Amini S, Choi H, Seche W, Blagojevic A, May N, Lefler BM, Davis SL, Elyahoodayan S, Tavousi P, May SJ, Caputo GA, Lowe TC, Hettinger J, Shahbazmohamadi S. Sustainability inspired fabrication of next generation neurostimulation and cardiac rhythm management electrodes via reactive hierarchical surface restructuring. MICROSYSTEMS & NANOENGINEERING 2024; 10:125. [PMID: 39251609 PMCID: PMC11384795 DOI: 10.1038/s41378-024-00754-w] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Received: 11/06/2023] [Revised: 06/05/2024] [Accepted: 06/23/2024] [Indexed: 09/11/2024]
Abstract
Over the last two decades, platinum group metals (PGMs) and their alloys have dominated as the materials of choice for electrodes in long-term implantable neurostimulation and cardiac rhythm management devices due to their superior conductivity, mechanical and chemical stability, biocompatibility, corrosion resistance, radiopacity, and electrochemical performance. Despite these benefits, PGM manufacturing processes are extremely costly, complex, and challenging with potential health hazards. Additionally, the volatility in PGM prices and their high supply risk, combined with their scarce concentration of approximately 0.01 ppm in the earth's upper crust and limited mining geographical areas, underscores their classification as critical raw materials, thus, their effective recovery or substitution worldwide is of paramount importance. Since postmortem recovery from deceased patients and/or refining of PGMs that are used in the manufacturing of the electrodes and microelectrode arrays is extremely rare, challenging, and highly costly, therefore, substitution of PGM-based electrodes with other biocompatible materials that can yield electrochemical performance values equal or greater than PGMs is the only viable and sustainable solution to reduce and ultimately substitute the use of PGMs in long-term implantable neurostimulation and cardiac rhythm management devices. In this article, we demonstrate for the first time how the novel technique of "reactive hierarchical surface restructuring" can be utilized on titanium-that is widely used in many non-stimulation medical device and implant applications-to manufacture biocompatible, low-cost, sustainable, and high-performing neurostimulation and cardiac rhythm management electrodes. We have shown how the surface of titanium electrodes with extremely poor electrochemical performance undergoes compositional and topographical transformations that result in electrodes with outstanding electrochemical performance.
Collapse
Affiliation(s)
- Shahram Amini
- Research and Development, Pulse Technologies Inc., Quakertown, PA, USA.
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, USA.
| | - Hongbin Choi
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, USA
| | - Wesley Seche
- Research and Development, Pulse Technologies Inc., Quakertown, PA, USA
| | | | - Nicholas May
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, USA
| | - Benjamin M Lefler
- Department of Materials Science and Engineering, Drexel University, Philadelphia, PA, USA
| | - Skyler L Davis
- Department of Metallurgical and Materials Engineering, Colorado School of Mines, Golden, CO, USA
| | - Sahar Elyahoodayan
- Department of Biomedical Engineering, University of Southern California, Los Angeles, CA, USA
| | - Pouya Tavousi
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, USA
| | - Steven J May
- Department of Materials Science and Engineering, Drexel University, Philadelphia, PA, USA
| | - Gregory A Caputo
- Department of Chemistry and Biochemistry, Rowan University, Glassboro, NJ, USA
| | - Terry C Lowe
- Department of Metallurgical and Materials Engineering, Colorado School of Mines, Golden, CO, USA
| | - Jeffrey Hettinger
- Department of Physics and Astronomy, Rowan University, Glassboro, NJ, USA
| | | |
Collapse
|
2
|
Ramesh V, Johny J, Jakobi J, Stuckert R, Rehbock C, Barcikowski S. Platinum-Iridium Alloy Nanoparticle Coatings Produced by Electrophoretic Deposition Reduce Impedance in 3D Neural Electrodes. Chemphyschem 2024; 25:e202300623. [PMID: 38842467 DOI: 10.1002/cphc.202300623] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/31/2023] [Revised: 05/14/2024] [Accepted: 05/28/2024] [Indexed: 06/07/2024]
Abstract
Platinum-based neural electrodes, frequently alloyed with Ir or W, are routinely used to treat neurological disorders. However, their performance is impaired by an increase in impedance that compromises long-term implant functionality. Though there are multiple coating techniques available to address this issue, electrode, and base material often exhibit a compositional mismatch, which impairs mechanical stability and may lead to toxicological side effects. In this work, we coated Pt wire electrodes with ligand-free electrostatically stabilized colloidal Pt90Ir10, Pt90W10, and Pt50W50 alloy nanoparticles (NPs) matching electrode compositions using electrophoretic deposition (EPD) with direct-current (DC) and pulsed-DC fields in aqueous medium. The generated alloy NPs exhibit a solid solution structure as evidenced by HR-TEM-EDX and XRD, though additional WOx phases were identified in the Pt50W50 samples. Consequently, coating efficiency was also impaired in the presence of high W mass fractions in the alloy NPs. Characterization of the NP coatings by cyclic voltammetry and impedance spectroscopy yielded a significant reduction of the impedance in the Pt90Ir10 sample in comparison to the control coated with Pt NPs. The electrochemical surface area (ECSA) of the PtW alloy coatings, on the other hand, was significantly reduced.
Collapse
Affiliation(s)
- Vaijayanthi Ramesh
- Technical Chemistry I, University of Duisburg-Essen, Center for Nanointegration Duisburg-Essen (CENIDE), Universitätsstr. 7, 45141, Essen, Germany
| | - Jacob Johny
- Technical Chemistry I, University of Duisburg-Essen, Center for Nanointegration Duisburg-Essen (CENIDE), Universitätsstr. 7, 45141, Essen, Germany
| | - Jurij Jakobi
- Technical Chemistry I, University of Duisburg-Essen, Center for Nanointegration Duisburg-Essen (CENIDE), Universitätsstr. 7, 45141, Essen, Germany
| | - Robert Stuckert
- Technical Chemistry I, University of Duisburg-Essen, Center for Nanointegration Duisburg-Essen (CENIDE), Universitätsstr. 7, 45141, Essen, Germany
| | - Christoph Rehbock
- Technical Chemistry I, University of Duisburg-Essen, Center for Nanointegration Duisburg-Essen (CENIDE), Universitätsstr. 7, 45141, Essen, Germany
| | - Stephan Barcikowski
- Technical Chemistry I, University of Duisburg-Essen, Center for Nanointegration Duisburg-Essen (CENIDE), Universitätsstr. 7, 45141, Essen, Germany
| |
Collapse
|
3
|
Shah DD, Carter P, Shivdasani MN, Fong N, Duan W, Esrafilzadeh D, Poole-Warren LA, Aregueta Robles UA. Deciphering platinum dissolution in neural stimulation electrodes: Electrochemistry or biology? Biomaterials 2024; 309:122575. [PMID: 38677220 DOI: 10.1016/j.biomaterials.2024.122575] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/31/2024] [Revised: 03/28/2024] [Accepted: 04/13/2024] [Indexed: 04/29/2024]
Abstract
Platinum (Pt) is the metal of choice for electrodes in implantable neural prostheses like the cochlear implants, deep brain stimulating devices, and brain-computer interfacing technologies. However, it is well known since the 1970s that Pt dissolution occurs with electrical stimulation. More recent clinical and in vivo studies have shown signs of corrosion in explanted electrode arrays and the presence of Pt-containing particulates in tissue samples. The process of degradation and release of metallic ions and particles can significantly impact on device performance. Moreover, the effects of Pt dissolution products on tissue health and function are still largely unknown. This is due to the highly complex chemistry underlying the dissolution process and the difficulty in decoupling electrical and chemical effects on biological responses. Understanding the mechanisms and effects of Pt dissolution proves challenging as the dissolution process can be influenced by electrical, chemical, physical, and biological factors, all of them highly variable between experimental settings. By evaluating comprehensive findings on Pt dissolution mechanisms reported in the fuel cell field, this review presents a critical analysis of the possible mechanisms that drive Pt dissolution in neural stimulation in vitro and in vivo. Stimulation parameters, such as aggregate charge, charge density, and electrochemical potential can all impact the levels of dissolved Pt. However, chemical factors such as electrolyte types, dissolved gases, and pH can all influence dissolution, confounding the findings of in vitro studies with multiple variables. Biological factors, such as proteins, have been documented to exhibit a mitigating effect on the dissolution process. Other biological factors like cells and fibro-proliferative responses, such as fibrosis and gliosis, impact on electrode properties and are suspected to impact on Pt dissolution. However, the relationship between electrical properties of stimulating electrodes and Pt dissolution remains contentious. Host responses to Pt degradation products are also controversial due to the unknown chemistry of Pt compounds formed and the lack of understanding of Pt distribution in clinical scenarios. The cytotoxicity of Pt produced via electrical stimulation appears similar to Pt-based compounds, including hexachloroplatinates and chemotherapeutic agents like cisplatin. While the levels of Pt produced under clinical and acute stimulation regimes were typically an order of magnitude lower than toxic concentrations observed in vitro, further research is needed to accurately assess the mass balance and type of Pt produced during long-term stimulation and its impact on tissue response. Finally, approaches to mitigating the dissolution process are reviewed. A wide variety of approaches, including stimulation strategies, coating electrode materials, and surface modification techniques to avoid excess charge during stimulation and minimise tissue response, may ultimately support long-term and safe operation of neural stimulating devices.
Collapse
Affiliation(s)
- Dhyey Devashish Shah
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, Australia
| | - Paul Carter
- Cochlear Ltd, Macquarie University, NSW, Australia
| | | | - Nicole Fong
- Cochlear Ltd, Macquarie University, NSW, Australia
| | - Wenlu Duan
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, Australia
| | - Dorna Esrafilzadeh
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, Australia
| | - Laura Anne Poole-Warren
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, Australia; The Tyree Foundation Institute of Health Engineering, University of New South Wales, Sydney, Australia.
| | | |
Collapse
|
4
|
Cuttaz EA, Bailey ZK, Chapman CAR, Goding JA, Green RA. Polymer Bioelectronics: A Solution for Both Stimulating and Recording Electrodes. Adv Healthc Mater 2024:e2304447. [PMID: 38775757 DOI: 10.1002/adhm.202304447] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/13/2023] [Revised: 03/31/2024] [Indexed: 06/01/2024]
Abstract
The advent of closed-loop bionics has created a demand for electrode materials that are ideal for both stimulating and recording applications. The growing complexity and diminishing size of implantable devices for neural interfaces have moved beyond what can be achieved with conventional metallic electrode materials. Polymeric electrode materials are a recent development based on polymer composites of organic conductors such as conductive polymers. These materials present exciting new opportunities in the design and fabrication of next-generation electrode arrays which can overcome the electrochemical and mechanical limitations of conventional electrode materials. This review will examine the recent developments in polymeric electrode materials, their application as stimulating and recording electrodes in bionic devices, and their impact on the development of soft, conformal, and high-density neural interfaces.
Collapse
Affiliation(s)
- Estelle A Cuttaz
- Department of Bioengineering, Imperial College London, London, SW7 2BX, UK
| | - Zachary K Bailey
- Department of Bioengineering, Imperial College London, London, SW7 2BX, UK
| | - Christopher A R Chapman
- Department of Bioengineering, Imperial College London, London, SW7 2BX, UK
- School of Engineering and Materials Science, Queen Mary University of London, London, E1 4NS, UK
| | - Josef A Goding
- Department of Bioengineering, Imperial College London, London, SW7 2BX, UK
| | - Rylie A Green
- Department of Bioengineering, Imperial College London, London, SW7 2BX, UK
| |
Collapse
|
5
|
Gryszel M, Jakešová M, Vu XT, Ingebrandt S, Głowacki ED. Elevating Platinum to Volumetric Capacitance: High Surface Area Electrodes through Reactive Pt Sputtering. Adv Healthc Mater 2024:e2302400. [PMID: 38758352 DOI: 10.1002/adhm.202302400] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/26/2023] [Revised: 05/14/2024] [Indexed: 05/18/2024]
Abstract
Platinum is the most widespread electrode material used for implantable biomedical and neuroelectronic devices, motivating exploring ways to improve its performance and understand its fundamental properties. Using reactive magnetron sputtering, PtOx is prepared, which upon partial reduction yields a porous thin-film form of platinum with favorable properties, notably record-low impedance values outcompeting other reports for platinum-based electrodes. It is established that its high electrochemical capacitance scales with thickness, in the way of volumetric capacitor materials like IrOx and poly(3,4-ethylenedioxythiophene), PEDOT. Unlike these two well-known analogs, however, it is found that PtOx capacitance is not caused by reversible pseudofaradaic reactions but rather due to high surface area. In contrast to IrOx, PtOx is not a reversible valence-change oxide, but rather a porous form of platinum. The findings show that this oxygen-containing form of Pt can place Pt electrodes on a level competitive with IrOx and PEDOT. Due to its relatively low cost and ease of preparation, PtOx can be a good choice for microfabricated bioelectronic devices.
Collapse
Affiliation(s)
- Maciej Gryszel
- Laboratory of Organic Electronics, Department of Science and Technology, Linköping University, Bredgatan 33, Norrköping, 60174, Sweden
| | - Marie Jakešová
- Bioelectronics Materials and Devices Laboratory, Central European Institute of Technology, Brno University of Technology, Purkyňova 123, Brno, 61200, Czech Republic
| | - Xuan Thang Vu
- Institute of Materials in Electrical Engineering 1, RWTH Aachen University, 52074, Aachen, Germany
| | - Sven Ingebrandt
- Institute of Materials in Electrical Engineering 1, RWTH Aachen University, 52074, Aachen, Germany
| | - Eric Daniel Głowacki
- Bioelectronics Materials and Devices Laboratory, Central European Institute of Technology, Brno University of Technology, Purkyňova 123, Brno, 61200, Czech Republic
| |
Collapse
|
6
|
Chen FD, Sharma A, Roszko DA, Xue T, Mu X, Luo X, Chua H, Lo PGQ, Sacher WD, Poon JKS. Development of wafer-scale multifunctional nanophotonic neural probes for brain activity mapping. LAB ON A CHIP 2024; 24:2397-2417. [PMID: 38623840 DOI: 10.1039/d3lc00931a] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 04/17/2024]
Abstract
Optical techniques, such as optogenetic stimulation and functional fluorescence imaging, have been revolutionary for neuroscience by enabling neural circuit analysis with cell-type specificity. To probe deep brain regions, implantable light sources are crucial. Silicon photonics, commonly used for data communications, shows great promise in creating implantable devices with complex optical systems in a compact form factor compatible with high volume manufacturing practices. This article reviews recent developments of wafer-scale multifunctional nanophotonic neural probes. The probes can be realized on 200 or 300 mm wafers in commercial foundries and integrate light emitters for photostimulation, microelectrodes for electrophysiological recording, and microfluidic channels for chemical delivery and sampling. By integrating active optical devices to the probes, denser emitter arrays, enhanced on-chip biosensing, and increased ease of use may be realized. Silicon photonics technology makes possible highly versatile implantable neural probes that can transform neuroscience experiments.
Collapse
Affiliation(s)
- Fu Der Chen
- Max Planck Institute of Microstructure Physics, Weinberg 2, 06120 Halle, Germany.
- Department of Electrical and Computer Engineering, University of Toronto, 10 King's College Road, Toronto, Ontario M5S 3G4, Canada
| | - Ankita Sharma
- Max Planck Institute of Microstructure Physics, Weinberg 2, 06120 Halle, Germany.
- Department of Electrical and Computer Engineering, University of Toronto, 10 King's College Road, Toronto, Ontario M5S 3G4, Canada
| | - David A Roszko
- Max Planck Institute of Microstructure Physics, Weinberg 2, 06120 Halle, Germany.
- Department of Electrical and Computer Engineering, University of Toronto, 10 King's College Road, Toronto, Ontario M5S 3G4, Canada
| | - Tianyuan Xue
- Max Planck Institute of Microstructure Physics, Weinberg 2, 06120 Halle, Germany.
- Department of Electrical and Computer Engineering, University of Toronto, 10 King's College Road, Toronto, Ontario M5S 3G4, Canada
| | - Xin Mu
- Max Planck Institute of Microstructure Physics, Weinberg 2, 06120 Halle, Germany.
- Department of Electrical and Computer Engineering, University of Toronto, 10 King's College Road, Toronto, Ontario M5S 3G4, Canada
| | - Xianshu Luo
- Advanced Micro Foundry Pte Ltd, 11 Science Park Road, Singapore Science Park II, 117685, Singapore
| | - Hongyao Chua
- Advanced Micro Foundry Pte Ltd, 11 Science Park Road, Singapore Science Park II, 117685, Singapore
| | - Patrick Guo-Qiang Lo
- Advanced Micro Foundry Pte Ltd, 11 Science Park Road, Singapore Science Park II, 117685, Singapore
| | - Wesley D Sacher
- Max Planck Institute of Microstructure Physics, Weinberg 2, 06120 Halle, Germany.
| | - Joyce K S Poon
- Max Planck Institute of Microstructure Physics, Weinberg 2, 06120 Halle, Germany.
- Department of Electrical and Computer Engineering, University of Toronto, 10 King's College Road, Toronto, Ontario M5S 3G4, Canada
| |
Collapse
|
7
|
Tsui CT, Mirkiani S, Roszko DA, Churchward MA, Mushahwar VK, Todd KG. In vitro biocompatibility evaluation of functional electrically stimulating microelectrodes on primary glia. Front Bioeng Biotechnol 2024; 12:1351087. [PMID: 38314352 PMCID: PMC10834782 DOI: 10.3389/fbioe.2024.1351087] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 12/06/2023] [Accepted: 01/10/2024] [Indexed: 02/06/2024] Open
Abstract
Neural interfacing devices interact with the central nervous system to alleviate functional deficits arising from disease or injury. This often entails the use of invasive microelectrode implants that elicit inflammatory responses from glial cells and leads to loss of device function. Previous work focused on improving implant biocompatibility by modifying electrode composition; here, we investigated the direct effects of electrical stimulation on glial cells at the electrode interface. A high-throughput in vitro system that assesses primary glial cell response to biphasic stimulation waveforms at 0 mA, 0.15 mA, and 1.5 mA was developed and optimized. Primary mixed glial cell cultures were generated from heterozygous CX3CR-1+/EGFP mice, electrically stimulated for 4 h/day over 3 days using 75 μm platinum-iridium microelectrodes, and biomarker immunofluorescence was measured. Electrodes were then imaged on a scanning electron microscope to assess sustained electrode damage. Fluorescence and electron microscopy analyses suggest varying degrees of localized responses for each biomarker assayed (Hoescht, EGFP, GFAP, and IL-1β), a result that expands on comparable in vivo models. This system allows for the comparison of a breadth of electrical stimulation parameters, and opens another avenue through which neural interfacing device developers can improve biocompatibility and longevity of electrodes in tissue.
Collapse
Affiliation(s)
- Christopher T. Tsui
- Department of Biomedical Engineering, University of Alberta, Edmonton, AB, Canada
- Neurochemical Research Unit, Department of Psychiatry, University of Alberta, Edmonton, AB, Canada
- Neuroscience and Mental Health Institute (NMHI), University of Alberta, Edmonton, AB, Canada
- Sensory Motor Adaptive Rehabilitation Technology (SMART) Network, University of Alberta, Edmonton, AB, Canada
| | - Soroush Mirkiani
- Neuroscience and Mental Health Institute (NMHI), University of Alberta, Edmonton, AB, Canada
- Sensory Motor Adaptive Rehabilitation Technology (SMART) Network, University of Alberta, Edmonton, AB, Canada
| | - David A. Roszko
- Neuroscience and Mental Health Institute (NMHI), University of Alberta, Edmonton, AB, Canada
- Sensory Motor Adaptive Rehabilitation Technology (SMART) Network, University of Alberta, Edmonton, AB, Canada
| | - Matthew A. Churchward
- Neurochemical Research Unit, Department of Psychiatry, University of Alberta, Edmonton, AB, Canada
- Neuroscience and Mental Health Institute (NMHI), University of Alberta, Edmonton, AB, Canada
- Sensory Motor Adaptive Rehabilitation Technology (SMART) Network, University of Alberta, Edmonton, AB, Canada
- Department of Biological and Environmental Sciences, Concordia University of Edmonton, Edmonton, AB, Canada
| | - Vivian K. Mushahwar
- Neuroscience and Mental Health Institute (NMHI), University of Alberta, Edmonton, AB, Canada
- Sensory Motor Adaptive Rehabilitation Technology (SMART) Network, University of Alberta, Edmonton, AB, Canada
- Division of Physical Medicine and Rehabilitation, Department of Medicine, University of Alberta, Edmonton, AB, Canada
| | - Kathryn G. Todd
- Department of Biomedical Engineering, University of Alberta, Edmonton, AB, Canada
- Neurochemical Research Unit, Department of Psychiatry, University of Alberta, Edmonton, AB, Canada
- Neuroscience and Mental Health Institute (NMHI), University of Alberta, Edmonton, AB, Canada
- Sensory Motor Adaptive Rehabilitation Technology (SMART) Network, University of Alberta, Edmonton, AB, Canada
| |
Collapse
|
8
|
Khosla H, Seche W, Ammerman D, Elyahoodayan S, Caputo GA, Hettinger J, Amini S, Feng G. Development of antibacterial neural stimulation electrodes via hierarchical surface restructuring and atomic layer deposition. Sci Rep 2023; 13:19778. [PMID: 37957282 PMCID: PMC10643707 DOI: 10.1038/s41598-023-47256-9] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 07/20/2023] [Accepted: 11/10/2023] [Indexed: 11/15/2023] Open
Abstract
Miniaturization and electrochemical performance enhancement of electrodes and microelectrode arrays in emerging long-term implantable neural stimulation devices improves specificity, functionality, and performance of these devices. However, surgical site and post-implantation infections are amongst the most devastating complications after surgical procedures and implantations. Additionally, with the increased use of antibiotics, the threat of antibiotic resistance is significant and is increasingly being recognized as a global problem. Therefore, the need for alternative strategies to eliminate post-implantation infections and reduce antibiotic use has led to the development of medical devices with antibacterial properties. In this work, we report on the development of electrochemically active antibacterial platinum-iridium electrodes targeted for use in neural stimulation and sensing applications. A two-step development process was used. Electrodes were first restructured using femtosecond laser hierarchical surface restructuring. In the second step of the process, atomic layer deposition was utilized to deposit conformal antibacterial copper oxide thin films on the hierarchical surface structure of the electrodes to impart antibacterial properties to the electrodes with minimal impact on electrochemical performance of the electrodes. Morphological, compositional, and structural properties of the electrodes were studied using multiple modalities of microscopy and spectroscopy. Antibacterial properties of the electrodes were also studied, particularly, the killing effect of the hierarchically restructured antibacterial electrodes on Escherichia coli and Staphylococcus aureus-two common types of bacteria responsible for implant infections.
Collapse
Affiliation(s)
- Henna Khosla
- Department of Mechanical Engineering, Villanova University, Villanova, PA, 19085, USA
| | - Wesley Seche
- Pulse Technologies Inc., Research and Development, Quakertown, PA, 18951, USA
| | - Daniel Ammerman
- Department of Chemistry and Biochemistry, Rowan University, Glassboro, NJ, 08028, USA
| | - Sahar Elyahoodayan
- Department of Biomedical Engineering, University of Southern California, Los Angeles, CA, 90089, USA
| | - Gregory A Caputo
- Department of Chemistry and Biochemistry, Rowan University, Glassboro, NJ, 08028, USA
| | - Jeffrey Hettinger
- Department of Physics and Astronomy, Rowan University, Glassboro, NJ, 08028, USA
| | - Shahram Amini
- Pulse Technologies Inc., Research and Development, Quakertown, PA, 18951, USA.
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, 06269, USA.
| | - Gang Feng
- Department of Mechanical Engineering, Villanova University, Villanova, PA, 19085, USA
| |
Collapse
|
9
|
Li SY, Tseng HY, Chen BW, Lo YC, Shao HH, Wu YT, Li SJ, Chang CW, Liu TC, Hsieh FY, Yang Y, Lai YB, Chen PC, Chen YY. Proof of Concept for Sustainable Manufacturing of Neural Electrode Array for In Vivo Recording. BIOSENSORS 2023; 13:280. [PMID: 36832046 PMCID: PMC9953957 DOI: 10.3390/bios13020280] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Figures] [Subscribe] [Scholar Register] [Received: 12/23/2022] [Revised: 02/01/2023] [Accepted: 02/13/2023] [Indexed: 06/18/2023]
Abstract
Increasing requirements for neural implantation are helping to expand our understanding of nervous systems and generate new developmental approaches. It is thanks to advanced semiconductor technologies that we can achieve the high-density complementary metal-oxide-semiconductor electrode array for the improvement of the quantity and quality of neural recordings. Although the microfabricated neural implantable device holds much promise in the biosensing field, there are some significant technological challenges. The most advanced neural implantable device relies on complex semiconductor manufacturing processes, which are required for the use of expensive masks and specific clean room facilities. In addition, these processes based on a conventional photolithography technique are suitable for mass production, which is not applicable for custom-made manufacturing in response to individual experimental requirements. The microfabricated complexity of the implantable neural device is increasing, as is the associated energy consumption, and corresponding emissions of carbon dioxide and other greenhouse gases, resulting in environmental deterioration. Herein, we developed a fabless fabricated process for a neural electrode array that was simple, fast, sustainable, and customizable. An effective strategy to produce conductive patterns as the redistribution layers (RDLs) includes implementing microelectrodes, traces, and bonding pads onto the polyimide (PI) substrate by laser micromachining techniques combined with the drop coating of the silver glue to stack the laser grooving lines. The process of electroplating platinum on the RDLs was performed to increase corresponding conductivity. Sequentially, Parylene C was deposited onto the PI substrate to form the insulation layer for the protection of inner RDLs. Following the deposition of Parylene C, the via holes over microelectrodes and the corresponding probe shape of the neural electrode array was also etched by laser micromachining. To increase the neural recording capability, three-dimensional microelectrodes with a high surface area were formed by electroplating gold. Our eco-electrode array showed reliable electrical characteristics of impedance under harsh cyclic bending conditions of over 90 degrees. For in vivo application, our flexible neural electrode array demonstrated more stable and higher neural recording quality and better biocompatibility as well during the 2-week implantation compared with those of the silicon-based neural electrode array. In this study, our proposed eco-manufacturing process for fabricating the neural electrode array reduced 63 times of carbon emissions compared to the traditional semiconductor manufacturing process and provided freedom in the customized design of the implantable electronic devices as well.
Collapse
Affiliation(s)
- Szu-Ying Li
- Department of Biomedical Engineering, National Yang Ming Chiao Tung University, No.155, Sec. 2, Linong St., Taipei 112304, Taiwan
| | - Hsin-Yi Tseng
- The Ph.D. Program in Medical Neuroscience, College of Medical Science and Technology, Taipei Medical University, No. 250 Wu-Xing St., Taipei 11031, Taiwan
| | - Bo-Wei Chen
- Department of Biomedical Engineering, National Yang Ming Chiao Tung University, No.155, Sec. 2, Linong St., Taipei 112304, Taiwan
| | - Yu-Chun Lo
- The Ph.D. Program in Medical Neuroscience, College of Medical Science and Technology, Taipei Medical University, No. 250 Wu-Xing St., Taipei 11031, Taiwan
| | - Huai-Hsuan Shao
- Department of Biomedical Engineering, National Yang Ming Chiao Tung University, No.155, Sec. 2, Linong St., Taipei 112304, Taiwan
| | - Yen-Ting Wu
- Department of Biomedical Engineering, National Yang Ming Chiao Tung University, No.155, Sec. 2, Linong St., Taipei 112304, Taiwan
| | - Ssu-Ju Li
- Department of Biomedical Engineering, National Yang Ming Chiao Tung University, No.155, Sec. 2, Linong St., Taipei 112304, Taiwan
| | - Ching-Wen Chang
- Department of Biomedical Engineering, National Yang Ming Chiao Tung University, No.155, Sec. 2, Linong St., Taipei 112304, Taiwan
| | - Ta-Chung Liu
- Department of Biomedical Engineering, National Yang Ming Chiao Tung University, No.155, Sec. 2, Linong St., Taipei 112304, Taiwan
| | - Fu-Yu Hsieh
- Franz Collection Inc., 13F, No. 167, Sec. 5, Ming Sheng E. Rd., Taipei 10589, Taiwan
| | - Yi Yang
- Department of Biomedical Engineering, Johns Hopkins University, No. 720 Rutland Ave., Baltimore, MD 21205, USA
| | - Yan-Bo Lai
- Department of Materials and Mineral Resources Engineering, National Taipei University of Technology, No. 1, Sec. 3, Zhongxiao E. Rd., Taipei 10608, Taiwan
| | - Po-Chun Chen
- Department of Materials and Mineral Resources Engineering, National Taipei University of Technology, No. 1, Sec. 3, Zhongxiao E. Rd., Taipei 10608, Taiwan
| | - You-Yin Chen
- Department of Biomedical Engineering, National Yang Ming Chiao Tung University, No.155, Sec. 2, Linong St., Taipei 112304, Taiwan
- Franz Collection Inc., 13F, No. 167, Sec. 5, Ming Sheng E. Rd., Taipei 10589, Taiwan
| |
Collapse
|
10
|
Ong S, Kullmann A, Mertens S, Rosa D, Diaz-Botia CA. Electrochemical Testing of a New Polyimide Thin Film Electrode for Stimulation, Recording, and Monitoring of Brain Activity. MICROMACHINES 2022; 13:1798. [PMID: 36296151 PMCID: PMC9611492 DOI: 10.3390/mi13101798] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Figures] [Subscribe] [Scholar Register] [Received: 08/20/2022] [Revised: 10/13/2022] [Accepted: 10/15/2022] [Indexed: 06/16/2023]
Abstract
Subdural electrode arrays are used for monitoring cortical activity and functional brain mapping in patients with seizures. Until recently, the only commercially available arrays were silicone-based, whose thickness and lack of conformability could impact their performance. We designed, characterized, manufactured, and obtained FDA clearance for 29-day clinical use (510(k) K192764) of a new thin-film polyimide-based electrode array. This study describes the electrochemical characterization undertaken to evaluate the quality and reliability of electrical signal recordings and stimulation of these new arrays. Two testing paradigms were performed: a short-term active soak with electrical stimulation and a 29-day passive soak. Before and after each testing paradigm, the arrays were evaluated for their electrical performance using Electrochemical Impedance Spectroscopy (EIS), Cyclic Voltammetry (CV) and Voltage Transients (VT). In all tests, the impedance remained within an acceptable range across all frequencies. The different CV curves showed no significant changes in shape or area, which is indicative of stable electrode material. The electrode polarization remained within appropriate limits to avoid hydrolysis.
Collapse
|
11
|
Amini S, Seche W, May N, Choi H, Tavousi P, Shahbazmohamadi S. Femtosecond laser hierarchical surface restructuring for next generation neural interfacing electrodes and microelectrode arrays. Sci Rep 2022; 12:13966. [PMID: 35978090 PMCID: PMC9385846 DOI: 10.1038/s41598-022-18161-4] [Citation(s) in RCA: 3] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/28/2022] [Accepted: 08/05/2022] [Indexed: 11/09/2022] Open
Abstract
Long-term implantable neural interfacing devices are able to diagnose, monitor, and treat many cardiac, neurological, retinal and hearing disorders through nerve stimulation, as well as sensing and recording electrical signals to and from neural tissue. To improve specificity, functionality, and performance of these devices, the electrodes and microelectrode arrays-that are the basis of most emerging devices-must be further miniaturized and must possess exceptional electrochemical performance and charge exchange characteristics with neural tissue. In this report, we show for the first time that the electrochemical performance of femtosecond-laser hierarchically-restructured electrodes can be tuned to yield unprecedented performance values that significantly exceed those reported in the literature, e.g. charge storage capacity and specific capacitance were shown to have improved by two orders of magnitude and over 700-fold, respectively, compared to un-restructured electrodes. Additionally, correlation amongst laser parameters, electrochemical performance and surface parameters of the electrodes was established, and while performance metrics exhibit a relatively consistent increasing behavior with laser parameters, surface parameters tend to follow a less predictable trend negating a direct relationship between these surface parameters and performance. To answer the question of what drives such performance and tunability, and whether the widely adopted reasoning of increased surface area and roughening of the electrodes are the key contributors to the observed increase in performance, cross-sectional analysis of the electrodes using focused ion beam shows, for the first time, the existence of subsurface features that may have contributed to the observed electrochemical performance enhancements. This report is the first time that such performance enhancement and tunability are reported for femtosecond-laser hierarchically-restructured electrodes for neural interfacing applications.
Collapse
Affiliation(s)
- Shahram Amini
- Research and Development, Pulse Technologies Inc., Quakertown, PA, 18951, USA.
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, 06269, USA.
| | - Wesley Seche
- Research and Development, Pulse Technologies Inc., Quakertown, PA, 18951, USA
| | - Nicholas May
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, 06269, USA
| | - Hongbin Choi
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, 06269, USA
| | - Pouya Tavousi
- UConn Tech Park, University of Connecticut, Storrs, CT, 06269, USA
| | - Sina Shahbazmohamadi
- Biomedical Engineering Department, University of Connecticut, Storrs, CT, 06269, USA
| |
Collapse
|
12
|
Tringides CM, Mooney DJ. Materials for Implantable Surface Electrode Arrays: Current Status and Future Directions. ADVANCED MATERIALS (DEERFIELD BEACH, FLA.) 2022; 34:e2107207. [PMID: 34716730 DOI: 10.1002/adma.202107207] [Citation(s) in RCA: 12] [Impact Index Per Article: 6.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 09/10/2021] [Revised: 10/26/2021] [Indexed: 06/13/2023]
Abstract
Surface electrode arrays are mainly fabricated from rigid or elastic materials, and precisely manipulated ductile metal films, which offer limited stretchability. However, the living tissues to which they are applied are nonlinear viscoelastic materials, which can undergo significant mechanical deformation in dynamic biological environments. Further, the same arrays and compositions are often repurposed for vastly different tissues rather than optimizing the materials and mechanical properties of the implant for the target application. By first characterizing the desired biological environment, and then designing a technology for a particular organ, surface electrode arrays may be more conformable, and offer better interfaces to tissues while causing less damage. Here, the various materials used in each component of a surface electrode array are first reviewed, and then electrically active implants in three specific biological systems, the nervous system, the muscular system, and skin, are described. Finally, the fabrication of next-generation surface arrays that overcome current limitations is discussed.
Collapse
Affiliation(s)
- Christina M Tringides
- Harvard Program in Biophysics, Harvard University, Cambridge, MA, 02138, USA
- Harvard-MIT Division in Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, MA, 02139, USA
- Wyss Institute for Biologically Inspired Engineering, Harvard University, Cambridge, MA, 02138, USA
| | - David J Mooney
- Wyss Institute for Biologically Inspired Engineering, Harvard University, Cambridge, MA, 02138, USA
- John A. Paulson School of Engineering and Applied Sciences, Harvard University, Cambridge, MA, 02138, USA
| |
Collapse
|
13
|
Hierarchical platinum–iridium neural electrodes structured by femtosecond laser for superwicking interface and superior charge storage capacity. Biodes Manuf 2021. [DOI: 10.1007/s42242-021-00160-5] [Citation(s) in RCA: 4] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 10/20/2022]
|
14
|
Ramesh V, Rehbock C, Giera B, Karnes JJ, Forien JB, Angelov SD, Schwabe K, Krauss JK, Barcikowski S. Comparing Direct and Pulsed-Direct Current Electrophoretic Deposition on Neural Electrodes: Deposition Mechanism and Functional Influence. LANGMUIR : THE ACS JOURNAL OF SURFACES AND COLLOIDS 2021; 37:9724-9734. [PMID: 34357777 DOI: 10.1021/acs.langmuir.1c01081] [Citation(s) in RCA: 4] [Impact Index Per Article: 1.3] [Reference Citation Analysis] [Abstract] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/13/2023]
Abstract
Electrophoretic deposition (EPD) of platinum nanoparticles (PtNPs) on platinum-iridium (Pt-Ir) neural electrode surfaces is a promising strategy to tune the impedance of electrodes implanted for deep brain stimulation in various neurological disorders such as advanced Parkinson's disease and dystonia. However, previous results are contradicting as impedance reduction was observed on flat samples while in three-dimensional (3D) structures, an increase in impedance was observed. Hence, defined correlations between coating properties and impedance are to date not fully understood. In this work, the influence of direct current (DC) and pulsed-DC electric fields on NP deposition is systematically compared and clear correlations between surface coating homogeneity and in vitro impedance are established. The ligand-free NPs were synthesized via pulsed laser processing in liquid, yielding monomodal particle size distributions, verified by analytical disk centrifugation (ADC). Deposits formed were quantified by UV-vis supernatant analysis and further characterized by scanning electron microscopy (SEM) with semiautomated interparticle distance analyses. Our findings reveal that pulsed-DC electric fields yield more ordered surface coatings with a lower abundance of particle assemblates, while DC fields produce coatings with more pronounced aggregation. Impedance measurements further highlight that impedance of the corresponding electrodes is significantly reduced in the case of more ordered coatings realized by pulsed-DC depositions. We attribute this phenomenon to the higher active surface area of the adsorbed NPs in homogeneous coatings and the reduced particle-electrode electrical contact in NP assemblates. These results provide insight for the efficient EPD of bare metal NPs on micron-sized surfaces for biomedical applications in neuroscience and correlate coating homogeneity with in vitro functionality.
Collapse
Affiliation(s)
- Vaijayanthi Ramesh
- Technical Chemistry I, Center for Nanointegration Duisburg-Essen (CENIDE), University of Duisburg-Essen, 45141 Essen, Germany
| | - Christoph Rehbock
- Technical Chemistry I, Center for Nanointegration Duisburg-Essen (CENIDE), University of Duisburg-Essen, 45141 Essen, Germany
| | - Brian Giera
- Center for Engineered Materials and Manufacturing, Lawrence Livermore National Laboratory, Livermore, California 94550, United States
| | - John J Karnes
- Center for Engineered Materials and Manufacturing, Lawrence Livermore National Laboratory, Livermore, California 94550, United States
| | - Jean-Baptiste Forien
- Center for Engineered Materials and Manufacturing, Lawrence Livermore National Laboratory, Livermore, California 94550, United States
| | - Svilen D Angelov
- Department of Neurosurgery, Hannover Medical School, 30625 Hannover, Germany
| | - Kerstin Schwabe
- Department of Neurosurgery, Hannover Medical School, 30625 Hannover, Germany
| | - Joachim K Krauss
- Department of Neurosurgery, Hannover Medical School, 30625 Hannover, Germany
| | - Stephan Barcikowski
- Technical Chemistry I, Center for Nanointegration Duisburg-Essen (CENIDE), University of Duisburg-Essen, 45141 Essen, Germany
| |
Collapse
|
15
|
Cuttaz EA, Chapman CAR, Syed O, Goding JA, Green RA. Stretchable, Fully Polymeric Electrode Arrays for Peripheral Nerve Stimulation. ADVANCED SCIENCE (WEINHEIM, BADEN-WURTTEMBERG, GERMANY) 2021; 8:2004033. [PMID: 33898185 PMCID: PMC8061359 DOI: 10.1002/advs.202004033] [Citation(s) in RCA: 29] [Impact Index Per Article: 9.7] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 10/21/2020] [Revised: 01/19/2021] [Indexed: 05/08/2023]
Abstract
There is a critical need to transition research level flexible polymer bioelectronics toward the clinic by demonstrating both reliability in fabrication and stable device performance. Conductive elastomers (CEs) are composites of conductive polymers in elastomeric matrices that provide both flexibility and enhanced electrochemical properties compared to conventional metallic electrodes. This work focuses on the development of nerve cuff devices and the assessment of the device functionality at each development stage, from CE material to fully polymeric electrode arrays. Two device types are fabricated by laser machining of a thick and thin CE sheet variant on an insulative polydimethylsiloxane substrate and lamination into tubing to produce pre-curled cuffs. Device performance and stability following sterilization and mechanical loading are compared to a state-of-the-art stretchable metallic nerve cuff. The CE cuffs are found to be electrically and mechanically stable with improved charge transfer properties compared to the commercial cuff. All devices are applied to an ex vivo whole sciatic nerve and shown to be functional, with the CE cuffs demonstrating superior charge transfer and electrochemical safety in the biological environment.
Collapse
Affiliation(s)
- Estelle A. Cuttaz
- Department of BioengineeringImperial CollegeSouth KensingtonLondonSW7 2AZUK
| | | | - Omaer Syed
- Department of BioengineeringImperial CollegeSouth KensingtonLondonSW7 2AZUK
| | - Josef A. Goding
- Department of BioengineeringImperial CollegeSouth KensingtonLondonSW7 2AZUK
| | - Rylie A. Green
- Department of BioengineeringImperial CollegeSouth KensingtonLondonSW7 2AZUK
| |
Collapse
|
16
|
Shekar RCMC, Hansen JHL. An evaluation framework for research platforms to advance cochlear implant/hearing aid technology: A case study with CCi-MOBILE. THE JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA 2021; 149:229. [PMID: 33514127 PMCID: PMC7803384 DOI: 10.1121/10.0002989] [Citation(s) in RCA: 6] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 08/12/2020] [Revised: 12/06/2020] [Accepted: 12/09/2020] [Indexed: 06/12/2023]
Abstract
Cochlear implants (CIs) and hearing aids (HAs) are advanced assistive hearing devices that perform sound processing to achieve acoustic to acoustic/electrical stimulation, thus enabling the prospects for hearing restoration and rehabilitation. Since commercial CIs/HAs are typically constrained by manufacturer design/production constraints, it is necessary for researchers to use research platforms (RPs) to advance algorithms and conduct investigational studies with CI/HA subjects. While previous CI/HA research platforms exist, no study has explored establishing a formal evaluation protocol for the operational safety and reliability of RPs. This study proposes a two-phase analysis and evaluation paradigm for RPs. In the acoustic phase 1 step, a signal processing acoustic space is explored in order to present a sampled set of audio input content to explore the safety of the resulting output electric/acoustic stimulation. In the parameter phase 2 step, the configurable space for realizable electrical stimulation pulses is determined, and overall stimulation reliability and safety are evaluated. The proposed protocol is applied and demonstrated using Costakis Cochlear Implant Mobile. Assessment protocol observations, results, and additional best practices for subsampling of the acoustic and parameter test spaces are discussed. The proposed analysis-evaluation protocol establishes a viable framework for assessing RP operational safety and reliability. Guidelines for adapting the proposed protocol to address variability in RP configuration due to experimental factors such as custom algorithms, stimulation techniques, and/or individualization are also considered.
Collapse
Affiliation(s)
- Ram C M C Shekar
- Cochlear Implant Processing Laboratory, Center for Robust Speech Systems (CRSS-CILab), Department of Electrical and Computer Engineering, Jonsson School of Engineering and Computer Science, University of Texas at Dallas, Richardson, Texas 75080, USA
| | - John H L Hansen
- Cochlear Implant Processing Laboratory, Center for Robust Speech Systems (CRSS-CILab), Department of Electrical and Computer Engineering, Jonsson School of Engineering and Computer Science, University of Texas at Dallas, Richardson, Texas 75080, USA
| |
Collapse
|
17
|
Apollo NV, Murphy B, Prezelski K, Driscoll N, Richardson AG, Lucas TH, Vitale F. Gels, jets, mosquitoes, and magnets: a review of implantation strategies for soft neural probes. J Neural Eng 2020; 17:041002. [PMID: 32759476 PMCID: PMC8152109 DOI: 10.1088/1741-2552/abacd7] [Citation(s) in RCA: 11] [Impact Index Per Article: 2.8] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/10/2023]
Abstract
Implantable neuroelectronic interfaces have enabled breakthrough advances in the clinical diagnosis and treatment of neurological disorders, as well as in fundamental studies of brain function, behavior, and disease. Intracranial electroencephalography (EEG) mapping with stereo-EEG (sEEG) depth electrodes is routinely adopted for precise epilepsy diagnostics and surgical treatment, while deep brain stimulation has become the standard of care for managing movement disorders. Intracortical microelectrode arrays for high-fidelity recordings of neural spiking activity have led to impressive demonstrations of the power of brain-machine interfaces for motor and sensory functional recovery. Yet, despite the rapid pace of technology development, the issue of establishing a safe, long-term, stable, and functional interface between neuroelectronic devices and the host brain tissue still remains largely unresolved. A body of work spanning at least the last 15 years suggests that safe, chronic integration between invasive electrodes and the brain requires a close match between the mechanical properties of man-made components and the neural tissue. In other words, the next generation of invasive electrodes should be soft and compliant, without sacrificing biological and chemical stability. Soft neuroelectronic interfaces, however, pose a new and significant surgical challenge: bending and buckling during implantation that can preclude accurate and safe device placement. In this topical review, we describe the next generation of soft electrodes and the surgical implantation methods for safe and precise insertion into brain structures. We provide an overview of the most recent innovations in the field of insertion strategies for flexible neural electrodes such as dissolvable or biodegradable carriers, microactuators, biologically-inspired support structures, and electromagnetic drives. In our analysis, we also highlight approaches developed in different fields, such as robotic surgery, which could be potentially adapted and translated to the insertion of flexible neural probes.
Collapse
Affiliation(s)
- Nicholas V Apollo
- Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neuroengineering and Therapeutics, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neurotrauma, Neurodegeneration, and Restoration, Corporal Michael J. Crescenz Veterans Affairs Medical Center, Philadelphia, Pennsylvania, 19104, United States of America
| | - Brendan Murphy
- Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neuroengineering and Therapeutics, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neurotrauma, Neurodegeneration, and Restoration, Corporal Michael J. Crescenz Veterans Affairs Medical Center, Philadelphia, Pennsylvania, 19104, United States of America
- These authors contributed equally
| | - Kayla Prezelski
- Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neuroengineering and Therapeutics, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neurotrauma, Neurodegeneration, and Restoration, Corporal Michael J. Crescenz Veterans Affairs Medical Center, Philadelphia, Pennsylvania, 19104, United States of America
- These authors contributed equally
| | - Nicolette Driscoll
- Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neuroengineering and Therapeutics, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neurotrauma, Neurodegeneration, and Restoration, Corporal Michael J. Crescenz Veterans Affairs Medical Center, Philadelphia, Pennsylvania, 19104, United States of America
| | - Andrew G Richardson
- Center for Neuroengineering and Therapeutics, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Department of Neurosurgery, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
| | - Timothy H Lucas
- Center for Neuroengineering and Therapeutics, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Department of Neurosurgery, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
| | - Flavia Vitale
- Department of Bioengineering, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neuroengineering and Therapeutics, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Center for Neurotrauma, Neurodegeneration, and Restoration, Corporal Michael J. Crescenz Veterans Affairs Medical Center, Philadelphia, Pennsylvania, 19104, United States of America
- These authors contributed equally
- Department of Neurology, University of Pennsylvania, Philadelphia, Pennsylvania 19104, United States of America
- Department of Physical Medicine & Rehabilitation, University of Pennsylvania, Philadelphia, Pennsylvania, 19104, United States of America
| |
Collapse
|
18
|
Baek DH, Ahn S, Kim HS, Kim DW. Fabrication of Donut-Type Neural Electrode for Visual Information as Well as Surface Electrical Stimulation. J Med Biol Eng 2020. [DOI: 10.1007/s40846-020-00540-9] [Citation(s) in RCA: 0] [Impact Index Per Article: 0] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/01/2022]
|
19
|
Dalrymple AN, Huynh M, Robles UA, Marroquin JB, Lee CD, Petrossians A, Whalen JJ, Li D, Parkington HC, Forsythe JS, Green RA, Poole-Warren LA, Shepherd RK, Fallon JB. Electrochemical and mechanical performance of reduced graphene oxide, conductive hydrogel, and electrodeposited Pt-Ir coated electrodes: an active in vitro study. J Neural Eng 2019; 17:016015. [PMID: 31652427 DOI: 10.1088/1741-2552/ab5163] [Citation(s) in RCA: 17] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/10/2023]
Abstract
OBJECTIVE To systematically compare the in vitro electrochemical and mechanical properties of several electrode coatings that have been reported to increase the efficacy of medical bionics devices by increasing the amount of charge that can be delivered safely to the target neural tissue. APPROACH Smooth platinum (Pt) ring and disc electrodes were coated with reduced graphene oxide, conductive hydrogel, or electrodeposited Pt-Ir. Electrodes with coatings were compared with uncoated smooth Pt electrodes before and after an in vitro accelerated aging protocol. The various coatings were compared mechanically using the adhesion-by-tape test. Electrodes were stimulated in saline for 24 hours/day 7 days/week for 21 d at 85 °C (1.6-year equivalence) at a constant charge density of 200 µC/cm2/phase. Electrodes were graded on surface corrosion and trace analysis of Pt in the electrolyte after aging. Electrochemical measurements performed before, during, and after aging included electrochemical impedance spectroscopy, cyclic voltammetry, and charge injection limit and impedance from voltage transient recordings. MAIN RESULTS All three coatings adhered well to smooth Pt and exhibited electrochemical advantage over smooth Pt electrodes prior to aging. After aging, graphene coated electrodes displayed a stimulation-induced increase in impedance and reduction in the charge injection limit (p < 0.001), alongside extensive corrosion and release of Pt into the electrolyte. In contrast, both conductive hydrogel and Pt-Ir coated electrodes had smaller impedances and larger charge injection limits than smooth Pt electrodes (p < 0.001) following aging regardless of the stimulus level and with little evidence of corrosion or Pt dissolution. SIGNIFICANCE This study rigorously tested the mechanical and electrochemical performance of electrode coatings in vitro and provided suitable candidates for future in vivo testing.
Collapse
|
20
|
Sikder KU, Shivdasani MN, Fallon JB, Seligman P, Ganesan K, Villalobos J, Prawer S, Garrett DJ. Electrically conducting diamond films grown on platinum foil for neural stimulation. J Neural Eng 2019; 16:066002. [DOI: 10.1088/1741-2552/ab2e79] [Citation(s) in RCA: 8] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/31/2022]
|
21
|
Harris AR, Allitt BJ, Paolini AG. Predicting neural recording performance of implantable electrodes. Analyst 2019; 144:2973-2983. [PMID: 30888346 DOI: 10.1039/c8an02214c] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.4] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/21/2022]
Abstract
Recordings of neural activity can be used to aid communication, control prosthetic devices or alleviate disease symptoms. Chronic recordings require a high signal-to-noise ratio that is stable for years. Current cortical devices generally fail within months to years after implantation. Development of novel devices to increase lifetime requires valid testing protocols and a knowledge of the critical parameters controlling electrophysiological performance. Here we present electrochemical and electrophysiological protocols for assessing implantable electrodes. Biological noise from neural recording has significant impact on signal-to-noise ratio. A recently developed surgical approach was utilised to reduce biological noise. This allowed correlation of electrochemical and electrophysiological behaviour. The impedance versus frequency of modified electrodes was non-linear. It was found that impedance at low frequencies was a stronger predictor of electrophysiological performance than the typically reported impedance at 1 kHz. Low frequency impedance is a function of electrode area, and a strong correlation of electrode area with electrophysiological response was also seen. Use of these standardised testing protocols will allow future devices to be compared before transfer to preclinical and clinical trials.
Collapse
Affiliation(s)
- Alexander R Harris
- ARC Centre of Excellence for Electromaterials Science, Intelligent Polymer Research Institute, University of Wollongong, NSW 2522, Australia.
| | | | | |
Collapse
|
22
|
Chen C, Ruan S, Bai X, Lin C, Xie C, Lee IS. Patterned iridium oxide film as neural electrode interface: Biocompatibility and improved neurite outgrowth with electrical stimulation. MATERIALS SCIENCE & ENGINEERING. C, MATERIALS FOR BIOLOGICAL APPLICATIONS 2019; 103:109865. [PMID: 31349419 DOI: 10.1016/j.msec.2019.109865] [Citation(s) in RCA: 17] [Impact Index Per Article: 3.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Received: 01/24/2019] [Revised: 06/01/2019] [Accepted: 06/03/2019] [Indexed: 01/19/2023]
Abstract
Iridium (Ir) thin film was deposited on patterned titanium substrate by direct-current (DC) magnetron sputtering, and then activated in sulfuric acid (H2SO4) through repetitive potential sweeps to form iridium oxide (IrOx) as neural electrode interface. The resultant IrOx film showed a porous and open morphology with aligned microstructure, exhibited superior electrochemical performance and excellent stability. The IrOx film supported neural stem cells (NSCs) attachment, proliferation and improved processes without causing toxicity. The patterned IrOx films offered a unique system to investigate the synergistic effects of topographical cue and electrical stimulation on neurite outgrowth. Electrical stimulation, when applied through patterned IrOx films, was found to further increase the neurite extension of neuron-like cells and significantly reorient the neurite alignment towards to the direction of stimulation. These results indicate that IrOx film, as electrode-tissue interface is highly stable and biocompatible with excellent electrochemical properties.
Collapse
Affiliation(s)
- Cen Chen
- College of Life Sciences and Medicine, Zhejiang Sci-Tech University, Hangzhou 310018, PR China; School of Pharmaceutical Sciences, Wenzhou Medical University, Wenzhou 325035, PR China; Institute of Natural Sciences, Yonsei University, Seoul 03722, Republic of Korea
| | - Shichao Ruan
- College of Life Sciences and Medicine, Zhejiang Sci-Tech University, Hangzhou 310018, PR China
| | - Xue Bai
- College of Life Sciences and Medicine, Zhejiang Sci-Tech University, Hangzhou 310018, PR China
| | - Chenming Lin
- College of Life Sciences and Medicine, Zhejiang Sci-Tech University, Hangzhou 310018, PR China
| | - Chungang Xie
- College of Life Sciences and Medicine, Zhejiang Sci-Tech University, Hangzhou 310018, PR China
| | - In-Seop Lee
- Institute of Natural Sciences, Yonsei University, Seoul 03722, Republic of Korea.
| |
Collapse
|
23
|
Cuttaz E, Goding J, Vallejo-Giraldo C, Aregueta-Robles U, Lovell N, Ghezzi D, Green RA. Conductive elastomer composites for fully polymeric, flexible bioelectronics. Biomater Sci 2019; 7:1372-1385. [DOI: 10.1039/c8bm01235k] [Citation(s) in RCA: 45] [Impact Index Per Article: 9.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/30/2023]
Abstract
Soft, flexible and stretchable conductive elastomers made of polyurethane and PEDOT:PSS blends were fabricated into fully polymeric implantable bioelectrode arrays.
Collapse
Affiliation(s)
- Estelle Cuttaz
- Department of Bioengineering
- Imperial College London
- London
- UK
- Medtronic Chair in Neuroengineering
| | - Josef Goding
- Department of Bioengineering
- Imperial College London
- London
- UK
| | | | | | - Nigel Lovell
- Graduate School of Biomedical Engineering
- UNSW
- Sydney 2052
- Australia
| | - Diego Ghezzi
- Medtronic Chair in Neuroengineering
- Center for Neuroprosthetics and Institute of Bioengineering
- School of Engineering
- École Polytechnique Fédérale de Lausanne
- Switzerland
| | - Rylie A. Green
- Department of Bioengineering
- Imperial College London
- London
- UK
| |
Collapse
|
24
|
Advances in Penetrating Multichannel Microelectrodes Based on the Utah Array Platform. ADVANCES IN EXPERIMENTAL MEDICINE AND BIOLOGY 2019; 1101:1-40. [PMID: 31729670 DOI: 10.1007/978-981-13-2050-7_1] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 01/10/2023]
Abstract
The Utah electrode array (UEA) and its many derivatives have become a gold standard for high-channel count bi-directional neural interfaces, in particular in human subject applications. The chapter provides a brief overview of leading electrode concepts and the context in which the UEA has to be understood. It goes on to discuss the key advances and developments of the UEA platform in the past 15 years, as well as novel wireless and system integration technologies that will merge into future generations of fully integrated devices. Aspects covered include novel device architectures that allow scaling of channel count and density of electrode contacts, material improvements to substrate, electrode contacts, and encapsulation. Further subjects are adaptations of the UEA platform to support IR and optogenetic simulation as well as an improved understanding of failure modes and methods to test and accelerate degradation in vitro such as to better predict device failure and lifetime in vivo.
Collapse
|
25
|
Chapman CAR, Aristovich K, Donega M, Fjordbakk CT, Stathopoulou TR, Viscasillas J, Avery J, Perkins JD, Holder D. Electrode fabrication and interface optimization for imaging of evoked peripheral nervous system activity with electrical impedance tomography (EIT). J Neural Eng 2018; 16:016001. [PMID: 30444215 DOI: 10.1088/1741-2552/aae868] [Citation(s) in RCA: 18] [Impact Index Per Article: 3.0] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/21/2022]
Abstract
OBJECTIVE Non-invasive imaging techniques are undoubtedly the ideal methods for continuous monitoring of neural activity. One such method, fast neural electrical impedance tomography (EIT) has been developed over the past decade in order to image neural action potentials with non-penetrating electrode arrays. APPROACH The goal of this study is two-fold. First, we present a detailed fabrication method for silicone-based multiple electrode arrays which can be used for epicortical or neural cuff applications. Secondly, we optimize electrode material coatings in order to achieve the best accuracy in EIT reconstructions. MAIN RESULTS The testing of nanostructured electrode interface materials consisting of platinum, iridium oxide, and PEDOT:pTS in saline tank experiments demonstrated that the PEDOT:pTS coating used in this study leads to more accurate reconstruction dimensions along with reduced phase separation between recording channels. The PEDOT:pTS electrodes were then used in vivo to successfully image and localize the evoked activity of the recurrent laryngeal fascicle from within the cervical vagus nerve. SIGNIFICANCE These results alongside the simple fabrication method presented here position EIT as an effective method to image neural activity.
Collapse
Affiliation(s)
- Christopher A R Chapman
- Department of Medical Physics and Biomedical Engineering, University College London, Gower Street, London WC1E 6BT, United Kingdom
| | | | | | | | | | | | | | | | | |
Collapse
|
26
|
Green R. Are ‘next generation’ bioelectronics being designed using old technologies? ACTA ACUST UNITED AC 2018. [DOI: 10.2217/bem-2018-0008] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.5] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 01/07/2023]
Affiliation(s)
- Rylie Green
- Department of Bioengineering, Imperial College London, London, SW7 2BP, UK
| |
Collapse
|
27
|
Shepherd RK, Villalobos J, Burns O, Nayagam DAX. The development of neural stimulators: a review of preclinical safety and efficacy studies. J Neural Eng 2018; 15:041004. [PMID: 29756600 PMCID: PMC6049833 DOI: 10.1088/1741-2552/aac43c] [Citation(s) in RCA: 35] [Impact Index Per Article: 5.8] [Reference Citation Analysis] [Abstract] [MESH Headings] [Grants] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
OBJECTIVE Given the rapid expansion of the field of neural stimulation and the rigorous regulatory approval requirements required before these devices can be applied clinically, it is important that there is clarity around conducting preclinical safety and efficacy studies required for the development of this technology. APPROACH The present review examines basic design principles associated with the development of a safe neural stimulator and describes the suite of preclinical safety studies that need to be considered when taking a device to clinical trial. MAIN RESULTS Neural stimulators are active implantable devices that provide therapeutic intervention, sensory feedback or improved motor control via electrical stimulation of neural or neuro-muscular tissue in response to trauma or disease. Because of their complexity, regulatory bodies classify these devices in the highest risk category (Class III), and they are therefore required to go through a rigorous regulatory approval process before progressing to market. The successful development of these devices is achieved through close collaboration across disciplines including engineers, scientists and a surgical/clinical team, and the adherence to clear design principles. Preclinical studies form one of several key components in the development pathway from concept to product release of neural stimulators. Importantly, these studies provide iterative feedback in order to optimise the final design of the device. Key components of any preclinical evaluation include: in vitro studies that are focussed on device reliability and include accelerated testing under highly controlled environments; in vivo studies using animal models of the disease or injury in order to assess efficacy and, given an appropriate animal model, the safety of the technology under both passive and electrically active conditions; and human cadaver and ex vivo studies designed to ensure the device's form factor conforms to human anatomy, to optimise the surgical approach and to develop any specialist surgical tooling required. SIGNIFICANCE The pipeline from concept to commercialisation of these devices is long and expensive; careful attention to both device design and its preclinical evaluation will have significant impact on the duration and cost associated with taking a device through to commercialisation. Carefully controlled in vitro and in vivo studies together with ex vivo and human cadaver trials are key components of a thorough preclinical evaluation of any new neural stimulator.
Collapse
Affiliation(s)
- Robert K Shepherd
- Bionics Institute, East Melbourne, Australia. Medical Bionics Department, University of Melbourne, Melbourne, Australia
| | | | | | | |
Collapse
|
28
|
Mueller M, de la Oliva N, Del Valle J, Delgado-Martínez I, Navarro X, Stieglitz T. Rapid prototyping of flexible intrafascicular electrode arrays by picosecond laser structuring. J Neural Eng 2018; 14:066016. [PMID: 28695839 DOI: 10.1088/1741-2552/aa7eea] [Citation(s) in RCA: 16] [Impact Index Per Article: 2.7] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 02/08/2023]
Abstract
OBJECTIVE Interfacing the peripheral nervous system can be performed with a large variety of electrode arrays. However, stimulating and recording a nerve while having a reasonable amount of channels limits the number of available systems. Translational research towards human clinical trial requires device safety and biocompatibility but would benefit from design flexibility in the development process to individualize probes. APPROACH We selected established medical grade implant materials like precious metals and Parylene C to develop a rapid prototyping process for novel intrafascicular electrode arrays using a picosecond laser structuring. A design for a rodent animal model was developed in conjunction with an intrafascicular implantation strategy. Electrode characterization and optimization was performed first in saline solution in vitro before performance and biocompatibility were validated in sciatic nerves of rats in chronic implantation. MAIN RESULTS The novel fabrication process proved to be suitable for prototyping and building intrafascicular electrode arrays. Electrochemical properties of the electrode sites were enhanced and tested for long-term stability. Chronic implantation in the sciatic nerve of rats showed good biocompatibility, selectivity and stable stimulation thresholds. SIGNIFICANCE Established medical grade materials can be used for intrafascicular nerve electrode arrays when laser structuring defines structure size in the micro-scale. Design flexibility reduces re-design cycle time and material certificates are beneficial support for safety studies on the way to clinical trials.
Collapse
Affiliation(s)
- Matthias Mueller
- Laboratory for Biomedical Microtechnology, Department of Microsystems Engineering (IMTEK), University of Freiburg, Georges-Koehler-Allee 102, D-79110 Freiburg, Germany
| | | | | | | | | | | |
Collapse
|
29
|
Won SM, Song E, Zhao J, Li J, Rivnay J, Rogers JA. Recent Advances in Materials, Devices, and Systems for Neural Interfaces. ADVANCED MATERIALS (DEERFIELD BEACH, FLA.) 2018; 30:e1800534. [PMID: 29855089 DOI: 10.1002/adma.201800534] [Citation(s) in RCA: 85] [Impact Index Per Article: 14.2] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Received: 01/24/2018] [Revised: 02/28/2018] [Indexed: 06/08/2023]
Abstract
Technologies capable of establishing intimate, long-lived optical/electrical interfaces to neural systems will play critical roles in neuroscience research and in the development of nonpharmacological treatments for neurological disorders. The development of high-density interfaces to 3D populations of neurons across entire tissue systems in living animals, including human subjects, represents a grand challenge for the field, where advanced biocompatible materials and engineered structures for electrodes and light emitters will be essential. This review summarizes recent progress in these directions, with an emphasis on the most promising demonstrated concepts, materials, devices, and systems. The article begins with an overview of electrode materials with enhanced electrical and/or mechanical performance, in forms ranging from planar films, to micro/nanostructured surfaces, to 3D porous frameworks and soft composites. Subsequent sections highlight integration with active materials and components for multiplexed addressing, local amplification, wireless data transmission, and power harvesting, with multimodal operation in soft, shape-conformal systems. These advances establish the foundations for scalable architectures in optical/electrical neural interfaces of the future, where a blurring of the lines between biotic and abiotic systems will catalyze profound progress in neuroscience research and in human health/well-being.
Collapse
Affiliation(s)
- Sang Min Won
- Department of Electrical and Computer Engineering, Frederick Seitz Materials Research Laboratory, University of Illinois at Urbana Champaign, Urbana, IL, 61801, USA
| | - Enming Song
- Department of Materials Science and Engineering, Frederick Seitz Materials Research Laboratory, University of Illinois at Urbana Champaign, Northwestern University, Evanston, IL, 60208, USA
| | - Jianing Zhao
- Department of Mechanical Science and Engineering, Frederick Seitz Materials Research Laboratory, University of Illinois at Urbana Champaign, Urbana, IL, 61801, USA
| | - Jinghua Li
- Department of Materials Science and Engineering, Frederick Seitz Materials Research Laboratory, University of Illinois at Urbana Champaign, Northwestern University, Evanston, IL, 60208, USA
| | - Jonathan Rivnay
- Department of Biomedical Engineering, Simpson Querrey Institute for Nanobiotechnology, Northwestern University, Evanston, IL, 60208, USA
| | - John A Rogers
- Center for Bio-Integrated Electronics, Department of Materials Science and Engineering, Biomedical Engineering, Chemistry, Mechanical Engineering, Electrical Engineering and Computer Science, and Neurological Surgery, Simpson Querrey Institute for Nano/biotechnology, McCormick School of Engineering and Feinberg School of Medicine, Northwestern University, Evanston, IL, 60208, USA
| |
Collapse
|
30
|
Staples NA, Goding JA, Gilmour AD, Aristovich KY, Byrnes-Preston P, Holder DS, Morley JW, Lovell NH, Chew DJ, Green RA. Conductive Hydrogel Electrodes for Delivery of Long-Term High Frequency Pulses. Front Neurosci 2018; 11:748. [PMID: 29375292 PMCID: PMC5768631 DOI: 10.3389/fnins.2017.00748] [Citation(s) in RCA: 21] [Impact Index Per Article: 3.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/16/2017] [Accepted: 12/22/2017] [Indexed: 11/17/2022] Open
Abstract
Nerve block waveforms require the passage of large amounts of electrical energy at the neural interface for extended periods of time. It is desirable that such waveforms be applied chronically, consistent with the treatment of protracted immune conditions, however current metal electrode technologies are limited in their capacity to safely deliver ongoing stable blocking waveforms. Conductive hydrogel (CH) electrode coatings have been shown to improve the performance of conventional bionic devices, which use considerably lower amounts of energy than conventional metal electrodes to replace or augment sensory neuron function. In this study the application of CH materials was explored, using both a commercially available platinum iridium (PtIr) cuff electrode array and a novel low-cost stainless steel (SS) electrode array. The CH was able to significantly increase the electrochemical performance of both array types. The SS electrode coated with the CH was shown to be stable under continuous delivery of 2 mA square pulse waveforms at 40,000 Hz for 42 days. CH coatings have been shown as a beneficial electrode material compatible with long-term delivery of high current, high energy waveforms.
Collapse
Affiliation(s)
- Naomi A Staples
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, NSW, Australia
| | - Josef A Goding
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, NSW, Australia.,Department of Bioengineering, Imperial College London, London, United Kingdom
| | - Aaron D Gilmour
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, NSW, Australia
| | - Kirill Y Aristovich
- Medical Physics and Biomedical Engineering, University College London, London, United Kingdom
| | - Phillip Byrnes-Preston
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, NSW, Australia
| | - David S Holder
- Medical Physics and Biomedical Engineering, University College London, London, United Kingdom
| | - John W Morley
- School of Medical Science, University of New South Wales, Sydney, NSW, Australia.,School of Medicine, Western Sydney University, Sydney, NSW, Australia
| | - Nigel H Lovell
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, NSW, Australia
| | | | - Rylie A Green
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, NSW, Australia.,Department of Bioengineering, Imperial College London, London, United Kingdom
| |
Collapse
|
31
|
Barriga-Rivera A, Bareket L, Goding J, Aregueta-Robles UA, Suaning GJ. Visual Prosthesis: Interfacing Stimulating Electrodes with Retinal Neurons to Restore Vision. Front Neurosci 2017; 11:620. [PMID: 29184478 PMCID: PMC5694472 DOI: 10.3389/fnins.2017.00620] [Citation(s) in RCA: 36] [Impact Index Per Article: 5.1] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 08/31/2017] [Accepted: 10/23/2017] [Indexed: 01/06/2023] Open
Abstract
The bypassing of degenerated photoreceptors using retinal neurostimulators is helping the blind to recover functional vision. Researchers are investigating new ways to improve visual percepts elicited by these means as the vision produced by these early devices remain rudimentary. However, several factors are hampering the progression of bionic technologies: the charge injection limits of metallic electrodes, the mechanical mismatch between excitable tissue and the stimulating elements, neural and electric crosstalk, the physical size of the implanted devices, and the inability to selectively activate different types of retinal neurons. Electrochemical and mechanical limitations are being addressed by the application of electromaterials such as conducting polymers, carbon nanotubes and nanocrystalline diamonds, among other biomaterials, to electrical neuromodulation. In addition, the use of synthetic hydrogels and cell-laden biomaterials is promising better interfaces, as it opens a door to establishing synaptic connections between the electrode material and the excitable cells. Finally, new electrostimulation approaches relying on the use of high-frequency stimulation and field overlapping techniques are being developed to better replicate the neural code of the retina. All these elements combined will bring bionic vision beyond its present state and into the realm of a viable, mainstream therapy for vision loss.
Collapse
Affiliation(s)
- Alejandro Barriga-Rivera
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, NSW, Australia
- Faculty of Engineering and Information Technologies, University of Sydney, Sydney, NSW, Australia
- Division of Neuroscience, University Pablo de Olavide, Sevilla, Spain
| | - Lilach Bareket
- Graduate School of Biomedical Engineering, University of New South Wales, Sydney, NSW, Australia
- Faculty of Engineering and Information Technologies, University of Sydney, Sydney, NSW, Australia
| | - Josef Goding
- Department of Bioengineering, Imperial College London, London, United Kingdom
| | | | - Gregg J. Suaning
- Faculty of Engineering and Information Technologies, University of Sydney, Sydney, NSW, Australia
| |
Collapse
|
32
|
Kumsa D, Steinke GK, Molnar GF, Hudak EM, Montague FW, Kelley SC, Untereker DF, Shi A, Hahn BP, Condit C, Lee H, Bardot D, Centeno JA, Krauthamer V, Takmakov PA. Public Regulatory Databases as a Source of Insight for Neuromodulation Devices Stimulation Parameters. Neuromodulation 2017; 21:117-125. [PMID: 28782181 DOI: 10.1111/ner.12641] [Citation(s) in RCA: 14] [Impact Index Per Article: 2.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 01/10/2017] [Revised: 06/06/2017] [Accepted: 06/14/2017] [Indexed: 11/27/2022]
Abstract
OBJECTIVE The Shannon model is often used to define an expected boundary between non-damaging and damaging modes of electrical neurostimulation. Numerous preclinical studies have been performed by manufacturers of neuromodulation devices using different animal models and a broad range of stimulation parameters while developing devices for clinical use. These studies are mostly absent from peer-reviewed literature, which may lead to this information being overlooked by the scientific community. We aimed to locate summaries of these studies accessible via public regulatory databases and to add them to a body of knowledge available to a broad scientific community. METHODS We employed web search terms describing device type, intended use, neural target, therapeutic application, company name, and submission number to identify summaries for premarket approval (PMA) devices and 510(k) devices. We filtered these records to a subset of entries that have sufficient technical information relevant to safety of neurostimulation. RESULTS We identified 13 product codes for 8 types of neuromodulation devices. These led us to devices that have 22 PMAs and 154 510(k)s and six transcripts of public panel meetings. We found one PMA for a brain, peripheral nerve, and spinal cord stimulator and five 510(k) spinal cord stimulators with enough information to plot in Shannon coordinates of charge and charge density per phase. CONCLUSIONS Analysis of relevant entries from public regulatory databases reveals use of pig, sheep, monkey, dog, and goat animal models with deep brain, peripheral nerve, muscle and spinal cord electrode placement with a variety of stimulation durations (hours to years); frequencies (10-10,000 Hz) and magnitudes (Shannon k from below zero to 4.47). Data from located entries indicate that a feline cortical model that employs acute stimulation might have limitations for assessing tissue damage in diverse anatomical locations, particularly for peripheral nerve and spinal cord simulation.
Collapse
Affiliation(s)
- Doe Kumsa
- Division of Biology, Chemistry and Materials Science, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, White Oak Federal Research Center, Silver Spring, MD, USA.,Medical Device Innovation Consortium, Minneapolis, MN, USA
| | - G Karl Steinke
- Neuromodulation Division, Boston Scientific Corporation, Valencia, CA, USA
| | - Gregory F Molnar
- Medical Device Innovation Consortium, Minneapolis, MN, USA.,Department of Neurology, Medical School, University of Minnesota, Minneapolis, MN, USA
| | - Eric M Hudak
- Department of Research & Technology, Advanced Bionics LLC, Valencia, CA, USA
| | - Fred W Montague
- Department of Biomedical Engineering, Case Western Reserve University, Cleveland, OH, USA
| | | | | | - Alan Shi
- Medtronic Plc, Minneapolis, MN, USA
| | - Benjamin P Hahn
- Neuromodulation Division, Boston Scientific Corporation, Valencia, CA, USA
| | - Chris Condit
- Implantable Electronic Systems Division, St. Jude Medical, Plano, TX, USA
| | - Hyowon Lee
- Weldon School of Biomedical Engineering, Birck Nanotechnology Center, Purdue University, West Lafayette, IN, USA
| | - Dawn Bardot
- Medical Device Innovation Consortium, Minneapolis, MN, USA
| | - Jose A Centeno
- Division of Biology, Chemistry and Materials Science, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, White Oak Federal Research Center, Silver Spring, MD, USA
| | - Victor Krauthamer
- Division of Biomedical Physics, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, White Oak Federal Research Center, Silver Spring, MD, USA
| | - Pavel A Takmakov
- Division of Biology, Chemistry and Materials Science, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration, White Oak Federal Research Center, Silver Spring, MD, USA
| |
Collapse
|
33
|
Leber M, Bhandari R, Mize J, Warren DJ, Shandhi MMH, Solzbacher F, Negi S. Long term performance of porous platinum coated neural electrodes. Biomed Microdevices 2017; 19:62. [DOI: 10.1007/s10544-017-0201-4] [Citation(s) in RCA: 15] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/28/2022]
|
34
|
Carnicer-Lombarte A, Lancashire HT, Vanhoestenberghe A. In vitro biocompatibility and electrical stability of thick-film platinum/gold alloy electrodes printed on alumina. J Neural Eng 2017; 14:036012. [PMID: 28272027 DOI: 10.1088/1741-2552/aa6557] [Citation(s) in RCA: 15] [Impact Index Per Article: 2.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/12/2022]
Abstract
OBJECTIVE High-density electrode arrays are a powerful tool in both clinical neuroscience and basic research. However, current manufacturing techniques require the use of specialised techniques and equipment, which are available to few labs. We have developed a high-density electrode array with customisable design, manufactured using simple printing techniques and with commercially available materials. APPROACH Electrode arrays were manufactured by thick-film printing a platinum-gold alloy (Pt/Au) and an insulating dielectric on 96% alumina ceramic plates. Arrays were conditioned in serum and serum-free conditions, with and without 1 kHz, 200 µA, charge balanced stimulation for up to 21 d. Array biocompatibility was assessed using an extract assay and a PC-12 cell contact assay. Electrode impedance, charge storage capacity and charge injection capacity were before and after array conditioning. MAIN RESULTS The manufactured Pt/Au electrodes have a highly porous surface and exhibit electrical properties comparable to arrays manufactured using alternative techniques. Materials used in array manufacture were found to be non-toxic to L929 fibroblasts by extract assay, and neuronal-like PC-12 cells adhered and extended neurites on the array surfaces. Arrays remained functional after long-term delivery of electrical pulses while exposed to protein-rich environments. Charge storage capacities and charge injection capacities increased following stimulation accounted for by an increase in surface index (real surface area) observed by vertical scanning interferometry. Further, we observed accumulation of proteins at the electrode sites following conditioning in the presence of serum. SIGNIFICANCE This study demonstrates the in vitro biocompatibility of commercially available thick-film printing materials. The printing technique is both simple and versatile, with layouts readily modified to produce customized electrode arrays. Thick-film electrode arrays are an attractive tool that may be implemented for general tissue engineering and neuroscience research.
Collapse
Affiliation(s)
- Alejandro Carnicer-Lombarte
- John Van Geest Centre for Brain Repair, Department of Clinical Neurosciences, University of Cambridge, Cambridge, CB2 0PY, United Kingdom
| | | | | |
Collapse
|
35
|
Lee YJ, Kim HJ, Kang JY, Do SH, Lee SH. Biofunctionalization of Nerve Interface via Biocompatible Polymer-Roughened Pt Black on Cuff Electrode for Chronic Recording. Adv Healthc Mater 2017; 6. [PMID: 28092438 DOI: 10.1002/adhm.201601022] [Citation(s) in RCA: 11] [Impact Index Per Article: 1.6] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 09/05/2016] [Revised: 11/11/2016] [Indexed: 12/21/2022]
Abstract
Peripheral nerve cuff electrodes with roughened Pt black (BPt) are coated with polyethylene glycol (PEG) and Nafion (NF). Although the influence of coated PEG and Nafion on roughened BPt on the electrical properties is weak, the cuff electrode with BPt/PEG and BPt/Nafion exhibits some very important properties. For example, it markedly decreases interfacial impedance, increases charge storage capacity (CSC) due to retaining the BPt surface structure, good stability without exfoliation in repetitive cyclic voltammetry scanning because it is protected by PEG or Nafion coating. In cell viability test, Nafion-coated BPt does not show cytotoxicity to rat Schwann cell line (S16) at 24 and 72 h with the Nafion coating ranging from 0.1 to 10 mg cm-2 . In addition, real-time polymerase chain reaction (PCR) analysis indicates that Schwann cell differentiation (S100 calcium-binding protein B, myelin basic protein, peripheral myelin protein 22), proliferation (proliferating cell nuclear antigen, cyclin-dependent kinase 1 (CDK1)), and adhesion molecules (neural cell adhesion molecule, laminin, fibronectin) are upregulated up to 5 mg cm-2 of Nafion. In animal study, the BPt/Nafion reduces infiltration of fibrotic tissue with high axonal maintenance with upregulation of proliferation (CDK1), adhesion (laminin, neuronal cell adhesion molecule), and neurotrophic factor receptor-related (gdnf family receptor alpha 1) mRNA expressions.
Collapse
Affiliation(s)
- Yi Jae Lee
- Center for BioMicrosystems; Korea Institute of Science and Technology; Seoul 02792 Republic of Korea
| | - Han-Jun Kim
- Department of Clinical Pathology; College of Veterinary Medicine; Konkuk University; Seoul 05029 Republic of Korea
| | - Ji Yoon Kang
- Center for BioMicrosystems; Korea Institute of Science and Technology; Seoul 02792 Republic of Korea
| | - Sun Hee Do
- Department of Clinical Pathology; College of Veterinary Medicine; Konkuk University; Seoul 05029 Republic of Korea
| | - Soo Hyun Lee
- Center for BioMicrosystems; Korea Institute of Science and Technology; Seoul 02792 Republic of Korea
| |
Collapse
|
36
|
Koenen S, Rehbock C, Heissler HE, Angelov SD, Schwabe K, Krauss JK, Barcikowski S. Optimizing in Vitro Impedance and Physico-Chemical Properties of Neural Electrodes by Electrophoretic Deposition of Pt Nanoparticles. Chemphyschem 2017; 18:1108-1117. [PMID: 28122149 DOI: 10.1002/cphc.201601180] [Citation(s) in RCA: 7] [Impact Index Per Article: 1.0] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/28/2016] [Indexed: 11/12/2022]
Abstract
Neural electrodes suffer from an undesired incline in impedance when in permanent contact with human tissue. Nanostructures, induced by electrophoretic deposition (EPD) of ligand-free laser-generated nanoparticles (NPs) on the electrodes are known to stabilize impedance in vivo. Hence, Pt surfaces were systematically EPD-coated with Pt NPs and evaluated for impedance as well as surface coverage, contact angle, electrochemically active surface area (ECSA) and surface oxidation. The aim was to establish a systematic correlation between EPD process parameters and physical surface properties. The findings clearly reveal a linear decrease in impedance with increasing surface coverage, which goes along with a proportional reduction of the contact angle and an increase in ECSA and surface oxidation. EPD process parameters, prone to yield surface coatings with low impedance, are long deposition times (40-60 min), while high colloid concentrations (>250 μg mL-1 ) and electric field strengths (>25 V cm-1 ) should be avoided due to detrimental NP assemblage effects.
Collapse
Affiliation(s)
- Sven Koenen
- Technical Chemistry I and, Center for Nanointegration Duisburg-Essen (CENIDE), University of Duisburg-Essen, Universitätsstrasse 7, 45141, Essen, Germany
| | - Christoph Rehbock
- Technical Chemistry I and, Center for Nanointegration Duisburg-Essen (CENIDE), University of Duisburg-Essen, Universitätsstrasse 7, 45141, Essen, Germany
| | - Hans E Heissler
- Department of Neurosurgery, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany
| | - Svilen D Angelov
- Department of Neurosurgery, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany
| | - Kerstin Schwabe
- Department of Neurosurgery, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany
| | - Joachim K Krauss
- Department of Neurosurgery, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany
| | - Stephan Barcikowski
- Technical Chemistry I and, Center for Nanointegration Duisburg-Essen (CENIDE), University of Duisburg-Essen, Universitätsstrasse 7, 45141, Essen, Germany
| |
Collapse
|
37
|
Barriga-Rivera A, Guo T, Yang CY, Abed AA, Dokos S, Lovell NH, Morley JW, Suaning GJ. High-amplitude electrical stimulation can reduce elicited neuronal activity in visual prosthesis. Sci Rep 2017; 7:42682. [PMID: 28209965 PMCID: PMC5314337 DOI: 10.1038/srep42682] [Citation(s) in RCA: 27] [Impact Index Per Article: 3.9] [Reference Citation Analysis] [Abstract] [Track Full Text] [Download PDF] [Figures] [Journal Information] [Subscribe] [Scholar Register] [Received: 05/27/2016] [Accepted: 01/13/2017] [Indexed: 12/13/2022] Open
Abstract
Retinal electrostimulation is promising a successful therapy to restore functional vision. However, a narrow stimulating current range exists between retinal neuron excitation and inhibition which may lead to misperformance of visual prostheses. As the conveyance of representation of complex visual scenes may require neighbouring electrodes to be activated simultaneously, electric field summation may contribute to reach this inhibitory threshold. This study used three approaches to assess the implications of relatively high stimulating conditions in visual prostheses: (1) in vivo, using a suprachoroidal prosthesis implanted in a feline model, (2) in vitro through electrostimulation of murine retinal preparations, and (3) in silico by computing the response of a population of retinal ganglion cells. Inhibitory stimulating conditions led to diminished cortical activity in the cat. Stimulus-response relationships showed non-monotonic profiles to increasing stimulating current. This was observed in vitro and in silico as the combined response of groups of neurons (close to the stimulating electrode) being inhibited at certain stimulating amplitudes, whilst other groups (far from the stimulating electrode) being recruited. These findings may explain the halo-like phosphene shapes reported in clinical trials and suggest that simultaneous stimulation in retinal prostheses is limited by the inhibitory threshold of the retinal ganglion cells.
Collapse
Affiliation(s)
| | - Tianruo Guo
- Graduate School of Biomedical Engineering, UNSW, Sydney, 2052, Australia
| | - Chih-Yu Yang
- Graduate School of Biomedical Engineering, UNSW, Sydney, 2052, Australia
| | - Amr Al Abed
- Graduate School of Biomedical Engineering, UNSW, Sydney, 2052, Australia
| | - Socrates Dokos
- Graduate School of Biomedical Engineering, UNSW, Sydney, 2052, Australia
| | - Nigel H Lovell
- Graduate School of Biomedical Engineering, UNSW, Sydney, 2052, Australia
| | - John W Morley
- School of Medicine, Western Sydney University, Sydney, 2753, Australia.,School of Medical Science, UNSW, Sydney, 2052, Australia
| | - Gregg J Suaning
- Graduate School of Biomedical Engineering, UNSW, Sydney, 2052, Australia.,Sydney Medical School, University of Sydney, 2000, Australia
| |
Collapse
|
38
|
Concept and Development of an Electronic Framework Intended for Electrode and Surrounding Environment Characterization In Vivo. SENSORS 2016; 17:s17010059. [PMID: 28042815 PMCID: PMC5298632 DOI: 10.3390/s17010059] [Citation(s) in RCA: 3] [Impact Index Per Article: 0.4] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Download PDF] [Figures] [Subscribe] [Scholar Register] [Received: 10/15/2016] [Revised: 12/22/2016] [Accepted: 12/24/2016] [Indexed: 11/17/2022]
Abstract
There has been substantial progress over the last decade towards miniaturizing implantable microelectrodes for use in Active Implantable Medical Devices (AIMD). Compared to the rapid development and complexity of electrode miniaturization, methods to monitor and assess functional integrity and electrical functionality of these electrodes, particularly during long term stimulation, have not progressed to the same extent. Evaluation methods that form the gold standard, such as stimulus pulse testing, cyclic voltammetry and electrochemical impedance spectroscopy, are either still bound to laboratory infrastructure (impractical for long term in vivo experiments) or deliver no comprehensive insight into the material’s behaviour. As there is a lack of cost effective and practical predictive measures to understand long term electrode behaviour in vivo, material investigations need to be performed after explantation of the electrodes. We propose the analysis of the electrode and its environment in situ, to better understand and correlate the effects leading to electrode failure. The derived knowledge shall eventually lead to improved electrode designs, increased electrode functionality and safety in clinical applications. In this paper, the concept, design and prototyping of a sensor framework used to analyse the electrode’s behaviour and to monitor diverse electrode failure mechanisms, even during stimulation pulses, is presented. We focused on the electronic circuitry and data acquisition techniques required for a conceptual multi-sensor system. Functionality of single modules and a prototype framework have been demonstrated, but further work is needed to convert the prototype system into an implantable device. In vitro studies will be conducted first to verify sensor performance and reliability.
Collapse
|
39
|
Kovach KM, Kumsa DW, Srivastava V, Hudak EM, Untereker DF, Kelley SC, von Recum HA, Capadona JR. High-throughput in vitro assay to evaluate the cytotoxicity of liberated platinum compounds for stimulating neural electrodes. J Neurosci Methods 2016; 273:1-9. [PMID: 27485087 DOI: 10.1016/j.jneumeth.2016.07.018] [Citation(s) in RCA: 12] [Impact Index Per Article: 1.5] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 03/09/2016] [Revised: 07/27/2016] [Accepted: 07/28/2016] [Indexed: 12/24/2022]
Abstract
BACKGROUND It is currently unclear how the platinum (Pt) species released from platinum-containing stimulating electrodes may affect the health of the surrounding tissue. This study develops an effective system to assess the cytotoxicity of any electrode-liberated Pt over a short duration, to screen systems before future in vivo testing. NEW METHOD A platinum electrode was stimulated for two hours under physiologically relevant conditions to induce the liberation of Pt species. The total concentration of liberated Pt species was quantified and the concentration found was used to develop a range of Pt species for our model system comprised of microglia and neuron-like cells. RESULTS Under our stimulation conditions (k=2.3, charge density of 57.7μC/cm2), Pt was liberated to a concentration of 1ppm. Interestingly, after 24h of Pt exposure, the dose-dependent cytotoxicity plots revealed that cell death became statistically significant at 10ppm for microglia and 20ppm for neuronal cells. However, in neuron-like cell cultures, concentrations above 1ppm resulted in significant neurite loss after 24h. COMPARISON WITH EXISTING METHODS To our knowledge, there does not exist a simple, in vitro assay system for assessing the cytotoxicity of Pt liberated from stimulating neural electrodes. CONCLUSIONS This work describes a simple model assay that is designed to be applicable to almost any electrode and stimulation system where the electrode is directly juxtaposed to the neural target. Based on the application, the duration of stimulation and Pt exposure may be varied.
Collapse
Affiliation(s)
- Kyle M Kovach
- Department of Biomedical Engineering, Case Western Reserve University, 2071 Martin Luther King Jr. Drive, Wickenden Bldg., Cleveland, OH 44106, USA; Advanced Platform Technology Center, Louis Stokes Cleveland Department of Veterans Affairs Medical Center, 10701 East Blvd, 151W/APT, Cleveland, OH 44106, USA
| | - Doe W Kumsa
- Department of Biomedical Engineering, Case Western Reserve University, 2071 Martin Luther King Jr. Drive, Wickenden Bldg., Cleveland, OH 44106, USA
| | - Vishnupriya Srivastava
- Department of Biomedical Engineering, Case Western Reserve University, 2071 Martin Luther King Jr. Drive, Wickenden Bldg., Cleveland, OH 44106, USA; Advanced Platform Technology Center, Louis Stokes Cleveland Department of Veterans Affairs Medical Center, 10701 East Blvd, 151W/APT, Cleveland, OH 44106, USA
| | - Eric M Hudak
- Department of Research & Technology, Advanced Bionics LLC, 28515 Westinghouse Place, Valencia, CA 91355, USA
| | - Darrel F Untereker
- Medtronic plc, Science and Technology, 710 Medtronic Parkway, Minneapolis, MN 55432, USA
| | - Shawn C Kelley
- Medtronic plc, Science and Technology, 710 Medtronic Parkway, Minneapolis, MN 55432, USA
| | - Horst A von Recum
- Department of Biomedical Engineering, Case Western Reserve University, 2071 Martin Luther King Jr. Drive, Wickenden Bldg., Cleveland, OH 44106, USA; Advanced Platform Technology Center, Louis Stokes Cleveland Department of Veterans Affairs Medical Center, 10701 East Blvd, 151W/APT, Cleveland, OH 44106, USA
| | - Jeffrey R Capadona
- Department of Biomedical Engineering, Case Western Reserve University, 2071 Martin Luther King Jr. Drive, Wickenden Bldg., Cleveland, OH 44106, USA; Advanced Platform Technology Center, Louis Stokes Cleveland Department of Veterans Affairs Medical Center, 10701 East Blvd, 151W/APT, Cleveland, OH 44106, USA.
| |
Collapse
|
40
|
Mueller M, Boehler C, Jaeger J, Asplund M, Stieglitz T. A double-sided fabrication process for intrafascicular parylene C based electrode arrays. ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY. ANNUAL INTERNATIONAL CONFERENCE 2016; 2016:2798-2801. [PMID: 28268899 DOI: 10.1109/embc.2016.7591311] [Citation(s) in RCA: 2] [Impact Index Per Article: 0.3] [Reference Citation Analysis] [Abstract] [MESH Headings] [Track Full Text] [Subscribe] [Scholar Register] [Indexed: 06/06/2023]
Abstract
After the development of a single-sided fabrication process for intrafascicular parylene C based electrode arrays tests showed that an increase in integration density can only be achieved by a double-side process. The process uses 25 μm thick platinum iridium foil, which is thinned down with the laser and sandwiched between two 10 μm thick parylene C layers. Utilizing a picosecond laser (355 nm Nd:YVO4) it was possible to fabricate 40 μm thick electrodes that can be implanted directly in the nerve without relying on additional support layers like chitosan or silk. The fabricated samples feature three 80 μm diameter electrodes on each side and a large ground electrode that is opened to both sides. Impedance mismatches from front to back side as a result of the fabrication process are compensated by electrochemical deposition of nanostructured platinum. This step makes it possible to bring the impedances of the small electrodes down to the range of just a few kΩ at 1 kHz and illustrate the additionally gained surface due to the picosecond laser ablation on the front side electrodes. The safely injectable charge per pulse was found to be 635.75 μC/cm2 for such coated electrodes. Optical investigations show that this fabrication process offers an alternative to established lithographic processes for thin and flexible electrode arrays in neural implants.
Collapse
|
41
|
Lancashire HT, Vanhoestenberghe A, Pendegrass CJ, Ajam YA, Magee E, Donaldson N, Blunn GW. Microchannel neural interface manufacture by stacking silicone and metal foil laminae. J Neural Eng 2016; 13:034001. [PMID: 27001943 DOI: 10.1088/1741-2560/13/3/034001] [Citation(s) in RCA: 9] [Impact Index Per Article: 1.1] [Reference Citation Analysis] [Abstract] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 11/11/2022]
Abstract
OBJECTIVE Microchannel neural interfaces (MNIs) overcome problems with recording from peripheral nerves by amplifying signals independent of node of Ranvier position. Selective recording and stimulation using an MNI requires good insulation between microchannels and a high electrode density. We propose that stacking microchannel laminae will improve selectivity over single layer MNI designs due to the increase in electrode number and an improvement in microchannel sealing. APPROACH This paper describes a manufacturing method for creating MNIs which overcomes limitations on electrode connectivity and microchannel sealing. Laser cut silicone-metal foil laminae were stacked using plasma bonding to create an array of microchannels containing tripolar electrodes. Electrodes were DC etched and electrode impedance and cyclic voltammetry were tested. MAIN RESULTS MNIs with 100 μm and 200 μm diameter microchannels were manufactured. High electrode density MNIs are achievable with electrodes present in every microchannel. Electrode impedances of 27.2 ± 19.8 kΩ at 1 kHz were achieved. Following two months of implantation in Lewis rat sciatic nerve, micro-fascicles were observed regenerating through the MNI microchannels. SIGNIFICANCE Selective MNIs with the peripheral nervous system may allow upper limb amputees to control prostheses intuitively.
Collapse
Affiliation(s)
- Henry T Lancashire
- University College London, John Scales Centre for Biomedical Engineering, Institute of Orthopaedics and Musculoskeletal Science, RNOHT, Stanmore HA7 4LP, UK
| | | | | | | | | | | | | |
Collapse
|
42
|
Gilmour AD, Woolley AJ, Poole-Warren LA, Thomson CE, Green RA. A critical review of cell culture strategies for modelling intracortical brain implant material reactions. Biomaterials 2016; 91:23-43. [PMID: 26994876 DOI: 10.1016/j.biomaterials.2016.03.011] [Citation(s) in RCA: 23] [Impact Index Per Article: 2.9] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 10/30/2015] [Revised: 02/29/2016] [Accepted: 03/06/2016] [Indexed: 02/07/2023]
Abstract
The capacity to predict in vivo responses to medical devices in humans currently relies greatly on implantation in animal models. Researchers have been striving to develop in vitro techniques that can overcome the limitations associated with in vivo approaches. This review focuses on a critical analysis of the major in vitro strategies being utilized in laboratories around the world to improve understanding of the biological performance of intracortical, brain-implanted microdevices. Of particular interest to the current review are in vitro models for studying cell responses to penetrating intracortical devices and their materials, such as electrode arrays used for brain computer interface (BCI) and deep brain stimulation electrode probes implanted through the cortex. A background on the neural interface challenge is presented, followed by discussion of relevant in vitro culture strategies and their advantages and disadvantages. Future development of 2D culture models that exhibit developmental changes capable of mimicking normal, postnatal development will form the basis for more complex accurate predictive models in the future. Although not within the scope of this review, innovations in 3D scaffold technologies and microfluidic constructs will further improve the utility of in vitro approaches.
Collapse
Affiliation(s)
- A D Gilmour
- Graduate School of Biomedical Engineering, The University of New South Wales, Sydney, NSW 2052, Australia.
| | - A J Woolley
- Graduate School of Biomedical Engineering, The University of New South Wales, Sydney, NSW 2052, Australia; Western Sydney University, Sydney, NSW, Australia
| | - L A Poole-Warren
- Graduate School of Biomedical Engineering, The University of New South Wales, Sydney, NSW 2052, Australia
| | - C E Thomson
- Department of Veterinary Medicine, University of Alaska, Fairbanks, AK 99775, USA
| | - R A Green
- Graduate School of Biomedical Engineering, The University of New South Wales, Sydney, NSW 2052, Australia
| |
Collapse
|
43
|
Green R, Abidian MR. Conducting Polymers for Neural Prosthetic and Neural Interface Applications. ADVANCED MATERIALS (DEERFIELD BEACH, FLA.) 2015; 27:7620-37. [PMID: 26414302 PMCID: PMC4681501 DOI: 10.1002/adma.201501810] [Citation(s) in RCA: 198] [Impact Index Per Article: 22.0] [Reference Citation Analysis] [Abstract] [Key Words] [MESH Headings] [Grants] [Track Full Text] [Subscribe] [Scholar Register] [Received: 04/16/2015] [Revised: 06/11/2015] [Indexed: 05/20/2023]
Abstract
Neural-interfacing devices are an artificial mechanism for restoring or supplementing the function of the nervous system, lost as a result of injury or disease. Conducting polymers (CPs) are gaining significant attention due to their capacity to meet the performance criteria of a number of neuronal therapies including recording and stimulating neural activity, the regeneration of neural tissue and the delivery of bioactive molecules for mediating device-tissue interactions. CPs form a flexible platform technology that enables the development of tailored materials for a range of neuronal diagnostic and treatment therapies. In this review, the application of CPs for neural prostheses and other neural interfacing devices is discussed, with a specific focus on neural recording, neural stimulation, neural regeneration, and therapeutic drug delivery.
Collapse
Affiliation(s)
- Rylie Green
- Graduate School of Biomedical Engineering, The University of New South Wales, Sydney, NSW 2052, Australia
| | - Mohammad Reza Abidian
- Biomedical Engineering Department, Materials Science & Engineering Department, Chemical Engineering Department, Materials Research Institute, Huck Institutes of the Life Sciences, Pennsylvania State University, University Park, PA 16802 (USA)
| |
Collapse
|
44
|
Chung T, Wang JQ, Wang J, Cao B, Li Y, Pang SW. Electrode modifications to lower electrode impedance and improve neural signal recording sensitivity. J Neural Eng 2015; 12:056018. [DOI: 10.1088/1741-2560/12/5/056018] [Citation(s) in RCA: 52] [Impact Index Per Article: 5.8] [Reference Citation Analysis] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Indexed: 12/25/2022]
|
45
|
Boehler C, Stieglitz T, Asplund M. Nanostructured platinum grass enables superior impedance reduction for neural microelectrodes. Biomaterials 2015; 67:346-53. [PMID: 26232883 DOI: 10.1016/j.biomaterials.2015.07.036] [Citation(s) in RCA: 88] [Impact Index Per Article: 9.8] [Reference Citation Analysis] [Abstract] [Key Words] [Track Full Text] [Journal Information] [Subscribe] [Scholar Register] [Received: 02/26/2015] [Revised: 07/17/2015] [Accepted: 07/20/2015] [Indexed: 11/27/2022]
Abstract
Micro-sized electrodes are essential for highly sensitive communication at the neural interface with superior spatial resolution. However, such small electrodes inevitably suffer from high electrical impedance and thus high levels of thermal noise deteriorating the signal to noise ratio. In order to overcome this problem, a nanostructured Pt-coating was introduced as add-on functionalization for impedance reduction of small electrodes. In comparison to platinum black deposition, all used chemicals in the deposition process are free from cytotoxic components. The grass-like nanostructure was found to reduce the impedance by almost two orders of magnitude compared to untreated samples which was lower than what could be achieved with conventional electrode coatings like IrOx or PEDOT. The realization of the Pt-grass coating is performed via a simple electrochemical process which can be applied to virtually any possible electrode type and accordingly shows potential as a universal impedance reduction strategy. Elution tests revealed non-toxicity of the Pt-grass and the coating was found to exhibit strong adhesion to the metallized substrate.
Collapse
Affiliation(s)
- C Boehler
- Department of Microsystems Engineering (IMTEK), University of Freiburg, Georges-Koehler-Allee 102, 79110 Freiburg, Germany.
| | - T Stieglitz
- Department of Microsystems Engineering (IMTEK), University of Freiburg, Georges-Koehler-Allee 102, 79110 Freiburg, Germany.
| | - M Asplund
- Department of Microsystems Engineering (IMTEK), University of Freiburg, Georges-Koehler-Allee 102, 79110 Freiburg, Germany.
| |
Collapse
|