Mozes FE, Tunnicliffe EM, Pavlides M, Robson MD. Influence of fat on liver T1 measurements using modified Look-Locker inversion recovery (MOLLI) methods at 3T.
J Magn Reson Imaging 2016;
44:105-11. [PMID:
26762615 PMCID:
PMC4982078 DOI:
10.1002/jmri.25146]
[Citation(s) in RCA: 47] [Impact Index Per Article: 5.9] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/26/2015] [Accepted: 12/21/2015] [Indexed: 01/23/2023] Open
Abstract
PURPOSE
To characterize the effect of fat on modified Look-Locker inversion recovery (MOLLI) T1 maps of the liver. The balanced steady-state free precession (bSSFP) sequence causes water and fat signals to have opposite phase when repetition time (TR) = 2.3 msec at 3T. In voxels that contain both fat and water, the MOLLI T1 measurement is influenced by the choice of TR.
MATERIALS AND METHODS
MOLLI T1 measurements of the liver were simulated using the Bloch equations while varying the hepatic lipid content (HLC). Phantom scans were performed on margarine phantoms, using both MOLLI and spin echo inversion recovery sequences. MOLLI T1 at 3T and HLC were determined in patients (n = 8) before and after bariatric surgery.
RESULTS
At 3T, with HLC in the 0-35% range, higher fat fraction values lead to longer MOLLI T1 values when TR = 2.3 msec. Patients were found to have higher MOLLI T1 at elevated HLC (T1 = 929 ± 97 msec) than at low HLC (T1 = 870 ± 44 msec).
CONCLUSION
At 3T, MOLLI T1 values are affected by HLC, substantially changing MOLLI T1 in a clinically relevant range of fat content. J. Magn. Reson. Imaging 2016;44:105-111.
Collapse