[Citation(s) in RCA: 14] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [What about the content of this article? (0)] [Affiliation(s)] [Abstract] [Key Words] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/05/2013] [Revised: 10/31/2013] [Accepted: 11/26/2013] [Indexed: 11/15/2022]
Abstract
INTRODUCTION
The platelet function analyzer (PFA)-100 is used in clinical practice to screen patients with bleeding diathesis and suspected defects of primary hemostasis. A new cartridge, INNOVANCE PFA P2Y, has been specifically developed to monitor patients' response to drugs inhibiting the platelet P2Y₁₂ receptor for ADP. In this study, we compared the ability of INNOVANCE PFA P2Y to detect congenital defects of the platelet P2Y₁₂ receptor to that of standard cartridge formulations currently in clinical use.
MATERIALS AND METHODS
We studied two patients with severe P2Y₁₂ deficiency, one patient with heterozygous P2Y₁₂ deficiency and one with dysfunctional P2Y₁₂ receptor. Closure times were measured using 3 cartridges: collagen/ADP, collagen/epinephrine, and INNOVANCE PFA P2Y. The results obtained in the four patients with P2Y₁₂ defects were compared to those obtained for 20 healthy controls.
RESULTS
In 2 patients with severe P2Y₁₂ deficiency, closure times of INNOVANCE PFA P2Y and collagen/ADP cartridges were >300 s, while those of collagen/epinephrine cartridge were variable (186s and >300 s). In the patient with dysfunctional P2Y₁₂, closure time of INNOVANCE PFA P2Y was >300 s, while closure times of collagen/ADP and collagen/epinephrine were normal. Closure times of all cartridges were normal in the patient with heterozygous P2Y₁₂ deficiency.
CONCLUSION
Our study provides the first evidence that INNOVANCE PFA P2Y cartridge is sensitive to congenital severe and moderate defects of the platelet P2Y₁₂ receptors.