Peoples T, Bailey T, Ronald B, Zisser HC, Janowski B, Huang S, Talbot C, Yang Q. Accuracy performance of the Medtronic NexSensor™ for 6 days in an inpatient setting using abdomen and buttocks insertion sites.
J Diabetes Sci Technol 2011;
5:358-64. [PMID:
21527106 PMCID:
PMC3125929 DOI:
10.1177/193229681100500224]
[Citation(s) in RCA: 6] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Submit a Manuscript] [Subscribe] [Scholar Register] [Indexed: 11/16/2022]
Abstract
BACKGROUND
Users of continuous glucose monitoring are concerned with product accuracy and choice of insertion site. The Medtronic NexSensor™ was evaluated for accuracy during 6 days of wear when inserted in the abdomen and buttocks areas.
METHODS
Adults (ages 18-75) with type 1 diabetes wore two sensors simultaneously for 6 days, one each inserted in the abdomen and buttocks. Subjects underwent a frequent blood sampling study for 12 hours, during which time reference blood glucose values were obtained every 15 minutes and compared to sensor values.
RESULTS
Sixty-three subjects were enrolled, and 61 subjects completed the study. The mean agreement rate between sensor and blood glucose values was 75.5% [95% confidence interval (CI), 69.5, 81.4] at the abdomen site, 73.8% (95% CI, 68.8, 78.8) at the buttocks site, and 75.6% (95% CI, 70.8, 80.4) when sensor and reference data were combined between sites. Over 90% of paired sensor-reference values on Clarke error grids were within the A and B ranges. The mean absolute relative differences were 17.1% at the abdomen site, 16.5% at the buttocks site, and 16.8% when sites were combined.
CONCLUSION
The NexSensor was accurate for inpatient, frequent-sample testing for 6 days when inserted into the abdomen and buttocks. The results of this study also provide evidence that both the abdomen and buttocks are suitable as sensor insertion sites.