1
|
Molendowska M, Mueller L, Fasano F, Jones DK, Tax CMW, Engel M. Giving the prostate the boost it needs: Spiral diffusion MRI using a high-performance whole-body gradient system for high b-values at short echo times. Magn Reson Med 2025; 93:1256-1272. [PMID: 39497447 DOI: 10.1002/mrm.30351] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/30/2024] [Revised: 09/30/2024] [Accepted: 10/01/2024] [Indexed: 12/29/2024]
Abstract
PURPOSE To address key issues of low SNR and image distortions in prostate diffusion MRI (dMRI) by means of using strong gradients, single-shot spiral readouts and an expanded encoding model for image reconstruction. METHODS Diffusion-weighted spin echo imaging with EPI and spiral readouts is performed on a whole-body system equipped with strong gradients (up to 250 mT/m). An expanded encoding model including static off-resonance, coil sensitivities, and magnetic field dynamics is employed for image reconstruction. The acquisitions are performed on a phantom and in vivo (one healthy volunteer and one patient with prostate cancer). The resulting images are compared to conventional dMRI EPI with navigator-based image reconstruction and assessed in terms of their congruence, SNR, tissue contrast, and quantitative parameters. RESULTS Using the expanded encoding model, high-quality images of the prostate gland are obtained across all b-values (up to 3 ms/μm2), clearly outperforming the results obtained with conventional image reconstruction. Compared to EPI, spiral imaging provides an SNR gain up to 45% within the gland and even higher in the lesion. In addition, prostate dMRI with single-shot spirals at submillimeter in-plane resolution (0.85 mm) is accomplished. CONCLUSION The combination of strong gradients and an expanded encoding model enables imaging of the prostate with unprecedented image quality. Replacing the commonly used EPI with spirals provides the inherent benefit of shorter echo times and superior readout efficiency and results in higher SNR, which is in particular relevant for considered applications.
Collapse
Affiliation(s)
- Malwina Molendowska
- Cardiff University Brain Research Imaging Centre (CUBRIC), Cardiff University, Cardiff, UK
- Medical Radiation Physics, Clinical Sciences Lund, Lund University, Lund, Sweden
| | - Lars Mueller
- Leeds Institute of Cardiovascular and Metabolic Medicine, University of Leeds, Leeds, UK
| | - Fabrizio Fasano
- Siemens Healthcare Ltd, Camberly, UK
- Siemens Healthcare GmbH, Erlangen, Germany
| | - Derek K Jones
- Cardiff University Brain Research Imaging Centre (CUBRIC), Cardiff University, Cardiff, UK
| | - Chantal M W Tax
- Cardiff University Brain Research Imaging Centre (CUBRIC), Cardiff University, Cardiff, UK
- Image Sciences Institute, University Medical Center Utrecht, Utrecht, The Netherlands
| | - Maria Engel
- Cardiff University Brain Research Imaging Centre (CUBRIC), Cardiff University, Cardiff, UK
| |
Collapse
|
2
|
Stuprich CM, Loh M, Nemerth JT, Nagel AM, Uder M, Laun FB. Velocity-compensated intravoxel incoherent motion of the human calf muscle. Magn Reson Med 2024; 92:543-555. [PMID: 38688865 DOI: 10.1002/mrm.30059] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/19/2023] [Revised: 01/15/2024] [Accepted: 02/03/2024] [Indexed: 05/02/2024]
Abstract
PURPOSE To determine whether intravoxel incoherent motion (IVIM) describes the blood perfusion in muscles better, assuming pseudo diffusion (Bihan Model 1) or ballistic motion (Bihan Model 2). METHODS IVIM parameters were measured in 18 healthy subjects with three different diffusion gradient time profiles (bipolar with two diffusion times and one with velocity compensation) and 17 b-values (0-600 s/mm2) at rest and after muscle activation. The diffusion coefficient, perfusion fraction, and pseudo-diffusion coefficient were estimated with a segmented fit in the gastrocnemius medialis (GM) and tibialis anterior (TA) muscles. RESULTS Velocity-compensated gradients resulted in a decreased perfusion fraction (6.9% ± 1.4% vs. 4.4% ± 1.3% in the GM after activation) and pseudo-diffusion coefficient (0.069 ± 0.046 mm2/s vs. 0.014 ± 0.006 in the GM after activation) compared to the bipolar gradients with the longer diffusion encoding time. Increased diffusion coefficients, perfusion fractions, and pseudo-diffusion coefficients were observed in the GM after activation for all gradient profiles. However, the increase was significantly smaller for the velocity-compensated gradients. A diffusion time dependence was found for the pseudo-diffusion coefficient in the activated muscle. CONCLUSION Velocity-compensated diffusion gradients significantly suppress the IVIM effect in the calf muscle, indicating that the ballistic limit is mostly reached, which is supported by the time dependence of the pseudo-diffusion coefficient.
Collapse
Affiliation(s)
- Christoph M Stuprich
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany
| | - Martin Loh
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany
| | - Johannes T Nemerth
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany
| | - Armin M Nagel
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany
| | - Michael Uder
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany
| | - Frederik B Laun
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany
| |
Collapse
|
3
|
Pistel M, Brock L, Laun FB, Erber R, Weiland E, Uder M, Wenkel E, Ohlmeyer S, Bickelhaupt S. Stability of Radiomic Features against Variations in Lesion Segmentations Computed on Apparent Diffusion Coefficient Maps of Breast Lesions. Diagnostics (Basel) 2024; 14:1427. [PMID: 39001317 PMCID: PMC11241112 DOI: 10.3390/diagnostics14131427] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/18/2024] [Revised: 06/18/2024] [Accepted: 06/24/2024] [Indexed: 07/16/2024] Open
Abstract
Diffusion-weighted imaging (DWI) combined with radiomics can aid in the differentiation of breast lesions. Segmentation characteristics, however, might influence radiomic features. To evaluate feature stability, we implemented a standardized pipeline featuring shifts and shape variations of the underlying segmentations. A total of 103 patients were retrospectively included in this IRB-approved study after multiparametric diagnostic breast 3T MRI with a spin-echo diffusion-weighted sequence with echoplanar readout (b-values: 50, 750 and 1500 s/mm2). Lesion segmentations underwent shifts and shape variations, with >100 radiomic features extracted from apparent diffusion coefficient (ADC) maps for each variation. These features were then compared and ranked based on their stability, measured by the Overall Concordance Correlation Coefficient (OCCC) and Dynamic Range (DR). Results showed variation in feature robustness to segmentation changes. The most stable features, excluding shape-related features, were FO (Mean, Median, RootMeanSquared), GLDM (DependenceNonUniformity), GLRLM (RunLengthNonUniformity), and GLSZM (SizeZoneNonUniformity), which all had OCCC and DR > 0.95 for both shifting and resizing the segmentation. Perimeter, MajorAxisLength, MaximumDiameter, PixelSurface, MeshSurface, and MinorAxisLength were the most stable features in the Shape category with OCCC and DR > 0.95 for resizing. Considering the variability in radiomic feature stability against segmentation variations is relevant when interpreting radiomic analysis of breast DWI data.
Collapse
Affiliation(s)
- Mona Pistel
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
- Siemens Healthineers AG, 91052 Erlangen, Germany
| | - Luise Brock
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Frederik Bernd Laun
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Ramona Erber
- Institute of Pathology, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Elisabeth Weiland
- MR Application Predevelopment, Siemens Healthineers AG, 91052 Erlangen, Germany
| | - Michael Uder
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Evelyn Wenkel
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
- Radiologie München, 80331 München, Germany
| | - Sabine Ohlmeyer
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Sebastian Bickelhaupt
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| |
Collapse
|
4
|
Basukala D, Mikheev A, Sevilimedu V, Gilani N, Moy L, Pinker-Domenig K, Thakur SB, Sigmund EE. Multisite MRI Intravoxel Incoherent Motion Repeatability and Reproducibility across 3 T Scanners in a Breast Diffusion Phantom: A BReast Intravoxel Incoherent Motion Multisite (BRIMM) Study. J Magn Reson Imaging 2024; 59:2226-2237. [PMID: 37702382 PMCID: PMC10932866 DOI: 10.1002/jmri.29008] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/17/2023] [Revised: 08/28/2023] [Accepted: 08/29/2023] [Indexed: 09/14/2023] Open
Abstract
BACKGROUND Monoexponential apparent diffusion coefficient (ADC) and biexponential intravoxel incoherent motion (IVIM) analysis of diffusion-weighted imaging is helpful in the characterization of breast tumors. However, repeatability/reproducibility studies across scanners and across sites are scarce. PURPOSE To evaluate the repeatability and reproducibility of ADC and IVIM parameters (tissue diffusivity (Dt), perfusion fraction (Fp) and pseudo-diffusion (Dp)) within and across sites employing MRI scanners from different vendors utilizing 16-channel breast array coils in a breast diffusion phantom. STUDY TYPE Phantom repeatability. PHANTOM A breast phantom containing tubes of different polyvinylpyrrolidone (PVP) concentrations, water, fat, and sponge flow chambers, together with an MR-compatible liquid crystal (LC) thermometer. FIELD STRENGTH/SEQUENCE Bipolar gradient twice-refocused spin echo sequence and monopolar gradient single spin echo sequence at 3 T. ASSESSMENT Studies were performed twice in each of two scanners, located at different sites, on each of 2 days, resulting in four studies per scanner. ADCs of the PVP and water were normalized to the vendor-provided calibrated values at the temperature indicated by the LC thermometer for repeatability/reproducibility comparisons. STATISTICAL TESTS ADC and IVIM repeatability and reproducibility within and across sites were estimated via the within-system coefficient of variation (wCV). Pearson correlation coefficient (r) was also computed between IVIM metrics and flow speed. A P value <0.05 was considered statistically significant. RESULTS ADC and Dt demonstrated excellent repeatability (<2%; <3%, respectively) and reproducibility (both <5%) at the two sites. Fp and Dp exhibited good repeatability (mean of two sites 3.67% and 5.59%, respectively) and moderate reproducibility (mean of two sites 15.96% and 13.3%, respectively). The mean intersite reproducibility (%) of Fp/Dp/Dt was 50.96/13.68/5.59, respectively. Fp and Dt demonstrated high correlations with flow speed while Dp showed lower correlations. Fp correlations with flow speed were significant at both sites. DATA CONCLUSION IVIM reproducibility results were promising and similar to ADC, particularly for Dt. The results were reproducible within both sites, and a progressive trend toward reproducibility across sites except for Fp. LEVEL OF EVIDENCE 2 TECHNICAL EFFICACY: Stage 1.
Collapse
Affiliation(s)
- Dibash Basukala
- Center for Advanced Imaging and Innovation (CAIR), Center for Biomedical Imaging, Department of Radiology, NYU Langone Health, New York, New York, USA
| | - Artem Mikheev
- Center for Advanced Imaging and Innovation (CAIR), Center for Biomedical Imaging, Department of Radiology, NYU Langone Health, New York, New York, USA
| | - Varadan Sevilimedu
- Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, New York, USA
| | - Nima Gilani
- Center for Advanced Imaging and Innovation (CAIR), Center for Biomedical Imaging, Department of Radiology, NYU Langone Health, New York, New York, USA
| | - Linda Moy
- Center for Advanced Imaging and Innovation (CAIR), Center for Biomedical Imaging, Department of Radiology, NYU Langone Health, New York, New York, USA
| | - Katja Pinker-Domenig
- Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, New York, USA
| | - Sunitha B. Thakur
- Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, New York, USA
| | - Eric E. Sigmund
- Center for Advanced Imaging and Innovation (CAIR), Center for Biomedical Imaging, Department of Radiology, NYU Langone Health, New York, New York, USA
| |
Collapse
|
5
|
Bäuchle TA, Stuprich CM, Loh M, Nagel AM, Uder M, Laun FB. Influence of Magnetic Field Strength on Intravoxel Incoherent Motion Parameters in Diffusion MRI of the Calf. Tomography 2024; 10:773-788. [PMID: 38787019 PMCID: PMC11126135 DOI: 10.3390/tomography10050059] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/11/2024] [Revised: 04/26/2024] [Accepted: 05/13/2024] [Indexed: 05/25/2024] Open
Abstract
Background: The purpose of this study was to investigate the dependence of Intravoxel Incoherent Motion (IVIM) parameters measured in the human calf on B0. Methods: Diffusion-weighted image data of eight healthy volunteers were acquired using five b-values (0-600 s/mm2) at rest and after muscle activation at 0.55 and 7 T. The musculus gastrocnemius mediale (GM, activated) was assessed. The perfusion fraction f and diffusion coefficient D were determined using segmented fits. The dependence on field strength was assessed using Student's t-test for paired samples and the Wilcoxon signed-rank test. A biophysical model built on the three non-exchanging compartments of muscle, venous blood, and arterial blood was used to interpret the data using literature relaxation times. Results: The measured perfusion fraction of the GM was significantly lower at 7 T, both for the baseline measurement and after muscle activation. For 0.55 and 7 T, the mean f values were 7.59% and 3.63% at rest, and 14.03% and 6.92% after activation, respectively. The biophysical model estimations for the mean proton-density-weighted perfusion fraction were 3.37% and 6.50% for the non-activated and activated states, respectively. Conclusions: B0 may have a significant effect on the measured IVIM parameters. The blood relaxation times suggest that 7 T IVIM may be arterial-weighted whereas 0.55 T IVIM may exhibit an approximately equal weighting of arterial and venous blood.
Collapse
Affiliation(s)
- Tamara Alice Bäuchle
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Christoph Martin Stuprich
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Martin Loh
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Armin Michael Nagel
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Michael Uder
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| | - Frederik Bernd Laun
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), 91054 Erlangen, Germany
| |
Collapse
|
6
|
Johnson JTE, Irfanoglu MO, Manninen E, Ross TJ, Yang Y, Laun FB, Martin J, Topgaard D, Benjamini D. In vivo disentanglement of diffusion frequency-dependence, tensor shape, and relaxation using multidimensional MRI. Hum Brain Mapp 2024; 45:e26697. [PMID: 38726888 PMCID: PMC11082920 DOI: 10.1002/hbm.26697] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/10/2023] [Revised: 03/28/2024] [Accepted: 04/12/2024] [Indexed: 05/13/2024] Open
Abstract
Diffusion MRI with free gradient waveforms, combined with simultaneous relaxation encoding, referred to as multidimensional MRI (MD-MRI), offers microstructural specificity in complex biological tissue. This approach delivers intravoxel information about the microstructure, local chemical composition, and importantly, how these properties are coupled within heterogeneous tissue containing multiple microenvironments. Recent theoretical advances incorporated diffusion time dependency and integrated MD-MRI with concepts from oscillating gradients. This framework probes the diffusion frequency,ω $$ \omega $$ , in addition to the diffusion tensor,D $$ \mathbf{D} $$ , and relaxation,R 1 $$ {R}_1 $$ ,R 2 $$ {R}_2 $$ , correlations. AD ω - R 1 - R 2 $$ \mathbf{D}\left(\omega \right)-{R}_1-{R}_2 $$ clinical imaging protocol was then introduced, with limited brain coverage and 3 mm3 voxel size, which hinder brain segmentation and future cohort studies. In this study, we introduce an efficient, sparse in vivo MD-MRI acquisition protocol providing whole brain coverage at 2 mm3 voxel size. We demonstrate its feasibility and robustness using a well-defined phantom and repeated scans of five healthy individuals. Additionally, we test different denoising strategies to address the sparse nature of this protocol, and show that efficient MD-MRI encoding design demands a nuanced denoising approach. The MD-MRI framework provides rich information that allows resolving the diffusion frequency dependence into intravoxel components based on theirD ω - R 1 - R 2 $$ \mathbf{D}\left(\omega \right)-{R}_1-{R}_2 $$ distribution, enabling the creation of microstructure-specific maps in the human brain. Our results encourage the broader adoption and use of this new imaging approach for characterizing healthy and pathological tissues.
Collapse
Affiliation(s)
- Jessica T. E. Johnson
- Multiscale Imaging and Integrative Biophysics Unit, National Institute on Aging, NIHBaltimoreMarylandUSA
| | - M. Okan Irfanoglu
- Quantitative Medical Imaging Section, National Institute of Biomedical Imaging and Bioengineering, National Institutes of HealthBethesdaMarylandUSA
| | - Eppu Manninen
- Multiscale Imaging and Integrative Biophysics Unit, National Institute on Aging, NIHBaltimoreMarylandUSA
| | - Thomas J. Ross
- Neuroimaging Research Branch, National Institute on Drug Abuse, National Institutes of HealthBaltimoreMarylandUSA
| | - Yihong Yang
- Neuroimaging Research Branch, National Institute on Drug Abuse, National Institutes of HealthBaltimoreMarylandUSA
| | - Frederik B. Laun
- Institute of Radiology, University Hospital Erlangen, Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU)ErlangenGermany
| | - Jan Martin
- Department of ChemistryLund UniversityLundSweden
| | | | - Dan Benjamini
- Multiscale Imaging and Integrative Biophysics Unit, National Institute on Aging, NIHBaltimoreMarylandUSA
| |
Collapse
|
7
|
Rauch J, Laun FB, Bachert P, Ladd ME, Kuder TA. Compensation of concomitant field effects in double diffusion encoding by means of added oscillating gradients. Magn Reson Imaging 2024; 105:133-141. [PMID: 37939973 DOI: 10.1016/j.mri.2023.11.006] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/17/2023] [Accepted: 11/04/2023] [Indexed: 11/10/2023]
Abstract
Maxwell or concomitant fields imprint additional phases on the transverse magnetization. This concomitant phase may cause severe image artifacts like signal voids or distort the quantitative parameters due to the induced intravoxel dephasing. In particular, double diffusion encoding (DDE) schemes with two pairs of bipolar diffusion-weighting gradients separated by a refocusing radiofrequency (RF) pulse are prone to concomitant field-induced artifacts. In this work, a method for reducing concomitant field effects in these DDE sequences based on additional oscillating gradients is presented. These oscillating gradient pulses obtained by constrained optimization were added to the original gradient waveforms. The modified sequences reduced the accumulated concomitant phase without significant changes in the original sequence characteristics. The proposed method was applied to a DDE acquisition scheme consisting of 60 pairs of diffusion wave vectors. For phantom as well as for in vivo experiments, a considerable increase in the signal-to-noise ratio (SNR) was obtained. For phantom measurements with a diffusion weighting of b = 2000 s/mm2 for each of the gradient pairs, an SNR increase of up to 40% was observed for a transversal slice that had a distance of 5 cm from the isocenter. For equivalent slice parameters, in vivo measurements in the brain of a healthy volunteer exhibited an increase in SNR of up to 35% for b = 750 s/mm2 for each weighting. These findings are supported by corresponding simulations, which also predict a positive effect on the SNR. In summary, the presented method leads to an SNR gain without additional RF refocusing pulses.
Collapse
Affiliation(s)
- Julian Rauch
- Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany; MPI for Nuclear Physics, Max-Planck-Society, Saupfercheckweg 1, 69117 Heidelberg, Germany; Faculty of Physics and Astronomy, Heidelberg University, Im Neuenheimer Feld 226, 69120 Heidelberg, Germany
| | - Frederik B Laun
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Maximiliansplatz 3, 91054 Erlangen, Germany
| | - Peter Bachert
- Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany; Faculty of Physics and Astronomy, Heidelberg University, Im Neuenheimer Feld 226, 69120 Heidelberg, Germany
| | - Mark E Ladd
- Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany; Faculty of Physics and Astronomy, Heidelberg University, Im Neuenheimer Feld 226, 69120 Heidelberg, Germany; Faculty of Medicine, Heidelberg University, Im Neuenheimer Feld 672, 69120 Heidelberg, Germany
| | - Tristan A Kuder
- Division of Medical Physics in Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany; Faculty of Physics and Astronomy, Heidelberg University, Im Neuenheimer Feld 226, 69120 Heidelberg, Germany.
| |
Collapse
|
8
|
Phonlakrai M, Ramadan S, Simpson J, Skehan K, Goodwin J, Trada Y, Martin J, Sridharan S, Gan LT, Siddique SH, Greer P. Non-contrast based approach for liver function quantification using Bayesian-based intravoxel incoherent motion diffusion weighted imaging: A pilot study. J Appl Clin Med Phys 2023; 24:e14178. [PMID: 37819022 PMCID: PMC10647975 DOI: 10.1002/acm2.14178] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 06/14/2023] [Revised: 09/18/2023] [Accepted: 09/26/2023] [Indexed: 10/13/2023] Open
Abstract
PURPOSE Liver cirrhosis disrupts liver function and tissue perfusion, detectable by magnetic resonance imaging (MRI). Assessing liver function at the voxel level with 13-b value intravoxel incoherent motion diffusion-weighted imaging (IVIM-DWI) could aid in radiation therapy liver-sparing treatment for patients with early impairment. This study aimed to evaluate the feasibility of IVIM-DWI for liver function assessment and correlate it with other multiparametric (mp) MRI methods at the voxel level. METHOD This study investigates the variability of apparent diffusion coefficient (ADC) derived from 13-b value IVIM-DWI and B1-corrected dual flip angle (DFA) T1 mapping. Experiments were conducted in-vitro with QIBA and NIST phantoms and in 10 healthy volunteers for IVIM-DWI. Additionally, 12 patients underwent an mp-MRI examination. The imaging protocol included a 13-b value IVIM-DWI sequence for generating IVIM parametric maps. B1-corrected DFA T1 pulse sequence was used for generating T1 maps, and Gadoxatate low temporal resolution dynamic contrast-enhanced (LTR-DCE) MRI was used for generating the Hepatic extraction fraction (HEF) map. The Mann-Whitney U test was employed to compare IVIM-DWI parameters (Pure Diffusion, Dslow ; Pseudo diffusion, Dfast ; and Perfusion Fraction, Fp ) between the healthy volunteer and patient groups. Furthermore, in the patient group, statistical correlations were assessed at a voxel level between LTR-DCE MRI-derived HEF, T1 post-Gadoxetate administration, ΔT1%, and various IVIM parameters using Pearson correlation. RESULTS For-vitro measurements, the maximum coefficient of variation of the ADC and T1 parameters was 12.4% and 16.1%, respectively. The results also showed that Fp and Dfast were able to distinguish between healthy liver function and mild liver function impairment at the global level, with p = 0.002 for Fp and p < 0.001 for Dfast . Within the patient group, these parameters also exhibited a moderate correlation with HEF at the voxel level. CONCLUSION Overall, the study highlighted the potential of Dfast and Fp for detecting liver function impairment at both global and pixel levels.
Collapse
Affiliation(s)
- Monchai Phonlakrai
- School of Health Sciences, College of Health, Medicine and WellbeingThe University of NewcastleNewcastleNSWAustralia
- School of Radiological TechnologyFaculty of Health Science TechnologyChulabhorn Royal AcademyBangkokThailand
| | - Saadallah Ramadan
- HMRI Imaging CentreHunter Medical Research InstituteNewcastleNSWAustralia
- College of Health, Medicine and WellbeingThe University of NewcastleNewcastleNSWAustralia
| | - John Simpson
- Radiation Oncology DepartmentCalvary Mater NewcastleNewcastleNSWAustralia
- School of Information and Physical Sciences, College of Engineering, Science and EnvironmentThe University of NewcastleNewcastleNSWAustralia
| | - Kate Skehan
- Radiation Oncology DepartmentCalvary Mater NewcastleNewcastleNSWAustralia
| | - Jonathan Goodwin
- Radiation Oncology DepartmentCalvary Mater NewcastleNewcastleNSWAustralia
- School of Information and Physical Sciences, College of Engineering, Science and EnvironmentThe University of NewcastleNewcastleNSWAustralia
| | - Yuvnik Trada
- Radiation Oncology DepartmentCalvary Mater NewcastleNewcastleNSWAustralia
- Faculty of Medicine and HealthSydney Medical SchoolThe University of SydneySydneyNSWAustralia
| | - Jarad Martin
- Radiation Oncology DepartmentCalvary Mater NewcastleNewcastleNSWAustralia
- School of Medicine and Public Health, College of Health, Medicine and WellbeingThe University of NewcastleNewcastleNSWAustralia
| | - Swetha Sridharan
- Radiation Oncology DepartmentCalvary Mater NewcastleNewcastleNSWAustralia
- School of Medicine and Public Health, College of Health, Medicine and WellbeingThe University of NewcastleNewcastleNSWAustralia
| | - Lay Theng Gan
- The Gastroenterology Department at John Hunter HospitalNewcastleNSWAustralia
| | | | - Peter Greer
- Radiation Oncology DepartmentCalvary Mater NewcastleNewcastleNSWAustralia
- School of Information and Physical Sciences, College of Engineering, Science and EnvironmentThe University of NewcastleNewcastleNSWAustralia
| |
Collapse
|
9
|
Johnson JT, Irfanoglu MO, Manninen E, Ross TJ, Yang Y, Laun FB, Martin J, Topgaard D, Benjamini D. In vivo disentanglement of diffusion frequency-dependence, tensor shape, and relaxation using multidimensional MRI. BIORXIV : THE PREPRINT SERVER FOR BIOLOGY 2023:2023.10.10.561702. [PMID: 37987005 PMCID: PMC10659440 DOI: 10.1101/2023.10.10.561702] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 11/22/2023]
Abstract
Diffusion MRI with free gradient waveforms, combined with simultaneous relaxation encoding, referred to as multidimensional MRI (MD-MRI), offers microstructural specificity in complex biological tissue. This approach delivers intravoxel information about the microstructure, local chemical composition, and importantly, how these properties are coupled within heterogeneous tissue containing multiple microenvironments. Recent theoretical advances incorporated diffusion time dependency and integrated MD-MRI with concepts from oscillating gradients. This framework probes the diffusion frequency, ω , in addition to the diffusion tensor, D , and relaxation, R 1 , R 2 , correlations. A D ( ω ) - R 1 - R 2 clinical imaging protocol was then introduced, with limited brain coverage and 3 mm3 voxel size, which hinder brain segmentation and future cohort studies. In this study, we introduce an efficient, sparse in vivo MD-MRI acquisition protocol providing whole brain coverage at 2 mm3 voxel size. We demonstrate its feasibility and robustness using a well-defined phantom and repeated scans of five healthy individuals. Additionally, we test different denoising strategies to address the sparse nature of this protocol, and show that efficient MD-MRI encoding design demands a nuanced denoising approach. The MD-MRI framework provides rich information that allows resolving the diffusion frequency dependence into intravoxel components based on their D ( ω ) - R 1 - R 2 distribution, enabling the creation of microstructure-specific maps in the human brain. Our results encourage the broader adoption and use of this new imaging approach for characterizing healthy and pathological tissues.
Collapse
Affiliation(s)
- Jessica T.E. Johnson
- Multiscale Imaging and Integrative Biophysics Unit, National Institute on Aging, NIH, Baltimore, MD, USA
| | - M. Okan Irfanoglu
- Quantitative Medical Imaging Section, National Institute of Biomedical Imaging and Bioengineering, National Institutes of Health, Bethesda, MD, USA
| | - Eppu Manninen
- Multiscale Imaging and Integrative Biophysics Unit, National Institute on Aging, NIH, Baltimore, MD, USA
| | - Thomas J. Ross
- Neuroimaging Research Branch, National Institute on Drug Abuse, National Institutes of Health, Baltimore, MD, USA
| | - Yihong Yang
- Neuroimaging Research Branch, National Institute on Drug Abuse, National Institutes of Health, Baltimore, MD, USA
| | - Frederik B. Laun
- Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany
| | - Jan Martin
- Department of Chemistry, Lund University, Lund, Sweden
| | | | - Dan Benjamini
- Multiscale Imaging and Integrative Biophysics Unit, National Institute on Aging, NIH, Baltimore, MD, USA
| |
Collapse
|
10
|
Bruckmaier F, Allert RD, Neuling NR, Amrein P, Littin S, Briegel KD, Schätzle P, Knittel P, Zaitsev M, Bucher DB. Imaging local diffusion in microstructures using NV-based pulsed field gradient NMR. SCIENCE ADVANCES 2023; 9:eadh3484. [PMID: 37595048 PMCID: PMC10438442 DOI: 10.1126/sciadv.adh3484] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 03/16/2023] [Accepted: 07/20/2023] [Indexed: 08/20/2023]
Abstract
Understanding diffusion in microstructures plays a crucial role in many scientific fields, including neuroscience, medicine, or energy research. While magnetic resonance (MR) methods are the gold standard for diffusion measurements, spatial encoding in MR imaging has limitations. Here, we introduce nitrogen-vacancy (NV) center-based nuclear MR (NMR) spectroscopy as a powerful tool to probe diffusion within microscopic sample volumes. We have developed an experimental scheme that combines pulsed gradient spin echo (PGSE) with optically detected NV-NMR spectroscopy, allowing local quantification of molecular diffusion and flow. We demonstrate correlated optical imaging with spatially resolved PGSE NV-NMR experiments probing anisotropic water diffusion within an individual model microstructure. Our optically detected PGSE NV-NMR technique opens up prospects for extending the current capabilities of investigating diffusion processes with the future potential of probing single cells, tissue microstructures, or ion mobility in thin film materials for battery applications.
Collapse
Affiliation(s)
- Fleming Bruckmaier
- Department of Chemistry, TUM School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany
| | - Robin D. Allert
- Department of Chemistry, TUM School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany
| | - Nick R. Neuling
- Department of Chemistry, TUM School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany
| | - Philipp Amrein
- Division of Medical Physics, Department of Diagnostic and Interventional Radiology, Faculty of Medicine, University of Freiburg, Freiburg, Germany
| | - Sebastian Littin
- Division of Medical Physics, Department of Diagnostic and Interventional Radiology, Faculty of Medicine, University of Freiburg, Freiburg, Germany
| | - Karl D. Briegel
- Department of Chemistry, TUM School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany
| | - Philip Schätzle
- Department of Sustainable Systems Engineering (INATECH), University of Freiburg, Emmy-Noether-Str. 2, 79110 Freiburg, Germany
| | - Peter Knittel
- Fraunhofer Institute for Applied Solid State Physics, Tullastr. 72, 79108 Freiburg, Germany
| | - Maxim Zaitsev
- Division of Medical Physics, Department of Diagnostic and Interventional Radiology, Faculty of Medicine, University of Freiburg, Freiburg, Germany
| | - Dominik B. Bucher
- Department of Chemistry, TUM School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany
- Munich Center for Quantum Science and Technology (MCQST), Schellingstr. 4, 80799 München, Germany
| |
Collapse
|
11
|
Møller JM, Boesen L, Hansen AE, Kettles K, Løgager V. Quantification of cross-vendor variation in ADC measurements in vendor-specific prostate MRI-protocols. Eur J Radiol 2023; 165:110942. [PMID: 37364483 DOI: 10.1016/j.ejrad.2023.110942] [Citation(s) in RCA: 5] [Impact Index Per Article: 2.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/09/2023] [Revised: 06/14/2023] [Accepted: 06/19/2023] [Indexed: 06/28/2023]
Abstract
PURPOSE The purpose of this study was to quantify the variability of Apparent Diffusion Coefficient (ADC) and test if there were statistically significant differences in ADC between MRI systems and sequences. METHOD With a two-chamber cylindrical ADC phantom with fixed ADC values (1,000 and 1,600x10-6 mm2/s) a single-shot (ss) Echo Planar Imaging (EPI), a multi-shot EPI, a reduced field of view DWI (zoom) and a Turbo Spin Echo DWI sequence were tested in six MRI systems from three vendors at 1.5 T and 3 T. Technical parameters were according to Prostate Imaging Reporting and Data System Version 2.1. ADC maps were calculated by vendor specific algorithms. Absolute and relative differences in ADC from the phantom-ADC were calculated and differences between sequences were tested. RESULTS At 3 T absolute differences from phantom given ADC (∼1,000 and ∼ 1,600x10-6 mm2/s) were -83 - 42x10-6 mm2/s (-8.3%-4.2%) and -48 - 15x10-6 mm2/s (-3%-0.9%), respectively and at 1.5 T absolute differences were -81 - 26x10-6 mm2/s (-2.6%-8.1%) and -74 - 67x10-6 mm2/s (-4.6%-4.2%), respectively. Significant statistical differences in ADC measurements were identified between vendors in all sequences except for ssEPI and zoom at 3 T in the 1,600x10-6 mm2/s phantom chamber. Significant differences were also identified between ADC measurements at 1.5 T and 3 T in some of the sequences and vendors, but not all. CONCLUSION The variation of ADC between different MRI systems and prostate specific DWI sequences is limited in this phantom study and without apparent clinical relevance. However, prospective multicenter studies of prostate cancer patients are needed for further investigation.
Collapse
Affiliation(s)
- Jakob M Møller
- Dep. of Radiology, Copenhagen University Hospital, Herlev-Gentofte, Denmark, Borgmester Ib Juuls vej 17, DK-2730 Herlev, Denmark.
| | - Lars Boesen
- Dep. of Urology, Copenhagen University Hospital, Herlev-Gentofte, Denmark
| | - Adam Espe Hansen
- Dep of radiology, Copenhagen University Hospital, Rigshospitalet and dep. of clinical medicine Copenhagen University, Copenhagen, Denmark
| | | | - Vibeke Løgager
- Dep. of Radiology, Copenhagen University Hospital, Herlev-Gentofte, Denmark
| |
Collapse
|
12
|
Fritz V, Martirosian P, Machann J, Thorwarth D, Schick F. Soy lecithin: A beneficial substance for adjusting the ADC in aqueous solutions to the values of biological tissues. Magn Reson Med 2023; 89:1674-1683. [PMID: 36458695 DOI: 10.1002/mrm.29543] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/29/2022] [Revised: 10/31/2022] [Accepted: 11/11/2022] [Indexed: 12/04/2022]
Abstract
PURPOSE To test soy lecithin as a substance added to water for the construction of MRI phantoms with tissue-like diffusion coefficients. The performance of soy lecithin was assessed for the useable range of adjustable ADC values, the degree of non-Gaussian diffusion, simultaneous effects on relaxation times, and spectral signal properties. METHODS Aqueous soy lecithin solutions of different concentrations (0%, 0.5%, 1%, 2%, 3% …, 10%) and soy lecithin-agar gels were prepared and examined on a 3 Tesla clinical scanner at 18.5° ± 0.5°C. Echoplanar sequences (b values: 0-1000/3000 s/mm2 ) were applied for ADC measurements. Quantitative relaxometry and MRS were performed for assessment of T1 , T2 , and detectable spectral components. RESULTS The presence of soy lecithin significantly restricts the diffusion of water molecules and mimics the nearly Gaussian nature of diffusion observed in tissue (for b values <1000 s/mm2 ). ADC values ranged from 2.02 × 10-3 mm2 /s to 0.48 × 10-3 mm2 /s and cover the entire physiological range reported on biological tissue. Measured T1 /T2 values of pure lecithin solutions varied from 2685/2013 to 668/133 ms with increasing concentration. No characteristic signals of soy lecithin were observed in the MR spectrum. The addition of agar to the soy lecithin solutions allowed T2 values to be well adjusted to typical values found in parenchymal tissue without affecting the soy lecithin-controlled ADC value. CONCLUSION Soy lecithin is a promising substance for the construction of diffusion phantoms with tissue-like ADC values. It provides several advantages over previously proposed substances, in particular a wide range of adjustable ADC values, the lack of additional 1 H-signals, and the possibility to adjust ADC and T2 values (by adding agar) almost independently of each other.
Collapse
Affiliation(s)
- Victor Fritz
- Section on Experimental Radiology, Department of Diagnostic and Interventional Radiology, University of Tübingen, Tübingen, Germany
| | - Petros Martirosian
- Section on Experimental Radiology, Department of Diagnostic and Interventional Radiology, University of Tübingen, Tübingen, Germany
| | - Jürgen Machann
- Section on Experimental Radiology, Department of Diagnostic and Interventional Radiology, University of Tübingen, Tübingen, Germany.,Institute for Diabetes Research and Metabolic Diseases of the Helmholtz Centre Munich at the University of Tübingen, Tübingen, Germany.,German Center for Diabetes Research (DZD), Neuherberg, Germany
| | - Daniela Thorwarth
- Section for Biomedical Physics, Department of Radiation Oncology, University of Tübingen, Tübingen, Germany
| | - Fritz Schick
- Section on Experimental Radiology, Department of Diagnostic and Interventional Radiology, University of Tübingen, Tübingen, Germany.,Cluster of Excellence iFIT (EXC 2180) "Image Guided and Functionally Instructed Tumor Therapies", University of Tübingen, Tübingen, Germany
| |
Collapse
|
13
|
Wiesmueller M, Wuest W, Mennecke A, May MS, Heiss R, Fuehres T, Janka R, Uder M, Doerfler A, Laun FB. Comparison of Diagnostic Performance and Image Quality between Topup-Corrected and Standard Readout-Segmented Echo-Planar Diffusion-Weighted Imaging for Cholesteatoma Diagnostics. Diagnostics (Basel) 2023; 13:diagnostics13071242. [PMID: 37046460 PMCID: PMC10093611 DOI: 10.3390/diagnostics13071242] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/15/2023] [Revised: 03/14/2023] [Accepted: 03/16/2023] [Indexed: 03/29/2023] Open
Abstract
This study compares the diagnostic performance and image quality of single-shot turbo spin-echo DWI (tseDWI), standard readout-segmented DWI (rsDWI), and a modified rsDWI version (topupDWI) for cholesteatoma diagnostics. Thirty-four patients with newly suspected unilateral cholesteatoma were examined on a 1.5 Tesla MRI scanner. Diagnostic performance was evaluated by calculating and comparing the sensitivity and specificity using histopathological results as the standard of reference. Image quality was independently reviewed by two readers using a 5-point Likert scale evaluating image distortions, susceptibility artifacts, image resolution, lesion conspicuity, and diagnostic confidence. Twenty-five cholesteatomas were histologically confirmed after surgery and originated in the study group. TseDWI showed the highest sensitivity with 96% (95% confidence interval (CI): 88–100%), followed by topupDWI with 92% (95% CI: 81–100%) for both readers. The sensitivity for rsDWI was 76% (95% CI: 59–93%) for reader 1 and 84% (95% CI: 70–98%) for reader 2, respectively. Both tseDWI and topupDWI revealed a specificity of 100% (95% CI: 66–100%) and rsDWI of 89% (95% CI: 52–100%). Both tseDWI and topupDWI showed fewer image distortions and susceptibility artifacts compared to rsDWI. Image resolution was consistently rated best for topupDWI, followed by rsDWI, which both outperformed tseDWI. TopupDWI and tseDWI showed comparable results for lesions’ conspicuity and diagnostic confidence, both outperforming rsDWI. Modified readout-segmented DWI using the topup-correction method is preferable to standard rsDWI and may be regarded as an accurate alternative to single-shot turbo spin-echo DWI in cholesteatoma diagnostics.
Collapse
|
14
|
Fang LK, Keenan KE, Carl M, Ojeda-Fournier H, Rodríguez-Soto AE, Rakow-Penner RA. Apparent Diffusion Coefficient Reproducibility Across 3 T Scanners in a Breast Diffusion Phantom. J Magn Reson Imaging 2023; 57:812-823. [PMID: 36029225 DOI: 10.1002/jmri.28355] [Citation(s) in RCA: 4] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/04/2022] [Revised: 07/04/2022] [Accepted: 07/05/2022] [Indexed: 11/06/2022] Open
Abstract
BACKGROUND To date, the accuracy and variability of diffusion-weighted MRI (DW-MRI) metrics have been reported in a limited number of scanner/protocol/coil combinations. PURPOSE To evaluate the reproducibility of DW-MRI estimates across multiple scanners and DW-MRI protocols and to assess the effects of using an 8-channel vs. 16-channel breast coil in a breast phantom. STUDY TYPE Prospective. PHANTOM Breast phantom containing tubes of water and differing polyvinylpyrrolidone (PVP) concentrations with apparent diffusion coefficients (ADCs) matching breast tissue. FIELD STRENGTH/SEQUENCE 3 T (three standard and one wide bore), three DW-MRI single-shot echo planar imaging protocols of varying acquired spatial resolution. ASSESSMENT Accuracy of estimated ADCs was assessed using percent differences (PD) relative to nuclear magnetic resonance spectroscopy-derived reference values. Coefficients of variation (CV) were calculated to determine variation across scanners. CVs based on the median standard deviation (CVM ) were used to evaluate tube-specific dispersion using 8- or 16-channel coils at a single scanner. ADCs of PVP-containing tubes were additionally normalized by the median water tube ADC to account for temperature effects. STATISTICAL TESTS Two-way repeated measures analysis of variance and post hoc tests were used to evaluate differences in ADC, CV, and CVM across scanners and protocols (α = 0.05). RESULTS ADCs were within 11% (interquartile range [IQR] 7%) of reference values and significantly improved to 2% (IQR 7%) after normalization to an internal water reference. Normalization significantly reduced interscanner variability of ADC estimates from 7% to 4%. DW-MRI protocol did not affect ADC accuracy; however, the clinical and higher-resolution clinical protocols resulted in the greatest (9%) and least (6%) interscanner variability, respectively. The 8- and 16-channel receive coils yielded similar accuracy (PD: 12% vs. 16%) and precision (CVM : 2.7% vs. 2.9%). DATA CONCLUSION Normalization by an internal reference improved interscanner ADC reproducibility. High-resolution protocols yielded comparably accurate and significantly less variable ADCs compared to a clinical standard protocol. EVIDENCE LEVEL 2 TECHNICAL EFFICACY: Stage 1.
Collapse
Affiliation(s)
- Lauren K Fang
- Department of Radiology, University of California-San Diego, La Jolla, California, USA
| | - Kathryn E Keenan
- National Institute of Science and Technology, Boulder, Colorado, USA
| | | | - Haydee Ojeda-Fournier
- Department of Radiology, University of California-San Diego, La Jolla, California, USA
| | - Ana E Rodríguez-Soto
- Department of Radiology, University of California-San Diego, La Jolla, California, USA
| | - Rebecca A Rakow-Penner
- Department of Radiology, University of California-San Diego, La Jolla, California, USA.,Department of Bioengineering, University of California-San Diego, La Jolla, California, USA
| |
Collapse
|
15
|
Sedlaczek OL, Kleesiek J, Gallagher FA, Murray J, Prinz S, Perez-Lopez R, Sala E, Caramella C, Diffetock S, Lassau N, Priest AN, Suzuki C, Vargas R, Giandini T, Vaiani M, Messina A, Blomqvist LK, Beets-Tan RGH, Oberrauch P, Schlemmer HP, Bach M. Quantification and reduction of cross-vendor variation in multicenter DWI MR imaging: results of the Cancer Core Europe imaging task force. Eur Radiol 2022; 32:8617-8628. [PMID: 35678860 PMCID: PMC9705481 DOI: 10.1007/s00330-022-08880-7] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/25/2022] [Revised: 03/25/2022] [Accepted: 05/12/2022] [Indexed: 11/25/2022]
Abstract
OBJECTIVES In the Cancer Core Europe Consortium (CCE), standardized biomarkers are required for therapy monitoring oncologic multicenter clinical trials. Multiparametric functional MRI and particularly diffusion-weighted MRI offer evident advantages for noninvasive characterization of tumor viability compared to CT and RECIST. A quantification of the inter- and intraindividual variation occurring in this setting using different hardware is missing. In this study, the MRI protocol including DWI was standardized and the residual variability of measurement parameters quantified. METHODS Phantom and volunteer measurements (single-shot T2w and DW-EPI) were performed at the seven CCE sites using the MR hardware produced by three different vendors. Repeated measurements were performed at the sites and across the sites including a traveling volunteer, comparing qualitative and quantitative ROI-based results including an explorative radiomics analysis. RESULTS For DWI/ADC phantom measurements using a central post-processing algorithm, the maximum deviation could be decreased to 2%. However, there is no significant difference compared to a decentralized ADC value calculation at the respective MRI devices. In volunteers, the measurement variation in 2 repeated scans did not exceed 11% for ADC and is below 20% for single-shot T2w in systematic liver ROIs. The measurement variation between sites amounted to 20% for ADC and < 25% for single-shot T2w. Explorative radiomics classification experiments yield better results for ADC than for single-shot T2w. CONCLUSION Harmonization of MR acquisition and post-processing parameters results in acceptable standard deviations for MR/DW imaging. MRI could be the tool in oncologic multicenter trials to overcome the limitations of RECIST-based response evaluation. KEY POINTS • Harmonizing acquisition parameters and post-processing homogenization, standardized protocols result in acceptable standard deviations for multicenter MR-DWI studies. • Total measurement variation does not to exceed 11% for ADC in repeated measurements in repeated MR acquisitions, and below 20% for an identical volunteer travelling between sites. • Radiomic classification experiments were able to identify stable features allowing for reliable discrimination of different physiological tissue samples, even when using heterogeneous imaging data.
Collapse
Affiliation(s)
- Oliver Lukas Sedlaczek
- Department of Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany.
- Division of Translational Medical Oncology, National Center for Tumor Diseases Heidelberg (NCT) and German Cancer Research Center (DKFZ), Im Neuenheimer Feld 460, 69120, Heidelberg, Germany.
- Department of Radiology, University Hospital Heidelberg, Heidelberg, Germany.
| | - Jens Kleesiek
- Department of Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany
| | | | - Jacob Murray
- Department of Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany
| | - Sebastian Prinz
- Department of Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany
| | - Raquel Perez-Lopez
- Department of Radiology, Vall d'Hebron University Hospital, Barcelona, Spain
| | - Evia Sala
- Department of Radiology, University of Cambridge and Cancer Research UK Cambridge Centre, Cambridge, UK
| | - Caroline Caramella
- Imaging Department, Gustave Roussy, BIOMAPS, UMR1281, INSERM, CEA, CNRS, Université Paris Saclay, Villejuif, Paris, France
| | - Sebastian Diffetock
- Imaging Department, Gustave Roussy, BIOMAPS, UMR1281, INSERM, CEA, CNRS, Université Paris Saclay, Villejuif, Paris, France
| | - Nathalie Lassau
- Imaging Department, Gustave Roussy, BIOMAPS, UMR1281, INSERM, CEA, CNRS, Université Paris Saclay, Villejuif, Paris, France
| | - Andrew N Priest
- Department of Radiology, University of Cambridge, Cambridge, UK
- Department of Radiology, Addenbrooke's Hospital, Cambridge, UK
| | - Chikako Suzuki
- Department of Radiation Physics and Nuclear Medicine, Karolinska University Hospital and Department of Molecular Medicine and Surgery, Karolinska Institutet, Solna, Sweden
| | - Roberto Vargas
- Department of Radiology, Karolinska University Hospital and Department of Molecular Medicine and Surgery, Karolinska Institutet, Solna, Sweden
| | - Tommaso Giandini
- Medical Physics Unit, Fondazione IRCCS Istituto Nazionale dei Tumori, Milan, Italy
| | - Marta Vaiani
- Department of Radiology, Fondazione IRCCS Istituto Nazionale dei Tumori, Milan, Italy
| | - Antonella Messina
- Department of Radiology, Fondazione IRCCS Istituto Nazionale dei Tumori, Milan, Italy
| | - Lennart K Blomqvist
- Department of Radiation Physics and Nuclear Medicine, Karolinska University Hospital and Department of Molecular Medicine and Surgery, Karolinska Institutet, Solna, Sweden
| | - Regina G H Beets-Tan
- Department of Radiology, The Netherlands Cancer Institute, Amsterdam, The Netherlands
| | - Petra Oberrauch
- Division of Translational Medical Oncology, National Center for Tumor Diseases Heidelberg (NCT) and German Cancer Research Center (DKFZ), Im Neuenheimer Feld 460, 69120, Heidelberg, Germany
| | - Heinz-Peter Schlemmer
- Department of Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany
- Division of Translational Medical Oncology, National Center for Tumor Diseases Heidelberg (NCT) and German Cancer Research Center (DKFZ), Im Neuenheimer Feld 460, 69120, Heidelberg, Germany
| | - Michael Bach
- Department of Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120, Heidelberg, Germany
| |
Collapse
|
16
|
Neri JP, Koff MF, Koch KM, Tan ET. Validating the accuracy of multispectral metal artifact suppressed diffusion-weighted imaging. Med Phys 2022; 49:6538-6546. [PMID: 35953390 PMCID: PMC9588535 DOI: 10.1002/mp.15925] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/25/2021] [Revised: 06/29/2022] [Accepted: 08/07/2022] [Indexed: 11/11/2022] Open
Abstract
BACKGROUND Diffusion-weighted imaging (DWI) provides quantitative measurement of random water displacement in tissue as calculated by the apparent diffusion coefficient (ADC). While heavily utilized in stroke and oncology applications, DWI is a promising tool to map microstructural changes in musculoskeletal applications including evaluation of synovial reactions resulting from total hip arthroplasty (THA). One major challenge facing the application of DWI in THA is the significant artifacts related to the conventional echo-planar imaging (EPI) readout used. Multispectral imaging (MSI) techniques, including the multiacquisition with variable resonance image combination (MAVRIC), have been shown to effectively reduce metallic susceptibility artifacts around total joint replacements to render clinically useful images. Recently, a 2D periodically rotated overlapping parallel line with enhanced reconstruction (PROPELLER) FSE acquisition that incorporates a diffusion preparation pulse with 2D-MAVRIC has been developed to mitigate both distortion and dropout artifacts. While there have been some preliminary assessments of DWI-MAVRIC, the repeatability of DWI-MAVRIC and the effects of key parameters, such as the number of spectral bins, are unknown. PURPOSE To evaluate the quantitative accuracy of DWI-MAVRIC as compared to conventional diffusion sequences. METHODS A diffusion phantom with different reference diffusivities (ADC = 113-1123 μm2 /s) was used. Scans were performed on two 1.5T MRI scanners. DWI-EPI and DWI-MAVRIC were acquired in both the axial and coronal planes. Three spatial offsets (0 cm, 10 cm left, and 10 cm right off iso-center) were used to evaluate effects of off-isocenter positioning. To assess intraday and interday repeatability, DWI-EPI and DWI-MAVRIC acquisitions were repeated on one scanner at same-day and 9-month intervals. To assess inter-scanner repeatability, DWI-EPI and DWI-MAVRIC acquisitions were compared between two scanners. ADC maps were generated with and without gradient nonlinearity correction (GNC). Linear regression, correlation, and error statistics were determined between calculated and reference ADC values. Bland-Altman plots were generated to evaluate intraday, interday, and interscanner repeatability. RESULTS DWI-MAVRIC had excellent correlation to reference values but at reduced linearity (r = 1.00, slope = 0.91-0.94) as compared to DWI-EPI (r = 1.00, slope = 0.99-1.01). A greater than 5% ADC bias was observed at the lowest ADC values, predominantly in the DWI-MAVRIC scans. ADC values did not vary with DWI-MAVRIC parameters. DWI-EPI acquisitions had intraday, interday, and interscanner repeatability of 3.18 μm2 /s, 19.2 μm2 /s, and 20.2 μm2 /s, respectively. DWI-MAVRIC acquisitions had inferior intraday, interday, and interscanner repeatability of 13.3 μm2 /s, 44.7 μm2 /s, 110 μm2 /s, respectively. Lower ADC errors were found at isocenter, as compared to the left and right positions. GNC reduced the absolute error by 0.31% ± 0.89%, 3.6% ± 1.4%, 0.65% ± 2.4% for the center, left, and right positions, respectively. CONCLUSIONS DWI-MAVRIC provides good linearity with respect to reference values and good intra- and interday repeatability.
Collapse
Affiliation(s)
- John P Neri
- MRI Research Laboratory, Hospital for Special Surgery, New York, New York, USA
| | - Matthew F Koff
- MRI Research Laboratory, Hospital for Special Surgery, New York, New York, USA
| | - Kevin M Koch
- Center for Imaging Research, Medical College of Wisconsin, Milwaukee, Wisconsin, USA
| | - Ek T Tan
- MRI Research Laboratory, Hospital for Special Surgery, New York, New York, USA
| |
Collapse
|
17
|
Führes T, Saake M, Lorenz J, Seuss H, Stemmer A, Benkert T, Uder M, Laun FB. Reduction of the cardiac pulsation artifact and improvement of lesion conspicuity in flow‐compensated diffusion images in the liver—A quantitative evaluation of postprocessing algorithms. Magn Reson Med 2022; 89:423-439. [PMID: 36089798 DOI: 10.1002/mrm.29427] [Citation(s) in RCA: 7] [Impact Index Per Article: 2.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/22/2022] [Revised: 08/05/2022] [Accepted: 08/08/2022] [Indexed: 11/10/2022]
Abstract
PURPOSE To enhance image quality of flow-compensated diffusion-weighted liver MRI data by increasing the lesion conspicuity and reducing the cardiac pulsation artifact using postprocessing algorithms. METHODS Diffusion-weighted image data of 40 patients with liver lesions had been acquired at 1.5 T. These data were postprocessed with 5 different algorithms (weighted averaging, p-mean, percentile, outlier exclusion, and exception set). Four image properties of the postprocessed data were evaluated for optimizing the algorithm parameters. These properties were the lesion to tissue contrast-to-noise ratio (CNR), the reduction of the cardiac pulsation artifact, the data consistency, and the vessel darkness. They were combined into a total quality score ( Q total , $$ {Q}_{\mathrm{total}}, $$ set to 1 for the trace-weighted reference image), which was used to rate the image quality objectively. RESULTS The weighted averaging algorithm performed best according to the total quality score ( Q total = 1.111 ± 0.067 $$ {Q}_{\mathrm{total}}=1.111\pm 0.067 $$ ). The further ranking was outlier exclusion algorithm ( Q total = 1.086 ± 0.061 $$ {Q}_{\mathrm{total}}=1.086\pm 0.061 $$ ), p-mean algorithm ( Q total = 1.045 ± 0.049 $$ {Q}_{\mathrm{total}}=1.045\pm 0.049 $$ ), percentile algorithm ( Q total = 1.012 ± 0.049 $$ {Q}_{\mathrm{total}}=1.012\pm 0.049 $$ ), and exception set algorithm ( Q total = 0.957 ± 0.027 $$ {Q}_{\mathrm{total}}=0.957\pm 0.027 $$ ). All optimized algorithms except for the exception set algorithm corrected the pulsation artifact and increased the lesion CNR. Changes in Q total $$ {Q}_{\mathrm{total}} $$ were significant for all optimized algorithms except for the percentile algorithm. Liver ADC was significantly reduced (except for the exception set algorithm), particularly in the left lobe. CONCLUSION Postprocessing algorithms should be used for flow-compensated liver DWI. The proposed weighted averaging algorithm seems to be suited best to increase the image quality of artifact-corrupted flow-compensated diffusion-weighted liver data.
Collapse
Affiliation(s)
- Tobit Führes
- Institute of Radiology, University Hospital Erlangen Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU) Erlangen Germany
| | - Marc Saake
- Institute of Radiology, University Hospital Erlangen Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU) Erlangen Germany
| | - Jennifer Lorenz
- Institute of Radiology, University Hospital Erlangen Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU) Erlangen Germany
| | - Hannes Seuss
- Institute of Radiology, University Hospital Erlangen Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU) Erlangen Germany
- Abteilung für Radiologie Klinikum Forchheim – Fränkische Schweiz Forchheim Germany
| | - Alto Stemmer
- MR Application Predevelopment Siemens Healthcare GmbH Erlangen Germany
| | - Thomas Benkert
- MR Application Predevelopment Siemens Healthcare GmbH Erlangen Germany
| | - Michael Uder
- Institute of Radiology, University Hospital Erlangen Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU) Erlangen Germany
| | - Frederik Bernd Laun
- Institute of Radiology, University Hospital Erlangen Friedrich‐Alexander‐Universität Erlangen‐Nürnberg (FAU) Erlangen Germany
| |
Collapse
|
18
|
Development of a standard phantom for diffusion-weighted magnetic resonance imaging quality control studies: A review. POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING 2022. [DOI: 10.2478/pjmpe-2022-0020] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/18/2022]
Abstract
Abstract
Various materials and compounds have been used in the design of diffusion-weighted magnetic resonance imaging (DWMRI) phantoms to mimic biological tissue properties, including diffusion. This review thus provides an overview of the preparations of the various DW-MRI phantoms available in relation to the limitations and strengths of materials/solutions used to fill them. The narrative review conducted from relevant databases shows that synthesizing all relevant compounds from individual liquids, gels, and solutions based on their identified strengths could contribute to the development of a novel multifunctional DW-MRI phantom. The proposed multifunctional material at varied concentrations, when filled into a multi-compartment Perspex container of cylindrical or spherical geometry, could serve as a standard DW-MRI phantom. The standard multifunctional phantom could potentially provide DW-MRI quality control test parameters in one study session.
Collapse
|
19
|
Teh I, Romero R. WA, Boyle J, Coll‐Font J, Dall'Armellina E, Ennis DB, Ferreira PF, Kalra P, Kolipaka A, Kozerke S, Lohr D, Mongeon F, Moulin K, Nguyen C, Nielles‐Vallespin S, Raterman B, Schreiber LM, Scott AD, Sosnovik DE, Stoeck CT, Tous C, Tunnicliffe EM, Weng AM, Croisille P, Viallon M, Schneider JE. Validation of cardiac diffusion tensor imaging sequences: A multicentre test-retest phantom study. NMR IN BIOMEDICINE 2022; 35:e4685. [PMID: 34967060 PMCID: PMC9285553 DOI: 10.1002/nbm.4685] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Figures] [Subscribe] [Scholar Register] [Received: 05/24/2021] [Revised: 11/19/2021] [Accepted: 12/24/2021] [Indexed: 05/23/2023]
Abstract
Cardiac diffusion tensor imaging (DTI) is an emerging technique for the in vivo characterisation of myocardial microstructure, and there is a growing need for its validation and standardisation. We sought to establish the accuracy, precision, repeatability and reproducibility of state-of-the-art pulse sequences for cardiac DTI among 10 centres internationally. Phantoms comprising 0%-20% polyvinylpyrrolidone (PVP) were scanned with DTI using a product pulsed gradient spin echo (PGSE; N = 10 sites) sequence, and a custom motion-compensated spin echo (SE; N = 5) or stimulated echo acquisition mode (STEAM; N = 5) sequence suitable for cardiac DTI in vivo. A second identical scan was performed 1-9 days later, and the data were analysed centrally. The average mean diffusivities (MDs) in 0% PVP were (1.124, 1.130, 1.113) x 10-3 mm2 /s for PGSE, SE and STEAM, respectively, and accurate to within 1.5% of reference data from the literature. The coefficients of variation in MDs across sites were 2.6%, 3.1% and 2.1% for PGSE, SE and STEAM, respectively, and were similar to previous studies using only PGSE. Reproducibility in MD was excellent, with mean differences in PGSE, SE and STEAM of (0.3 ± 2.3, 0.24 ± 0.95, 0.52 ± 0.58) x 10-5 mm2 /s (mean ± 1.96 SD). We show that custom sequences for cardiac DTI provide accurate, precise, repeatable and reproducible measurements. Further work in anisotropic and/or deforming phantoms is warranted.
Collapse
Affiliation(s)
- Irvin Teh
- Leeds Institute of Cardiovascular and Metabolic MedicineUniversity of LeedsLeedsUK
| | - William A. Romero R.
- Univ Lyon, INSA‐Lyon, Université Claude Bernard Lyon 1UJM‐Saint Etienne, CNRS, Inserm, CREATIS UMR 5220, U1294, F‐42023Saint EtienneFrance
| | - Jordan Boyle
- School of Mechanical EngineeringUniversity of LeedsLeedsUK
| | - Jaume Coll‐Font
- Cardiovascular Research Center and A. A. Martinos Center for Biomedical ImagingMassachusetts General Hospital and Harvard Medical SchoolBostonMassachusettsUSA
| | - Erica Dall'Armellina
- Leeds Institute of Cardiovascular and Metabolic MedicineUniversity of LeedsLeedsUK
| | - Daniel B. Ennis
- Division of RadiologyVA Palo Alto Health Care SystemPalo AltoCaliforniaUSA
- Department of RadiologyStanford UniversityStanfordCaliforniaUSA
| | - Pedro F. Ferreira
- Cardiovascular Magnetic Resonance UnitThe Royal Brompton and Harefield NHS Foundation TrustLondonUK
- National Heart and Lung InstituteImperial College LondonLondonUK
| | - Prateek Kalra
- Department of RadiologyThe Ohio State University Wexner Medical CenterColumbusOhioUSA
| | - Arunark Kolipaka
- Department of RadiologyThe Ohio State University Wexner Medical CenterColumbusOhioUSA
| | - Sebastian Kozerke
- Institute for Biomedical EngineeringUniversity and ETH ZurichZurichSwitzerland
| | - David Lohr
- Department of Cardiovascular ImagingComprehensive Heart Failure CenterWürzburgGermany
| | | | - Kévin Moulin
- Department of RadiologyStanford UniversityStanfordCaliforniaUSA
| | - Christopher Nguyen
- Cardiovascular Research Center and A. A. Martinos Center for Biomedical ImagingMassachusetts General Hospital and Harvard Medical SchoolBostonMassachusettsUSA
| | - Sonia Nielles‐Vallespin
- Cardiovascular Magnetic Resonance UnitThe Royal Brompton and Harefield NHS Foundation TrustLondonUK
- National Heart and Lung InstituteImperial College LondonLondonUK
| | - Brian Raterman
- Department of RadiologyThe Ohio State University Wexner Medical CenterColumbusOhioUSA
| | - Laura M. Schreiber
- Department of Cardiovascular ImagingComprehensive Heart Failure CenterWürzburgGermany
| | - Andrew D. Scott
- Cardiovascular Magnetic Resonance UnitThe Royal Brompton and Harefield NHS Foundation TrustLondonUK
- National Heart and Lung InstituteImperial College LondonLondonUK
| | - David E. Sosnovik
- Cardiovascular Research Center and A. A. Martinos Center for Biomedical ImagingMassachusetts General Hospital and Harvard Medical SchoolBostonMassachusettsUSA
| | - Christian T. Stoeck
- Institute for Biomedical EngineeringUniversity and ETH ZurichZurichSwitzerland
| | - Cyril Tous
- Department of Radiology, Radiation‐Oncology and Nuclear Medicine and Institute of Biomedical EngineeringUniversité de MontréalMontréalCanada
| | - Elizabeth M. Tunnicliffe
- Radcliffe Department of MedicineUniversity of OxfordOxfordUK
- Oxford NIHR Biomedical Research CentreOxfordUK
| | - Andreas M. Weng
- Department of Diagnostic and Interventional RadiologyUniversity Hospital WürzburgWürzburgGermany
| | - Pierre Croisille
- Univ Lyon, INSA‐Lyon, Université Claude Bernard Lyon 1UJM‐Saint Etienne, CNRS, Inserm, CREATIS UMR 5220, U1294, F‐42023Saint EtienneFrance
| | - Magalie Viallon
- Univ Lyon, INSA‐Lyon, Université Claude Bernard Lyon 1UJM‐Saint Etienne, CNRS, Inserm, CREATIS UMR 5220, U1294, F‐42023Saint EtienneFrance
| | - Jürgen E. Schneider
- Leeds Institute of Cardiovascular and Metabolic MedicineUniversity of LeedsLeedsUK
| |
Collapse
|
20
|
Amouzandeh G, Chenevert TL, Swanson SD, Ross BD, Malyarenko DI. Technical Note: Temperature and concentration dependence of water diffusion in polyvinylpyrrolidone solutions. Med Phys 2022; 49:3325-3332. [PMID: 35184316 PMCID: PMC9090959 DOI: 10.1002/mp.15556] [Citation(s) in RCA: 4] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/19/2021] [Revised: 01/27/2022] [Accepted: 02/08/2022] [Indexed: 11/26/2022] Open
Abstract
Objective The goal of this work is to provide temperature and concentration calibration of water diffusivity in polyvinylpyrrolidone (PVP) solutions used in phantoms to assess system bias and linearity in apparent diffusion coefficient (ADC) measurements. Method ADC measurements were performed for 40 kDa (K40) PVP of six concentrations (0%, 10%, 20%, 30%, 40%, and 50% by weight) at three temperatures (19.5°C, 22.5°C, and 26.4°C), with internal phantom temperature monitored by optical thermometer (±0.2°C). To achieve ADC measurement and fit accuracy of better than 0.5%, three orthogonal diffusion gradients were calibrated using known water diffusivity at 0°C and system gradient nonlinearity maps. Noise‐floor fit bias was also controlled by limiting the maximum b‐value used for ADC calculation of each sample. The ADC temperature dependence was modeled by Arrhenius functions of each PVP concentration. The concentration dependence was modeled by quadratic function for ADC normalized by the theoretical water diffusion values. Calibration coefficients were obtained from linear regression model fits. Results Measured phantom ADC values increased with temperature and decreasing PVP concentration, [PVP]. The derived Arrhenius model parameters for [PVP] between 0% and 50%, are reported and can be used for K40 ADC temperature calibration with absolute ADC error within ±0.016 μm2/ms. Arrhenius model fit parameters normalized to water value scaled with [PVP] between 10% and 40%, and proportional change in activation energy increased faster than collision frequency. ADC normalization by water diffusivity, DW, from the Speedy–Angell relation accounted for the bulk of temperature dependence (±0.035 μm2/ms) and yielded quadratic calibration for ADCPVP/DW = (12.5 ± 0.7) ·10−5·[PVP]2 − (23.2 ± 0.3)·10−3·[PVP]+1, nearly independent of PVP molecular weight and temperature. Conclusion The study provides ground‐truth ADC values for K40 PVP solutions commonly used in diffusion phantoms for scanning at ambient room temperature. The described procedures and the reported calibration can be used for quality control and standardization of measured ADC values of PVP at different concentrations and temperatures.
Collapse
Affiliation(s)
| | | | | | - Brian D. Ross
- Department of Radiology University of Michigan Ann Arbor MI USA
| | | |
Collapse
|
21
|
Ohlmeyer S, Laun FB, Bickelhaupt S, Palm T, Janka R, Weiland E, Uder M, Wenkel E. Ultra-High b-Value Diffusion-Weighted Imaging-Based Abbreviated Protocols for Breast Cancer Detection. Invest Radiol 2021; 56:629-636. [PMID: 34494995 DOI: 10.1097/rli.0000000000000784] [Citation(s) in RCA: 13] [Impact Index Per Article: 3.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/17/2022]
Abstract
OBJECTIVES Contrast-enhanced (CE) magnetic resonance imaging (MRI) is the most effective imaging modality for breast cancer detection. A contrast agent-free examination technique would be desirable for breast MRI screening. The purpose of this study was to evaluate the capability to detect and characterize suspicious breast lesions with an abbreviated, non-contrast-enhanced MRI protocol featuring ultra-high b-value diffusion-weighted imaging (DWI) compared with CE images. MATERIALS AND METHODS The institutional review board-approved prospective study included 127 female subjects with different clinical indications for breast MRI. Magnetic resonance imaging examinations included DWI sequences with b-values of 1500 s/mm2 (b1500) and 2500 s/mm2 (b2500), native T1- and T2-weighted images, and CE sequences at 1.5 T and 3 T scanners. Two reading rounds were performed, including either the b1500 or the b2500 DWI in consecutive assessment steps: (A) maximum intensity projections (MIPs) of DWI, (B) DWI and apparent diffusion coefficient maps, (C) as (B) but with additional native T1- and T2-weighted images, and (D) as (C) but with additional CE images (full-length protocol). Two readers independently determined the presence of a suspicious lesion. Histological confirmation was obtained for conspicuous lesions, whereas the full MRI data set was obtained for inconspicuous and clearly benign lesions. Statistical analysis included calculation of diagnostic accuracy and interrater agreement via the intraclass correlation coefficient. RESULTS The cohort comprised 116 cases with BI-RADS 1 findings and 138 cases with BI-RADS ≥2 findings, including 38 histologically confirmed malignancies. For (A), breasts without pathological findings could be recognized with high diagnostic accuracy (negative predictive value, ≥97.0%; sensitivity, ≥92.1% for both readers), but with a limited specificity (≥58.3%; positive predictive value, ≥28.6%). Within the native readings, approach (C) with b2500 performed best (negative predictive value, 99.5%; sensitivity, 97.4%; specificity, 88.4%). The intraclass correlation coefficient was between 0.683 (MIP b1500) and 0.996 (full protocol). CONCLUSIONS A native abbreviated breast MRI protocol with advanced high b-value DWI might allow nearly equivalent diagnostic accuracy as CE breast MRI and seems to be well suited for lesion detection purposes.
Collapse
Affiliation(s)
- Sabine Ohlmeyer
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg
| | - Frederik Bernd Laun
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg
| | - Sebastian Bickelhaupt
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg
| | - Theresa Palm
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg
| | - Rolf Janka
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg
| | | | - Michael Uder
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg
| | - Evelyn Wenkel
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg
| |
Collapse
|
22
|
Saake M, Seuß H, Riexinger A, Bickelhaupt S, Hammon M, Uder M, Laun FB. Image Quality and Detection of Small Focal Liver Lesions in Diffusion-Weighted Imaging: Comparison of Navigator Tracking and Free-Breathing Acquisition. Invest Radiol 2021; 56:579-590. [PMID: 33813572 DOI: 10.1097/rli.0000000000000776] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/24/2022]
Abstract
OBJECTIVES The aim of this study was to compare intraindividual diffusion-weighted imaging (DWI) of the liver acquired with free breathing (FB) versus navigator triggering (NT) for assessing small focal liver lesions (FLLs) in noncirrhotic patients. MATERIALS AND METHODS Patients with known or suspected multiple FLLs were prospectively included, and spin-echo echo-planar DWI with NT and FB acquisition was performed (b-values, 50 and 800 s/mm2 [b50 and b800]). NT and FB DWI sequences with similar acquisitions times were used. Liver and lesion signal-to-noise ratios were measured at b800. The DWI scans were analyzed independently by 2 readers. Liver edge delineation, presence of stair-step artifacts, vessel sharpness, severity of cardiac motion artifacts, overall image quality, and lesion conspicuity were rated with 5-point Likert scales. Small and large FLLs (ie, <1 cm or ≥1 cm) were rated separately for lesion conspicuity. The FLL detectability was estimated by comparing the number of lesions visible with FB to those visible with NT. RESULTS Forty-three patients were included in the study. The FB acquisition performed better in terms of severity of cardiac motion artifacts. The NT performed better in terms of liver edge delineation and vessel sharpness. Little difference was found for stair-step artifact, overall image quality, and conspicuity of large FLL, whereas the conspicuity of small FLL was better for NT. For small FLL, both readers found more lesions with NT in 11 cases at b800. For large FLL, this effect was much less pronounced (1 case at b800 reported by 1 of the readers). The mean liver and lesion signal-to-noise ratios were 16.8/41.5 and 19.8/38.4 for NT/FB, respectively. CONCLUSIONS Small FLL detection is better with NT. Large FLL detection by FB and NT is similarly good. We conclude that NT should be used.
Collapse
Affiliation(s)
- Marc Saake
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen
| | | | - Andreas Riexinger
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen
| | - Sebastian Bickelhaupt
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen
| | - Matthias Hammon
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen
| | - Michael Uder
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen
| | - Frederik B Laun
- From the Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen
| |
Collapse
|
23
|
Li X, Bai Y, Liao Y, Xin SX. Assessment of the effects of mimicking tissue microstructural properties on apparent diffusion coefficient and apparent exchange rate in diffusion MRI via a series of specially designed phantoms. Magn Reson Med 2021; 87:292-301. [PMID: 34435698 DOI: 10.1002/mrm.28990] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/15/2021] [Revised: 08/04/2021] [Accepted: 08/09/2021] [Indexed: 11/11/2022]
Abstract
PURPOSE Diffusion MRI provides a valuable tool for imaging tissue microstructure. However, due to the lack of related experimental methods and specially designed phantoms, no experimental study has been conducted yet to quantitatively assess the effects of membrane permeability, intracellular volume fraction (IVF), and intracellular diffusivity on the apparent diffusion coefficient (ADC) obtained from diffusion weighted imaging (DWI), and the effects of membrane permeability on the apparent exchange rate (AXR) obtained from filter exchange imaging (FEXI). METHODS A series of phantoms with three adjustable parameters was designed to mimic tissue microstructural properties including membrane permeability, IVF, and intracellular diffusivity. Quantitative experiments were conducted to assess the effects of these properties on ADC and AXR. DWI scans were performed to obtain axial and radial ADC values. FEXI scans were performed to obtain AXR values. RESULTS Axial ADC values range from 1.148 μm2 /ms to 2.157 μm2 /ms, and radial ADC values range from 0.904 μm2 /ms to 2.067 μm2 /ms. Radial ADC decreased with a decrease in fiber permeability. Decreased axial and radial ADC values with increased intra-fiber volume fraction, and increased polyvinylpyrrolidone (PVP) concentration of the intra-fiber space were observed. AXR values range from 2.1 s-1 to 4.9 s-1 . AXR increases with fiber permeability. CONCLUSION The proposed phantoms can quantitatively evaluate the effects of mimicking tissue microstructural properties on ADC and AXR. This new phantom design provides a potential method for further understanding the biophysical mechanisms underlying the change in ADC and diffusion exchange.
Collapse
Affiliation(s)
- Xiaodong Li
- Laboratory of Biophysics, School of Medicine, South China University of Technology, Guangzhou, Guangdong, China
| | - Yafei Bai
- Laboratory of Biophysics, School of Medicine, South China University of Technology, Guangzhou, Guangdong, China
| | - Yupeng Liao
- Laboratory of Biophysics, School of Medicine, South China University of Technology, Guangzhou, Guangdong, China
| | - Sherman Xuegang Xin
- Laboratory of Biophysics, School of Medicine, South China University of Technology, Guangzhou, Guangdong, China
| |
Collapse
|
24
|
Iima M, Kataoka M, Honda M, Ohashi A, Ohno Kishimoto A, Ota R, Uozumi R, Urushibata Y, Feiweier T, Toi M, Nakamoto Y. The Rate of Apparent Diffusion Coefficient Change With Diffusion Time on Breast Diffusion-Weighted Imaging Depends on Breast Tumor Types and Molecular Prognostic Biomarker Expression. Invest Radiol 2021; 56:501-508. [PMID: 33660629 DOI: 10.1097/rli.0000000000000766] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/25/2022]
Abstract
INTRODUCTION The aim of this study was to investigate the variation of apparent diffusion coefficient (ADC) values with diffusion time according to breast tumor type and prognostic biomarkers expression. MATERIALS AND METHODS A total of 201 patients with known or suspected breast tumors were prospectively enrolled in this study, and 132 breast tumors (86 malignant and 46 benign) were analyzed. Diffusion-weighted imaging scans with 2 diffusion times were acquired on a clinical 3-T magnetic resonance imaging scanner using oscillating and pulsed diffusion-encoding gradients (effective diffusion times, 4.7 and 96.6 milliseconds) and b values of 0 and 700 s/mm2. Diagnostic performances to differentiate malignant and benign breast tumors for ADC values at short and long diffusion times (ADCshort and ADClong), ΔADC (the rate of change in ADC values with diffusion time), ADC0-1000 (ADC value from a standard protocol), and standard reading including dynamic contrast-enhanced magnetic resonance imaging (BI-RADS) were investigated. The correlations of ADCshort, ADClong, and ΔADC values with hormone receptor expression and breast cancer subtypes were also analyzed. RESULTS The ADC values were lower, and ΔADC was higher in malignant tumors compared with benign tumors. The specificity of ADC values at all diffusion times and ΔADC values for differentiating malignant and benign breast tumors was superior to that of BI-RADS (87.0%-95.7% vs 73.9%), whereas the sensitivity was inferior (87.2%-90.7% vs 100%). Lower ADCshort and ADC0-1000 in ER-positive compared with ER-negative cancers (false discovery rate [FDR]-adjusted P = 0.037 and 0.018, respectively) and lower ADCshort, ADClong, and ADC0-1000 in progesterone receptor-positive compared with progesterone receptor-negative cancers (FDR-adjusted P = 0.037, 0.036, and 0.018, respectively) were found. Ki-67-positive cancers had larger ΔADCs than Ki-67-negative cancers (FDR-adjusted P = 0.018). CONCLUSIONS The ADC values vary with different diffusion time and vary in correlation with molecular biomarkers, especially Ki-67. Those results suggest that the diffusion time, which should be reported, might be a useful parameter to consider for breast cancer management.
Collapse
Affiliation(s)
| | - Masako Kataoka
- From the Department of Diagnostic Imaging and Nuclear Medicine, Kyoto University Graduate School of Medicine
| | - Maya Honda
- From the Department of Diagnostic Imaging and Nuclear Medicine, Kyoto University Graduate School of Medicine
| | | | | | - Rie Ota
- From the Department of Diagnostic Imaging and Nuclear Medicine, Kyoto University Graduate School of Medicine
| | - Ryuji Uozumi
- Department of Biomedical Statistics and Bioinformatics, Kyoto University Graduate School of Medicine, Kyoto, Japan
| | | | | | - Masakazu Toi
- Department of Breast Surgery, Kyoto University Graduate School of Medicine, Kyoto, Japan
| | - Yuji Nakamoto
- From the Department of Diagnostic Imaging and Nuclear Medicine, Kyoto University Graduate School of Medicine
| |
Collapse
|
25
|
Bhuiyan EH, Dewdney A, Weinreb J, Galiana G. Feasibility of diffusion weighting with a local inside-out nonlinear gradient coil for prostate MRI. Med Phys 2021; 48:5804-5818. [PMID: 34287937 DOI: 10.1002/mp.15100] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 04/22/2020] [Revised: 04/04/2021] [Accepted: 06/23/2021] [Indexed: 11/09/2022] Open
Abstract
PURPOSE Prostate cancer remains the 2nd leading cancer killer of men, yet it is also a disease with a high rate of overtreatment. Diffusion weighted imaging (DWI) has shown promise as a reliable, grade-sensitive imaging method, but it is limited by low image quality. Currently, DWI quality image is directly related to low gradient amplitudes, since weak gradients must be compensated with long echo times. METHODS We propose a new type of MRI accessory, an "inside-out" and nonlinear gradient, whose sole purpose is to deliver diffusion encoding to a region of interest. Performance was simulated in OPERA and the resulting fields were used to simulate DWI with two compartment and kurtosis models. Experiments with a nonlinear head gradient prove the accuracy of DWI and ADC maps diffusion encoded with nonlinear gradients. RESULTS Simulations validated thermal and mechanical safety while showing a 5 to 10-fold increase in gradient strength over prostate. With these strengths, lesion CNR in ADC maps approximately doubled for a range of anatomical positions. Proof-of-principle experiments show that spatially varying b-values can be corrected for accurate DWI and ADC. CONCLUSIONS Dedicated nonlinear diffusion encoding hardware could improve prostate DWI.
Collapse
Affiliation(s)
| | | | - Jeffrey Weinreb
- Department of Radiology and Biomedical Imaging, Yale School of Medicine, USA
| | - Gigi Galiana
- Department of Radiology and Biomedical Imaging, Yale School of Medicine, USA
| |
Collapse
|
26
|
Avram AV, Sarlls JE, Basser PJ. Whole-Brain Imaging of Subvoxel T1-Diffusion Correlation Spectra in Human Subjects. Front Neurosci 2021; 15:671465. [PMID: 34177451 PMCID: PMC8232058 DOI: 10.3389/fnins.2021.671465] [Citation(s) in RCA: 11] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/23/2021] [Accepted: 05/14/2021] [Indexed: 12/12/2022] Open
Abstract
T1 relaxation and water mobility generate eloquent MRI tissue contrasts with great diagnostic value in many neuroradiological applications. However, conventional methods do not adequately quantify the microscopic heterogeneity of these important biophysical properties within a voxel, and therefore have limited biological specificity. We describe a new correlation spectroscopic (CS) MRI method for measuring how T1 and mean diffusivity (MD) co-vary in microscopic tissue environments. We develop a clinical pulse sequence that combines inversion recovery (IR) with single-shot isotropic diffusion encoding (IDE) to efficiently acquire whole-brain MRIs with a wide range of joint T1-MD weightings. Unlike conventional diffusion encoding, the IDE preparation ensures that all subvoxel water pools are weighted by their MDs regardless of the sizes, shapes, and orientations of their corresponding microscopic diffusion tensors. Accordingly, IR-IDE measurements are well-suited for model-free, quantitative spectroscopic analysis of microscopic water pools. Using numerical simulations, phantom experiments, and data from healthy volunteers we demonstrate how IR-IDE MRIs can be processed to reconstruct maps of two-dimensional joint probability density functions, i.e., correlation spectra, of subvoxel T1-MD values. In vivo T1-MD spectra show distinct cerebrospinal fluid and parenchymal tissue components specific to white matter, cortical gray matter, basal ganglia, and myelinated fiber pathways, suggesting the potential for improved biological specificity. The one-dimensional marginal distributions derived from the T1-MD correlation spectra agree well with results from other relaxation spectroscopic and quantitative MRI studies, validating the T1-MD contrast encoding and the spectral reconstruction. Mapping subvoxel T1-diffusion correlations in patient populations may provide a more nuanced, comprehensive, sensitive, and specific neuroradiological assessment of the non-specific changes seen on fluid-attenuated inversion recovery (FLAIR) and diffusion-weighted MRIs (DWIs) in cancer, ischemic stroke, or brain injury.
Collapse
Affiliation(s)
- Alexandru V Avram
- Eunice Kennedy Shriver National Institute of Child Health and Human Development, National Institutes of Health, Bethesda, MD, United States.,Center for Neuroscience and Regenerative Medicine, The Henry M. Jackson Foundation for the Advancement of Military Medicine, Bethesda, MD, United States
| | - Joelle E Sarlls
- National Institute of Neurological Disorders and Stroke, National Institutes of Health, Bethesda, MD, United States
| | - Peter J Basser
- Eunice Kennedy Shriver National Institute of Child Health and Human Development, National Institutes of Health, Bethesda, MD, United States
| |
Collapse
|
27
|
Wiesmueller M, Wuest W, May MS, Ellmann S, Heiss R, Saake M, Janka R, Uder M, Laun FB. Comparison of Readout-Segmented Echo-Planar Imaging and Single-Shot TSE DWI for Cholesteatoma Diagnostics. AJNR Am J Neuroradiol 2021; 42:1305-1312. [PMID: 33926901 DOI: 10.3174/ajnr.a7112] [Citation(s) in RCA: 8] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/10/2020] [Accepted: 01/23/2021] [Indexed: 01/18/2023]
Abstract
BACKGROUND AND PURPOSE The high diagnostic value of DWI for cholesteatoma diagnostics is undisputed. This study compares the diagnostic value of readout-segmented echo-planar DWI and single-shot TSE DWI for cholesteatoma diagnostics. MATERIALS AND METHODS Thirty patients with newly suspected cholesteatoma were examined with a dedicated protocol, including readout-segmented echo-planar DWI and single-shot TSE DWI at 1.5T. Acquisition parameters of both diffusion-weighted sequences were as follows: b=1000 s/mm,2 axial and coronal section orientations, and section thickness of 3 mm. Image quality was evaluated by 2 readers on a 5-point Likert scale with respect to lesion conspicuity, the presence of susceptibility artifacts mimicking cholesteatomas, and overall subjective image quality. Sensitivity and specificity were calculated using histology results as the gold standard. RESULTS Twenty-five cases of histologically confirmed cholesteatomas were included in the study group. Lesion conspicuity was higher and fewer artifacts were found when using TSE DWI (both P < .001). The overall subjective image quality, however, was better with readout-segmented DWI. For TSE DWI, the sensitivity for readers 1 and 2 was 92% (95% CI, 74%-99%) and 88% (95% CI, 69%-97%), respectively, while the specificity for both readers was 80% (95% CI, 28%-99%). For readout-segmented DWI, the sensitivity for readers 1 and 2 was 76% (95% CI, 55%-91%) and 68% (95% CI, 46%-85%), while the specificity for both readers was 60% (95% CI, 15%-95%). CONCLUSIONS The use of TSE DWI is advisable for cholesteatoma diagnostics and preferable over readout-segmented DWI.
Collapse
Affiliation(s)
- M Wiesmueller
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.) .,Image Science Institute (M.W., W.W., M.S.M., R.H., M.S., R.J., M.U.), University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany
| | - W Wuest
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.).,Image Science Institute (M.W., W.W., M.S.M., R.H., M.S., R.J., M.U.), University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany
| | - M S May
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.).,Image Science Institute (M.W., W.W., M.S.M., R.H., M.S., R.J., M.U.), University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany
| | - S Ellmann
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.)
| | - R Heiss
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.).,Image Science Institute (M.W., W.W., M.S.M., R.H., M.S., R.J., M.U.), University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany
| | - M Saake
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.).,Image Science Institute (M.W., W.W., M.S.M., R.H., M.S., R.J., M.U.), University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany
| | - R Janka
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.).,Image Science Institute (M.W., W.W., M.S.M., R.H., M.S., R.J., M.U.), University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany
| | - M Uder
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.).,Image Science Institute (M.W., W.W., M.S.M., R.H., M.S., R.J., M.U.), University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany
| | - F B Laun
- From the Institute of Radiology (M.W., W.W., M.S.M., S.E., R.H., M.S., R.J., M.U., F.B.L.)
| |
Collapse
|
28
|
Respiratory Motion Mitigation and Repeatability of Two Diffusion-Weighted MRI Methods Applied to a Murine Model of Spontaneous Pancreatic Cancer. ACTA ACUST UNITED AC 2021; 7:66-79. [PMID: 33704226 PMCID: PMC8048371 DOI: 10.3390/tomography7010007] [Citation(s) in RCA: 1] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/12/2020] [Accepted: 02/02/2021] [Indexed: 12/31/2022]
Abstract
Respiratory motion and increased susceptibility effects at high magnetic fields pose challenges for quantitative diffusion-weighted MRI (DWI) of a mouse abdomen on preclinical MRI systems. We demonstrate the first application of radial k-space-sampled (RAD) DWI of a mouse abdomen using a genetically engineered mouse model of pancreatic ductal adenocarcinoma (PDAC) on a 4.7 T preclinical scanner equipped with moderate gradient capability. RAD DWI was compared with the echo-planar imaging (EPI)-based DWI method with similar voxel volumes and acquisition times over a wide range of b-values (0.64, 535, 1071, 1478, and 2141 mm2/s). The repeatability metrics are assessed in a rigorous test-retest study (n = 10 for each DWI protocol). The four-shot EPI DWI protocol leads to higher signal-to-noise ratio (SNR) in diffusion-weighted images with persisting ghosting artifacts, whereas the RAD DWI protocol produces relatively artifact-free images over all b-values examined. Despite different degrees of motion mitigation, both RAD DWI and EPI DWI allow parametric maps of apparent diffusion coefficients (ADC) to be produced, and the ADC of the PDAC tumor estimated by the two methods are 1.3 ± 0.24 and 1.5 ± 0.28 × 10-3 mm2/s, respectively (p = 0.075, n = 10), and those of a water phantom are 3.2 ± 0.29 and 2.8 ± 0.15 × 10-3 mm2/s, respectively (p = 0.001, n = 10). Bland-Altman plots and probability density function reveal good repeatability for both protocols, whose repeatability metrics do not differ significantly. In conclusion, RAD DWI enables a more effective respiratory motion mitigation but lower SNR, while the performance of EPI DWI is expected to improve with more advanced gradient hardware.
Collapse
|
29
|
Hagiwara A, Fujita S, Ohno Y, Aoki S. Variability and Standardization of Quantitative Imaging: Monoparametric to Multiparametric Quantification, Radiomics, and Artificial Intelligence. Invest Radiol 2020; 55:601-616. [PMID: 32209816 PMCID: PMC7413678 DOI: 10.1097/rli.0000000000000666] [Citation(s) in RCA: 73] [Impact Index Per Article: 14.6] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/20/2020] [Accepted: 01/28/2020] [Indexed: 12/19/2022]
Abstract
Radiological images have been assessed qualitatively in most clinical settings by the expert eyes of radiologists and other clinicians. On the other hand, quantification of radiological images has the potential to detect early disease that may be difficult to detect with human eyes, complement or replace biopsy, and provide clear differentiation of disease stage. Further, objective assessment by quantification is a prerequisite of personalized/precision medicine. This review article aims to summarize and discuss how the variability of quantitative values derived from radiological images are induced by a number of factors and how these variabilities are mitigated and standardization of the quantitative values are achieved. We discuss the variabilities of specific biomarkers derived from magnetic resonance imaging and computed tomography, and focus on diffusion-weighted imaging, relaxometry, lung density evaluation, and computer-aided computed tomography volumetry. We also review the sources of variability and current efforts of standardization of the rapidly evolving techniques, which include radiomics and artificial intelligence.
Collapse
Affiliation(s)
- Akifumi Hagiwara
- From the Department of Radiology, Juntendo University School of Medicine, Tokyo
| | | | - Yoshiharu Ohno
- Department of Radiology, Fujita Health University School of Medicine, Toyoake, Aichi, Japan
| | - Shigeki Aoki
- From the Department of Radiology, Juntendo University School of Medicine, Tokyo
| |
Collapse
|
30
|
Winter L, Silemek B, Petzold J, Pfeiffer H, Hoffmann W, Seifert F, Ittermann B. Parallel transmission medical implant safety testbed: Real‐time mitigation of RF induced tip heating using time‐domain E‐field sensors. Magn Reson Med 2020; 84:3468-3484. [DOI: 10.1002/mrm.28379] [Citation(s) in RCA: 5] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/08/2019] [Revised: 05/19/2020] [Accepted: 05/22/2020] [Indexed: 12/15/2022]
Affiliation(s)
- Lukas Winter
- Physikalisch‐Technische Bundesanstalt (PTB) Braunschweig and Berlin Germany
| | - Berk Silemek
- Physikalisch‐Technische Bundesanstalt (PTB) Braunschweig and Berlin Germany
| | - Johannes Petzold
- Physikalisch‐Technische Bundesanstalt (PTB) Braunschweig and Berlin Germany
| | - Harald Pfeiffer
- Physikalisch‐Technische Bundesanstalt (PTB) Braunschweig and Berlin Germany
| | - Werner Hoffmann
- Physikalisch‐Technische Bundesanstalt (PTB) Braunschweig and Berlin Germany
| | - Frank Seifert
- Physikalisch‐Technische Bundesanstalt (PTB) Braunschweig and Berlin Germany
| | - Bernd Ittermann
- Physikalisch‐Technische Bundesanstalt (PTB) Braunschweig and Berlin Germany
| |
Collapse
|
31
|
Farrher E, Grinberg F, Kuo LW, Cho KH, Buschbeck RP, Chen MJ, Chiang HH, Choi CH, Shah NJ. Dedicated diffusion phantoms for the investigation of free water elimination and mapping: insights into the influence of T 2 relaxation properties. NMR IN BIOMEDICINE 2020; 33:e4210. [PMID: 31926122 DOI: 10.1002/nbm.4210] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 05/09/2018] [Revised: 09/16/2019] [Accepted: 10/11/2019] [Indexed: 06/10/2023]
Abstract
Conventional diffusion-weighted (DW) MRI suffers from free water contamination due to the finite voxel size. The most common case of free water contamination occurs with cerebrospinal fluid (CSF) in voxels located at the CSF-tissue interface, such as at the ventricles in the human brain. Another case refers to intra-tissue free water as in vasogenic oedema. In order to avoid the bias in diffusion metrics, several multi-compartment methods have been introduced, which explicitly model the presence of a free water compartment. However, fitting multi-compartment models in DW MRI represents a well known ill conditioned problem. Although during the last decade great effort has been devoted to mitigating this estimation problem, the research field remains active. The aim of this work is to introduce the design, characterise the NMR properties and demonstrate the use of two dedicated anisotropic diffusion fibre phantoms, useful for the study of free water elimination (FWE) and mapping models. In particular, we investigate the recently proposed FWE diffusion tensor imaging approach, which takes explicit account of differences in the transverse relaxation times between the free water and tissue compartments.
Collapse
Affiliation(s)
- Ezequiel Farrher
- Institute of Neuroscience and Medicine 4, Medical Imaging Physics, Forschungszentrum Jülich, Jülich, Germany
| | - Farida Grinberg
- Institute of Neuroscience and Medicine 4, Medical Imaging Physics, Forschungszentrum Jülich, Jülich, Germany
| | - Li-Wei Kuo
- Institute of Biomedical Engineering and Nanomedicine, National Health Research Institutes, Miaoli, Taiwan
- Institute of Medical Device and Imaging, National Taiwan University College of Medicine, Taipei, Taiwan
| | - Kuan-Hung Cho
- Institute of Biomedical Engineering and Nanomedicine, National Health Research Institutes, Miaoli, Taiwan
| | - Richard P Buschbeck
- Institute of Neuroscience and Medicine 4, Medical Imaging Physics, Forschungszentrum Jülich, Jülich, Germany
| | - Ming-Jye Chen
- Institute of Biomedical Engineering and Nanomedicine, National Health Research Institutes, Miaoli, Taiwan
| | - Husan-Han Chiang
- Institute of Biomedical Engineering and Nanomedicine, National Health Research Institutes, Miaoli, Taiwan
| | - Chang-Hoon Choi
- Institute of Neuroscience and Medicine 4, Medical Imaging Physics, Forschungszentrum Jülich, Jülich, Germany
| | - N Jon Shah
- Institute of Neuroscience and Medicine 4, Medical Imaging Physics, Forschungszentrum Jülich, Jülich, Germany
- Department of Neurology, Faculty of Medicine, RWTH Aachen University, Aachen, Germany
- JARA BRAIN Translational Medicine, RWTH Aachen University, Aachen, Germany
- Institute of Neuroscience and Medicine 11,JARA, Forschungszentrum Jülich, Jülich, Germany
| |
Collapse
|
32
|
Aggarwal M, Smith MD, Calabresi PA. Diffusion-time dependence of diffusional kurtosis in the mouse brain. Magn Reson Med 2020; 84:1564-1578. [PMID: 32022313 DOI: 10.1002/mrm.28189] [Citation(s) in RCA: 21] [Impact Index Per Article: 4.2] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/01/2019] [Revised: 01/06/2020] [Accepted: 01/07/2020] [Indexed: 12/28/2022]
Abstract
PURPOSE To investigate diffusion-time dependency of diffusional kurtosis in the mouse brain using pulsed-gradient spin-echo (PGSE) and oscillating-gradient spin-echo (OGSE) sequences. METHODS 3D PGSE and OGSE kurtosis tensor data were acquired from ex vivo brains of adult, cuprizone-treated, and age-matched control mice with diffusion-time (tD ) ~ 20 ms and frequency (f) = 70 Hz, respectively. Further, 2D acquisitions were performed at multiple times/frequencies ranging from f = 140 Hz to tD = 30 ms with b-values up to 4000 s/mm2 . Monte Carlo simulations were used to investigate the coupled effects of varying restriction size and permeability on time/frequency-dependence of kurtosis with both diffusion-encoding schemes. Simulations and experiments were further performed to investigate the effect of varying number of cycles in OGSE waveforms. RESULTS Kurtosis and diffusivity maps exhibited significant region-specific changes with diffusion time/frequency across both gray and white matter areas. PGSE- and OGSE-based kurtosis maps showed reversed contrast between gray matter regions in the cerebellar and cerebral cortex. Localized time/frequency-dependent changes in kurtosis tensor metrics were found in the splenium of the corpus callosum in cuprizone-treated mouse brains, corresponding to regional demyelination seen with histological assessment. Monte Carlo simulations showed that kurtosis estimates with pulsed- and oscillating-gradient waveforms differ in their sensitivity to exchange. Both simulations and experiments showed dependence of kurtosis on number of cycles in OGSE waveforms for non-zero permeability. CONCLUSION The results show significant time/frequency-dependency of diffusional kurtosis in the mouse brain, which can provide sensitivity to probe intrinsic cellular heterogeneity and pathological alterations in gray and white matter.
Collapse
Affiliation(s)
- Manisha Aggarwal
- Russell H. Morgan Department of Radiology and Radiological Science, Johns Hopkins University School of Medicine, Baltimore, Maryland
| | - Matthew D Smith
- Departments of Neurology and Neuroscience, Johns Hopkins University School of Medicine, Baltimore, Maryland
| | - Peter A Calabresi
- Departments of Neurology and Neuroscience, Johns Hopkins University School of Medicine, Baltimore, Maryland
| |
Collapse
|
33
|
Simultaneous Multislice Echo Planar Imaging for Accelerated Diffusion-Weighted Imaging of Malignant and Benign Breast Lesions. Invest Radiol 2019; 54:524-530. [DOI: 10.1097/rli.0000000000000560] [Citation(s) in RCA: 30] [Impact Index Per Article: 5.0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/21/2022]
|
34
|
Liao Y, Lechea N, Magill AW, Worthoff WA, Gras V, Shah NJ. Correlation of quantitative conductivity mapping and total tissue sodium concentration at 3T/4T. Magn Reson Med 2019; 82:1518-1526. [DOI: 10.1002/mrm.27787] [Citation(s) in RCA: 10] [Impact Index Per Article: 1.7] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 11/23/2018] [Revised: 04/02/2019] [Accepted: 04/07/2019] [Indexed: 01/15/2023]
Affiliation(s)
- Yupeng Liao
- Institute of Neuroscience and Medicine (INM‐4), Forschungszentrum Jülich Jülich Germany
| | - Nazim Lechea
- Institute of Neuroscience and Medicine (INM‐4), Forschungszentrum Jülich Jülich Germany
| | - Arthur W. Magill
- Institute of Neuroscience and Medicine (INM‐4), Forschungszentrum Jülich Jülich Germany
| | - Wieland A. Worthoff
- Institute of Neuroscience and Medicine (INM‐4), Forschungszentrum Jülich Jülich Germany
| | - Vincent Gras
- Institute of Neuroscience and Medicine (INM‐4), Forschungszentrum Jülich Jülich Germany
| | - N. Jon Shah
- Institute of Neuroscience and Medicine (INM‐4), Forschungszentrum Jülich Jülich Germany
- Institute of Neuroscience and Medicine (INM‐11) JARA, Forschungszentrum Jülich Jülich Germany
- JARA‐BRAIN‐Translational Medicine Aachen Germany
- Department of Neurology RWTH Aachen University Aachen Germany
- Monash Biomedical Imaging, School of Psychology Monash University Melbourne Australia
| |
Collapse
|
35
|
Iima M, Yamamoto A, Kataoka M, Yamada Y, Omori K, Feiweier T, Togashi K. Time‐dependent diffusion MRI to distinguish malignant from benign head and neck tumors. J Magn Reson Imaging 2018; 50:88-95. [DOI: 10.1002/jmri.26578] [Citation(s) in RCA: 18] [Impact Index Per Article: 2.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/30/2018] [Revised: 10/25/2018] [Accepted: 10/27/2018] [Indexed: 12/23/2022] Open
Affiliation(s)
- Mami Iima
- Department of Diagnostic Imaging and Nuclear MedicineGraduate School of Medicine, Kyoto University Kyoto Japan
- Hakubi Center for Advanced ResearchKyoto University Kyoto Kyoto Japan
| | - Akira Yamamoto
- Department of Diagnostic Imaging and Nuclear MedicineGraduate School of Medicine, Kyoto University Kyoto Japan
| | - Masako Kataoka
- Department of Diagnostic Imaging and Nuclear MedicineGraduate School of Medicine, Kyoto University Kyoto Japan
| | - Yosuke Yamada
- Department of PathologyGraduate School of Medicine, Kyoto University Kyoto Kyoto Japan
| | - Koichi Omori
- Department of Otolaryngology, Head and Neck SurgeryGraduate School of Medicine, Kyoto University Kyoto Kyoto Japan
| | | | - Kaori Togashi
- Department of Diagnostic Imaging and Nuclear MedicineGraduate School of Medicine, Kyoto University Kyoto Japan
| |
Collapse
|
36
|
Fieremans E, Lee HH. Physical and numerical phantoms for the validation of brain microstructural MRI: A cookbook. Neuroimage 2018; 182:39-61. [PMID: 29920376 PMCID: PMC6175674 DOI: 10.1016/j.neuroimage.2018.06.046] [Citation(s) in RCA: 59] [Impact Index Per Article: 8.4] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/31/2017] [Revised: 06/08/2018] [Accepted: 06/13/2018] [Indexed: 12/24/2022] Open
Abstract
Phantoms, both numerical (software) and physical (hardware), can serve as a gold standard for the validation of MRI methods probing the brain microstructure. This review aims to provide guidelines on how to build, implement, or choose the right phantom for a particular application, along with an overview of the current state-of-the-art of phantoms dedicated to study brain microstructure with MRI. For physical phantoms, we discuss the essential requirements and relevant characteristics of both the (NMR visible) liquid and (NMR invisible) phantom materials that induce relevant microstructural features detectable via MRI, based on diffusion, intra-voxel incoherent motion, magnetization transfer or magnetic susceptibility weighted contrast. In particular, for diffusion MRI, many useful phantoms have been proposed, ranging from simple liquids to advanced biomimetic phantoms consisting of hollow or plain microfibers and capillaries. For numerical phantoms, the focus is on Monte Carlo simulations of random walk, for which the basic principles, along with useful criteria to check and potential pitfalls are reviewed, in addition to a literature overview highlighting recent advances. While many phantoms exist already, the current review aims to stimulate further research in the field and to address remaining needs.
Collapse
Affiliation(s)
- Els Fieremans
- Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, NY, USA.
| | - Hong-Hsi Lee
- Center for Biomedical Imaging, Department of Radiology, New York University School of Medicine, New York, NY, USA
| |
Collapse
|