.1, 72.8 ± 8.2, and 64.9 ± 0.8 mg. When ambient temperature was set to 20°C, condensation with the RT202 was statistically significantly greater than with the SLH at all flow settings (P < .001). CONCLUSIONS Ambient temperature statistically significantly influenced the amount of condensation in the tubes.
Collapse
Affiliation(s)
- Yusuke Chikata
- Medical Equipment Center, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima 770-8503, Japan
| | - Kazuaki Unai
- Faculty of Medicine, Tokushima University, 3-18-15 Kuramoto, Tokushima 770-8503, Japan
| | - Masayo Izawa
- Emergency and Critical Care Medicine, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima 770-8503, Japan
| | - Nao Okuda
- Emergency and Critical Care Medicine, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima 770-8503, Japan
| | - Jun Oto
- Emergency and Critical Care Medicine, Tokushima University Graduate School of Medicine, 3-18-15 Kuramoto, Tokushima 770-8503, Japan
| | - Masaji Nishimura
- Emergency and Critical Care Medicine, Tokushima University Graduate School of Medicine, 3-18-15 Kuramoto, Tokushima 770-8503, Japan.
| |
Collapse
|
6
|
Chikata Y, Izawa M, Okuda N, Itagaki T, Nakataki E, Onodera M, Imanaka H, Nishimura M. Humidification performance of two high-flow nasal cannula devices: a bench study. Respir Care 2015; 59:1186-90. [PMID: 24368861 DOI: 10.4187/respcare.02932] [Citation(s) in RCA: 28] [Impact Index Per Article: 3.1] [Reference Citation Analysis] [What about the content of this article? (0)] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022]
Abstract
INTRODUCTION Delivering heated and humidified medical gas at 20-60 L/min, high-flow nasal cannula (HFNC) creates low levels of PEEP and ameliorates respiratory mechanics. It has become a common therapy for patients with respiratory failure. However, independent measurement of heat and humidity during HFNC and comparison of HFNC devices are lacking. METHODS We evaluated 2 HFNC (Airvo 2 and Optiflow system) devices. Each HFNC was connected to simulated external nares using the manufacturer's standard circuit. The Airvo 2 outlet-chamber temperature was set at 37°C. The Optiflow system incorporated an O2/air blender and a heated humidifier, which was set at 40°C/3. For both systems, HFNC flow was tested at 20, 40, and 50 L/min. Simulating spontaneous breathing using a mechanical ventilator and TTL test lung, we tested tidal volumes (VT) of 300, 500, and 700 mL, and breathing frequencies of 10 and 20 breaths/min. The TTL was connected to the simulated external nares with a standard ventilator circuit. To prevent condensation, the circuit was placed in an incubator maintained at 37°C. Small, medium, and large nasal prongs were tested. Absolute humidity (AH) of inspired gas was measured at the simulated external nares. RESULTS At 20, 40, and 50 L/min of flow, respective AH values for the Airvo 2 were 35.3 ± 2.0, 37.1 ± 2.2, and 37.6 ± 2.1 mg/L, and for the Optiflow system, 33.1 ± 1.5, 35.9 ± 1.7, and 36.2 ± 1.8 mg/L. AH was lower at 20 L/min of HFNC flow than at 40 and 50 L/min (P < .01). While AH remained constant at 40 and 50 L/min, at 20 L/min of HFNC flow, AH decreased as VT increased for both devices. CONCLUSIONS During bench use of HFNC, AH increased with increasing HFNC flow. When the inspiratory flow of spontaneous breathing exceeded the HFNC flow, AH was influenced by VT. At all experimental settings, AH remained > 30 mg/L.
Collapse
|
7
|
Kitaoka A, Chikata Y, Hayashi M, Onodera M, Nishimura M. Treatment with extracorporeal membrane oxygenation at Tokushima University Hospital. J Crit Care 2015. [DOI: 10.1016/j.jcrc.2015.04.068] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
|
8
|
Chikata Y, Unai K, Onodera M, Nishimura M. Amount of condensation in the limb of high-flow nasal cannula therapy: A bench study. J Crit Care 2015. [DOI: 10.1016/j.jcrc.2015.04.063] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/26/2022]
|
9
|
Shibata E, Nagai K, Takeuchi R, Noda Y, Makino T, Chikata Y, Hann M, Yoshimoto S, Ono H, Ueda S, Tamaki M, Murakami T, Matsuura M, Abe H, Doi T. Re-evaluation of Pre-pump Arterial Pressure to Avoid Inadequate Dialysis and Hemolysis: Importance of Prepump Arterial Pressure Monitoring in Hemodialysis Patients. Artif Organs 2015; 39:627-34. [DOI: 10.1111/aor.12448] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/30/2022]
Affiliation(s)
- Eriko Shibata
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Kojiro Nagai
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Risa Takeuchi
- Department of Hemodialysis; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Yasuhiro Noda
- Department of Hemodialysis; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Tomomi Makino
- Department of Hemodialysis; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Yusuke Chikata
- Department of Emergency and Critical Care; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Michael Hann
- Department of Graduate Medical Education; Naval Medical Center San Diego; San Diego CA USA
| | - Sakiya Yoshimoto
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Hiroyuki Ono
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Sayo Ueda
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Masanori Tamaki
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Taichi Murakami
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Motokazu Matsuura
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Hideharu Abe
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| | - Toshio Doi
- Department of Nephrology; Graduate School of Medicine; Health-Bioscience Institute; The University of Tokushima; Tokushima Japan
| |
Collapse
|
10
|
Abstract
BACKGROUND Magnetic resonance imaging (MRI) is indispensable for diagnosing brain and spinal cord abnormalities. Magnetic components cannot be used during MRI procedures; therefore, patient support equipment must use MRI-compatible materials. However, little is known of the performance of MRI-compatible ventilators. METHODS At commonly used settings, we tested the delivered tidal volume (V(T)), F(IO2), PEEP, and operation of the high-inspiratory-pressure-relief valves of 4 portable MRI-compatible ventilators (Pneupac VR1, ParaPAC 200DMRI, CAREvent MRI, iVent201) and one ICU ventilator (Servo-i). Each ventilator was set in volume control/continuous mandatory ventilation mode. Breathing frequency and V(T) were tested at 10 breaths/min and 300, 500, and 700 mL, respectively. The Pneupac VR1 has fixed V(T) and frequency combinations, so it was tested at V(T) = 300 mL and 20 breaths/min, V(T) = 500 mL and 12 breaths/min, and V(T) = 800 mL and 10 breaths/min. F(IO2) was 0.6 and 1.0. At the air-mix setting, F(IO2) was fixed at 0.5 with the Pneupac VR1, 0.45 with the ParaPAC 200DMRI, and 0.6 with the CAREvent MRI. PEEP was set at 5 and 10 cm H2O, and pressure relief was set at 30 and 40 cm H2O. RESULTS V(T) error varied widely among ventilators (-28.1 to 25.5%). As V(T) increased, error decreased with the Pneupac VR1, ParaPAC 200DMRI, and CAREvent MRI (P < .05). F(IO2) error ranged from -13.3 to 25.3% at 0.6 (or air mix). PEEP error varied among ventilators (-29.2 to 42.5%). Only the Servo-i maintained V(T), F(IO2), and PEEP at set levels. The pressure-relief valves worked in all ventilators. CONCLUSIONS None of the MRI-compatible ventilators maintained V(T), F(IO2), and PEEP at set levels. Vital signs of patients with unstable respiratory mechanics should be monitored during transport and MRI.
Collapse
Affiliation(s)
- Yusuke Chikata
- Medical Equipment Center, Tokushima University Hospital, Tokushima, Japan
| | - Nao Okuda
- Emergency and Critical Care Medicine, Tokushima University Graduate School, Tokushima, Japan
| | - Masayo Izawa
- Emergency and Critical Care Medicine, Tokushima University Graduate School, Tokushima, Japan
| | - Mutsuo Onodera
- Emergency and Critical Care Medicine, Tokushima University Graduate School, Tokushima, Japan
| | - Masaji Nishimura
- Emergency and Critical Care Medicine, Tokushima University Graduate School, Tokushima, Japan.
| |
Collapse
|
11
|
Chikata Y, Izawa M, Imanaka H, Okuda N, Nishimura M. Correction: Temperature of gas delivered from ventilators. J Intensive Care 2013; 1:16. [PMID: 25960876 PMCID: PMC4424707 DOI: 10.1186/2052-0492-1-16] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.2] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Track Full Text] [Download PDF] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/05/2013] [Accepted: 12/05/2013] [Indexed: 11/10/2022] Open
Affiliation(s)
- Yusuke Chikata
- Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima City 770-8503, Japan ; Medical Equipment Center, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima City 770-8503, Japan
| | - Masayo Izawa
- Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima City 770-8503, Japan
| | - Hideaki Imanaka
- Emergency and Disaster Medicine, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima City 770-8503, Japan
| | - Nao Okuda
- Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima City 770-8503, Japan
| | - Masaji Nishimura
- Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima City 770-8503, Japan
| |
Collapse
|
12
|
Abstract
Background Although heated humidifiers (HHs) are the most efficient humidifying device for mechanical ventilation, some HHs do not provide sufficient humidification when the inlet temperature to the water chamber is high. Because portable and home-care ventilators use turbines, blowers, pistons, or compressors to inhale in ambient air, they may have higher gas temperature than ventilators with piping systems. We carried out a bench study to investigate the temperature of gas delivered from portable and home-care ventilators, including the effects of distance from ventilator outlet, fraction of inspiratory oxygen (FIO2), and minute volume (MV). Methods We evaluated five ventilators equipped with turbine, blower, piston, or compressor system. Ambient air temperature was adjusted to 24°C ± 0.5°C, and ventilation was set at FIO2 0.21, 0.6, and 1.0, at MV 5 and 10 L/min. We analyzed gas temperature at 0, 40, 80, and 120 cm from ventilator outlet and altered ventilator settings. Results While temperature varied according to ventilators, the outlet gas temperature of ventilators became stable after, at the most, 5 h. Gas temperature was 34.3°C ± 3.9°C at the ventilator outlet, 29.5°C ± 2.2°C after 40 cm, 25.4°C ± 1.2°C after 80 cm and 25.1°C ± 1.2°C after 120 cm (P < 0.01). FIO2 and MV did not affect gas temperature. Conclusion Gas delivered from portable and home-care ventilator was not too hot to induce heated humidifier malfunctioning. Gas soon declined when passing through the limb.
Collapse
Affiliation(s)
- Yusuke Chikata
- Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima City, 770-8503 Japan ; Medical Equipment Center, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima City, 770-8503 Japan
| | - Mutsuo Onodera
- Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima City, 770-8503 Japan
| | - Hideaki Imanaka
- Emergency and Disaster Medicine, Tokushima University Hospital, 2-50-1 Kuramoto, Tokushima City, 770-8503 Japan
| | - Masaji Nishimura
- Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima City, 770-8503 Japan
| |
Collapse
|
13
|
Chikata Y, Oto J, Onodera M, Nishimura M. Humidification performance of humidifying devices for tracheostomized patients with spontaneous breathing: a bench study. Respir Care 2013; 58:1442-8. [PMID: 23386732 DOI: 10.4187/respcare.02093] [Citation(s) in RCA: 9] [Impact Index Per Article: 0.8] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/05/2022]
Abstract
BACKGROUND Heat and moisture exchangers (HMEs) are commonly used for humidifying respiratory gases administered to mechanically ventilated patients. While they are also applied to tracheostomized patients with spontaneous breathing, their performance in this role has not yet been clarified. We carried out a bench study to investigate the effects of spontaneous breathing parameters and oxygen flow on the humidification performance of 11 HMEs. METHODS We evaluated the humidification provided by 11 HMEs for tracheostomized patients, and also by a system delivering high-flow CPAP, and an oxygen mask with nebulizer heater. Spontaneous breathing was simulated with a mechanical ventilator, lung model, and servo-controlled heated humidifier at tidal volumes of 300, 500, and 700 mL, and breathing frequencies of 10 and 20 breaths/min. Expired gas was warmed to 37°C. The high-flow CPAP system was set to deliver 15, 30, and 45 L/min. With the 8 HMEs that were equipped with ports to deliver oxygen, and with the high-flow CPAP system, measurements were taken when delivering 0 and 3 L/min of dry oxygen. After stabilization we measured the absolute humidity (AH) of inspired gas with a hygrometer. RESULTS AH differed among HMEs applied to tracheostomized patients with spontaneous breathing. For all the HMEs, as tidal volume increased, AH decreased. At 20 breaths/min, AH was higher than at 10 breaths/min. For all the HMEs, when oxygen was delivered, AH decreased to below 30 mg/L. With an oxygen mask and high-flow CPAP, at all settings, AH exceeded 30 mg/L. CONCLUSIONS None of the HMEs provided adequate humidification when supplemental oxygen was added. In the ICU, caution is required when applying HME to tracheostomized patients with spontaneous breathing, especially when supplemental oxygen is required.
Collapse
Affiliation(s)
- Yusuke Chikata
- Department of Emergency and Critical Care Medicine, The University of Tokushima Graduate School, Tokushima, Japan
| | | | | | | |
Collapse
|
14
|
Chikata Y, Imanaka H, Ueta M, Nishimura M. Humidification during high-frequency oscillatory ventilation for adults: a bench study. Med Sci Monit 2010; 16:MT89-MT93. [PMID: 21119590] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [MESH Headings] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 05/30/2023] Open
Abstract
BACKGROUND High-frequency oscillatory ventilation (HFOV) has recently been applied to acute respiratory distress syndrome patients. However, the issue of humidification during HFOV has not been investigated. In a bench study, we evaluated humidification during HFOV for adults to test if adequate humidification was achieved in 2 different HFOV systems. MATERIAL/METHODS We tested 2 brands of adult HFOV ventilators, the R100 (Metran, Japan) and the 3100B (SensorMedics, CA), under identical bias flow. A heated humidifier consisting of porous hollow fiber (Hummax II, Metran) was set for the R100, and a passover-type heated humidifier (MR850, Fisher & Paykel) was set for the 3100B, while inspiratory heating wire was applied to both systems. Each ventilator was connected to a lung model in an incubator. Absolute humidity, relative humidity and temperature at the airway opening were measured using a hygrometer under a variety of ventilatory settings: 3 stroke volumes/amplitudes, 3 frequencies, and 2 mean airway pressures. RESULTS The R100 ventilator showed higher absolute humidity, higher relative humidity, and lower temperature than the 3100B. In the R100, as stroke volume and frequency increased, absolute humidity and temperature increased. In the 3100B, amplitude, frequency, and mean airway pressure minimally affected absolute humidity and temperature. Relative humidity was almost 100% in the R100, while it was 80.5±2.3% in the 3100B. CONCLUSIONS Humidification during HFOV for adults was affected by stroke volume and frequency in the R100, but was not in the 3100B. Absolute humidity was above 33 mgH_2 O/L in these 2 systems under a range of settings.
Collapse
Affiliation(s)
- Yusuke Chikata
- Department of Emergency and Critical Care Medicine, University of Tokushima Graduate School, Tokushima, Japan
| | | | | | | |
Collapse
|
15
|
Chikata Y, Imanaka H, Onishi Y, Ueta M, Nishimura M. Humidification during high-frequency oscillation ventilation is affected by ventilator circuit and ventilatory setting. Paediatr Anaesth 2009; 19:779-83. [PMID: 19624365 DOI: 10.1111/j.1460-9592.2009.03068.x] [Citation(s) in RCA: 7] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/29/2022]
Abstract
BACKGROUND High-frequency oscillation ventilation (HFOV) is an accepted ventilatory mode for acute respiratory failure in neonates. As conventional mechanical ventilation, inspiratory gas humidification is essential. However, humidification during HFOV has not been clarified. In this bench study, we evaluated humidification during HFOV in the open circumstance of ICU. Our hypothesis is that humidification during HFOV is affected by circuit design and ventilatory settings. METHODS/MATERIALS We connected a ventilator with HFOV mode to a neonatal lung model that was placed in an infant incubator set at 37 degrees C. We set a heated humidifier (Fisher & Paykel) to obtain 37 degrees C at the chamber outlet and 40 degrees C at the distal temperature probe. We measured absolute humidity and temperature at the Y-piece using a rapid-response hygrometer. We evaluated two types of ventilator circuit: a circuit with inner heating wire and another with embedded heating element. In addition, we evaluated three lengths of the inspiratory limb, three stroke volumes, three frequencies, and three mean airway pressures. RESULTS The circuit with embedded heating element provided significantly higher absolute humidity and temperature than one with inner heating wire. As an extended tube lacking a heating wire was shorter, absolute humidity and temperature became higher. In the circuit with inner heating wire, absolute humidity and temperature increased as stroke volume increased. CONCLUSION Humidification during HFOV is affected by circuit design and ventilatory settings.
Collapse
Affiliation(s)
- Yusuke Chikata
- Department of Emergency and Critical Care Medicine, The University of Tokushima Graduate School, Tokushima 770-8503, Japan
| | | | | | | | | |
Collapse
|
16
|
Chikata Y. The effect of cornin on DNA synthesis in mammalian cells: on the chemical properties of cornin extracted from muscle and cornea. Nihon Seirigaku Zasshi 1971; 33:266-7. [PMID: 5104606] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [MESH Headings] [Subscribe] [Scholar Register] [Indexed: 01/13/2023]
|
17
|
Chikata Y. [The effect of cornin on viral replication and DNA synthesis in mammalian cells]. Nihon Seirigaku Zasshi 1970; 32:803-12. [PMID: 4993703] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [What about the content of this article? (0)] [MESH Headings] [Subscribe] [Scholar Register] [Indexed: 01/13/2023]
|